BS-300 Service Manual (v5.0)
BS-300 Service Manual (v5.0)
BS-300 Service Manual (v5.0)
Service Manual
All rights
For this Service Manual, the issued Date is 2010-04 (Version: 5.0).
,
,
,
,
,
are the
registered trademarks or trademarks owned by Mindray in China and other countries.
All other trademarks that appear in this manual are used only for editorial purposes
without the intention of improperly using them. They are the property of their
respective owners.
the electrical installation of the relevant room complies with the applicable
national and local requirements;
Upon request, Mindray may provide, with compensation, necessary circuit diagrams,
calibration illustration list and other information to help qualified technician to
maintain and repair some parts, which Mindray may define as user serviceable.
WARNING:
It is important for the hospital or organization that employs this
equipment to carry out a reasonable service/maintenance plan.
Neglect of this may result in machine breakdown or injury of human
health.
NOTE:
This equipment is to be operated only by medical professionals trained
and authorized by Mindray or Mindray-authorized distributors.
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY
OR FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any
transportation or other charges or liability for direct, indirect or consequential
damages or delay resulting from the improper use or application of the product or the
use of parts or accessories not approved by Mindray or repairs by people other than
Mindray authorized personnel.
This warranty shall not extend to:
any Mindray product from which Mindray's original serial number tag or product
identification markings have been altered or removed;
Return Policy
Return Procedure
In the event that it becomes necessary to return this product or part of this product to
Mindray, the following procedure should be followed:
Obtain return authorization: Contact the Mindray Service Department and obtain
a Customer Service Authorization (Mindray) number. The Mindray number must
appear on the outside of the shipping container. Returned shipments will not be
accepted if the Mindray number is not clearly visible. Please provide the model
number, serial number, and a brief description of the reason for return.
Freight policy: The customer is responsible for freight charges when this
product is shipped to Mindray for service (this includes customs charges).
Return address: Please send the part(s) or equipment to the address offered by
Customer Service department.
ii
Company Contact
Manufacturer:
Address:
Tel:
Fax:
iii
Preface
Who Should Read This Manual
This manual is written for service professionals authorized by Mindray.
Then
WARNING:
BIOHAZARD:
CAUTION:
NOTE:
Graphics
All graphics, including screens and printout, are for illustration purpose only and
must not be used for any other purposes.
EC Representative
Name:
Address:
Tel:
+49 40 2513174
Fax:
+49 40 255726
iv
Contents
Preface........................................................................................................................................... i
Who Should Read This Manual .............................................................................................. i
Conventions Used in This Manual .......................................................................................... i
1
Specifications.................................................................................................................... 1-1
1.1
1.2
1.3
1.4
3.2
3.3
3.4
3.5
3.6
Contents
6
6.2
7
7.6
7.7
7.8
7.9
7.10
10.3
10.4
10.5
1 Specifications
1.1
System Feature
1.2
1-1
Specifications
Washing function: automatically washing interiors and exteriors of reagent probes;
carryover no more than 0.1%
Mixing bar: for single-reagent tests, it functions after sample dispensing; for
double-reagent tests, it functions after the dispensing of the sample and the
second reagent.
1.3
1.4
Others
1-2
2 System Installation
NOTE:
The analyzer should be installed or moved to another place by
Mindray-authorized personnel only.
2.1
2.2
Installation Procedure
1
Go to the installation field, and then check the delivery list for acceptance.
Install the four handles on the four angles of the analyzer. Move the analyzer
to the installation field, fix the casters, and then remove the handles.
Open the front plate, and check whether cable connections are loose. Open
the top plate, check whether the probe assemblies, reagent disk and sample
disk are intact and in good performance.
Connect the communication cable, power cable, grounding wire, waste tank
and deionized water tank. Install the used-cuvette bucket, reagent probe,
sample probe and mixing bar.
Use syringe to inject water into the filter, and connect filter to the cap
assembly of the DI water tank.
10
Put reaction cuvettes in the feeder. Remember to check whether the surfaces
of the cuvettes are smooth. In case of any bump, remove it before loading the
cuvette to the compartment. Do not touch the light transmission part of the
cuvette in which the colorimetric reading is taken.
11
Load acid and alkaline detergents to positions 46 and 47, and distilled water
to positions 49 on the reagent disk. Load distilled water to position 60 on the
sample disk.
12
13
2-1
System Installation
14
Select the [System/Status] menu, and then observe the system status and
record it in the table below:
Feeder status
Unconnected
Reaction
temperature
disk
Reagent
temperature
disk
Full
Half full
Empty
Ambient temperature
Waste tank status
Abnormal (full)
Detergent status
Abnormal (empty)
Normal (available)
Printer status
No printer
Normal
Unconnected
Unconnected
Unconnected
Unconnected
Loading/unloading unit
Unconnected
Temperature
unit
Unconnected
control
Idle
Running
Lamp
Network connection
Unconnected
Wavelength
Dark current
340
405
450
510
546
578
630
670
700
Reference light system
15
Select the [System/Maintenance] menu. Then select the Motion tab page,
and implement all sub-steps of each unit to see whether they are normal. In
case of any exception, adjust it.
16
Wash the interiors and exteriors of the sample probe, reagent probe and
mixing bar for several times to make the fluid circuit filled.
2-2
System Installation
17
Set parameters for test, reagent and calibration under the Parameters menu.
18
Request for calibration and samples, run and then debug the results.
19
ALT
CREA
BUN
Target value
2sd range
Test value 1
Test value 2
Test value 3
Test value 4
Test value 5
Test value 6
Test value 7
Test value 8
Test value 9
Test value 10
20
Training
Can the user complete daily tests?
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
2-3
3 System Descriptions
The BS-300 analyzer consists of the analyzing unit, operation unit and output unit.
The analyzing unit consists of the dispensing system, feeder, temperature control
system, photometric system and fluid system.
3.1
Dispensing System
The dispensing system consists of the probe assemblies (including the reagent
probe assembly, sample probe assembly and mixing bar assembly), reagent disk,
sample disk and reaction disk.
3-1
System Descriptions
3-2
System Descriptions
The step motors control the disk assemblies, and the synchronizing belts serve as
the gearing.
There is a build-in bar code scanner to the left of the sample disk. The scanner is
optional.
3-3
System Descriptions
3.2
Feeder
The feeder consists of the feeder assemblies and the manipulator. It is designated to
send cuvette segments to the reaction disk, take out the used ones and abandon
them to the used-cuvette bucket.
3-4
System Descriptions
There are five transducers that are shown in the figure below.
The no-cuvette transducer is used to detect whether there is a cuvette segment at
the loading position. The insufficient-cuvette transducer is used for determine
whether there are less than 10 reaction cuvettes in the compartment or not. If yes,
the analyzer will give a prompt.
Pressure Transducer
No-Cuvette
Transducer
Cuvette-Pushing
Limit Transducer
Insufficient-Cuvette
Transducer
Cuvette-taking limit
transducer
3.2.2 Manipulator
Two step motors (horizontal and vertical) supply power for horizontal and vertical
movements of the manipulator.
The upper finger and lower finger are same in their structures. They work together to
replace used cuvette segments with new ones.
The manipulator runs in a relatively complicated way. There are four transducers on
it: vertical transducer, horizontal transducer and two finger transducers.
3-5
System Descriptions
3.3
3-6
System Descriptions
Upper heater: square in shape, 220/110VAC, 125W
Lower heater: ring in shape, 220/110VAC, 350W
Total power: 475W.
The function of heaters is to compensate the heat for incubating the reagent and for
maintaining the temperature of the temperature-controlled chamber.
Fans are used in series in the temperature-controlled chamber. It makes the air
circulating in the chamber, and enhances the convective heat exchange. There are
four fans in the chamber. All have the alarm function.
The temperature transducer feeds back the air temperature at the position several
millimeters from the bottom of the reaction cuvette.
The overheat protection switch is to switch off the power when the temperature
controller does work and the temperature-controlled chamber reaches 55, so as to
avoid overheat or fire. When the temperature-controlled chamber becomes 35,
this switch will automatically be reset.
3-7
System Descriptions
3.4
Photometric System
The photometric system consists of a measurement photometric system and a
reference photometric system. The measurement photometric system provides 9
monochromatic lights to measure the absorbance of the reacting liquid in the rotating
reaction cuvettes. The reference photometric system compensates the
measurement photometric system to make the measurement more accurate.
3-8
System Descriptions
Tungsten-halogen
lamp
Biconvex
lens
Main
fiber
Plano
convex
lens Cuvette
Plano
Filterconvex
lens Photodiode
Fibers
.
.
.
Measurement
lights
Photodiode
light
Reference
3.5
Fluid System
The fluid system is shown in the following figure.
3-9
System Descriptions
As shown in the figure above, the fluid system consists of interior washing and
exterior washing.
The syringe assembly controls the aspiration volume by controlling the travel of the
sample/reagent syringe. It is the core part of the fluid system.
