Module 7 Calibration
Module 7 Calibration
•Hardware/DeviceConfiguration
•Actuator bench-set
Valve actuators provide force to move control valve trim. For precise positioning of a
control valve, there must be a calibrated relationship between applied force and valve
position.
Most pneumatic actuators exploit Hooke’s Law to translate applied air pressure to valve
stem position.
F=kx
Where,
F=Force applied to spring in Newton’s (metric) or pounds (English)
k=Constant of elasticity, or spring constant in Newton’s per meter (metric) or pounds per
foot (English)
x=Displacement of spring in meters (metric) or feet (English)
Hooke’s Law is a linear function, which means that spring motion will be linearly
related to applied force from the actuator element (piston or diaphragm). Since the
working area of a piston or diaphragm is constant, the relationship between actuating
fluid pressure and force will be a simple proportion (F=PA). By algebraic substitution,
we may alter Hooke’s Law to include pressure and area:
F= k x PA= k x
Solving for spring compression as a function of pressure, area, and spring constant:
x=PA/k
There are really only two mechanical adjustments that need to be made when
coupling a pneumatic diaphragm actuator to a sliding – stem valve: the stem
connector and the spring adjuster.
•The stem connector mechanically joins the sliding stems of both actuator and
valve body so they move together as one stem.
•This connector must be adjusted so neither the actuator nor the valve trim
prevents full travel of the valve trim:
•Once the stem length has been properly set by adjusting the stem connector, the
spring adjuster must be set for the proper bench set pressure. This is the pneumatic
signal pressure required to lift the plug of the seat. For an air-to-open control valve
with a 3 to15PSI signal range, the bench set pressure would be 3PSI.
Bench set is a very important parameter for a control valve because it establishes
the seating pressure of the plug when the valve is fully closed.
Consult the manufacturer’s instructions when adjusting the bench set pressure for
any sliding-stem control valve.
These instructions will typically guide you through both the stem connector and
the spring adjuster procedures, to ensure both parameters are correctly set.
•Split-ranging
There are many process control applications in industry where it is desirable to have
multiple control valves respond to the output of a common controller.
Control valves configured to follow the command of the same controller are said to be
split-ranged, or sequenced.
Split-ranged control valves may take different forms of sequencing.
A few different modes of control valve sequencing are commonly seen in industry:
complementary, exclusive, and progressive.
Complementary valve sequencing
The first is a mode where two valves serve to proportion a mixture of two fluid streams,
such as this example where base and pigment liquids are mixed together to form colored
paint:
Both base and pigment valves operate from the same 3to15PSI pneumatic signal output
by the I/P transducer (AY), but one of the valves is Air-To-Open while the other is Air-To-
Close.
Figure 1. 5Complementary valve sequence
•Exclusive valve sequencing
Other applications for split-ranged control valves call for a form of valve
sequencing where both valves are fully closed at a 50% controller output signal,
with one valve opening fully as the controller output drives toward 100% and the
other valve opening fully as the controller output goes to 0%.
The nature of this valve sequencing is to have an “either-or” throttled path for
process fluid.
That is, either process fluid flows through one valve or through the other, but never
through both at the same time.
A practical example of this form of split-ranging is in reagent feed to a pH
neutralization process, where the PH value of process liquid is brought closer to
neutral by the addition of either acid or caustic:
Figure 1. 7 Exclusive valve sequencing
•Progressive valve sequencing
A third form of control valve sequencing is used to expand the operating range of
flow control for some fluid beyond that which a single control valve could muster.
Once again pH control provides a suitable example to illustrate an application of this
form of sequencing.
1.Instrumentation and control devices and their standard
Instrumentation and control devices are used to measure and control process
variables in various industries.
These variables include pressure, temperature, humidity, flow, pH, force,
and speed.
The field of instrumentation and control engineering is interdisciplinary and
requires knowledge of chemistry, mechanics, electricity and magnetism,
electronics, microcontrollers and microprocessors, software languages,
process control, pneumatics and hydraulics principles, and communications.
The primary objective of instrumentation is to measure the process variables accurately.
Control devices are used to maintain the process variables at a desired set point.
Instruments are devices that measure or manipulate variables such as flow,
temperature, level, or pressure.
They can be classified into different types based on various criteria, such as
location (in-field or panel), power source (pneumatic or electronic), output signal
(analog or digital), or measurement parameter (pressure, temperature, flow, level,
etc.) .
Control devices are mechanical, electro-mechanical, or electronic devices that
use input signals to change conditions or values in processes or oversee access
to buildings, gated areas, etc. Controllers generally receive voltage inputs from
sources, analyze the inputs, and then oversee condition changes via signal
outputs.
Instrumentation and control engineering (ICE) is a branch of engineering that studies the
measurement and control of process variables using instruments and software tools. ICE
involves the design and implementation of systems that incorporate sensors, transmitters,
controllers, actuators, and displays
Recorders
Recording precipitation automatically has the advantage that it can provide better time
resolution than manual measurements, and it is possible to reduce the evaporation and
wetting losses. Three types of automatic precipitation recorders are in general use, namely
•The weighing recording type,
• The tilting or tipping-bucket type and
•The float types.
Only the weighing type is satisfactory for measuring all kinds of precipitation, the use of the
other two types being for the most part limited to the measurement of rainfall.
