1
IR SPECTROSCOPY
Manu Jose First year M.Pharm Pharmaceutical Analysis Dept. National College of Pharmacy
OVERVIEW
INTRODUCTION THEORY FACTORS AFFECTING VIBRATIONAL FREQUENCY
SAMPLING TECHNIQUES
INSTRUMENTATION
INTERPRETATION
APPLICATION
Introduction
Spectroscopy
Spectroscopy is the measurement and interpretation of EMR absorbed or emitted,
when the organic molecule or atoms or ions of a
sample move from one energy state to another.
Definition
IR spectroscopy can be defined as a method for the identification of substances based on their
absorption of IR wavelength.
IR RADIATION
Electromagnetic spectrum
g-rays
X-rays
UV
IR
Microwave
Radio
Visible
Wave number Near IR Mid IR Far IR 14000-4000cm-1 4000-400cm-1 400-10cm-1
Wave length 0.8-2.5m 2.5-25m 25-1000m
Theory
IR radiation in the range 10,000- 100 cm-1 is absorbed. Frequency of absorption depends on relative masses, the force constant, and geometry of atoms. Converted to energy of molecular vibration. The absorption is quantized. Vibrational spectra is appear as bands.
A molecule absorbs a selected frequencies of IR.
Absorption corresponding to 8-40 KJ/mol
Not have sufficient energy to cause the excitation
of electrons, it can cause vibrations.
10
Criteria for absorption
Not all bonds are capable of absorbing IR. 1. Dipole moment To absorb IR, a molecule must undergo net change in dipole moment as it vibrate. Electric field of radiation can interact with molecule- change amplitude of vibration. Eg: Hydrochloride.
2. Correct wavelength
Applied IR frequency =Natural frequency of vibration
11
Calculation of stretching frequency:
By Hooks law;
v - vibrational frequency k - force constant = (m1.m2)/(m1+m2) m1 and m2 mass of atom 1 and 2 respectively.
12
Molecular vibrations
13
The relative positions of atoms in a molecule are not fixed but fluctuate continuously as a
consequence of vibrations and rotations.
Two types of fundamental vibrations are there,
Stretching
Bending
14
Stretching vibrations
Involves continuous change in inter atomic distance along the axis.
Two types:
Symmetrical stretching-stretching/ compression in symmetrical way.
Asymmetric stretching-one bond is stretching while other is compressing.
15
symmetrical
H
C
H
asymmetrical
H
C
H
16
Bending vibrations:
Involves change in bond angle between bonds with a common atom or the movement of group of atoms to the reminder of molecule.
Two types:
In-plane
Out-of plane
17
In plane bending vibrations Scissoring- atoms move towards and away from each other.
H C H
Rocking- The structural unit swings back and forth in the plane.
H C H
18
Out of plane bending vibrations
Wagging: The structural unit swings back and forth out of the plane
Twisting: structural unit rotate about the bond which joins to the rest of the molecule
H C H
H C
19
Degree of freedom
Each atom has 3 degrees of freedom N-atom molecule there will be 3N degree of freedom. Translation - the movement of the entire molecule while the positions of the atoms relative to each other remain
fixed
Rotational transitions inter-atomic distances remain constant but the entire molecule rotates with respect to three mutually perpendicular axes.
Translation Non linear linear 3 3 Rotational 3 2
20
Number of vibrational degree of freedom
For molecules having N atoms in them Nonlinear molecules= 3N - 6 Eg : H2O (3x3)-6=3
Linear molecules= 3N -5
Eg : N2 (3x2)-5=1
21
Why not 3N-6/3N-5 bands in IR Spectrum?
The theoretical number of fundamental vibrations (absorption frequencies) will seldom be observed. Overtone and combined tone increase the number of bands.
22
Reasons for decrease in number of bands are-
Bands which are so close that they coalesce.
Bands which are too weak to be observed.
Occurrence of a degenerate band.
Lack of change in molecular dipole moment.
Fundamental frequencies outside 4000-400 cm-1
23
Eg:
CO2 is linear 3x3-5 =4 degree of freedom. But
Symmetric stretching- no change in dipole Bending vibrations are equivalent (degenerate)
24
Factors influencing vibrational frequencies
Calculated value of frequency of absorption for a
particular bond is never exactly equal to its experimental value.
Electronic effects inductive and field effect Ring size Hydrogen bonding Vibrational coupling
25
Inductive effect:
Electro negative/ Stronger bond Negative inductive Increased force constant Wave number rises Acetone chloroacetone Dichloroacetone Tetrachloroacetone H-CHO Electro positive/ electron attracting Positive inductive effect Weaker bond Decrease force constant Wave number decreases CH3-CHO 1745 Wave no cm-1 1715 1725 1740 1778 1750
electron withdrawing
26
Field effect:
Lone pair of electrons on two atoms influences
each other through space, can change vibrational
frequency of both groups.
