0% found this document useful (0 votes)
981 views45 pages

02.physics of Mammography

Mammography is a specialized radiographic examination aimed at detecting breast pathology and cancer, utilizing low radiation doses and dedicated equipment to enhance image quality. Key features of breast cancer visible in mammograms include irregular masses, microcalcifications, and architectural distortions. The document discusses the importance of specific x-ray tube designs, breast compression, and anti-scatter grids in optimizing mammography for accurate diagnosis while minimizing patient exposure.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPT, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
981 views45 pages

02.physics of Mammography

Mammography is a specialized radiographic examination aimed at detecting breast pathology and cancer, utilizing low radiation doses and dedicated equipment to enhance image quality. Key features of breast cancer visible in mammograms include irregular masses, microcalcifications, and architectural distortions. The document discusses the importance of specific x-ray tube designs, breast compression, and anti-scatter grids in optimizing mammography for accurate diagnosis while minimizing patient exposure.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPT, PDF, TXT or read online on Scribd
You are on page 1/ 45

Physics of Mammography

References
The Essential Physics of Medical Imaging, Second Edition
JT. Bushberg et. al
1
Mammography

Radiographic examination specially designed for


detecting breast pathology and cancer

Low cost, low radiation dose procedure

Has the sensitivity for early detection and


improved treatment

2
Mammographic features of breast cancer

1. Masses, particularly with irregular or speculated


(strands of tissue radiating out from an ill-defined
mass, producing a stellate appearance) margins

2. Clusters of micro calcifications (specks of calcium


hydroxyapatite)

3. Architectural distortions of breast structures

3
Mammographic Features of Breast Cancer

4
Why general plain radiography equipment are not
suitable in mammography?

Why dedicated x-ray machines and x-ray


detectors are required in mammography?

To improve contrast and resolution,


To decrease the radiation dose to the patient

5
Mammography Facts

Breast is composed of fatty tissue, glandular tissue and a


50/50 combination of both
Attenuation differences between normal and cancerous
tissues is small. It is highest at very low x-ray energies (10
– 15 keV) and is poor at higher energies (>35 keV).

6
Mammography Facts

Low x-ray energies provide the best contrast


7
Mammography Challenges

Detection of micro calcifications is important because


high correlation of calcification patterns with disease

Low x-ray energies provide the best differential


attenuation between the tissues (photoelectric effect)

However high absorption results in a high breast dose


and long exposure time

8
Mammography
Detect minute calcifications
Minimize dose
Enhance low contrast detection

Imposes extreme requirements on mammographic


equipment and detectors

Led to the refinement of dedicated x-ray equipment,


specialize x-ray tubes, compression devices, anti
scatter grids, photo timers, and detector systems

9
Improvements in Mammography

Contrast and details are significantly improved


10
Screening Mammography
Craniocaudal (CC) and mediolateral oblique
(MLO) views of each breast are routinely acquired

11
Modern Mammography

Dedicated Mammography
equipment

Specialized x-ray tubes

Breast compression devices

Optimized screen/film
detector systems

12
Mammography Equipment

13
X-ray Tube Design

Cathode and Filament Circuit


Dual filaments in a focusing cup
0.3 mm (contact) and 0.1 mm (magnification) focal spot
sizes
Small focal spot minimizes geometric blurring and
maintains spatial resolution
Low operating voltage below 35 kVp

Tube currents
100 mA (± 25 mA) for large focal spot
25 mA (± 10 mA) for small focal spot
14
X-ray Tube Design

Molybdenum (Mo) and Rhodium (Rh) anodes (targets)


are used

Characteristic x-ray production is the major reason for


choosing Molybdenum and Rhodium

For Molybdenum at 17.5 and 19.6 keV

For Rhodium at 20.2 and 22.7 keV

Source to image distance (SID) of 65 cm is used


15
X-ray Tube Design

X-ray tube is tilted to


Minimize the effective focal spot size
Provide adequate field coverage
16
X-ray Tube Design

Heel effect – Lower x-ray intensity on the anode side


of the field due to attenuation through the target
To reduce the above and to get more uniform
exposure cathode-anode axis is placed from the chest
wall to the nipple
17
X-ray Tube Design

This orientation also minimizes equipment bulk near


the patient’s head for easy positioning

18
Tube Port

1 mm thick Beryllium used as the tube port

Beryllium provides both low attenuation and good


structural integrity

19
Beam Quality
Monoenergetic x-rays of 15 to 25 keV are optimal
choice to achieve high subject contrast at lowest
possible radiation dose

Compromises of polychromatic spectra


High energy x-rays in the bremsstrahlung spectrum
diminish subject contrast

Low energy x-rays in the bremsstrahlung spectrum


have inadequate penetration and contribute to patient
dose without providing a useful image

