Comparative Accuracy and Reliability of Three Electronic Apex Locators in Determining the Apical Constriction of Molar Canals: A Micro-CT Evaluation
Abstract
:1. Introduction
2. Materials and Methods
2.1. Sample Selection
2.2. Sample Preparation
2.3. Measurement of the Working Length (WL)
2.4. Micro-CT Scanning and Determining Distance from Apical Constriction
2.5. First Scan—Pre-WL Measurement
2.6. Second Scan—Post-WL Measurement
2.7. Data Analysis
3. Results
3.1. Accuracy of Electronic Apex Locators (EALs)
3.2. Deviation from Apical Constriction
3.3. Reliability of EAL Readings
4. Discussion
5. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Conflicts of Interest
References
- Ricucci, D.; Langeland, K. Apical Limit of Root Canal Instrumentation and Obturation, Part 2. A Histological Study. Int. Endod. J. 1998, 31, 394–409. [Google Scholar] [CrossRef] [PubMed]
- Sjogren, U.; Hagglund, B.; Sundqvist, G.; Wing, K. Factors Affecting the Long-Term Results of Endodontic Treatment. J. Endod. 1990, 16, 498–504. [Google Scholar] [CrossRef]
- Torabinejad, M.; Walton, R.E. Endodontics: Principles and Practice, 4th ed.; Elsevier: Amsterdam, The Netherlands, 2015. [Google Scholar]
- Cohen, S.; Hargreaves, K.M. Pathways of the Pulp, 11th ed.; Elsevier: Amsterdam, The Netherlands, 2016. [Google Scholar]
- Ali, R.; Okechukwu, N.C.; Brunton, P.; Nattress, B. An Overview of Electronic Apex Locators: Part 1. Br. Dent. J. 2013, 214, 155–158. [Google Scholar] [CrossRef] [PubMed]
- Heo, M.-S.; Han, D.-H.; An, B.-M.; Huh, K.-H.; Yi, W.-J.; Lee, S.-S.; Choi, S.-C. Effect of Ambient Light and Bit Depth of Digital Radiograph on Observer Performance in Determination of Endodontic File Positioning. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. Endod. 2008, 105, 239–244. [Google Scholar] [CrossRef]
- Custer, L.E. Exact Methods of Locating the Apical Foramen. J. Natl. Dent. Assoc. 1918, 5, 815–819. [Google Scholar] [CrossRef]
- Sunada, I. New Method for Measuring the Length of the Root Canal. J. Dent. Res. 1962, 41, 375–387. [Google Scholar] [CrossRef]
- Kobayashi, C.; Suda, H. New Electronic Canal Measuring Device Based on the Ratio Method. J. Endod. 1994, 20, 111–114. [Google Scholar] [CrossRef]
- Shirazi, Z.; Al-Jadaa, A.; Saleh, A.R. Electronic Apex Locators and Their Implications in Contemporary Clinical Practice: A Review. Open Dent. J. 2023, 17, 1–10. [Google Scholar] [CrossRef]
- ESE. Quality Guidelines for Endodontic Treatment: Consensus Report of the European Society of Endodontology. Int. Endod. J. 2006, 39, 921–930. [Google Scholar] [CrossRef]
- AAE. Treatment Standards: Executive Summary. Am. Assoc. Endodontists 2019, 31, 333–340. [Google Scholar]
- Nasiri, K.; Wrbas, K.-T. Accuracy of Different Generations of Apex Locators in Determining Working Length; a Systematic Review and Meta-Analysis. Saudi Dent. J. 2022, 34, 11–20. [Google Scholar] [CrossRef] [PubMed]
- Martins, J.N.R.; Marques, D.; Mata, A.; Caramês, J. Clinical Efficacy of Electronic Apex Locators: Systematic Review. J. Endod. 2014, 40, 759–777. [Google Scholar] [CrossRef] [PubMed]
