Why use an 11° solids content. Modern breweries Why is the reading of lab and
measurement? can now measure this parameter process sometimes different?
Forward scattered light is particle inline and thus provide the ability to The beer in the pipe is homogeneous,
size sensitive and its strength is in release beer automatically. under pressure, and at a constant
detecting particles such as yeast, temperature. In the lab, the samples
trub and Kieselguhr, efficiently Why use a 0° absorption have time to change, solids may
detecting filter breaks and filtrate measurement? precipitate, and temperature changes
turbidity caused by particles, even Using the 0° absorption channel allows will affect the solubility of colloids and
at extremely low concentrations. measurements at higher turbidity lev- with that the measurement. In addi-
11° measurements are very sensi- els, well beyond the range of the 11° tion, differences in the optical congu-
tive in this regard and correlate well or 90° results, allowing you to monitor rations of the instruments themselves,
to actual non-dissolved solids con- and control the lter pre-coat cycle as along with the calibration methods
tent, essential for proper filtration well. This provides an additional oppor- used, have an inuence
on compara-
control and optimization. tunity for lter optimization. tive results.
Control is Power.
Filtration is one of the most important steps in the brewing process. The optek
ptek
Haze Control DTF16 allows you to measure and control this crucial process and
provides real-time quality assessment. In addition to releasing beer by instrument,
rument,
there are tremendous opportunities to reduce beer loss, media usage and
production costs while increasing ltration capacity and ensuring consis-
tent product quality.
TOP 1
Factory Zero Point standards currently in use in process detectors. This design does not rely on
The power behind the optek Haze plants today. This calibration is stable any coatings and maintains a clean-
Control DTF16 is the factory zero point. for the life of the instrument and does able, sanitary design. Combine this
No longer is a questionable zero media not need recalibration, providing a low ingenious design with the legendary
(process water) or offine zero proce- cost of ownership. This ensures precise, optek signal amplication stability, dual
dure required. Using known technical repeatable and reliable measurements beam ratio detectors and advanced
reference solutions and purest zero across multiple instruments and loca- Haze Control rmware and you get a
media, optek developed a factory cali- tions. User selectable units of measure drift and calibration-free process tur-
bration and a precise zero point. This and built in correlation tables provide bidity analyzer with excellent long-time
drift-free factory zero point eliminates unrivalled exibility to the brewery. performance..
the need for regular zeroing of the
instrument saving you time and money Drift-free by Design Information on Demand
and does not depend on the operator. Baseline drift of scattered-light sensors Designed for ease of use, the Haze
It also ensures that every instrument is has long been a problematic issue. Control software has seven languages
base lined identically and precisely. This This is usually caused by unwanted to choose from. In addition to the four
is invaluable when relying on multiple scattered-light, or stray light, from transmitted results, the Haze Control
instrument locations. reections
in the sensor assembly from has a built in data logger providing you
the light source itself, worn articial with continuous trending. This data can
Factoryy Calibration coatings, or external light (i.e. sight be displayed locally and downloaded to
Each and every optek Haze Control glasses). This stray light in
inuences
uences a computer allowing
DTF16 is factory calibrated to EBC the measurement in production personnel
pr
standards which an un- to review process con-
correlate to sistencies or track down
sistenc
all other any historica
historical problems.
Haze Control units and ranges Range of Measurement Haze Control DTF16
Unit EBC Correlation 90° side scatter 11° forward scatter 0° absorption
EBC 1 0-25 0-25 0-500
FTU 4 = 1 EBC 0-100 0-100 0-2000
NTU 4 = 1 EBC 0-100 - -
ASBC-FTU 69 = 1 EBC 0-1725 0-1725 0-34500
Helms 40 = 1 EBC 0-1000 - -
ppm (DE) 6.4 1 EBC* - 0-200 -
* non linear correlation
The instrument may be set by the operator to various lower ranges which can be displayed and transmitted simultaneously.