3.6
ISE
Calibrant B
Pump W
Pump A
Calibrant A
Pump B
Waste
3-10
System Descriptions
3-11
4 Functions of Boards
The analyzer is integrated with the following boards:
Reagent
refrigeration board
4.1
ISE module
Step motors
DC motors
Pumps, valves
Heaters of reaction
disk
Reagent pre-heater
Lamp
Peltiers
transducers
4-1
Functions of Boards
receiving signals of fluid level detection and bump collision through the interface
with the level detection boards.
detecting signals from temperature transducers and controlling temperature of the
reaction disk and reagent preheating.
receiving signals from position transducers, deionized water transducer and waste
transducer and controlling the transducers.
controlling the built-in sample bar code scanner, reading the data and sending it to
the PC.
controlling the ISE module, reading the results and sending them to the PC.
4.2
to 13 step motors
to 2 DC motors
to 2 solenoid valves
to 3 pumps
to 2 heaters of
reaction disk
4.3
to lamp power
to 2 magnets
to reagent
preheating
4.4
4-2
Functions of Boards
Heat sink system for the whole device: 3 lamp-cooling fans, 4 temperature control
fans of the reaction disk.
4.5
4.6
4.7
Connector
Connected with
J91
Front transducer
J92
Back transducer
J93
Intermediate transducer
J94
No-cuvette transducer
J95
Pressure transducer
J96
J97
J98
Cuvette-loading button
J99
Connected with
J101
J102
J103
4-3
Functions of Boards
4.8
Connector
Connected with
J108
J104
J105
J106
J107
J109
J110
4.9
Connector
Connected with
J200
J201
J202
J203
J204
J205
J206
J207
AC/DC conversion;
Supplying the +390V and VDD voltage to the 24V board and the 12V&5V board;
Supplying stable 12V voltage for the lamp;
Supplying control signals of the analyzing unit switch to control the C12V, 5V and
24V outputs.
The 24V board converts the 390VDC current transmitted from the PFC board to the
separated 24VDC outputs through the forward converter.
4-4
Functions of Boards
The 12V&5V board converts the 390VDC voltage from the PFC board to B12V,
C12V and 5V voltages through the forward converter.
The ISE power supply board converts the 12V of the reagent refrigeration board to
24V that provides power supply for the ISE unit (optional).
4-5
BIOHAZARD:
Wear gloves and lab coat and, if necessary, goggles.
Dispose of the waste in accordance with your local or national
guidelines for biohazard waste disposal.
CAUTION:
Please use Mindray-recommended consumables. Other consumables
may decrease the system performance.
Refer to the BS-300 Chemistry Analyzer Operation Manual for details about
5.1
5-1
Lens
Screw
Lens Seat
Gasket
Filter
Shield Box
Gasket
Photoelectric
Conversion Board
Flat Spring
Supporting Sleeve
WARNING:
Before operating, ensure to place the POWER (main switch) to OFF.
Unscrew (counter clockwise) the two cap screws on the screen gland
whose wavelength is to be replaced.
Open the cover of the A/D conversion board, and unplug the plug
corresponding to certain wavelength.
5-2
Take out the photoelectric conversion board and the supporting sleeve.
Keep the photoelectric conversion board upward, and loosen the two
retaining screws on it.
Keep the photoelectric conversion board upward, and pull the photoelectric
amplification board out of the supporting sleeve.
Unpack the new supporting sleeve containing the optical assembly. Be sure
to keep the top side (where to assemble the photoelectric amplification
board) of the supporting sleeve upward.
Install the original photoelectric conversion board onto the new supporting
sleeve, and then fasten the two retaining screws.
5-3
WARNING:
When replacing the light filter assembly, do not touch the optical
assembly in the supporting sleeve and the photoelectric receiving tube
of the photoelectric conversion board by hand.
The light filters and the photoelectric conversion boards are in
one-to-one relationship. Do not disarrange them.
5.2
WARNING:
Before operating, ensure to place the POWER (main switch) to OFF.
Unscrew the four screws on the supporting pillars of the cuvette feeder, and
remove the cuvette feeder.
5-4
Unscrew the four screws on the reaction disk cover, and open the reaction
disk cover.
Attention should be paid to the power cable of the upper heater when the
reaction disk cover is being opened.
Take out two cuvette segments in a diagonal with the needle-nose pliers to
make two spaces for disassemble the colorimetric disk.
5-5
Unscrew three M3 cap screws, and then disassemble the colorimetric disk.
Take out the optical fibers one by one, and fix the optical components in the
colorimetric clamp by fastening the screws slightly.
Screw of the reference light
Draw out all the nine optical fibers from the reaction compartment.
5-6
Loosen the M3 cap screw (used for retaining optical fibers) on the lamp
housing, and then draw out the optical fibers.
Put nine of the ten branches of the new optical fiber into the reaction disk
from its bottom one by one, loosen the retaining screw, insert the optical
fiber to the end, and then fasten the retaining screw.
10
11
12
13
14
CAUTION:
When replacing the optical fiber, ensure that its bending radius is no
less than 20cm. Otherwise, the optical fiber will be damaged.
5.3
5-7
Disassemble the cuvette feeder, and adjust the circular position of the
reaction disk (through the initial position transducer of the reaction disk) to
the standard position (The finger of the lower arm points to the center of the
slot between reaction cuvette segments).
Vertically adjust the manipulator the standard position (The finger of the
lower arm can work on reaction cuvette well, and the finger support of the
lower arm should be 0.15mm above the reaction cuvette.).
5-8
5.4
Run the manipulator to the position for taking new cuvettes, and
horizontally adjust the reaction cuvettes in the feeder. When catching a
cuvette, ensure a clearance of 0.4mm ~ 0.6mm between the finger of the
upper arm and the cuvette, and center them.
Adjust the working position of the reagent probe (To minimize the
cumulative error, ensure that the reagent probe return to the initial position
before each adjustment.) as follows:
A
Adjust the washing position of the reagent probe. Ensure the reagent
probe is in the center of the wash well and the probe pin is 5mm
away from the bottom of the wash well. If necessary, adjust the
position of the wash well.
Adjust the reagent probes position on the outer circle of the reagent
disk. Ensure the reagent probe is in the center of the hole of the outer
circle.
Adjust the reagent probes position on the inner circle of the reagent
disk. Ensure the reagent probe is in the center of the hole of the inner
circle.
Adjust the working position of the sample probe (To minimize the
cumulative error, ensure that the sample probe return to the initial position
before each adjustment.) as follows:
A
5-9
5.5
1
Adjust the washing position of the sample probe. Ensure the sample
probe is in the center of the wash well and the probe pin is 5mm
away from the bottom of the wash well. If necessary, adjust the
position of the wash well.
Adjust the sample probes position on the outer circle of the sample
disk. Ensure the sample probe is in the center of the hole of the outer
circle.
Adjust the sample probes position on the inner circle of the sample
disk. Ensure the sample probe is in the center of the hole of the inner
circle.
Switch off the analyzing unit first, and then rotate the mixing arm to
see if its rotation radius is proper. If the mixing bar cannot reach the
center of the reaction cuvette, adjust the position of the mixing bar on
the mixing arm properly.
Disassemble the mixing arm, and then switch on the analyzing unit to
run the reaction disk and stop it at any cuvette. Hold the mixing arm
by hand, and fix it at the center of the reaction cuvette.
Adjust the washing position of the mixing bar. Ensure the mixing bar
is in the center of the wash well and about 5mm away from the
bottom of the wash well. If necessary, adjust the position of the wash
well.
5-10
Inject water into the new filter assembly through the big adapter by using
the syringe. When water wells up from the small adapter, the injection is
completed. Purpose: increase the weight of the filter assembly to make
it sink to the bottom of the water tank.
Connect the new filter assembly to the tubes of the cap assembly.
5-11
Put the assembly into the water tank and make sure that the filter
sink smoothly to the tank bottom and does not twist with the floater
connecting rod. Tighten the cap of the DI water until secure.
2)
Place the Main Power and Power of the analyzer to ON. Turn on
the computer and start the operating software.
3)
Reset 10 times (click Motion and then select Others, click Reset
Mechnical Parts) to expel the air in the filter assemby and the
tubes.
4)
Check for large amount of buubles in the outlet tube. If yes, continue the
resetting process; if not, the air expelling is completed.
5-12
5.6
ISE module
Reagent module to be
placed here
Pump B
Pump A
Pump W
5-13
If you want to replace the W1, unscrew the screw of the ISE module and
take off the cover.
Put out the fluidic tubing W2 and insert it into a container such as a cup that
is used to contain the waste solution flowing from the W2.
If you want to replace W1, pull out its two adapters directly from the waste
pump tubing adapter and the right angle adapter that is fixed to the
compression plate.
Right angle adapter
Note that when pulling out the fluidic tubing adapter, in order not to release
the right angle adapter, you can hold the right angle adapter with a finger.
After that, install a new W1. For W2 exchange, it just needs to replace the
W2 with a new one.
10
If you want to replace the W2, replace it with a new one. Otherwise, insert
the W2 back to the reagent module.
11
12
13
Select the Others tab and click the Download Settings button.
14
15
Click the A purge button to observe if there is solution leaking out. If there
is solution leaking out, repeat from the step 4 to install the tubing again.
16
Install the cover of the ISE module if it has been taken off.
17
5-14
Take off the panel under the sample disk and you can see the shielding box
and the peristaltic pumps.
Put the fluidic tubing W2 in a container such as a cup that is used to contain
the fluid flowing out from the W2.
