Biosensors and Applications-1
Biosensors and Applications-1
• Yang, V.C. and T.T. Ngo. 2000. Biosensors and Their Applications. Kluwer
Academic/Plenum Publishers, New York, NY. Ligler, F.S. and Rowe Taitt, C.A.
2002.
Voltage
Current
Charge
Amplitude
Frequency
Phase
Digital code
Digital voltmeter LED 3 digit display
Analog signals
Digital signals
Example for How the BP monitoring instrument developed
Electronic Sphygmomanometer
Aneroid Sphygmomanometer
Pressure gauge:
that transforms the signal resulting from the interaction of the analyte with the
biological responsible for the display of the results in a user-friendly way.
Physical or Electronic
Transducers
Chemical changes signals
Selective ion
Ion Field effect transistor determination
Voltage
Mechanical (Potentiometric)
Voltage
Thermal (Potentiometric)
Peizoelectric Pressure sensors
Classification of biosensors based on the transducer element & bio-recognition
element
Transducer Microprocessor
Filter
Amplifier
Digital
Analog
• label-based sensing involves the use of a label and the optical signal is
then generated by a colorimetric, fluorescent or luminescent method.
• which in turn is dependent on the concentration gradient and hence the bulk
oxygen concentration
Sensors are devices that register a physical, chemical, or biological change and convert that
into a measurable signal.
The sensor contains a recognition element that enables the selective response to a particular
analyte or a group of analytes, thus minimizing interferences from other sample components
.
Another main component of a sensor is the transducer or the detector device that produces a
signal. A signal processor collects, amplifies, and displays the signal.
• Electrochemical biosensors, a subclass of chemical sensors, combine the sensitivity,
tissues or receptors) that selectively reacts with the target analyte and produces an
electrical signal that is related to the concentration of the analyte being studied.
Electrochemical biosensors can be divided into two main categories based on the
nature of the biological recognition process
1. Biocatalytic devices
Biocatalytic devices incorporate enzymes, whole cells or tissue slices that
recognize the target analyte and produce electroactive species. Special emphasis
will be placed on enzyme electrodes for the detection of glucose, lactose, and
xanthine.
2. Affinity sensors:
Affinity sensors rely on a selective binding interaction between the analyte and a
biological component such as an antibody, nucleic acid, or a receptor.
Biocatalytic sensors:
• Such sensors are typically easy to use, compact, and inexpensive devices.
vs. flow conditions or bulk sample solution vs. a micro drop detected using a
microelectrode.
Enzymes, globular proteins composed mainly of the 20 naturally occurring amino acids
that catalyze biochemical reactions, are the oldest and still most commonly used
specificity for their substrate molecule and can detect individual substances in a
Researchers also had to find ways to manage the enzyme adsorption that could lead
The enzyme is the most critical component of the enzyme electrode since it provides
the selectivity for the sensor and catalyzes the formation of the electroactive product
for detection.
The electroactive product can be monitored directly using amperometry, in which the
produced current is measured in response to an applied, constant voltage.
The activity of the immobilized enzyme depends on solution parameters and electrode
design. The rapid enzymatic catalysis can also sometimes provide significant signal
amplification in a biosensor.
• Development of biocatalytic sensors for medical applications, primarily blood glucose monitoring
starting in the late 1960s, was the main driving force for this research area.
• Electrodes coated with glucose oxidase (GOx) have been widely used in detection of glucose
• These amperometric sensors became known as the first-generation biosensors or Clark oxygen
electrodes and were soon implemented by Updike and Hicks, who constructed the first functional
biocatalytic sensor for glucose.
that is among the most important enzymes in biosensor applications and industrial
processes. GOx is highly specific for β-D-glucose, which can be detected via the
following reactions.
In eqn (1) the prosthetic group of the enzyme, FAD, is reduced and glucose is
oxidized to d-D-gluconolactone.
. During the oxidation of H2O2 at a working electrode two electrons are transferred
directly to the electrode, resulting in the current response of the enzyme electrode.
These first-generation sensors required the ample and constant presence of ambient
oxygen as a co-substrate for the enzyme to function optimally.
However, oxygen is not very soluble in aqueous solutions and can therefore limit the
currents produced in the presence of the analyte.
• Direct redox reactions between enzymes and electrodes are very rare because most proteins tend to
denature at the electrode surface and many direct electron transfer reactions are slow and irreversible.
• However, a limited number of enzymes such as horseradish peroxidase have proven capable of direct
electron transfer between the enzyme active site’s prosthetic group and the electrode.
