HPLC 2009
HPLC 2009
HPLC 2009
Chromatography (HPLC)
Learning outcomes
At the end of the class we should be able:
Liquid Chromatography
ISOCRATIC Pump
Detector
Injector
Mobile
phase
Column
Instrument Components
In terms of online LC, the choice of phase separation
processes is more limited than in traditional HPLC
Ion exchange – high strength salt buffers
Size exclusion – not very useful for small molecules,
not very selective
LC/MS is dominated by reversed phase (95%)
chromatography, particularly C18
Normal phase chromatography may still be useful for
some compounds (e.g. small highly organic
molecules), including those that have poor solubility
in water and partially aqueous mobile phases
Normal Phase and Reversed Phase
CH3
H3C CH3
7 Carbons
8 Carbons
6 Carbons
Particle Diameter
Has a greater effect on resolution than length
Small particles – short analysis times
Short columns with small particles ideal
5m is standard size
3m better, but restricted range of packings
available
Downside is high back pressure and issues
with retention of small particles inside the
column, blockages
Sensitivity Considerations and Column ID
1mm ID column
2mm ID column
4.6mm ID column
Normal Phase HPLC
Peak height
% of peak area
Shutdown
Slowly bring the flow rate on the pump back to
zero.
Shut off the detector, pump and degasser
sequentially. Leave solvent line in the solvent
reservoir.
Important Hints,Tips, and Reminders...
Be sure to filter all HLPC solvents with the HPLC filters
and glass ware.
Prime the pump before using the instrument upon a
solvent change!
Practical Hints and Troubleshooting
The relevant
operation for
this HPLC...
(from www.
http://www.alltechweb.com/Technical/ColumnCare/Manual_Insert1.htm)
Regeneration Procedure
Noise Spikes Bubbles passing through cell Degas mobile phase and/or apply
backpressure to the sample cell
High pressure fitting not making Tighten Fittings
an air tight seal
Excessive Noise Sample cell windows are contaminated Flush cell with solvents to remove
contamination
Bubble trapped in sample cell Increase backpressure on flow cell
UV lamp is sputtering Replace UV lamp
Pump flow pulsing or inconsistent Shut off pump. If baseline quiet:
service pump
Excessive Drift Insufficient warm-up Allow detector to warm-up for 30 min.
Sample cell is contaminated Flush cell with solvents to remove
contamination
Column is not equilibrated Allow time for column to equilibrate
UV absorbing components in mobile phase Replace solvents with HPLC/UV grade
(Excerpts taken from the Beckman 110B Solvent Delivery Module Manual)