If you want to replace A1 or A2, click the A purge button. If you want to
replace B1 or B2, click the B purge button.
Repeat this step for several times until the received data indicates that the
Calibrant A or Calibrant B has air bubble in it.
10
11
12
If the system software is not running, start it. Otherwise, enter the System
Maintenance screen of the system software, select the Others tab and
click the Download Settings button.
13
14
If you have replaced A1 or A2, click the A purge button. If you have
replaced B1 or B2, click the B purge button.
Repeat this step for several times until the received data indicates that the
Calibrant A or Calibrant B has no air bubble in it.
Pinch one adapter of the pump tubing and take it out from the tubing shelf.
Then take off the whole tubing.
Tubing shelf
5-15
Put one adapter of the new pump tubing onto the tubing shelf and make the
tubing around the pump ahead, then put the other adapter onto the shelf
with a little strength.
Click the Pumps button to see if the calibration of pump is correct. If not,
repeat the upper steps.
Cable
Pump motor
Put a new pump on the pump shelf, tighten the screws, connect the cable
and pump motor, and then put the pump tubing around the pump head.
Click the Pumps button to see if the calibration of pump is correct. If not,
repeat the upper steps.
Take off the panel under the sample disk and you can see the shielding box
and the pumps easily.
5-16
Unscrew the screw of the ISE module and take off the cover.
10
11
12
Unscrew the four screws that fix the ISE module to the shielding box. Then
take out the module from the shielding box.
13
Put a new ISE module to the shielding box. Then connect the cable and
install the electrodes sequentially. At last mount the right angle adapter to
the compression plate.
14
15
16
Select the Others tab and click the Download Settings button.
17
18
Click the A purge button to see if there is solution leaking out. If there is,
take off the electrodes and install them again.
19
5-17
6 Software Introduction
The software of BS-300 analyzer is composed of the system software and the control
software.
6.1
System Software
The system software can, according to the requirements and inputs of users,
generate a work schedule (instruction sequence), control the units of analyzing unit in
the sequence of instructions in the work schedule, receive photoelectric data,
response messages or execution results from the analyzing unit, output them to the
PC screen or the printer. With these outputs, users can obtain correct test results.
According to different functions, the entire PC software system can be divided into the
following parts:
System Initialization
This part includes the initialization of the PC operating system, the initialization of the
communication between the PC and analyzing unit, and the controls of analyzing unit
reset.
Control system
This part includes the formation of the work schedule, instruction data sending and
receiving.
GUI
This part includes the requests for tests (routine tests, emergency tests, calibration
tests and QC tests), the observation of test status (status of the reaction disk, sample
disk and reagent disk), test management, calibration management, QC management,
result query, alarm management and the help system.
Shutdown Processing
This part includes the resets of sub-units.
6-1
Software Introduction
other operation. In this case, users cannot switch over to any other program or
print the current screen.
Check whether there is any username and password of the
maintenance/debugging personnel in the registry. The system software has the
debugging function. Only the authorized maintenance/debugging personnel can
enter the debugging window and maintain or debug the system and the host. For
the confidentiality, the username and password are saved in the registry. If they
are unavailable in the registry, the system will re-write the default username and
password into the registry.
6.2
Control Software
The analyzing unit can be functionally divided into the following units: photoelectric
unit, reaction disk unit, reagent disk unit, sample disk unit, reagent probe syringe unit,
sample probe syringe unit, mixing bar unit, loading and manipulator unit, temperature
control unit, fluid tubing unit and reagent refrigeration unit. The functions of those
units are listed in the following table.
6-2
Software Introduction
Unit
Function
Main unit
Photoelectric
unit
Reaction
unit
disk
Reagent
unit
disk
Sample
unit
disk
Reagent probe
syringe unit
Sample probe
syringe unit
Loading
and
manipulator
unit
6-3
Software Introduction
Unit
Temperature
control unit
Function
Controls the reaction disk temperature, reagent probe
preheating temperature and fluid tubing.
Note: The reaction disk temperature should be kept at 37,
the heating cavity of the reagent probe should be preheated
to 45 .
Reagent
refrigeration
unit
6-4
and
controls
its
7 Service Flow
WARNING:
Before disassembling or assembling the analyzing unit, ensure the
POWER is placed to OFF.
The probe tip is sharp and can cause puncture wounds. To prevent
injury, exercise caution when working around the probe.
When you disassemble or replace any board, ensure to wear antistatic
gloves.
BIOHAZARD:
Wear gloves and lab coat and, if necessary, goggles.
7.1
The
reagent
probe
cannot
detect
the
surface
on the
reagent
disk
Surface
detection
failure of
reagent
probe(Th
e reagent
probe
does not
detect
the
surface
on
the
reaction
disk.)
The surface
detection
signal fails
to be
generated
by
capacitance
change.
The
reagent
probe misdetected
the surface
7-1
Service Flow
7.2
The sample
probe cannot
detect the
surface at the
reaction disk.
No reagent in the reaction disk: The reagent probe misdetects the surface, aspirates air and dispenses into the
reaction cuvette. Therefore, the reagent unit CPU fails to
make any judgment. In this case, the sample probe cannot
detect the surface when dispensing the sample into this
reaction cuvette.
Possible cause: Surface detection failure of the reagent probe
insufficient reagent: The reagent probe aspirates less reagent
than the set volume. As a result, the sample probe cannot
detect the surface at the preset height.
Possible cause: The reagent is prepared in the bottle, the
reagent syringe is not fixed, leakage occurs in the tubing of
the reagent probe, or there are bubbles in the reagent.
Position correction parameter error of the sample probe: In
case of such an error, the sample probe cannot reach the
surface in the steps.
Troubleshooting: Re-correct the sample probe position.
Surface detection signal transmission failure: When the probe
touches the surface and stops, the indicator of the surface
detection board is on for about 2 seconds. The output signal
of the surface detection board switches normally, but there is
always a host alarm indicating that no surface is detected.
Possible cause: The patch cord for detecting the sample
surface is disconnected, or the connector is not well
connected.
Surface
detection
failure of
the
sample
probe
The sample
probe cannot
detect the
surface at the
reaction disk.
The sample
probe
misdetects
the
surface
(In
this
case,
there will be
no alarm, but
test
results
will become
abnormal. On
the reaction
curve,
you
can see that
no sample is
added.)
7-2
Service Flow
7.3
Troubleshootin
g for the liquid
dropping from
the reagent
probe and
sample probe.
Drips drop from the sample probe
due to surface detection failures. In
this case, most liquid drop on the
area between the reaction disk and
the wash well.
7-3
Service Flow
7.4
Failing to
Detect the
Surface of
the Water
for
Washing
Exteriors
7-4
Service Flow
7.5
Abnormal Results
All test
results are
abnormal
Some results
are higher,
and some are
lower, with
bad
repeatability.
Bad repeatability
Observe the reaction curve, and find the possible
causes.
Some test results are
abnormal.
All test results are approaching to 0.
Observe the reaction curve first.
Possible cause: The lamp base of corresponding
wavelength exceeds 65535; the photoelectric
detection unit fails to detect the change of
absorbance, reagent failure and optical fiber
disconnection.
7-5
Service Flow
7-6
Service Flow
Replace lamb
YES
NO
YES
Replace the
photoelectric
transducer assembly.
NO
YES
Clean the
photometric system
(such as the lens)
7.6
YES
Note:
Do not adjust the photoelectric gain parameter in case you have no idea about the
cause for the alarm indicating insufficient light intensity of the lamp.
7-7
Service Flow
NO
NO
yes
Replace the
temperature
transducer.
YES
7.7
7-8
Service Flow
After that, print or snap the temperature curve and then return it to
the headquarters together with the service work report.
Open the left side plate of the BS-300 analyzing unit to expose the
main control board.
7.8
If the bar code scanner cannot emit laser, you can service it according to the following
steps.
7-9
Service Flow
1
Place sample tubes with bar code labels that meet the requirements
mentioned in the operation manual.
Enter the Scanner Setup & Maintenance screen of the system software.
Enter the sample position range in the Sample Position field and click the
Step Scan button. Observe if the scanner can emit laser. If not, click the
Laser on button. If the scanner still cannot emit laser, go to the next step.
If the bar code scanner emits laser normally but is failed in reading, you can service it
according to the following steps.
1
Place sample tubes with bar code labels that meet the requirements
mentioned in the operation manual.
See if the window on the sample department is stained. If so, clean the
window with absolute alcohol-dipped gauze and go to the next step. If not,
go to the step 8.
Enter the Scanner Setup & Maintenance screen of the system software.
Enter the sample position range in the Sample Position field and click the
Step Scan button. Observe the scanning procedures to see if the scanner
can read the bar codes. If not, go to the next step.
10
11
Click the Scan button to see if the scanner can read the bar codes. If not,
go to the next step.
12
13
7-10
Service Flow
Bracket 1
Bracket 2
14
7.9
If the scanner cannot read the bar codes successfully, replace the bar code
scanner. If the new scanner cannot read the bar code, you need to check
the bar code labels the user used.
Feeder Failure
No-Cuvette
Transducer
Cuvette-Pushing
Limit Transducer
Insufficient-Cuvette
Transducer
7-11
Cuvette-taking limit
transducer
Service Flow
Feeder failure
symptom
7-12
Service Flow
Troubleshooting:
1. Lubricate the
lead screw after
cleaning it.