- Kim, E.; Lee, S.-J. Electronic Apex Locator. Dent. Clin. N. Am. 2004, 48, 35–54. [Google Scholar] [CrossRef] [PubMed]
- Gordon, M.P.J.; Chandler, N.P. Electronic Apex Locators. Int. Endod. J. 2004, 37, 425–437. [Google Scholar] [CrossRef]
- Welk, A.R.; Baumgartner, J.C.; Marshall, J.G. An in Vivo Comparison of Two Frequency-Based Electronic Apex Locators. J. Endod. 2003, 29, 497–500. [Google Scholar] [CrossRef]
- Fouad, A.F.; Reid, L.C. Effect of Using Electronic Apex Locators on Selected Endodontic Treatment Parameters. J. Endod. 2000, 26, 364–367. [Google Scholar] [CrossRef]
- Villas-Bôas, M.H.; Bernardineli, N.; Cavenago, B.C.; Marciano, M.; Del Carpio-Perochena, A.; de Moraes, I.G.; Duarte, M.H.; Bramante, C.M.; Ordinola-Zapata, R. Micro-Computed Tomography Study of the Internal Anatomy of Mesial Root Canals of Mandibular Molars. J. Endod. 2011, 37, 1682–1686. [Google Scholar] [CrossRef]
- Piasecki, L.; Carneiro, E.; da Silva Neto, U.X.; Westphalen, V.P.D.; Brandão, C.G.; Gambarini, G.; Azim, A.A. The Use of Micro-Computed Tomography to Determine the Accuracy of 2 Electronic Apex Locators and Anatomic Variations Affecting Their Precision. J. Endod. 2016, 42, 1263–1267. [Google Scholar] [CrossRef]
- Pascon, E.A.; Marrelli, M.; Congi, O.; Ciancio, R.; Miceli, F.; Versiani, M.A. An Ex Vivo Comparison of Working Length Determination by 3 Electronic Apex Locators. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. Endod. 2009, 108, e147–e151. [Google Scholar] [CrossRef]
- Tsesis, I.; Blazer, T.; Ben-Izhack, G.; Taschieri, S.; Del Fabbro, M.; Corbella, S.; Rosen, E. The Precision of Electronic Apex Locators in Working Length Determination: A Systematic Review and Meta-Analysis of the Literature. J. Endod. 2015, 41, 1818–1823. [Google Scholar] [CrossRef]
- Vasconcelos, B.C.D.; Araújo, R.B.R.; Luna-Cruz, S.M.; Duarte, M.A.H.; Fernandes, C.A.D.O. In Vivo Accuracy of Two Electronic Foramen Locators Based on Different Operation Systems. Braz. Dent. J. 2014, 25, 12–16. [Google Scholar] [CrossRef]
- Dentsply Sirona Operating Instructions Intego. Intego/Intego Pro Intego Ambidextrous/Intego Pro Ambidextrous; Dentsply Sirona: Charlotte, NC, USA, 2018; Available online: https://www.dmiequipment.ie/wp-content/uploads/2019/05/Intego-User-Manual.pdf (accessed on 25 July 2024).
- Sirona, D. Sinius/Sinius CS/Sinius TS Operating Instructions English; Dentsply Sirona: Charlotte, NC, USA, 2018. [Google Scholar]
- Schell, S.; Judenhofer, M.S.; Mannheim, J.G.; Hülber-J, M.; Löst, C.; Pichler, B.J.; ElAyouti, A. Validity of Longitudinal Sections for Determining the Apical Constriction. Int. Endod. J. 2017, 50, 706–712. [Google Scholar] [CrossRef]
- Connert, T.; Judenhofer, M.S.; Hülber-J, M.; Schell, S.; Mannheim, J.G.; Pichler, B.J.; Löst, C.; ElAyouti, A. Evaluation of the Accuracy of Nine Electronic Apex Locators by Using Micro-CT. Int. Endod. J. 2018, 51, 223–232. [Google Scholar] [CrossRef] [PubMed]
- Saatchi, M.; Iravani, S.; Akhavan Khaleghi, M.; Mortaheb, A. Influence of Root Canal Curvature on the Accuracy of Root ZX Electronic Foramen Locator: An In Vitro Study. Iran. Endod. J. 2017, 12, 173–178. [Google Scholar] [PubMed]