• The active site of an enzyme that allows the selective targeting of an analyte is usually buried within
the enzyme’s tertiary protein structure, near the centroid of the protein.
• Therefore, the electrons produced in the enzyme-catalyzed reaction cannot always be easily and
rapidly transferred to the electrode surface thereby limiting the electrical communication between the
enzyme and the transducer.
• The widely accepted Marcus theory of electron transfer states that electron transfer decays
exponentially with distance.
• Therefore enzymes often require some assistance with electron transfer to the transducer surface.
• Artificial redox mediators are small, soluble molecules capable of undergoing rapid and reversible
redox reactions, which shuttle electrons between the redox center at the active site of the enzyme and
the electrode surface.
• Mediators are re-oxidized at relatively low potentials and generate a current when they come in
contact with the working electrode
• Mediators should ideally be nontoxic, independent of the pH, stable in both the oxidized and
reduced forms, and unreactive with oxygen.
• Although many organic compounds are capable of acting as enzyme mediators, organometallic
redox compounds are the most common.
• Examples of previously used mediators include quinones, organic conducting salts, dyes,
ruthenium complexes, ferrocene, and ferricyanide derivatives.
• Mediated enzyme electrodes had a much better sensor performance than the first-generation
biosensors mainly due to eliminating the O2 dependence and being able to control the
concentration of the oxidizing agent in the biosensor.
• Hand selecting the oxidizing agent for the sensor also allowed more suitable oxidation potentials
to be used for the amperometric sensors. These mediated enzyme electrodes were named second-
generation biosensors.
• Third-generation biosensors have the biorecognition component coupled with the electrode by co-
immobilizing the enzyme and the mediator at an electrode surface.
• This can be achieved by direct electrical contact between the enzyme and the electrode,
immobilizing the enzyme and mediator in a conducting polymer, or ‘wiring’ the enzyme to the
electrode by immobilizing it in a redox polymer.
• The co-immobilization prevents the mediators from diffusing out of the biosensor film.
• The co-immobilized mediators, or the flexible surrounding redox polymer, help to transport
electrons between the enzyme’s active site and the working electrode surface in an array of rapid
electron relays and hence generate high current densities.
• The enzymes immobilized in flexible redox polymers that are covalently attached to the electrode
have been called ‘wired enzymes’.
• The 3rd-generation sensors are ideal for repeated measurements since neither mediator nor enzyme
need to be added.
• This self-contained nature also lowers the cost per measurement and opens up possibilities for
continuously monitoring the analytes.
Voltammetric and amperometric techniques are characterized by applying a potential to a working (or indicator) electrode
versus a reference electrode and measuring the current. The current is a result of electrolysis by means of an
electrochemical reduction or oxidation at the working electrode. The electrolysis current is limited by the mass
The term voltammetry is used for those techniques in which the potential is scanned over a set potential range. The current
response is usually a peak or a plateau that is proportional to the concentration of analyte. Voltammetric methods include
linear sweep voltammetry, cyclic voltammetry, hydrodynamic voltammetry, differential pulse voltammetry, square-wave
voltammetry, ac voltammetry, polarography, and stripping voltammetry. These methods have a wide dynamic range, and
The technique is implemented by stepping the potential directly to the desired value and
then measuring the current, or holding the potential at the desired value and flowing
samples across the electrode as in flow injection analysis.
The frequency is varied over a wide range to obtain the impedance spectrum.
The in-phase and out-of-phase current responses are then determined to obtain the
resistive and capacitive components of impedance, respectively.
Impedance methods are powerful because they are capable of sampling electron
transfer at high frequency and
mass transfer at low frequency. Impedimetric detection is primarily used for affinity
biosensors.
Impedance biosensors
Affinity biosensors
Most applications of immunoassays (IA) and immunosensors with electrochemical detection were
initially developed at research laboratories due to the level of expertise required, time, and the high initial
However, the cost of immunological reagents continues to decrease with recent developments in
Many of the early radio immunoassays were developed for biomedical applications such as detecting
hormones and disease related proteins, but applications in environmental, agricultural, processed food and
beverage areas, and to detect harmful chemical and biological agents in national defense.
Biorecognition and immunochemical reactions. IgG antibodies (Ab), large Y-shaped
glycoproteins of MW E
150 kDa, are produced by a host in response to the presence of a foreign molecule
called antigen (Ag).
Antigens are anything that the body recognizes as foreign such as chemical
compounds, proteins, and particulate matter (dust, pollen, etc.).