2. Replace
relevant
assemblies.
1. Adjust the
positions of the
upper hand and
the
feeder/
reaction disk.
2. Replace the
manipulator (Its
position must be
adjusted).
1. Press the
cuvette-taking
key after adding
cuvette
segments
into
the feeder.
2. Adjust the
positions of the
lower hand and
reaction disk.
3. Replace the
manipulator (Its
position must be
adjusted)
Troubleshooting:
1. Replace the
manipulator (Its
position must be
adjusted).
Manipulator
Failure
7-13
Service Flow
Digit 3: mV out of range for Calibrant A in Sample Cycle, or Calibrant B in Urine
Cycle.
Digit 4: Noise in mV of Calibrant B in Calibration Cycle, or Sample mV during a
Sample Cycle.
Digit 5: mV Noise in Calibrant A of the Calibration Cycle, or Sample Cycle, mV
Noise in Calibrant B of the Urine Cycle.
Digit 6: mV Drift of calibrant A.
Digit 7: Slope of Sample value is out of range.
Error
Digit 1
Digit 2
Digit 3
Digit 4
Digit 5
Digit 6
Digit 7
Air/
Hardwa
re
mV Out
Calib/S
ample
mV Out
Cal A in
Calib/S
ample
Mode,
Cal B in
Urine
Mode
mV
Noise
Cal
B/Samp
le
mV
Noise
Cal A in
Calib/S
ample
Mode,
Cal B in
Urine
Mode
Cal
A
Drift
in
Sample
Mode,
Cal B in
Urine
Mode
Out of
Slope/
Machin
e
Ranges
Air in Calibrant A
Air in Calibrant B
Air in Cleaner
Air in Segment
Pump Calibration
No Flow
Bubble Detector
Reagent
Read
Chip
Reagent
Write
Chip
Store Calibration
Value Error
Command Error
No Error
Na+
Air
Sample/Urine
in
7-14
Service Flow
Error
Digit 1
Digit 2
Digit 3
Digit 4
Digit 5
Digit 6
Digit 7
Air/
Hardwa
re
mV Out
Calib/S
ample
mV Out
Cal A in
Calib/S
ample
Mode,
Cal B in
Urine
Mode
mV
Noise
Cal
B/Samp
le
mV
Noise
Cal A in
Calib/S
ample
Mode,
Cal B in
Urine
Mode
Cal
A
Drift
in
Sample
Mode,
Cal B in
Urine
Mode
Out of
Slope/
Machin
e
Ranges
K+
Na+, K+
Li+
Li+, Na+
Li+, K+
Cl-
Cl-, Na+
Cl-, K+
Cl-, Li+
Cl-, Li+, K+
The following table displays the causes and the correction measures of some errors.
Note that if the correction measure is the replacement of board, bubble detector or
other components in the ISE unit, you have to replace the whole ISE unit. You can
refer to the chapter 5 for detailed information about some replacement procedures.
Symptom
System
respond
Problem
does
not
Correction
1. No power
2. Communication failure
Replace board.
Replace board.
7-15
Service Flow
Symptom
Problem
Correction
Low Slope
1. Misalignment of electrodes
2. Deterioration of calibrator
solutions
3. Deterioration of
electrode (low slope)
Replace
and test.
sensing
Single electrode
electrodes
5. Deterioration of reference
electrode
Replace reference
and retest.
1. Deterioration
electrode
Replace
and test.
problem
of
sensing
problem
electrode
electrodes
1. Deterioration of reference
electrode
Replace reference
and retest.
electrode
2. Deterioration
electrode
of
sensing
Purge
Calibrant
A
recalibrate the Module.
2. Deterioration of reference
electrode.
Replace reference
and retest.
7-16
and
electrode
Service Flow
Symptom
Problem
Correction
3.
Electrical
spike
environmental source.
from
Air in Sample
Air in Sample
Calibrant A
and
3.
Bubble
malfunctioning.
detector
4.Waste
pump
malfunctioning.
Air in Calibrant B and
Air in Calibrant A
is
is
7-17
Service Flow
Symptom
Problem
Correction
2. Tubing is disconnected.
7-18
Replace tubing
8 Mechanical Structure
Synchronous belts
Coder sensors
8-1
Mechanical Structure
8-2
Mechanical Structure
Lamp assembly
8-3
Mechanical Structure
Reaction disk TEMP.
control assembly
Optical assembly
8-4
Mechanical Structure
Washing assemblies
Reagent syringe
Sample syringe
8-5
Valves
Pumps
Mechanical Structure
A/D conversion board
8-6
Mechanical Structure
Loading system
Manipulator
assembly
8-7
Mechanical Structure
8-8
Service Tools
S/N
TOOL
Quantity
Specification
Unit
Allen Key
1 set
mm
Phillips screwdriver
1 set
mm
Flat-bladed screwdriver
1 set
mm
1 set
2.5, 3.0
mm
Tweezers
1 pair
small
Multimeter
common
Movable Wrench
common
Electric iron
30
VA
Stainless steel
0.5
(diameter),
(length)
10
Knife
common
11
Pincers
common
9.2
15
mm
Parts
S/N
P/N
Name
BA30-20-06448
BA30-20-06449
BA30-20-06450
BA34-20-63601
BA30-20-06452
BA30-20-06453
BA30-20-06454
BA30-20-06455
BA30-20-06458
10
BA30-20-06459
11
BA34-20-63602
12
BA30-20-06464
13
BA30-20-06468
9-1
P/N
Name
14
BA30-20-06476
15
BA30-20-06477
16
BA30-20-06480
17
BA30-20-06481
18
BA30-20-06484
19
BA30-20-06485
20
BA34-20-63651
21
BA30-20-06523
22
BA30-20-15291
23
BA30-20-15292
24
BA34-20-63603
25
BA34-20-63631
26
BA30-20-15295
27
BA30-20-15296
28
A21-000010---
Tubing
29
BA30-10-15051
Single-way valve
30
0040-10-32303
Tubing
31
0040-10-32304
Connector
32
0040-10-32305
Connector
33
M90-100014---
34
M90-100015---
35
M90-100030---
M90-100050---
M90-100051---
38
M90-100071---
Tubing
39
BA30-30-15169
5-way connector
40
BA30-10-06666
Step motor
41
BA30-10-15041
42
BA30-10-15258
43
BA34-30-63559
44
BA30-30-06312
45
BA34-30-63613
36
37
9-2
P/N
Name
46
BA30-10-06365
47
BA30-30-06691
48
BA30-30-15228
49
BA30-10-06367
Filter 340
50
BA30-30-06673
51
BA30-30-15229
52
BA30-10-06368
Filter 405
53
BA30-30-06675
54
BA30-30-15230
55
BA30-10-06369
Filter 450
56
BA30-30-06677
57
BA30-30-15231
58
BA30-10-06370
Filter 510
59
BA30-30-06679
60
BA30-30-15232
61
BA30-10-06371
Filter 546
62
BA30-30-06681
63
BA30-30-15233
64
BA30-10-06372
Filter 578
65
BA30-30-06683
66
BA30-30-15234
67
BA30-10-06373
Filter 630
68
BA30-30-15235
69
BA30-30-15235
70
BA30-10-06374
Filter 670
71
BA30-30-06687
72
BA30-30-15236
73
BA30-10-06375
Filter 700
74
BA30-30-06689
75
BA30-30-06565
76
BA34-30-63637
Syringe assembly
77
3100-10-49438
100ul Syringe
78
BA30-10-06651
500ul Syringe
79
BA30-10-15115
Step motor
9-3
P/N
Name
80
BA30-20-06594
81
BA30-21-15161
82
BA30-21-15162
83
BA30-30-06634
84
BA30-21-15159
85
BA30-21-15160
86
BA30-30-06639
87
BA30-30-06641
88
0040-10-32302
Connector
89
0040-10-32303
Washer
90
BA30-21-15311
91
BA30-30-06702
92
BA30-30-06715
93
BA30-30-06752
94
BA30-30-06304
95
BA30-30-06753
96
BA30-30-06761
Heat-block assembly
97
BA30-30-06768
98
BA30-30-14978
99
BA30-30-06770
100
BA34-20-63653
101
BA30-21-06488
102
BA30-21-06490
103
BA30-21-06491
104
BA30-21-06492
105
BA34-20-63648
106
BA30-21-06489
107
BA30-30-15034
108
BA30-30-15095
109
BA30-30-06780
110
BA30-30-15113
111
BA30-21-15155
112
BA30-21-15156
113
BA30-30-15166
9-4
P/N
Name
114
BA30-30-15167
115
BA30-30-15284
116
BA34-30-63557
117
BA33-30-35098
118
BA33-30-35078
PFC Board
119
BA33-30-35082
24V Board
120
BA33-30-35080
12V&5V Board
121
TSB1-20-20399
Fan
122
BA34-20-63655
Manipulator assembly
123
BA33-30-35223
124
BA34-20-63622
125
BA34-20-63614
126
BA30-30-15029
127
BA34-20-63638
9-5
10
10.1 General
The maintenance and test software of the BS-300 Chemistry Analyzer is the
software that helps maintenance personnel debug and maintain the BS-300
Chemistry Analyzer. Maintenance personnel can, through this software, check
whether the probes, mixing bar, disks, fluidic system and photometric system of
the BS-300 Chemistry Analyzer work normally, modify parameters, and execute
individual commands or combined commands.