- ElAyouti, A.; Connert, T.; Dummer, P.; Löst, C. A Critical Analysis of Research Methods and Experimental Models to Study Working Length Determination and the Performance of Apex Locators—A Narrative Review with Recommendations for the Future. Int. Endod. J. 2022, 55, 281–294. [Google Scholar] [CrossRef] [PubMed]
- Dawson, V.S.; Fransson, H.; Isberg, P.-E.; Bjørndal, L.; Dawson, V.S.; Fransson, H.; Frisk, F.; Jonasson, P.; Kvist, T.; Markvart, M.; et al. Further Interventions after Root Canal Treatment Are Most Common in Molars and Teeth Restored with Direct Restorations: A 10–11-Year Follow-Up of the Adult Swedish Population. J. Endod. 2024, 50, 766–773. [Google Scholar] [CrossRef]
- Piasecki, L.; José Dos Reis, P.; Jussiani, E.I.; Andrello, A.C. A Micro-Computed Tomographic Evaluation of the Accuracy of 3 Electronic Apex Locators in Curved Canals of Mandibular Molars. J. Endod. 2018, 44, 1872–1877. [Google Scholar] [CrossRef]
- Saatchi, M.; Rahimi, I.; Khademi, A.; Farhad, A.R.; Nekoofar, M.H.; Dummer, P.M.H. Influence of Tooth Length on the Accuracy of the Root ZX Electronic Apical Foramen Locator: An Ex Vivo Study. Acta Odontol. Scand. 2014, 73, 101–106. [Google Scholar] [CrossRef]
- Lipski, M.; Trąbska-Świstelnicka, M.; Woźniak, K.; Dembowska, E.; Droździk, A. Evaluation of Alginate as a Substitute for Root-Surrounding Tissues in Electronic Root Canal Measurements. Aust. Endod. J. J. Aust. Soc. Endodontology Inc 2013, 39, 155–158. [Google Scholar] [CrossRef]
- Wolf, T.G.; Krauß-Mironjuk, A.; Wierichs, R.J.; Briseño-Marroquín, B. Influence of Embedding Media on the Accuracy of Working Length Determination by Means of Apex Locator: An Ex Vivo Study. Sci. Rep. 2021, 11, 3340. [Google Scholar] [CrossRef]
- De-Deus, G.; Cozer, V.; Souza, E.M.; Silva, E.J.N.L.; Belladonna, F.G.; Simões-Carvalho, M.; Versiani, M.A. Clinical Accuracy and Precision of 3 Multifrequency Electronic Apex Locators Assessed through Micro–Computed Tomographic Imaging. J. Endod. 2023, 49, 487–495. [Google Scholar] [CrossRef] [PubMed]
- Bland, J.M.; Altman, D.G. Statistical Methods for Assessing Agreement between Two Methods of Clinical Measurement. Lancet 1986, 1, 307–310. [Google Scholar] [CrossRef] [PubMed]
- Cury, M.T.S.; Vasques, A.M.V.; Bueno, C.R.E.; Machado, T.; Trizzi, J.Q.; Santana, V.d.S.; Dezan-Júnior, E. Accuracy of Root ZXII, E-PEX and FIND Apex Locators in Teeth with Vital Pulp: An in Vivo Study. Braz. Oral Res. 2021, 35, e080. [Google Scholar] [CrossRef]
- Srivastava, S.; Alharbi, H.M.; Soliman, M.; Eldwakhly, E.; Abdelhafeez, M.M. Influence of Open Apex on Working Length Determination Using Cone-Beam Computed Tomography and Apex Locators: A Comparative In Vitro Study. Biomed Res. Int. 2022, 2022, 3811983. [Google Scholar] [CrossRef]
- Dennis, D.; Farahanny, W.; Prasetia, W.; Batubara, F.Y. Biological and Mechanical Principles of Chemomechanical Preparation in Root Canal Therapy: A Review. Int. J. Clin. Dent. 2021, 14, 187–198. [Google Scholar]