20
Beam Quality

Optimal x-ray energy is achieved by use of specific


target materials and filters to remove the low energy x-
rays

Added tube filters of the same element as the target


reduce the low-energy and high-energy x-rays in the
x-ray spectrum and allow transmission of
characteristic x-ray energies

Common target/filters in mammography Mo/Mo,


Mo/Rh, Rh/Rh
21
Molybdenum Spectra

22
Molybdenum Spectra

23
Mo/Mo Target Filter

Use for thin breasts

Provides 17-20 keV effective energy spectrum

24
Mo/Rh Target Filter

Common for imaging thicker and denser breasts


Produces slightly higher effective energy than
Mo/Mo.
Provides 20-23 keV effective energy leading to
increased penetration for thick and/or dense breasts
25
Rh/Rh Target Filter

Provides the highest effective energy beam (2 to 3


keV higher)

Useful for the thickest and densest breasts

26
Collimation
Fixed size metal apertures or variable field size
shutters are used to collimate the x-ray beam

Field size matches the 18×24 cm or 24×30 cm film


cassette sizes

Collimator light and mirror define the x-ray field

Useful x-ray field must extend to the chest wall


edge without field cutoff

27
X-ray Generator
Design and function are similar to a standard X-
ray generator
High frequency generators are the standard for
mammography

Differences
Lower power rating (3 kW)
Lower kVp (22 – 40 kVp)
Automatic Exposure Control (AEC) circuit

28
Automatic Exposure Control (AEC)
Called also as a photo timer
AEC sensor is located underneath the cassette in
mammography unlike conventional radiography

29
Automatic Exposure Control (AEC)

If the transmission of photons is insufficient to


trigger the AEC switch after an extended exposure
time, a backup timer is there to terminate the
exposure

For a retake, a higher kVp must be selected for


greater beam penetrability

30
Breast Compression

Breast compression is necessary


Reduces overlapping anatomy and decreases the
breast thickness
Less scatter, more contrast, less geometric
blurring of the anatomic structures, less motion
and lower radiation dose to the tissues
31
Breast Compression

Breast is compressed using a low attenuating


lexan paddle attached to a compression device

Flat, 90° paddle provides uniform density image

Principal drawback of compression is patient


discomfort 32
Spot Compression

Spot compression uses small paddles

Can get better compression over a small area


33
Scatter Radiation
Degrades subject contrast

d
Amount of scatter increases
with breast thickness and
breast area and is relatively
constant with kVp (25-35
kVp)

Without scatter rejection,


only 50 to 70% of the
inherent subject contrast will
be detected
34
Antiscatter Grid

Used to reject scatter

Placed between the breast and the image receptor

Linear grids with a grid ratio 4:1 to 5:1 are typical.


Some manufacturers use cellular grids

Higher grid ratios provide greater x-ray scatter


removal but also give higher dose
35
Anti-Scatter Grids
Aluminum and carbon fiber are typical interspace
materials.

Carbon fiber is preferred (lower Z) because


Aluminum attenuates too many of the low energy
x-rays used in mammography

Grid frequencies range from 30 to 50 lines/cm for


moving grids and up to 80 lines/cm for stationary
grids

36
Anti-Scatter Grids

Bucky factor is the ratio of exposure with the grid


compared to the exposure without the grid for the
same film optical density

For mammography, Bucky factor is about 2 to 3,


so breast dose is doubled or tripled, but image
contrast improves by 40%

37
Air Gaps
Use of an air gap between the
patient and the screen-film d
reduces the amount of scatter
detection

Grids not used in


magnification, air gap used

Reduction of the breast dose is


offset by the shorter focal spot
to skin distance

38
Magnification
Small focal spot is used
d
1.5× to 2.0× magnification is
used

Advantages
Reduction of scatter

Increased resolution of the


image by the magnification
factor
39
Mammography Screen Film Systems

40
Mammography Screen Film System
Cassettes are made of low attenuation carbon fiber

Have a single high definition phosphor screen

Used with a single emulsion film

Terbium activated gadolinium oxysulfide


(Gd2O2S:Tb) is the most commonly used screen
phosphor

41
Mammography Screen Film System

Screen is positioned in the back of the cassette

X-rays pass through the cassette cover and film


before interacting with the phosphor

Because of exponential attenuation, x-rays are more


likely to interact near the phosphor surface closer to
the film emulsion 42
Mammography Screen Film System

Reduces distance traveled by light

Minimizes the light spreading and preserve spatial


resolution

X-ray absorption at depth in the screen produces


broader light distribution and reduces resolution
43
Mammography Screen Film System
For 100 speed mammography screen film, limiting spatial
resolution is about 15 – 20 lp/mm.

Occasionally, extended time processing is used


– (reduces dose and increases contrast)

Screen and film are designed to be matched for optimal


image quality

Mixing of screens and films is not recommended because of


potential variations in speed and other characteristics

44
Thank You
Next lesson will be on equipment accessories used in
mammography examination

45

You might also like