Run the NewDebug.exe file to enter the Maintenance and Test window. See
Figure 10-1.
Display Area
As shown in Figure 10-1, there are 4 tabs in the Maintenance and Test window:
Command
PARA and Speed
Temperature
Photoelectric
10-1
Note:
The buttons displayed in gray in software interface are not
functional.
Display area
At the bottom of the window is the display area for displaying the frame
information and fault/error information. The upper box is for frame information, and
the lower one is for fault/error information.
There are five types of frame information, including Result, Echo, Data,
Command and Warning.
Select the check box in front of frame information to display relevant frame
information. Click the check box again to de-select, and the frame information will
not be displayed.
Click the Clear button to clear the information displayed in relevant box. The two
Clear buttons are respectively for two boxes.
Setup button
Click the Setup button at up right corner to show a dialog as shown in Figure 10-2,
which is used to change communication port and set baud rate and check sum
type of the port. Check sum type includes odd, even and none.
Baud Rate
About button
Clicking the About button at up right corner of the interface, you can view the
version information of this software.
In the following sections, the 4 tabs mentioned above will be detailed respectively.
10-2
Note:
The units of parameters involved in this software are respectively:
Time: 50ms;
Sample (reagent) aspirating/dispensing volume: steps of motors;
Temperature: .
Temperature control coefficient: 0.01.
10.2 Command
In the Command tab, you can send commands to the analyzing unit and check
whether probes, mixing bar, disks, fluidic system, photometric system, ISE unit
and built-in bar code scanner work normally.
Macroinstruction Area
As shown in Figure 10-3, the Command tab mainly consists of two parts: Single
Command Area and Macroinstruction Area.
Main Unit
Reagent Unit
Sample Unit
10-3
Load Unit
Temp Unit
Photoelectric Unit
ISE Module
Sample Bar Code Module
Note:
After starting this software, you must click the Download
parameters to Analyzing Unit button in the Main Unit sub-tab.
Otherwise, the software will not operate the analyzing unit
successfully.
After clicking the Enable modifying parameters button in the Main
Unit sub-tab, you can modify the parameters in the PARA and
Speed tab.
Main Unit
Command
What it does
Download parameters
Analyzing Unit
to
Configures running
analyzing unit.
parameters
for
the
Enable
parameters
modifying
Disable
parameters
modifying
Reagent Unit
Command
What it does
10-4
What it does
Reagent
position
disk
rotate
to
10-5
What it does
Sample
position
disk
rotate
to
10-6
What it does
Load Unit
Command
What it does
Unload cup
fingers
Moves the
position.
To H-position
cuvettes
for
placing
To V-position
cuvettes
for
placing
10-7
manipulator
to
vertical
initial
What it does
Temp Unit
Command
What it does
Temperature control on
Turns
on
controller(s).
the
specified
temperature
as
Turns
off
controller(s).
the
specified
temperature
as
Photoelectric Unit
Command
What it does
Turn on light
10-8
What it does
ISE Module
Command
What it does
Handshake
Maintain
Detect Bubble
Cleaning Period
Calibrate
Correct Pumps
Measure Serum
A Purge
B Purge
<DSPA>
<SWBC>
<SWPC>
<CKSM>
<DVON>
<MVON>
<DVFF>
What it does
Set
Select a symbology from the pull down list and click this
button. If Valid is selected, the symbology will be set as the
one that the scanner can recognize. Otherwise, it will be
the one that can not be recognized.
10-9
What it does
Send
Enter an instruction into the edit box and click this button to
send it to the analyzing unit.
Laser On
Laser Off
Handshake
The bar code scanner shakes hands with the main unit.
Scan
After entering the start position No. and the end one into
Start and End edit boxes respectively, click this button to
scan bar codes on the selected positions.
Macroinstruction
Commands
Button
Parameters of the
current command
10-10
Note:
After all commands in each macroinstruction are executed, the
system counts it as one time.
When the check box on the right of the macroinstruction pulldown
list is selected, the selected macroinstruction will be executed for
once; otherwise, the macroinstruction will be executed cyclically.
For each macroinstruction, the commands of it are executed one by one in the
sequence in which they are listed. When the set delay expires after a command is
executed, the next command is executed immediately. In case of an error in the
execution process, the macroinstruction will be stopped.
Following are instructions to the action implemented by each macroinstruction.
Macroinstruction
Command
What it does
A,B purge
B purge
A purge
10-11
probe
probe
Command
What it does
Test Loader
Mix instruction
Reagent
probe
reaction disk
to
Reagent
probe
vertical init
to
10-12
probe
probe
Urine
test
diluted)
(auto
Command
What it does
Sample
probe
reaction disk
to
Sample
probe
vertical init
to
Reagent
disk
in
position 50 in next
period
Dispensing 270l
urine diluent
of
Dispensing 30l
urine sample
of
First period
Second period
Third period
Idle period
Measuring.
Idle period
Measuring.
Idle period
Measuring.
First period
Second period
Third period
Idle period
Measuring.
Idle period
Measuring.
Idle period
Measuring.
10-13
Buttons
As shown in Figure 10-5, the PARA and Speed tab mainly consists of two parts as
unit parameter and several operating buttons.
Unit Name includes Main unit, Sample unit, Reagent unit, Photoelectric unit,
Temp unit and Load unit.
At this screen, you can not only inquire and configure each unit parameters, but
also save and read relevant data. Now lets introduce them respectively.
10-14
Click the Config button to configure the parameter for relevant unit of the
analyzing unit.
Click the Config All button to configure all parameters of the selected unit
for relevant units of the analyzing unit.
Save data
1
Input a filename and click the Save button, you can save parameters of the
currently selected unit to a .txt file.
Read Data
1
Select the relevant .txt file and click Open button, you can configure
parameters in the file for relevant unit of the analyzing unit.
10-15
Unit
Default
H-pos.
for
washing
reagent probe
Whole
step
42
Whole
step
-42
Whole
step
215
Whole
step
165
Position correction of
reagent disk inner circle
Micro
step
10
Position correction of
reagent disk outer circle
Micro
step
10
Parameter
Unit
Default
Bar
code
position
correction of sample
disk inner circle
Micro
step
18
Whole
step
102
Whole
step
202
Whole
step
165
Sample unit
10-16
Unit
Default
Whole
step
-106
Whole
step
490
Position correction of
sample disk inner circle
Micro
step
10
Position correction of
sample disk outer circle
Micro
step
10
Whole
step
163
Parameter
Unit
Default
Manipulator h-position
for placing cuvettes
Whole
step
70
Manipulator h-position
for taking cuvettes
Whole
step
259
Manipulator v-position
for placing cuvettes
Whole
step
70
Manipulator v-position
for taking cuvettes
Whole
step
768
Position correction
manipulator in vertical
when it takes a
segment from the
compartment.
Whole
step
68
Load unit
10-17
of the
direction
cuvette
cuvette
Unit
Default
Whole
step
70
Unit
Default
Photoelectric unit
Parameter
Photoelectric
parameter 405
gain
127
Photoelectric
parameter 450
gain
127
Photoelectric
parameter 510
gain
127
Photoelectric
parameter 546
gain
127
Photoelectric
parameter 578
gain
127
Photoelectric
parameter 630
gain
127
Photoelectric
parameter 670
gain
127
Photoelectric
parameter 700
gain
127
Photoelectric
gain
parameter
reference
light
127
Photoelectric
parameter 340
127
Micro
step
25
gain
Start
position
of
photoelectric collection
10-18
10.4 Temperature
Figure 10-1 Temperature tab
Reagent
Temperature Curve
Control Panel
Reaction Tray
Temperature Curve
Control Panel
The two temperature curves are controlled respectively through two control panels
on the right.
Meanings of the parameters on the control panels are as follows.
Parameter
What it does
Value
YRange
XRange
Temp
Period
After setting Xrange, Yrange and Temp on the control panel, click Confirm button
to implement them on relevant temperature curve.
There are 4 more buttons in each of the two control panels:
Start: After setting Period, clicking it, you can start collecting the temperature
data;
10-19
10.5 Photoelectric
In the Photoelectric tab, the values of photoelectric data and photoelectric data
curve of the selected wavelength are displayed.
Display difference
between the above two
Through the panel, you can set relevant parameters of photoelectric test data and
photoelectric curve.
Nine wavelengths (340, 405, 450, 510, 546, 578, 630, 670 and 700) and a base
wavelength are listed on the control panel. You can select one of them to display
its photoelectric data curve.
Parameter
What it does
MinMax
YRange
XRange
CupIndex
Dozes-reagent
Dozes-sample
Period
10-20
What it does
Times
Mixer stir
Through the control panel, you can select parameters like Wavelength, MinMax,
YRange, CupIndex and Period. Moreover, you can select Mixer stir as required.
After setting them, click the Confirm button to confirm the settings.