Abs are produced by specialized B lymphocyte cells of the immune system and can
usually be found in blood
serum, tissue fluids, and membranes of vertebrates.
The production of Abs against low molecular weight analytes (MW o 1000 g mol1)
called haptens is more challenging and often requires coupling the hapten to a
carrier protein with a spacer molecule before an immune response can be provoked
in the host animal.
The four most commonly used types of labeled immunoassays are:
Enzyme Immunoassay (EIA) or Enzyme-linked immunosorbent assay
(ELISA):
• used to visualize and quantify antigens.
• They use an antibody conjugated to an enzyme to bind the antigen, and the
enzyme converts a substrate into an observable end product. The substrate
may be either a chromogen or a fluorogen.
Radioimmunoassay (RIA)
• an immunoassay that uses radiolabeled molecules in a stepwise formation
of immune complexes.
Fluoro immnoassay (FIA)
• In the ultraviolet spectrum. After incubation with the antigen, the antibody-antigen
complexes are isolated, and the fluorescent intensity is measured.
• The intensity read by this type of immunoassay is linearly correlated to the amount
of antigen present.
Chemiluminescence immunoassay (CLIA)
• which is the emission of light based on a particular chemical reaction, i.e the
formation of the antibody-antigen binding complex.
PNPP- p-Nitrophenyl Phosphate, NBT : nitro blue tetrazolium/ 5-Bromo-4-chloro-3-
indolyl phosphate , ABTS: 2,2′-azino-bis(3ethylbenzothiazoline-6-sulfonic acid), TMB :
3,3′,5,5′-Tetramethylbenzidine.
Immunoblots
Immunostaining uses antibodies to detect an antigen in cells
or tissue. The major benefit of immunostaining in
immunohistochemistry (IHC) is the ability to see the desired
target in a tissue sample while maintaining the spatial
context and tissue architecture.
Phalloidin: coupled with fluorophore
• ELISA assay uses the coupling of antigens and antibodies and relies on the
specificity and affinity of antibodies for antigens.
• label-based sensing involves the use of a label and the optical signal is
then generated by a colorimetric, fluorescent or luminescent method.
IO I
Thermal light sources
Lasers
• that this shift of the wavelength of maximum spectral power density, λmax , is
inversely proportional to the absolute temperature of the glowing body,
Incandescent lamps
10 lm/W.
Life time =1000 h
Higher filament temperatures
filaments are made of tungsten as well, but the bulb is filled with iodine or
bromine vapor.
• Sodium (Na) lamps emit yellow light because of the Na D lines, a narrow
-spaced line doublet at 589.0 nm (D2 ) and 589.6 nm (D1 ).
• During power - up, however, they emit red light due to the neon added to
the gas in the bulb.
• Mercury has a strong emission line at 253.7 nm. Mercury tubes, also called
backlight tubes, emit ultraviolet ( UV ) light.
• sharp energy levels become disturbed so far that they form quasicontinuous energy bands.
• bands are still separated by forbidden zones, so - called gaps
• electronic transition from the lower edge of the conduction band to the
upper edge of the valence band will lead to the emission of a light
quantum.
• energy will equal the width of the energy gap, E G :
• h ⋅ν = EG
• Organic Light Emitting Diode s ( OLED s) are based on recombination processes
in thin organic layers with thicknesses between 80 nm and 200 nm.
OLEDs may reach lumi-nous effi cacies of more than 100 lm/W
Laser:
• laser light is strongly concentrated in one beam that hardly changes its diameter over long distances, a
radiant intensity of, for example, 1 mW and a beam diameter of 3 mm will amount to a light intensity (the
radiant power per unit surface) of I ≈ 300 W/m 2 .
Photo detectors
• evaluation of the information that the received light carries can start, it has to be transformed into electrical
signals.
• which light incident onto a surface produces a current or a voltage are therefore processes around which photo
detectors can be designed.
E kin = h ⋅ν −W
• As the current amplification or gain, g , can reach 107 for a supply voltage of around 2 kV, PMTs must not be
illuminated with high intensities.
• This would result in high-current discharges between the dynodes and to a
destruction of the PMT.
Dark current, Id , is caused by:
• for example, thermionic emission from the electrodes, field emission and by radioactivity from the
environment, cosmic rays, or from the materials used in the PMT.
• This dark current is usually in the range of 1 pA and increases with increasing supply voltage.
• I noise = (2e Id g Δf) 1/2
Photodiodes
• utilize the effect of photons onto the charge carriers in the depletion zone in the pn - junction of a semiconductor
diode:
• when photons are absorbed, they create electron – hole pairs.