Communication
mA-OUT Relays
DTF16 AS16-N AS16-F mA-IN Remote-IN
Haze Control 0/4-20 mA 0-50 V AC
11°/90° NIR-Absorp. Color 0/4-20 mA 18-29 V DC
incl. Failsafe 0-75 V DC
HC4301 x - - 2 3 - -
x x -
HC4402 4 3 - -
x - x
HC4321 x - - 2 3 2 7
x x -
HC4422 4 3 2 7
x - x
HC4351 x x - 2 3
HC4452 x - x 4 3
08 I Filter Control – From Line to Lab
TOP 1
filler
www.optek.com
Line and Lab: The 11° forward angle is very sensitive Using the optek DT9011 the sample is
One optical design to particle size so it is an indicator of placed in a water bath and measured
Two of optek newest products are abnormal particulate such as Kiesel- 250 times during one rotation, fol-
specically designed for the beverage guhr and yeast. In addition it corre- lowed by advanced data analysis. This
industry. The optek DT9011 is an “In lates very well to actual non-dissolved eliminates the effects of bottle color
the Bottle” bench top turbidimeter. The content while the 90° side-scatter and shape, as well as scratches, seams
process sensor DTF16 is intended as a angle provides sensitivity to colloids and other imperfections in the sample
true real-time QA/QC turbidity analyzer and haze. Using the 11° forward scat- bottle. 90°
that can eliminate lab assessments ter angle is more specic than 25°and
all together. Both products measure at eliminates false high readings due to
both a forward scatter angle of 11° as haze, because it is nearly independant
well as the side scatter method of 90°. of colloidal turbidity. Any ltration is-
sues, whether recipe, process or equip-
Both results are displayed and record- ment related can now be diagnosed 11°
blend
TOP 2
separator
www.optek.com
Outlet (Centrate) Control Feed Control
Centrifugal separators equipped with Adding a second NIR abssorption optek C4000 Photometric Converter
a NIR absorption based photometer based photometer AF16-N or AS16-N
AF16-N or AS16-N on the outlet, can to the feed line, and measuring incom- yeast back into the claried process
eliminate needless discharges and ing loads directly, allows immediate re- stream. With this instrumentation
initiate them only when solids (yeast) sponse to varying process conditions, constant beer quality can be ensured.
carry-over is detected in the centrate including diverting high solids slugs to Control the turbidity inline without the
stream. Counting the frequency of prevent plugging up a separator bowl. need for sample taking and manual
these discharges is an indicator of in- One mishap like this costs more than dosing and save time and money.
coming solids loads, which can be used the analyzers used to prevent it. Of
to adjust the ow rate to allow maxi- course proper beer/yeast interface Concentrate Control
mum separation performance in high detection upstream will also prevent The concentrate stream of a separator
load conditions, and allows maximum unacceptably high solids loads from can be equipped with a NIR absorp-
throughput at low load conditions. shutting down a separator. In some tion based photometer AS16-N or
Optionally, a scattered light sensor breweries, claried
streams are bright AF16-N using a very short optical path
TF16-N can be installed instead of a enough for release. In this case a length to accurately correlate absorp-
NIR absorption sensor to ensure low- scattered light sensor TF16-N would tion measurements directly to weight-
est measuring ranges, this obviously be used for turbidity QA as well as for percent. This enables yield measure-
depends on your application. separator control. ment and control of product quality.
TOP 3
www.optek.com
Fermentation
In fermenting, AS16-N probes have the
unique ability of mounting directly in
the fermentation vessel or directly in a optek C4000 Photometric Converter
bypass line via different adaptors. Prop-
erly located, the entire fermentation
cycle can be monitored in real-time, in the tank outlet, or in the recovered or Beer / yeast interface, yeast recovery
allowing optimization of the process. spent yeast line. When the yeast is re- also typically uses the AS16-N inline
Tanks can be released automati- moved from the beer the turbidimeter sensor. However, for basic interface
cally rather than waiting hours for lab will determine the interface between applications, where the interface detec-
samples to indicate proper occulation. the yeast and beer. In the past this tion is the only objective, optek offers
Feedback on any recipe inuences
of was accomplished by time / volume or a lower cost probe style sensor in the
yeast cell growth can also be seen. In visual inspection. Since the set points AS56-N. Its lower power lamp mod-
larger cellars, saving 6-10 hours per are adjustable the optimal cut-off level ule also allows for up to four AS56-N
fermentation cycle can add up quickly, can be obtained to provide the most series probes to be multiplexed in one
adding valuable fermentation capacity, ef
fcient operation. This reduces prod- C4000 based system, thus making the
without adding a single fermenter. uct loss and provides a more uniform measurement point cost even more
product for ltration.
attractive.
Beer / Yeast Interface,
Yeast Recovery
Many breweries are using large vertical
fermenters or multi-purpose tanks.