There are 6 more buttons in the panel:
Start: Clicking it, you can start collecting the temperature data;
Stop: Clicking it, you can stop collecting the temperature data;
Freeze: Clicking it, you can freeze the displayed curve;
Resume: Clicking it, you can resume the frozen curve;
Save: Clicking it, you can save the photoelectric data of all current wavelengths
as a .txt file;
Load: Clicking it, you can read the data in a .txt file saved with photoelectric
data.
10-21
Appendix A
Power supply
assembly
Reagent
refrigeration
board
Main control
board
Power drive
board
A/D conversion
board
Probes
connection
board
Feeder
connection
board
A-1
1
3
4
J38
Cable
BA30-20-15305
4 red
1 VCC
3yellow
3 12V
2black
4 GND
6
3
1blue
3 red 5V
2 white FAN
2
Cable
BA30-20-06480
1 black GND
1
J42
P352
J87
Cable
BA30-21-15182
2 PRTD_REF
black
3 PRTD_T1
red
Cable
BA30-20-06462
12&5V board
BA33-30-35080
6 12GND
Reset button
2100-10-07946
Reagent refrigeration
board
BA30-30-06312
J44
2red
Cable
BA30-20-06484
4 red WASTE_FULL
3 blue GND
J33
4 red
3 white
1 black
J47
2 blue
4 red
J35
1 green
1 green
Cable
BA30-20-15291
J34
3red
3
Cable
BA30-20-15292
3
3 white
2 blue
2red
Cable
BA30-20-15290
1
J203
2 black
J45
48 47
50 49
J204
1
1 black
48 47
50 49
42 41
44 43
38 37
36 35
40 39
34 33
28 27
32 31
26 25
30 29
24 23
18 17
22 21
20 19
16 15
14 13
12 11
46 45
46 45
44 43
42 41
40 39
38 37
36 35
34 33
32 31
30 29
28 27
26 25
J8
24 23
22 21
12
13
9
4
J41
2
1
4
3
1 NC
1 NC
3 blue
RXD 2
2 green
TXD 3
4 NC
ISE module
5 black
6 NC
NC 4
GND 5
NC 6
8 white
RTS 7
7 yellow
CTS 8
9 NC
NC 9
+
C
E
C17: reagent probe rotation +
C
position transducer
BA30-21-15156
E
+
C19: reagent syringe position
C
transducer
BA30-21-15161
E
C24: sample syringe position +
C
transducer
BA30-21-15162
E
+
C13: reaction disk initial
C
position transducer
BA30-21-15154
E
+
C14: reaction disk coder
C
transducer
BA30-21-15153
E
C18: reagent probe vertical
position transducer
BA30-21-15155
Cable
BA34-20-63597
8 10 12 14 16 18 20 22 24 26
7 9 11 13 15 17 19 21 23 25
Cable
BA30-20-06476 VCC
RU_PHO
GND
GND
VCC
RR_PHO
GND
GND
VCC
RS_PHO
GND
GND
VCC
SS_PHO
GND
GND
VCC
REAC_PHO
GND
GND
VCC
REACC_PHO
GND
GND
1red
4white
2black
3green
1red
4white
2black
3green
1red
4white
2black
3green
1red
4white
2black
3green
1red
4white
2black
3green
1red
4white
2black
3green
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
2
J37
4
3
6
5
8 10 12 14 16 18 20 22 24 26 28 30 32 34
7 9 11 13 15 17 19 21 23 25 27 29 31 33
Cable
BA30-20-06477
+
C
E
+
C
E
+
C
E
+
C
E
+
C
E
+
C
E
+
C
E
+
C
E
1red
4white
2black
3green
1red
4white
2black
3green
1red
4white
2black
3green
1red
4white
2black
3green
1red
4white
2black
3green
1red
4white
2black
3green
4white
2black
3green
1red
4white
2black
3green
VCC
STIRR_PHO
GND
GND
VCC
STIRU_PHO
GND
GND
VCC
RTC_PHO
GND
GND
VCC
RT_PHO
GND
GND
VCC
SU_PHO
GND
GND
VCC
SR_PHO
GND
GND
VCC
ST_PHO
GND
GND
VCC
STC_PHO
GND
GND
A-2
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
2
J32
4
3
6
5
8 10
7
2
J43
Cable
BA30-20-06485
1 PRESSURE_PHO 1
2 FRONT_PHO 2
3 BUTTON_S
3
4 BACK_PHO 4
VCC
5
5
6 MIDDLE_PHO 6
GND
7
7
8 LOADE_PHO 8
GND
9
9
10 GND
10
4
3
6
5
8 10
7
Cable
BA34-20-63651
URAM_PHO 1
1
2 LOADX_PHO 2
3 DRAM_PHO
3
4
LAODY_PHO 4
5
5
VCC
6
MAGU_PHO 6
GND
7
7
8
MAGD_PHO 8
GND
9
9
10
10
GND
J31
8 10
GND
NC
4
8 10
J99
Feeder connection
board
BA30-30-06314
8 10
J107
Manipulator connection
board
BA30-30-06316
24
10
23
9
22
8
21
7
20
ISP_RESET
NC
NC
NC
5
18
19
4
17
3
16
2
15
1
14
J46
Cable
BA30-20-06552
NC
RXD232
TXD232
25
PC COM
ISP_PSEN
11
J39
3
8
3
7
2
1 +15V
14 +15V
2 -15V
15 -15V
3 VCC
16 VCC
4 15GND
17 15GND
5 GND
18 AD_BUSY
6 AD_DIN
19 AD_CLK
7 GND
20 AD_RC
8 GND
21 CH_A3
9 DCP_EN
22 CH_A2
10 DCP_CLK
23 CH_A1
11 DCP_DIN
24 CH_A0
12 GND
25 GND
13 GND
6
J7
Cable
BA34-20-63652
18 17
20 19
16 15
14 13
J9
J40
1 GND
NC 15
12 11
NC 14
NC 13
NC 12
NC 8
NC 11
NC 7
4 NC
NC 5
Cable
BA30-20-06481
orange 9
9 NC
2 LOAD_DC+
3 LOAD_DC4 MAG_UP+
5 MAG_UP6 MAG_DOWNN+
7 MAG_DOWNN8 SP_VALVE_IN
9 RP_VALVE_IN
10 P_PUMP_IN
11 P_PUMP_OUT
12 P_PUMP_WASTE
13 PELTIER
14 STIR_DC
15 LAMP_CTRL
16 STIR_PUMP
17 STIR-VALVE
18 T-RELAY
19 GND
20 REAC-CLK
21 REAC-DIR
22 LOADX-CLK
23 LOADX-DIR
24 LOADY-DIR
25 LOADY-CLK
26 ST-CLK
27 ST-DIR
28 RT-CLK
29 RT-DIR
30 STIRR-CLK
31 STIRR-DIR
32 STIRU-CLK
33 STIRU-DIR
34 GND
35 SR-CLK
36 SR-DIR
37 SU-CLK
38 SU-DIR
39 SS-CLK
40 SS-DIR
41 RR-CLK
42 RR-DIR
43 RU-CLK
44 RU-DIR
45 RS-CLK
46 RS-DIR
47 GND
48 VCC
49 VCC
50 VCC
yellow 10
8 TRIG
white 6
7 CTS
6 RTS
J205
BNC
C25: deionized water
transducer
BA30-21-15163
black 4
blue 2
1 white GND
5 GND
BNC
green 3
2 yellow WASH_EMPTY
red 1
2 RXD
3 TXD
J20
1 VCC
Cable
BA34-20-63598
2
3
4
5
6
15
14
J61
1
16
3
17
4
18
5
20
19
6
21
8
22
9
23
10
8
9
10
SHIELD
25
24
11
12
13
+15V
AGND
GND
2
1
4
3
5
J54
6
4
6
1
3
5
J55
J61
J39
1
2
3
1 +15V
14 +15V
2 -15V
15 -15V
3 VCC
16 VCC
4 15GND
17 15GND
5 GND
18 AD_BUSY
6 AD_DIN
19 AD_CLK
7 GND
20 AD_RC
8 GND
21 CH_A3
9 DCP_EN
22 CH_A2
10 DCP_CLK
23 CH_A1
11 DCP_DIN
24 CH_A0
12 GND
25 GND
13 GND
3
J53
VCC
1
J52
J62
J51
2
4
6
1
3
5
J56
A-3
2
4
6
1
3
5
J57
2
4
6
1
3
5
Cable
BA30-20-06552
14 15 16 17 18 19 20 21 22 23 24 25
9 10 11 12 13
J58
2
4
6
1
3
5
J59
2
4
6
1
3
5
J60
1
2
3
4
5
6
Cable
BA30-20-06545 1
2
3
4
5
6
Cable
BA30-20-06544 1
2
3
4
5
6
Cable
BA30-20-06543 1
