Module 01 - Updated_v1[1]
Module 01 - Updated_v1[1]
Power
Supply
• Broad Definition
Device that receives and responds to a
stimulus
• Natural Systems
> Usually respond with signals, having electrochemical character
> Physical nature is based on ion transport
• Man-made Systems
> Information is transmitted and processed in electrical form
> Based on electron transport
• Direct Sensor
> Converts stimulus into an electrical signal
o Example: Thermopile
– Temperature to Voltage
• Complex Sensor
> Converts stimulus into a non-electrical signal before output
electrical signal
> Needs more than one transducer
o Example: Chemical Sensor
– Chemical reaction to Heat to Voltage
> Has at least one direct sensor
• Passive Sensor
> Does not need any additional energy source
> Directly generates electric signal in response
to external stimulus
> Example: Thermocouple, Photodiode, Piezoelectric sensor
• Active Sensor
> Require external source for operation, usually called excitation
signal
> Sometimes called parametric, because the sensor’s own properties
change in response to the external effect
> Example: Thermistor, Resistive strain gauge
• Absolute Sensor
> Detects stimulus in reference to absolute scale
irrespective of measuring conditions
> Example: Thermistor
o Resistivity related to absolute temperature scale (K)
• Relative Sensor
> Produces signal that relates to some special case
> Example: Thermocouple
o Output voltage function of temperature gradient
• Transducers:
Convert one type of energy into another
• Example: Microphone
> Acoustic waves Variable Magnetic Field Electrical energy
• Transducer has nothing to do with perception or sensing
• Types of Characteristics
1.Static Characteristics
2.Dynamic Characteristics
3.Environmental Characteristics
• Transfer function
• Full scale input
• Full scale output
• Non-linearity
• Accuracy
• Calibration
• Hysteresis
• Threshold, Saturation, Offset, Deadband
• Repeatability, Reliability
• Resolution
• Impedance
• Excitation
• Sensitivity:
Transfer
> Linear functions, function:
> Non-linear functions, Y=1.043 +
0.020 X
Regression =
0.998
• In many cases, non-linear function is considered linear over a
limited range and the sensor is approximated as a linear
function over that range. This is called piecewise
approximation
• NOTE: Decibels do not measure absolute values, but ratio of values only
a
a l Dat Terminal Points
Re
• Example:
> A piezoelectric pressure sensor with 100 kPa F.S.I, and 10 Ω F.S.O.
> If the sensor provides an output of 10.05 Ω for 100 kPa input, then
inaccuracy can be
o ± 500 Pa - In terms of measured value
o ± 0.5% - In percent of input span
o ± 0.05 Ω - In terms of output signal
• Systematic Errors:
> Interfering or modifying variables (eg. Temperature)
> Drift (eg. Changes in chemical structure or mechanical stress)
> Measurement process changes the measurand (eg. Loading effects)
> Transmission process changes the signal (eg. Attenuation)
> Human observations (eg. Parallax error)
• Ways to avoid Systematic Errors:
> Triangulation
> Regular calibration
• Random Errors:
> Environmental noise (eg. Background noise picked by microphone)
> Transmission noise (eg. 60Hz hum)
• Ways to avoid Systematic Errors:
> Taking repeated measurements
> Increasing sample size
• How to calibrate:
> Transfer function of the sensor needs to be found during calibration
> Apply inverse transfer function to compute applied input value from
the measured output
• Example:
> Consider a sensor that will measure pressure (P) and produce an
voltage (V) with a transfer function: ;
> To compute a and b, we need at least two values:
and
> So, and
> From this, input pressure and be identified from output voltage
using inverse transfer function:
• Error calculation:
> Two stimuli S1 and S2 applied, output is A1 and A2
> At S1 – no error, S2 – error of –Δ
> New intercept a1 is calculated by
• Zero Drift
Zero Drift
> Calibration gradually shifts due to
slipping, zero setting, etc.
• Span/Sensitivity Drift
> Proportional change in the transfer function
Span/Sensitivity Drift
• Example:
> When object moves from left to right, a displacement sensor
produces a voltage difference of 20 mV from that when the object
moves from right to left
> Assuming sensitivity of sensor is 10 mV/mm,
hysteresis error is 2 mm (in terms of measured value).
• Saturation:
> Operating limit after which the sensor will not
d
respond to change in input stimuli (Maximum)
ol
n
ti o
sh
• Threshold:
re
ra
Th
tu
> Operating limit until which the sensor will not
Sa
respond to change in input stimuli (Minimum) Offset
• Offset:
> Measured output, when no stimulus is applied
> Also, considered baseline in most measurements
• Dead Band:
> Region where the output may remain at
certain value without any change in
measurement
• Associated parameter
> Reproducibility - Ability to produce same
output under different conditions
• Reliability Indices
> Failure Rate (λ)
> Mean Time to Failure (MTTF)
> Mean Time Between Failure (MTBF)
• Bathtub Curve
> Tool used to model the reliability
of a unit (or system) over the
units entire life
• Early Failures:
> High number of failures
> Due to errors in design
or manufacturing
> Decreasing over time, as products
with defects are removed from the population
• Useful lifetime:
> Lowest number of failures
> Due to random effects like
external stress, mishandling
of product, etc.
>
• Wear-out failures
> Increasing number of failures
> Aging of components
> Due to corrosion, oxidation or fatigue
• Infinitesimal resolution:
> Continuous resolution
> Condition when there are
no measurable steps
in the output signal
• Example:
> Consider a tire pressure monitoring system:
> Case1: If pressure drops from 30psi to 29.9psi, and the sensor can
accurately measure and display this change, then the resolution of
the sensor is 0.1psi (Higher resolution)
> Case2: If there is a pressure drop from 30psi to 29.9psi and the
sensor can not measure this change. Instead, the output changes
only when the pressure drops to 29psi, then the resolution is 1psi.
(Lower resolution)
1.Input Impedance
> To measure, a fixed resistor is used and voltage V1 and V2
are measured. So,
> Input impedance of CUT is
2.Output Impedance
> To measure, a fixed load resistor (RL) is used and output
voltage is measured first with full load (VL), then without load
(V)
> Output impedance of CUT is
• Example:
> Maximum current through thermistor
in still air is 50 μA
in water is 200 μA
• Response time
• Measuring lag
• Fidelity
• Dynamic error
• Transient Response
• Speed of Response
> Speed with which sensor responds to a change in stimulus
• Response time
> Time taken by the sensor to reach final stable value (steady-
state) from zero, once stimulus is applied
> Depending on application, stable value
can be a specified percentage of
true value
> Eg: For portable instruments, time taken to
reach within 0.3% of final value, while for
switch board instruments, it can be
within 1% of final value
Sensor Value
• Two types of Lag
Measuring Lag
1.Retardation type
> Response begins immediately after
a change in stimulus has occurred
Dead Time
2.Time Delay
> Response begins after a dead-zone
after a change in stimulus has occurred
Sensor Value
100%
50%
Delay time
0%
100%
90%
50%
10%
0%
Rise time
Mp
100%
50%
Peak time
0%
100%
98%
50%
Settling time
0%
• Overshoot (Mp)
> Maximum peak deviation from the final or steady-state value
100% Overshoot
50%
0%
• Stability
• Environmental conditions
• Temperature effects
• Self-heating
• Storage Conditions
> Nonoperating environmental limits to which a sensor may be
subjected during a specified period without permanently altering the
performance Electrode