One of the main advantages of these
tanks is avoiding extra product trans-
fers. However, settled yeast must be
removed. A turbidity sensor is installed
TOP 5
www.optek.com
Beer Color is Important the software features
ures of the C4000
The use of precision color sensors photometric converter.
erter.
before the llers can provide color The optek sensor AF26, dual channel
dosing control and quality control absorption based photometer
photometer, is ideal
based on specic color aspects of the for this application. Simultaneously,
brands. Single or dual beam colorim- the reference channel can monitor
eters congured
to monitor to specic
trub and cold break if the AF26 is stra-
visible wavelengths can offer con- tegically located after the wort cooler.
tinuous inline monitoring to minimize
sampling, lab analysis. A consistent Color Dosing
visual impression for the end custom- The color of some beers will be
ers is as important as the possibillity adjusted using i.e. malt extract or rye
for product identication. malt beer. Dosage can be controlled
using an optek sensor. The color of the optek AS16 Single Channel Absorption Probe
Wort Color beer is measured and automated sig-
Wort color is also measured using nals i.e. relays can be used to trigger
absorption. However, beer color is pumps to start dosing the color into product is essential to maintain a
measured using visible light (VIS) at the beer stream. After the dosage, the protable
business; however the costs
430 nm. In the brewhouse this poses color can be veried
by a secondary go beyond the product alone. Water as
a problem as background turbidity sensor. Color of dark and bright beers, the push-out medium is also a com-
(solids) will also absorb light at as well as the color of high gravity modity and has an inherent cost to
430 nm. To compensate for this beer after blending can be controlled. procure and treat. Product and water,
with optek you can measure at two along with yeast, going needlessly
wavelengths, one at 430 nm and a Interface Detection / down the drain are a large contributor
reference wavelength in the NIR (Near Phase Separation to plant ef
fuent and the costs to treat
Infrared). Precise process interface detection it. Local municipalities may also be in-
with a single channel VIS absorption volved; applying charges to the volume
Subtracting the absorbance signal based photometer (optek AS56-F of ef
fuent and even nes if solids load
of the reference channel from the or AF56-F) is the easiest and least or BOD limits are exceeded. Gaining
primary, visible channel signal gives a expensive technique available to brew- control of plant interface functions not
pure color measurement. This result is ers today to reduce product losses. In only makes sense, it’s also environ-
then correlated to ASBC or EBC, using a large brewery, the sheer number of mentally friendly.
change-over operations can cost the
brewer thousands of hectoliters of
product and by-products each week.
In today’s economy, recovering this
14 I Principles of Measurement
Lamp Detector(s)
www.optek.com
Sensor AF16
VIS- and NIR-Absorption,
Abs.
Single Channel Concentration and
Color Measurement
Sensor AF26
VIS-Absorption, Abs.
Turbidity Compensation
11°
90°
Sensor DTF16
Triple Beam Scattered-light Optical Design
11°/ 90° Scattered-light
with Light Compensation and additional
Absorption Measurement for high 11°
concentration
Abs.
11°
I 15
www.optek.com
optek Advantage:
Control is Power
Today, more than ever, the need for
process optimization, product recov-
ery, cost reduction, and reduced waste
is the key to a successful business.
Real-time data from reliable process
scale analyzers from optek is essential
to gain complete process control, real-
ize optimization potential and ensure
protability.
Germany Singapore
optek-Danulat GmbH optek-Danulat Pte. Ltd.
Emscherbruchallee 2 25 Int’l Business Park
45356 Essen / Germany #05-109-f German Centre
Phone: +49-201-63409-0 Singapore 609916
inline control Fax: +49-201-63409-999
E-Mail: [email protected]
Phone: +65 6562 8292
Fax: +65 6562 8293
E-Mail: [email protected]
USA
optek-Danulat Inc. Russian Federation
N118 W18748 Bunsen Drive optek-Danulat
t k GmbH
Germantown WI 53022 / USA Liter A, Room 911, 26-N
Phone: +1 262 437 3600 Kolomjazhskij Prospekt 33
Toll free call: +1 800 371 4288 197341 St. Petersburg
Fax: +1 262 437 3699 Phone: +7 812 380 72 47
E-Mail: [email protected] Fax: +7 812 380 72 05
E-Mail: [email protected]
Netherlands
optek-Danulat bv
Grote Brugse Grintweg 12a
4005 AH Tiel / Netherlands
Phone: +31 -344 -683800
Fax: +31 -344 -653950
E-Mail: [email protected]
Please visit our website for contacts of our local distributors in other countries.
www.optek.com