2
3
4
5
6
Cable
BA30-20-06542 1
2
3
4
5
6
Cable
BA30-20-06541 1
2
3
4
5
6
Cable
BA30-20-06540 1
2
3
4
5
6
Cable
BA30-20-06539
Cable
BA30-20-06538 1
2
3
4
5
6
Cable
BA30-20-06537 1
2
3
4
5
6
Cable
BA30-20-06536 1
2
3
4
5
6
REF
+15V Photoelectric conversion
board (REF)
AGND
BA30-30-06691
AGND
AGND_SHILED
-15V
700nm
+15V Photoelectric conversion
board (700nm)
AGND
BA30-30-06689
AGND
AGND_SHILED
-15V
670nm
+15V Photoelectric conversion
board (670nm)
AGND
BA30-30-06687
AGND
AGND_SHILED
-15V
630nm
+15V Photoelectric conversion
board (630nm)
AGND
BA30-30-06685
AGND
AGND_SHILED
-15V
578nm
+15V Photoelectric conversion
board (578nm)
AGND
BA30-30-06683
AGND
AGND_SHILED
-15V
546nm
+15V Photoelectric conversion
AGND
board (546nm)
AGND
BA30-30-06681
AGND_SHILED
-15V
510nm
+15V Photoelectric conversion
board (510nm)
AGND
BA30-30-06679
AGND
AGND_SHILED
-15V
AGND_SHILED
-15V
450nm
+15V Photoelectric conversion
board (450nm)
AGND
BA30-30-06677
AGND
405nm
+15V Photoelectric conversion
board (405nm)
AGND
BA30-30-06675
AGND
AGND_SHILED
-15V
340nm
+15V Photoelectric conversion
board (340nm)
AGND
BA30-30-06673
AGND
AGND_SHILED
-15V
2
4
6
1
3
5
MAGU-
MAGU+
MAGD-
MAGD+
BA30-20-06464
B
R
LOADING
CONNCETING BD
BA34-30-63648
J97
DCDC+
BA30-20-06463
J18
1
DC-
DC+
J24
BA30-20-15293
BA30-20-15289
1
B
W
Y
B
R
W
Y
2 M15INNER PUMP
1 2000-10-06120
B
R
2
BA30-20-06465
J22
3
4
24V
GND
J10
24V
J5
POWER DRIVE BD
BA34-30-63559
J20
12V
5V
GND
12 22
3
4
2 2
25 26
72
39
13
83
0
3
73
49
14
84
40
1 BLC
2 R
M4LIFT MOTOR OF
M20:
20:MIXING MODULE M3:HORIZENTAL MOTOR
LIFT
OF LIFT
ROTATORY MOTOR
BA31-21-568164
BA31-21-56814
BA33-21-55743
J7 4
9 0
M6:REACTION
M21:
21:LIFT MOTOR OF
MIXING
COMPARTMENT MOTOR
BA31-21-56816
BA31-21-56816
BA30
1 GND
BA3030-2020-06481
2 LOAD_DC+
3 LOAD_DC4 MAG_UP+
5 MAG_UP6 MAG_DOWNN+
7 MAG_DOWNN8 SP_VALVE_IN
9 RP_VALVE_IN
10 P_PUMP_IN
11 P_PUMP_OUT
12 P_PUMP_WASTE
13 PELTIER
14 STIR_DC
15 LAMP_CTRL
16 STIR_PUMP
17 STIR-VALVE
18 T-RELAY
19 GND
20 REAC-CLK
21 REAC-DIR
22 LOADX-CLK
23 LOADX-DIR
24 LOADY-DIR
25 LOADY-CLK
26 ST-CLK
27 ST-DIR
28 RT-CLK
29 RT-DIR
30 STIRR-CLK
31 STIRR-DIR
32 STIRU-CLK
33 STIRU-DIR
34 GND
35 SR-CLK
36 SR-DIR
37 SU-CLK
38 SU-DIR
39 SS-CLK
40 SS-DIR
41 RR-CLK
42 RR-DIR
43 RU-CLK
44 RU-DIR
45 RS-CLK
46 RS-DIR
47 GND
48 VCC
49 VCC
50 VCC
BA30-20-15352
J11 4
BA30-20-15354
M7:REAGENT
M11
11SAMPLE
COMPARTMENT MOTOR COMPARTMENT MOTOR
BA31-21-56816
BA31-21-56816
1 J16
1 J6 4
1 J14 4
BA34-20-63775
J9 4
24
3
4
4 4
5
4 46
74 85
12GND
1 J15
1
BA30-20-15358
B
BA34-20-63756
2
2
2
3
4
3
3 3
6
5
3 3
BA34-20-63630
3
3
15 16
71 82
29 20
GND
BA30-20-15357
W
6
P350
12&5V BD
BA33-30-35080
P405
1
BA3
BA33-2020-35185
M12:
12:ROTAORY MOTOR
OF SAMPLE PROBE
BA31-21-56814
A-4
9 4
1 1
11 2
1
3
4
1 1
15 16
71 82
J40
29 20
12 22
4
3
2 2
25 26
72
39
13
83
0
3
2
3
3
4
3 3
6
5
3 3
73 84
0
49 4
14 24
4
3
4 4
5
4 46
74 85
9 0
CPU BD
BA34-30-63557
VCC 2
GND 1
12V 4
4 R
12GND 3
J12 4
2 B
BA30-20-15347
P350
GND 5
GND 6
2 W
1 W
24V 7
3 O
P403
24V
24V BD
BA33
33BA
33-3030-35082
J3 1
24V BD
BA33-30-35082
9 4
1 1
11 2
1
3
4
1 1
BA30-20-15353
4 J106
J17
J13
J19 1
M9:LIFT MOTOR OF
REAGENT PROBE
BA31-21-56816
J23 1
2
3
4
BA30-20-15348
M8:ROTATORY MOTOR
OF REAGENT PROBE
BA31-21-56814
BA30-20-15349
M14:
14:MOTOR OF SAMPLE
SYRINGE
BA31-21-56814
M13:
13:LIFT MOTOR OF
SAMPLE PROBE
BA31-21-56816
3
2
J8
BA30-20-15351
W
J21 1
J
1
LIFT
COONCTING BD
BA34-30-63650
J4
J
2
PROBES
CONNECTING BD
BA30
BA3030-3030-15284
J206
J207
M10:
10:MOTOR OF
REAGENT SYRINGE
BA31-21-56814
BA30-20-06468
LAMB 12V
CONTROL CABLE
BA33-20-35186
BA34-20-63755
W
M16OUT PUMP
2000-10-06120
M17SAMPLE VALVE
BA30-21-06469
2
1
M18REAGENT
VALVE
BA30-21-06470
B
R
P6
1
PFC BD
BA33-30-35078
J80
J78
J82
J83
J84
12V
GND
12V
1 blue
2 red
12V
1 red
2 blue
12V
GND
2
2 red
1 blue
1
J85
12&5V board
BA33-30-35080
2 black
3 red
4 black
COOLER
12V
GND
GND
COOLER
GNDFAN
12FAN
Cable
BA30-20-06480
1 red
2 blue
2 1
3
3 red
2 white
J87
1black
2
2 red
1
1 blue
J77
12V
2 red
1 red
GND
2 blue
1 blue
12V
2 red
2
J76
1 red
2 blue
3 red
Cable
BA30-20-06525
1 blue
GND
12V
1
5V
J75
3 red
1 red
2 blue
FAN
2 blue
GND
3 yellow
1 yellow
12V
3 red
FAN
J71
1 red
2 blue
FAN
2 blue
3
2 white
1 yellow
GND
3 yellow
3 red
12V
2 blue
2
J72
1 red
FAN
3
GND
2 blue
1 yellow
GND
1black
J73
3 yellow
2 blue
FAN
3
3
3 red
12V
1 red
J74
J42
M37-M38: Fans on sides of lamp
housing
2100-20-08114
A-5
Cable
BA30-20-15308
4 yellow
P353
5 yellow
1 blue
5
12V
2 blue
GND
J79
12V
1 yellow
12V
J86
GND
GND
12V
Cable
BA30-20-06524
12V
J81
J89
1 red
3 yellow
M30-M31: PELTIER
2100-20-06633
Cable
BA30-20-06535
1 red
Cable
BA30-20-06529
GND
2 blue
12V
1 red
2 red
GND
2 blue
Cable
BA30-20-06523
1 blue
1
J88
GND
2 blue
3 white
6 orange
3 white 12V
J1
J2
1 24V
2 GND
ISE
unit
Enclosure
grounded
POWER
M07-00061S---
E yellowish green
Power jack
Reagent refrigeration
board
PFC board
BA30-30-06312
BA33-30-35078
12&5V board
BA33-30-35080
P353
4
4
5
1
2
3
6
yellow
yellow
blue
blue
white
orange
Cable
BA30-20-15308
N light blue
J86
12V 6
12V 5
GND 4
GND 3
6
5
4
P404
P1
1
GNDFAN 2
12FAN 1
1 brown L
Cable
BA34-20-63696
2
3
3 light blue N
P3
P350
1 blue
3
4
1
2
J17
GND 7
3 yellow
12V 4
4 red
VCC 3
1 yellow
yellow
3 white
white
1 blue
2 black
3 yellow
4 red
12GND 6
GND 4
12V 3
VCC 1
P406
5
Cable
BA33-20-35092
5 brown
1 blue
7 brown