Eg: O- haloacetophenone
CH3 C = O: :
Electrostatic repulsion due to lone pair off electrons
Change the state of hybridization of C=O
X:
:
Absorption occurs at higher wave number
27
Ring size
Ring strain causes large shift to higher frequency
7 membered ring 1702 cm-1 6 membered ring 1715 cm-1 5 membered ring 1745 cm-1 4 membered ring 1780 cm-1 3 membered ring 1850 cm-1
Increased s character
28
Hydrogen bonding
Remarkable downward frequency shift. Stronger H bond- greater the shift towards lower wave number.
Free alcohol H bonded alcohol - sharp OHstr band - broad OHstr band
OH of phenol condensed to polymeric wide
envelope of O-Hstr
29
Vibrational coupling
Isolated C-H bond only one str frequency
For methylene (-CH2-) two coupled vibrations of
different frequency. ( asymmetric and symmetric)
Fundamental+ fundamental
Amide
1600- 1700cm-1
C=O str N-H def
Fundamental + overtone
Aldehydes
2800- 2700 cm-1
C-H str C-H def
30
Sampling techniques
31
1) Solid samples:
Mull technique Pressed pellet technique Solids run in solutions Solid film
2) Liquid samples 3) Gaseous sample
32
Mull technique: Grinding 2-5mg of sample in smooth agitate mortar. Powdered sample + Nujol paste Paste between the two plates of salt. The oil has few absorption bands at 2857, 1449 and 1389 cm-1
33
Pressed pellet technique:
1mg Sample+ 100mg KBr powder
pressed in hydraulic press transparent disc inserted in sample holder.
34
Solids films: Amorphous solid samples melted between salt plates allowed to form solid film. For qualitative purpose.
Solids run in solution Solid + suitable solvent solution Kept in cells for liquids Solvents- non associated solvents- CS2, CCl4,
35
Liquid sample:
36
Gaseous sample:
Vapors in specially designed cells.
End walls made up of
Sodium chloride.
37
Instrumentation
38
Dispersive infra red spectrometers
Instrument produce a beam of IR radiation Mirrors divide it into 2 parallel beam of equal intensity. Sample is placed in one beam, other is used for reference. The beam is then passed through monochromator which disperse into a continuous spectrum. The detector senses the ratio b/w intensities of sample and reference beam. Signals is amplified and recorded.
39
Dispersive infra red spectrometers
40
Radiation sources
Consist of inert solid Heated electrically
Nernst glower Globar source Mercury arc Incandescent wire source Tungsten filament
41
Nernst glower
Composed of rare earth oxides- zirconium, yttrium and thorium Hollow tube (2-5cm x 1-3mm) Platinum leads at one end Large negative temp. coefficient. Emit radiation over wide range and remains steady over a long period.
Disadvantages: Fragile Auxiliary heater Over heating
42
Globar source
Rod of sintered silicon carbide. (5cm x 5mm). Positive coefficient of resistance. Self starting and electrically heated Enclosed in water cooled brass tube
Less intense.
43
Mercury arc
For far IR region High pressure Hg arc, enclosed in quartz jacketed tube, at 1 atm Passage of electricity through vapor internal
plasma source IR radiation
44
Incandescent wire loop
Tightly wound spiral of Nichrome wire. No water cooling Less maintenance Less intense than other sources.
Tungsten filament lamp
Tightly wound spiral of tungsten wire. For near IR
45
Sample holders
Constructed of rock salt. Path length is adjusted with Teflon. Filled and emptied with hypodermic needles.
Foggy due to moisture.
Care:
Moisture free samples Fingers should not be come in contact Prevent contamination with silicones
46
Monochromator
To select desired frequency from radiation source.
Prism monochromator Grating monochromator
Material used: Halogen salt
47
Detectors
Thermal detectors
(IR heating potential difference amt of radiation)
Thermocouple Bolometer Golay cell
Photo detectors
48
Thermocouple
49
Bolometer
50
Golay cell
51
FT Optical System Diagram
Light source
He-Ne gas laser
(ceramic) Beam splitter Movable mirror Sample chamber
(DLATGS) Fixed mirror Interferometer Detector
52
53
Michelson interferometer FT-IR
Moving mirror
IR Source Stationary mirror
54
Constructive and destructive interferences
55
Comparison
Dispersion Spectrometer
The dispersion Spectrometer takes several minutes to measure an IR spectrum Also the detector receives only a few % of the energy of original light source.
FT-IR
FT-IR takes only a few seconds, to measure an IR spectrum. The detector receives up to 50% of the energy of original light source.
56
Spectrum
57
IR spectrum
Each dip- band 100% transmittance- no absorption. Scale of spectrum is not entirely linear.
58
Interaction of IR Radiation with Organic Molecules
Primary use is to detect functional group. All organic functional groups are made up of multiple bonds and therefore they will show multiple IR bands. Capable of giving sufficient information about the structure of a compound.