- Fortea, L.; Sanz-Serrano, D.; Luz, L.-B.; Bardini, G.; Mercade, M. Update on Chelating Agents in Endodontic Treatment: A Systematic Review. J. Clin. Exp. Dent. 2024, 16, e516–e538. [Google Scholar] [CrossRef]
- Chandler, N.; Chellappa, D. Lubrication during Root Canal Treatment. Aust. Endod. J. 2019, 45, 106–110. [Google Scholar] [CrossRef]
- Boessler, C.; Peters, O.A.; Zehnder, M. Impact of Lubricant Parameters on Rotary Instrument Torque and Force. J. Endod. 2007, 33, 280–283. [Google Scholar] [CrossRef]
- Cheung, G.S.P.; Shen, Y.; Darvell, B.W. Effect of Environment on Low-Cycle Fatigue of a Nickel–Titanium Instrument. J. Endod. 2007, 33, 1433–1437. [Google Scholar] [CrossRef]
- Shin, H.-S.; Yang, W.-K.; Kim, M.-R.; Ko, H.-J.; Cho, K.-M.; Park, S.-H.; Kim, J.-W. Accuracy of Root ZX in Teeth with Simulated Root Perforation in the Presence of Gel or Liquid Type Endodontic Irrigant. Restor. Dent. Endod. 2012, 37, 149. [Google Scholar] [CrossRef]
- Yahya, E.M.; Alchalabi, A.S.; Alkhalidi, E.F. Effect of Different Electroconductive Root Canal Irrigations on the Accuracy of Different Apex Locators: An In Vitro Comparative Study. J. Int. Soc. Prev. Community Dent. 2024, 14, 211–217. [Google Scholar] [CrossRef]
- Prasad, A.B.; Srivastava, H.; Srivastava, A.A.; Raisingani, D. An in Vitro Evaluation of the Accuracy of Two Electronic Apex Locators to Determine Working Length in the Presence of Various Irrigants. Ethiop. J. Health Sci. 2016, 26, 457. [Google Scholar] [CrossRef]
- Bhagat, E.; Agnihotri, Y.; Das, A.; Das, S.; Panda, S.; Hota, S. Accuracy of an Electronic Apex Locator (EAL) in Measuring the Working Length of Root Canals in Comparison With Radiographs: An In Vivo Study. Cureus 2024, 16, e60283. [Google Scholar] [CrossRef] [PubMed]
- Jha, P.; Nikhil, V.; Raj, S.; Ravinder, R.; Mishra, P. Accuracy of Electronic Apex Locator in the Presence of Different Irrigating Solutions. Endodontology 2021, 33, 232. [Google Scholar] [CrossRef]
- da Silva, T.M.; Alves, F.R. Ex Vivo Accuracy of Root Zx Ii, Root Zx Mini and RomiApex A-15 Apex Locators in Extracted Vital Pulp Teeth. J. Contemp. Dent. Pract. 2014, 15, 312–314. [Google Scholar] [CrossRef]
- Stoll, R.; Urban-Klein, B.; Roggendorf, M.J.; Jablonski-Momeni, A.; Strauch, K.; Frankenberger, R. Effectiveness of Four Electronic Apex Locators to Determine Distance from the Apical Foramen. Int. Endod. J. 2010, 43, 808–817. [Google Scholar] [CrossRef]
- Swapna, D.; Krishna, A.; Patil, A.; Rashmi, K.; Pai, V.; Ranjini, M. Comparison of Third Generation versus Fourth Generation Electronic Apex Locators in Detecting Apical Constriction: An in Vivo Study. J. Conserv. Dent. 2015, 18, 288. [Google Scholar] [CrossRef]
- D’Assunção, F.L.C.; de Albuquerque, D.S.; Salazar-Silva, J.R.; de Queiroz Ferreira, L.C.; Bezerra, P.M. The Accuracy of Root Canal Measurements Using the Mini Apex Locator and Root ZX-II: An Evaluation in Vitro. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. Endodontol. 2007, 104, e50–e53. [Google Scholar] [CrossRef]