• As the charge carriers remain inside the material, this is called the internal photoeffect.
• An applied reverse voltage forces the charge carriers to drift toward the external electrodes
• This mechanism can be understood as an inversion of the working principle of LEDs and diode lasers.
• that enable the charge carriers to travel faster across the junction provide higher reverse voltages.
• One of these architectures is a sequence of a p - doped, an undoped (intrinsic), and an n - doped semiconductor
material, a so - called PIN diode.
• proportional to the number of incident photons, that is, to the incident radiant
power, PL . With the reverse current, IS , with UT = kT / e , and with the spectral
sensitivity Sλ , it is
Photo resistors
• The resulting free electrons (and their hole partners) conduct electricity, thereby
lowering resistance.
• The resistance range and sensitivity of a photo-resistor can substantially differ among
dissimilar devices.
• Ge:Cu photoconductors are among the best far-infrared detectors available, and are used for infrared
astronomy spectroscopy and infrared spectroscopy.
Optical biosensors use the interaction of light (electromagnetic radiation/waves) with matter to measure the
concentration of a species.
The interaction can be measured through adsorption, scattering (absorbing and re-emitting the light), refractive index
changes and fluorescence.
Optical sensors have already been established as a staple of the healthcare industry, ranging from traditional sensors such
as pulse oximeters to modern photoplethysmograms for heart rate monitoring in wearable devices such as the Apple
Watch and Fitbit.
• pathogen detection
• cancer diagnosis
• environmental monitoring
• DNA sensing
This can be organic such as enzymes, antibodies, aptamers, cells, or tissues (Chen
and Wang, 2020), or inorganic such as molecularly imprinted polymers (MIPs)
Interactions between the bio-recognition element and the analyte lead to a signal
through interactions with light, which can originate from a light source such as a
laser or LED.
These interactions include fluorescence, absorption, refraction index or light
scattering.
Finally, a signal processing unit converts the measured signal into readable
information for the users, which in this case is the concentration of the analyte.
Optical biosensors can be broadly classified as label-free (detection signal is generated
from the interaction of the analyte with the transducer) or label-based (uses a label, and the
optical signal is generated by a colorimetric or luminescent method).
To improve the measured signal, other optical elements, such as lenses for focusing the
beam or filters for removing noise, may also be included.
Some of the prominent technologies that have dominated the field of optical biosensors
include (1) spectrophotometry, (2) fluorimetry, (3) Surface Enhanced Raman Spectroscopy
(SERS), (4) chemiluminescence, and (5) Surface Plasmon Resonance (SPR)
• Spectrophotometry measures the absorption of light as it passes through a solution. Spectrophotometry
• By using the Beer-Lambert law, which states that the absorbance is directly proportional to the concentration of the
analyte (Ball, 2006), the concentration of analyte in a solution can be quantified.
• These include visible spectrometry (also known as colorimetry), UV spectrometry, and infra-red spectrometry .
• Traditionally, colorimetry has been used for qualitative purposes, e.g. in lateral flow immunoassays as
aforementioned.
• However, by measuring the intensity of visible light at a specific wavelength, quantitative applications are also
possible.
• While drug molecules may not inherently absorb nor emit light in the visible spectrum, colorimetric approaches are
made possible by conjugating signal transducers (e.g. gold or silver nanoparticles) to biorecognition elements.
Various research groups have modified the surface of noble metal, such as gold and silver,
nanoparticles with small molecules that interact favourably to the target analyte.
For instance, amidosulfonic acid-capped silver nanoparticles (ASA-AgNPs) were used for
the visible spectrophotometric detection of lamotrigine, an antiepileptic drug.
The amine groups of amidosulfonic acid interact with lamotrigine through cooperative
hydrogen bonding.
These modified AgNPs are normally dispersed and give a yellow color, but in the presence
of lamotrigine, the ASA-AgNPs aggregate and turn red.
The color change was monitored by measuring the ratio of absorbance at 450 nm–390 nm
after 30 min of incubation of the ASA-AgNPs and sample using a UV-2550
spectrophotometer.
Consequently, biologically derive constructs, such as antibodies, enzymes, and aptamers, have been explored as highly
selective and sensitive bioreceptors.
For example, gold nanoparticles (AuNPs) were employed in conjunction with digoxin aptamers for the quantification of
digoxin.
This causes the color of the solution to turn blue. Here, the
absorbance at 520 nm was measured using a Synergy H4
microplate reader.