3 blue
8 brown
4 blue
P405
P6
Cable
BA30-20-15305
2
N 3 light blue
Cable
BA33-20-35189
2 blue
Cable
BA33-21-35187 (for 220V) / BA33-21-35094 (for 110V)
Cable
BA33-21-35188 (for 220V) / BA33-21-35095 (for 110V)
Cable
BA33-20-35093
12GND 8
2 black
L 1 brown
24V board
BA33-30-35082
L brown
2
1
1 red
2 blue
Cable
BA33-20-35186 2 red
1 blue
J21
P5
2
1
2 black
1 red
Cable
BA33-20-35091
Cable
BA33-20-35185
J18
2 red
1 blue
1 blue
2
1
2
1
P403
J38
3
2 red
2
M32: lamp
12VGND
LAMP12V
A-6
Cable
BA34-20-63600
4 orange
3 orange
2 white
GND 6
1 white
GND 5
24V 1
24V 2
J17
2
Power drive
board
BA34-30-63559
1 yellow GND
M28 reaction
temperature switch
J22
1
2
1
2
1
2
3
1
A-7
7
8
1
2
2
3
J204
3
1
2
3
4
red 4
4 red
J203
white 3
4
3
3 white
blue 2
J200
2 blue
4 red 12V
4
3
2red
1
2
2red
2red
J66
3 black SP_LEVEL
Cable
BA30-20-06547
1black
1black
2red
Cable
BA30-20-06546
1black
1black
C30: Preheating
temperature transducer
BA30-10-06631
M29: Preheater
BA30-10-15115
GND
PROBE
GND
PROBE
2 blue SP_RAM_PHO
Cable
BA30-20-15295
Cable
BA30-20-15292
green 1
1 green
1 green AGND
4
8 red 12V
J35
7 black RP_LEVEL
red 4
2
6 blue RP_RAM_PHO
J66
4 red
white 3
3 white
J205
blue 2
5 green AGND
4 purple
3 brown
Cable
BA30-20-15294
2 blue
Cable
BA30-20-15291
green1
1 green
J206
J34
red 2
2 orange
1 yellow
2 red
Cable
BA30-20-15296
3 red
J45
J201
Cable
BA30-20-15290
black 1
2 black
red 2
blue 1
J207
2 red
Cable
BA34-20-63604
1
2
1 blue
J24
J202
red 2
1 black
2 red
Cable
BA34-20-63603
blue1
1 blue
2 red
Cable
BA33-21-35151
1 blue
2 red
Cable
BA33-21-35152
1 blue
MAGU-
3 blue
MAGD+
2 yellow
MAGD-
1 white
3 blue
2 yellow
1 white
MAGU-
MAGD+
4
3
GND
GND
VCC
LOADY_PHO
J107
2
GND
GND
7
9
J110
8
10
VCC
URAM_PHO
LOADX_PHO
DRAM_PHO
LOADY_PHO
VCC
MAGU_PHO
GND
MAGD_PHO
GND
GND
4 red
3 white
Cable
BA33-21-35153
2 black
red
white
black
1 green
green
4 red
3 white
Cable
BA33-21-35154
2 black
red
white
black
1 green
4 red
3 white
GND
2 blue
GND
1 black
MAGU_PHO
green
Cable
BA33-20-35150
red
white
blue
black
J103
J109
VCC
MAGD_PHO
GND
GND
1
2
3
4
5
6
7
8
9
10
DRAM_PHO
Cable
BA30-20-06486
GND
1
2
3
4
5
6
7
8
9
10
GND
2
4
blue
J108
red
J102
URAM_PHO
LOADX_PHO
J43
URAM_PHO
LOADX_PHO
DRAM_PHO
LOADY_PHO
VCC
MAGU_PHO
GND
MAGD_PHO
GND
GND
VCC
MAGD-
blue
J101
MAGU+
J105
MAGD-
MAGU+
Cable
J106
BA30-20-06463
4
red
4 red
4
MAGD+
J3
MAGU+
MAGU-
J104
red
A-8
GND
GND
4 red
3 white
2 blue
1 black
Cable
BA33-20-35149
red
white
blue
black
Cable
BA30-20-06490
4 red
3 white
white
2 black
black
1 green
4
1
VCC
BACK_PHO
J32
PRESSURE_PHO
FRONT_PHO
BUTTON_S
BACK_PHO
VCC
MIDDLE_PHO
GND
LOADE_PHO
GND
GND
1
3
5
2
4
6
10
1
2
3
4
5
6
7
8
9
10
1
2
3
4
5
6
7
8
9
10
GND
VCC 4
PRESSURE_PHO
FRONT_PHO
BUTTON_S
MIDDLE_PHO
3
BACK_PHO
VCC
GND 2
MIDDLE_PHO
GND 1
GND
LOADE_PHO
J94
GND
GND
VCC 4
10
3 white
3 white
GND
GND
1 green
J95
PRESSURE_PHO
DC+
2
DC1
GND
GND
DC+ DCJ97
J96
1 black
2 red
2 red
Cable
BA30-20-06494
1 blue
Cable
BA30-20-06464
A-9
DCDC+
red
white
black
C10: No-cuvette
transducer
BA30-21-06489
green
Cable
BA30-20-06488
4 red
C9: Insufficient-cuvette
transducer
BA30-21-06492
green
Cable
BA30-20-06489
2 black
VCC
black
1 green
red
white
2 black
4 red
C8: Cuvette-taking
limit transducer
BA30-21-06491
green
Cable
BA30-20-06492
black
1 green
4 red
red
white
2 black
LOADE_PHO
4 red
3 white
J98
J19
Cable
BA30-20-06491
J93
2
GND
J99
Cable
BA30-20-06485
C7: Cuvette-pushing
limit transducer
BA30-21-06490
green
J92
J91
red
3 white
2 black
1 green
red
white
black
green
C11: Pressure
transducer
BA30-21-06488
Appendix B
B.1
TP
TP1
TOP/BOTTOM
TOP
Descriptions
The signal at 340nm
TP2
TOP
TP3
TOP
TP4
TOP
TP5
TOP
TP6
TOP
TP7
TOP
TP8
TOP
TP9
TOP
TP10
TOP
TP11
TOP
TP12
TOP
GND
TP13
TOP
B.2
TP
TP1
TOP/BOTTOM
TOP
Descriptions
Signal ground (SHIELD) from the main control board
TP2
TOP
TP3
TOP
TP4
TOP
B.3
TP
TP1
TOP/BOTTOM
TOP
Descriptions
The temperature switch signal (low)
TP2
TOP
TP3
TOP
Reference ground
TP4
TOP
TP5
TOP
B.4
TP
TP1
Descriptions
TxD of Sample Unit
B-1
TOP/BOTTOM
BOTTOM&TOP
Descriptions
RxD of Sample Unit
TP3
BOTTOM&TOP
TP4
BOTTOM&TOP
TP5
BOTTOM&TOP
TP6
BOTTOM&TOP
TP7
BOTTOM&TOP
TP8
BOTTOM&TOP
TP9
BOTTOM&TOP
TP10
BOTTOM&TOP
TP11
BOTTOM&TOP
TP12
BOTTOM&TOP
TP13
BOTTOM&TOP
TP14
BOTTOM&TOP
TP15
BOTTOM&TOP
GND
TP16
BOTTOM&TOP
TP17
BOTTOM&TOP
TP18
BOTTOM&TOP
TP19
BOTTOM&TOP
TP20
BOTTOM&TOP
GNDTT GND of
TP21
BOTTOM&TOP
VCC 5V Power
TP22
BOTTOM&TOP
TP23
BOTTOM&TOP
TP24
BOTTOM&TOP
TP25
BOTTOM&TOP
GND
TP26
BOTTOM&TOP
TP27
BOTTOM&TOP
TP28
BOTTOM&TOP
TP29
BOTTOM&TOP
TP30
BOTTOM&TOP
TP31
BOTTOM&TOP
TP32
BOTTOM&TOP
TP33
BOTTOM&TOP
TP34
BOTTOM&TOP
TP35
BOTTOM&TOP
TP36
BOTTOM&TOP
TP37
BOTTOM&TOP
TP38
BOTTOM&TOP
B-2
15V
B.5
TP
TP1
TOP/BOTTOM
BOTTOM
Descriptions
The switch waveforms of Q3 and Q4
TP2
BOTTOM
TP3
BOTTOM
TP4
BOTTOM
TP5
BOTTOM
TP6
BOTTOM
TP7
BOTTOM
TP8
BOTTOM
TP9
BOTTOM
TP10
BOTTOM
TP11
BOTTOM
TP12
BOTTOM
TP13
BOTTOM
TP14
BOTTOM
B.6
TP
TP1
TOP/BOTTOM
BOTTOM
Descriptions
The DC output (390V)
TP2
BOTTOM
TP3
BOTTOM
TP4
BOTTOM
TP5
BOTTOM
TP6
BOTTOM
TP7
BOTTOM
TP8
BOTTOM
TP9
BOTTOM
TP10
BOTTOM
TP11
BOTTOM
TP12
BOTTOM
TP13
BOTTOM
TP14
BOTTOM
TP15
BOTTOM
TP16
BOTTOM
B-3
B.7
TP
TP1
TOP/BOTTOM
BOTTOM
Descriptions
The DC output (390V)
TP2
BOTTOM
TP3
BOTTOM
TP4
BOTTOM
TP5
BOTTOM
TP6
BOTTOM
TP7
BOTTOM
TP8
BOTTOM
TP9
BOTTOM
TP10
BOTTOM
TP11
BOTTOM
TP12
BOTTOM
TP13
BOTTOM
B-4