- D’Assunção, F.L.C.; Albuquerque, D.S.; Salazar-Silva, J.R.; Dos Santos, V.C.; Sousa, J.C.N. Ex Vivo Evaluation of the Accuracy and Coefficient of Repeatability of Three Electronic Apex Locators Using a Simple Mounting Model: A Preliminary Report. Int. Endod. J. 2010, 43, 269–274. [Google Scholar] [CrossRef]
- Monteiro Czornobay, L.F.; de Oliveira Rocha, A.; Leite, C.C.; Menezes dos Anjos, L.; Ribeiro, J.S.; Pacheco de Castro Henriques, B.A.; Cardoso, M.; Machado, R.; da Silveira Teixeira, C.; da Fonseca Roberti Garcia, L. A Global Overview on Electronic Apex Locator Use From 1968 to 2023: A Bibliometric Analysis. J. Endod. 2024, 50, 925–933. [Google Scholar] [CrossRef]
- López-Maekawa, A.; García Rupay, C.R. Accuracy of Four Apex Locators in Premolars with Root Resorption: An In Vitro Study. Iran. Endod. J. 2022, 17, 20–26. [Google Scholar] [CrossRef] [PubMed]
- Shekarchizade, N.; Farhad, A.; Khalifezade, S. The The Accuracy of Three Apex Locators in Determining the Location of Strip Root Perforation in Different Environments. Iran. Endod. J. 2021, 16, 184–188. [Google Scholar] [CrossRef] [PubMed]
- El Hachem, R.; Wassef, E.; Mokbel, N.; Abboud, R.; Zogheib, C.; El Osta, N.; Naaman, A. Influence of the Maxillary Sinus on the Accuracy of the Root ZX Apex Locator: An Ex Vivo Study. Dent. J. 2019, 7, 3. [Google Scholar] [CrossRef]
Device/Medium | Mean ±Std. Dev. (mm) | %95 Confidence Interval | p-Value | |
---|---|---|---|---|
Lower Limit | Upper Limit | |||
Root ZX mini DRY | 0.456 ± 0.437 a | 0.391 | 0.520 | <0.0001 |
Root ZX mini EDTA | 0.503 ± 0.449 a | 0.437 | 0.569 | |
Root ZX II DRY | 0.471 ± 0.453 a | 0.404 | 0.537 | |
Root ZX II EDTA | 0.484 ± 0.458 a | 0.416 | 0.551 | |
Sirona Integrated DRY | 0.704 ± 0.559 b | 0.621 | 0.786 | |
Sirona Integrated EDTA | 0.733 ± 0.501 b | 0.659 | 0.807 |
Source of Variance | Sum of Squares | df | Mean Square | F | Sig. * |
---|---|---|---|---|---|
Device | 38.447 | 2 | 19.224 | 39.351 | <0.0001 |
Medium | 0.007 | 1 | 0.007 | 0.015 | 0.90 |
Device * Medium | 0.362 | 2 | 0.181 | 0.371 | 0.69 |
Error | 515.877 | 1056 | 0.489 | ||
Total | 578.307 | 1062 | |||
Corrected Total | 554.694 | 1061 |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2024 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Barakat, R.M.; Almohareb, R.A.; Alharbi, A.O.; Alhazmi, A.; Alomar, R. Comparative Accuracy and Reliability of Three Electronic Apex Locators in Determining the Apical Constriction of Molar Canals: A Micro-CT Evaluation. J. Clin. Med. 2024, 13, 5199. https://doi.org/10.3390/jcm13175199
Barakat RM, Almohareb RA, Alharbi AO, Alhazmi A, Alomar R. Comparative Accuracy and Reliability of Three Electronic Apex Locators in Determining the Apical Constriction of Molar Canals: A Micro-CT Evaluation. Journal of Clinical Medicine. 2024; 13(17):5199. https://doi.org/10.3390/jcm13175199
Chicago/Turabian StyleBarakat, Reem M., Rahaf A. Almohareb, Arwa O. Alharbi, Asma Alhazmi, and Reem Alomar. 2024. "Comparative Accuracy and Reliability of Three Electronic Apex Locators in Determining the Apical Constriction of Molar Canals: A Micro-CT Evaluation" Journal of Clinical Medicine 13, no. 17: 5199. https://doi.org/10.3390/jcm13175199