0% found this document useful (0 votes)
6 views24 pages

Module 02 - HF Generator

The document provides an overview of the MAGNUM C HF X-ray generator, including its power supply requirements, UPS options, and key features such as a digital interface for radiography. It details the generator's components, voltage specifications, and exposure factors necessary for operation. Additionally, it includes practical activities for measuring line and phase voltages in a facility.

Uploaded by

Dheresa Ligaba
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPTX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
6 views24 pages

Module 02 - HF Generator

The document provides an overview of the MAGNUM C HF X-ray generator, including its power supply requirements, UPS options, and key features such as a digital interface for radiography. It details the generator's components, voltage specifications, and exposure factors necessary for operation. Additionally, it includes practical activities for measuring line and phase voltages in a facility.

Uploaded by

Dheresa Ligaba
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPTX, PDF, TXT or read online on Scribd
You are on page 1/ 24

MAGNUM C HF X-RAY GENERATOR

1
Power Supply

Power supply unit:


• An electrical device that supplies electric power to an
electrical load. It can be AC, DC, Single phase or 3
phase power

Single phase power supply:


• A single wire delivering current to load & a single
neutral wire caring current back to the source.

Three phase power source:


• A three wire ac power circuit with each phase 120
electrical degrees apart.

2
Distribution Transformer
∆-Y Connection:

Note:
 Line voltage is the potential difference between
two phases
 Phase voltage is the potential difference
3
Activity

• What is the line voltage in your facility ?


• What is the phase voltage in your facility ?
• What is neutral to ground voltage in your facility ?

• Use DVM to check

4
1 Phase & 3 Phase Voltages
Measured Values:

Star connection

5
UPS
Offline/Standby UPS:
• The most basic model designed for use in small,
non critical applications that require protection
against momentary power lose.

6
Cont…
Online / Double Conversion UPS:
• A robust solution for providing continuous & clean
power to critical equipment

7
Cont…

Advantage of UPS Disadvantage of UPS

• Smooth operation • Complex structure

• Backup batteries • Expensive in price

• Surge protection • Need periodic

• Stable power maintenance


• Need cooling system

8
MAGNUM C HF Generator

• MAGNUM C® is a 100 kHz high frequency x-ray


generator
• This x-ray generator can add a digital interface
for digital radiography equipment
• It consists main power cabinet and membrane
control console with digital interface
• The control console allows the operator to select
the technique factors, image receptors & to
initiate an x-ray exposure
9
Magnum HF Generator Power

Power Requirements (40 - 80kw)


 Line Voltage: 400 VAC ± 10%, 3P
 Line Frequency: 50/60 Hz.
 Nominal Current : ≤5 Amps.
 Momentary Current : 105 Amps
 Momentary Power : 100kVA.

10
Magnum HF Generator Features
• Integral low speed starter or optional dual-
speed starter on some models
• 24 VDC, 110, or 220 VAC power source for
Bucky (fused at 0.8 amps)
• 24 VAC 150 watts power source for collimator
lamp
• 24 VDC 45 watts power source for system
locks
• Optional AEC & DAP (Dose-Area Product)
11
Magnum HF Generator

12
Line Voltage Tapping
• Set the auxiliary transformer primary
voltage based on your nominal line voltage:

13
Generator Components

14
Cont…

HV Connector HV Cable & Socket

15
Cont...
Beam Limiting Device:

Halogen Lamp:

A light source for collimation


(Usually 24V, 150Watt)

Mirror:

A reflecting mirror is used to keep


the lamp out of the x ray path & to
have concentrated/focused light

Lead Shutters:

Used to control the area of light


field
16
Cont...

X-Ray Exposure Factors:


• Kilovolt (kv) 40 – 150kv
• Milliamperage (mA) 25-500 mA
• Exposure time (ms) 40 – 800 ms
• Source to image distance (SID)

Note:
 KVp will control the penetrating power/energy of the x ray
beam
 mAs will control the number of x rays & increase image
quality 17
Cont...
Tube control panel:
1. Control tube vertical
movement
2. Control tube angular
rotation
3. Control tube transversal
movement
4. Control tube longitudinal
movement
5. Tube rotation angle indicator
gauge
18
Interconnection Diagram

19
Cont...

20
Cont...

21
DC Bus & Power Distribution

22
System ‘On’

Generator Control Board

23
Question ?