Annunciator associated with the installed devices
In industrial process control, an annunciator panel is a system to alert operators of alarm
conditions in the plant
Process switches
Another type of instrument commonly seen in measurement and control
systems is the process switches. The purpose of a switch is to turn on and o ff a
device like heaters, motors, valves etc… with varying process conditions
Signal generator:
A signal generatoris an electronic device that generates repeating or non-repeating
electronic signals in either the analog or the digital domain. It is generally used in
designing, testing, troubleshooting, and repairing electronic or electro acoustic
devices, though it often has artistic uses as well.
There are many different types of signal generators with different purposes and
applications and at varying levels of expense.
These types include function generators, RF and microwave signal generators, pitch
generators, arbitrary waveform generators, digital pattern generators and frequency
generators.
A function generator is a device which produces simple repetitive waveforms.
Oscilloscope
An oscilloscope is a laboratory instrument commonly used to display and analyze the
waveform of electronic signals. In effect, the device draws a graph of the instantaneous
signal voltage as a function of time.
Low Pressure Gauges: - Low pressure gauges operated by a capsule system and
only suitable for use on air and some gases. Pressure ranges from -2.5-0 kPa up to
0-60 kpa pressure or vacuum.
Test Gauge: - Precision test gauges manufactured to the highest standard in quality.
Used for the testing of industrial gauges or equipment of the same standard. Safety
Pattern Pressure Gauges: - These gauges are generally used within the gas industry
are designed with operator safety in mind in case of a bourdon tube rupture and
that no projectiles will blow out from the front of the gauge
UNIT 2. Calibrate Instrumentation and Control Devices
Data loggers
Electric meters
Multi-meters
Oscilloscopes
Frequency counters
Insulation Testers
Loop testers etc
Mechanical Calibration: Mechanical instruments are prone to drift as a result of
repeated use, mechanical stress, and exposure to fluctuating air conditions etc.,
The following are some of the most commonly tested mechanical calibration
instruments:
Accelerometers
Scales/Balances
Force Gauges & Load Cells
Micrometers, Vernier, and height gauges
Screwdrivers & Torque Wrenches
Sets of Weight and Mass
Flow Calibration: A flow meter (also known as a flow sensor) is a device that
measures the linear or non-linear, mass or volumetric flow rate of a liquid or gas.
The four most common types of flow meters that need calibration are:
Thermal Mass Flowmeters
Laminar flowmeters
Gas and Air Rotameters
Turbine meters.
Pressure calibration is a critical operation performed in a variety of industries where
measurement equipment is required to monitor process performance and safety,
with gas and hydraulic pressure being the most common measurements
The following are some examples of pressure devices that are
calibrated on a regular basis:
•Digital Pressure Gauges
•Digital Indicators
•Transducers
•Transmitters
•Analogue Pressure Gauges
•Barometers
Test Gauges .
Temperature calibration is undertaken and carried out in a controlled
environment in all processes where temperature readings play an important role
for the equipment’s to run without interruption.
The following are some examples of equipment that need temperature
calibration on a regular basis:
Data Acquisition Systems
Thermometers/Thermocouples
Dial Thermometers
Chambers/Furnaces
Infrared Meters
PRTs and Thermistors
Thermal Cameras
1.Calibration Procedures in Linear, Non-Linear and Discrete
Instruments
Calibration refers to the adjustment of an instrument so its output accurately
corresponds to its input throughout a specified range.
Procedures for efficiently calibrating different types of instruments
Linear instruments
The simplest calibration procedure for an analog, linear instrument is the so-called zero-and-
span method. The method is as follows:
Apply the lower-range value stimulus to the instrument, wait for it to stabilize
Move the “zero” adjustment until the instrument registers accurately at this point
Apply the upper-range value stimulus to the instrument, wait for it to stabilize
Move the “span” adjustment until the instrument registers accurately at this point
Repeat steps 1 through 4 as necessary to achieve good accuracy at both ends of the range.
Nonlinear instruments
The calibration of inherently nonlinear instruments is much more challenging than for linear
instruments. No longer are two adjustments (zero and span) sufficient, because more than
two points are necessary to define a curve.
Examples of nonlinear instruments include expanded-scale electrical meters, square root
characterizers, and position-characterized control valves.
Discrete instruments
The word “discrete” means individual or distinct. In engineering, a “discrete” variable or
measurement refers to a true-or-false condition. Thus, a discrete sensor is one that is only
able to indicate whether the measured variable is above or below a specified setpoint.
a.Calibration of instrument with their procedure
Owing to the physical limitations of measuring devices and the system under study, every
practical measurement will always have some errors. Several types of errors occur in a
measurement system. These include;
Static Errors:
They are caused by limitations of the measuring device or the physical laws governing its
behavior.
Dynamic Errors:
They are caused by the instrument not responding fast enough to follow the changes in
measured variable. A practical example can be seen in a situation where the room
thermometer does not show the correct temperature until several minutes after the
temperature has reached a steady value.
Random Errors:
These may be due to causes which cannot be readily established; could also be caused by
random variations in the system under study.
Basic Steps in Instrument Calibration
There are three basic steps involved in the calibration of an instrument. These include:
To collect measured values (Output values) of standard values (input values) provided by
a standard input reference.
To complete verification/calibration tables for upscale and down scale values (5 or 3
points)
To calculate the error on the output signal and to compare the result with the expected
accuracy.
Five Point Calibration Basics
In a five point calibration for an instrument, the output is measured at 0%, 25%, 50%,
75% & 100% of the calibration range of the instrument. In the five-point calibration
process, output readings are taken for upscale and down scale values of the calibration
range to determine the repeatability and hysteresis of the instrument. In a five-point
calibration, LRV = 0% Input; URV = 100%