59
ALKANES
General structure :
R- CH2- R
Wave number
C-H (str) 3000-2840 cm-1
C-H (def)
1470-1450 cm-1
C-H (rock)
1370-1350 cm-1
In long chain alkanes
C-H (rock)
725-720 cm-1
60
Eg: Octane
61
ALKENES
General structure : R-C=C-R
Wave number
=C-H str =C-H def C=C str
3100-3000 cm-1 1000-650 cm-1 1680-1640 cm-1
62
Eg: 1-Octene
63
ALKYNES
General structure : R- C C - R
Wave number - C C - H str 3333-3267 cm-1
- C C - H def
700-610 cm-1
-C C str
2260-2100 cm-1
64
Eg: 1-Hexyne
65
Aromatic Compounds
General structure : Ar-R Wave number
C -H str
3100-3000cm-1
C -C str (in ring)
1600-1585cm-1 1500-1400cm-1
C-H bend
900-675cm-1
66
Eg: Toluene
Alkyl Halides
General structure : R-X
Wave number
C-H str 3000-2850 cm-1
67
C-H def
1300-1150 cm-1
C-X str eg: C-Cl str C-Br str
850-515 cm-1 850-550 cm-1 690-515 cm-1
68
Eg: 1-Bromopropane
69
Alcohols
General structure : R-OH
Wave number
O-H str
3500-3200 cm-1
C-O str
1260-1050 cm-1
70
Eg: Ethanol
71
KETONES
General structure : R-CO-R
Wave number
C=O str
aliphatic ketones
1715 cm-1
,-unsaturated ketones
1685-1666 cm-1
72
Eg: 2-Butanone
73
Aldehydes
General structure : R-CHO
Wave number
H-C=Ostr
C=O str
2830-2695 cm-1
aliphatic aldehyde
1740-1720 cm-1
,-unsaturated aldehydes
1710-1685 cm-1
74
Eg: Butaraldehyde
75
76
CARBOXYLIC ACIDS
General structure : R-COOH
Wave number O-H str 3300-2500 cm-1
C=O str
1760-1690 cm-1
C-O str
1320-1210 cm-1
O-H bend
1440-1395 cm-1 950-910 cm-1
77
Eg: Hexanoic acid
78
ESTERS
General structure : R-COOR
Wave number
C=O str
aliphatic esters
1750-1735cm-1
,-unsaturated
1730-1715cm-1
C-O str
1300-1100cm-1
79
Eg: Ethyl acetate
80
Amines
General structure : R- NH2 Wave number N-H str 1O amines 3400-3250cm-1 3400-3300cm-1 3330-3250cm-1 3350-3310cm-1
2Oamines
3O amines
N-H bend (1O only) N-H wag C-N str aromatic amines aliphatic amines
no bands
1650-1580cm-1 910-665cm-1 1335-1250cm-1 1250-1020cm-1
81
Eg: Aniline
82
NITRO GROUP
General structure: R-NO2
Wave number
N-O str (asymmetric)
1550-1475cm-1
N-O str(symmetric)
1360-1290cm-1
83
Eg: Nitromethane
84
85
Recent Advancements
86
Attenuated Total Reectance Spectroscopy ( ATR Spectroscopy)
Utilizes the phenomenon of Total Internal Reection Angle of incidents > critical angle The beam penetrates a fraction of a wavelength beyond the reecting surface.
87
when a material that selectively absorbs radiation, the beam
loses energy at the wavelength
where the material absorbs. The resultant attenuated radiation is measured. plotted as a function of wavelength and gives rise to the absorption spectral characteristics of the sample.
88
Applications
89
Quantitative analysis
Based on the determination of the concentration of one of the functional group of the compound being estimated. Concentration is determined using Beer- lamberts law A = I1/I0 = abc A c ;
a and b are constant.
90
Qualitative analysis
By comparison with reference spectra. Group frequency region: (3700-1500cm-1)
Hydrogen bonding region
By str vibration between hydrogen and other atom (O-H)
Triple bond region
Triple bond produce a peak in the region 2700-1850 cm-1
Double bond region
Compound having double bond- peak at 1850- 1500cm-1
Finger print region: (1500-700cm-1)
Identity of a compound Due to bending vibrations
91
92
Miscellaneous applications
Study of progress of reaction. Detection of impurities. Tautomerism. Functional group identification.
Presence of water in sample.
Industrial applications.
93
Conclusion
Infrared spectroscopy is certainly one of the most important analytical techniques available to todays scientists for the determination of identity and molecular structure of organic compounds
94
References
95
William Kemp; Organic Spectroscopy; 3rd edition; page no:20-94. Robert M Silverstein; Spectrometric Identification of Organic Compounds; 6th edition; page no:71-109. Skoog, Holler, Crouch; Instrumental Analysis; 5th edition; page no:477- 528. John R Dyer; Application of Absorption Spectroscopy; page no:23-33.
96
H.Kaur; Instrumental Method of Chemical Analysis; 6th edition; page no:181-184,192-195. P.S Kalsi Spectroscopy of organic compound; 6th
edition, page no 67,74.
97