Fluorimetry
Fluorescence occurs when a molecule absorbs electromagnetic radiation and re-emits
light at a different wavelength; the wavelength of the re-emitted light is characteristic of
the molecule.
The intensity of the re-emitted light, measured using a fluorimeter or plate reader, is
subsequently used to determine the concentration of the molecule.
Like visible spectroscopy, most drug molecules are not fluorophores and cannot be
inherently measured through fluorimetry.
Other systems may pair the fluorophore with a quenching agent, which is a substance
that reduces the fluorescence intensity of the fluorophore.
Fluorescent energy meter:
Other systems may pair the fluorophore with a quenching agent, which is a substance that reduces the
fluorescence intensity of the fluorophore.
In these, the distance between the fluorophore and the quenching agent is altered as a result of the biorecognition
event. This can occur in several ways, such as a conformational change in the biorecognition element due to
analyte binding, or displacement of the fluorophore/quenching agent by the analyte.
The concentration of the analyte is directly or indirectly proportional to the fluorescence intensity depending on
the system’s native state, i.e. the former if the biorecognition event causes the distance between the fluorophore
and quenching agent to increase.
transformations.
• total heat evolution is proportional to the molar enthalpy change and to the
• resulting temperature change (ΔT) is dependent on the total heat capacity (CS) of
the system.
• the sum of all the reaction enthalpies that determines the sensitivity of
the assay
• highly exothermic reaction doubles the sensitivity and reduces the deleterious effects
that hydrogen peroxide has on enzyme activity.
• gain in enthalpy change by selecting the most suitable buffer can be as much as
50 kJ mol−1.
• Small amounts of organic solvents (around 5%, v/v) present in the aqueous
buffer increases the temperature (double times the registered temperature
changes by increasing the total enthalpy change in the reaction.
Immuno-biosensors Based on Thermistors
• total heat evolution is proportional to the molar enthalpy change and to the total
number of moles of product molecules created in the reaction.
where ,
Q – total heat,
nP – moles of product,
H – molar enthalpy change.
capacity (CS) of the system including the heat capacity of the solvent.
Bio-recognition elements in thermistors
• several simple, inexpensive devices for the determination of glucose, urea and
other biological compounds.
• Choice of coupling technique - the properties of the support material and the
• Pores should be large enough for the enzyme to be immobilised within the support
matrix.
• for the substrate to easily access the catalytic site of the surface-bound enzyme.
• such as functional groups available for immobilisation and the size of the enzyme
• can therefore be useful for obtaining the maximum activity of the bound enzyme within
the support matrix.
required to obtain and maintain the maximum activity of the support-bound enzyme.
Alternative methods:
• were less specific than the pure enzyme and responded to a number of interfering
compounds that are usually present in food samples.
Physical and Chemical property of the support: on which the enzyme is immobilised
has a crucial role for retaining the catalytic activity of the enzyme.
Choosing the support material some basic criteria should be fulfilled:
(1) the support must possess mechanical stability high enough to withstand physical stress
the sample that would disturb the measurements of the enzymatic reaction .
• controlled pore glass (CPG) which is available is different pore and particle size ranges
• CPG offers high enzyme-binding capacity, good mechanical, chemical and microbial
• the bio-recognition element (such as enzyme) can be covalently + nylon and glut
ethyleneimine.
• crushed into smaller pieces in order to increase surface area and then coated with
chemically polymerised pyrrole for better electrical conductivity. But the binding
capacity is very less.
• enzymes covalently attached to amine-CPG with a pore size in the range 500–
2000˚A and particle size around 80 mesh (0.18 mm) activated with
glutaraldehyde.
• Schiff’s base formation between the aldehyde group and the amine group of the
enzyme
Cell immobilisation
• small dialysis/filtration unit was included into the construction to remove blood
cells.
• allows a large number of whole blood samples to be injected without any sign
of clogging.
• 14 mm ×6 mm ×0.4 mm silicon
chip
mixing
• Using the combined setup the mixture of metals such as arsenic and
organophosphate pesticides
Surface Plasma Resonance
It can determine the kinetic on and off rates for the interaction of a biomolecule with
a ligand in real time.
The change of the incident angle required for SPR is defined as SPR response in the unit of response unit
(RU).
The main objectives of SPR are:
i) to quantify proteins and other biological molecules at surfaces.
vi) to interface with mass spectrometers provides discovery-based research in proteomic studies
Utility of Biacore systems:
Biacore systems are used primarily in pharmaceutical development, quality control and basic life science
research.
Measurements with Biacore systems provide valuable information in a number of application areas:
• To screen the binding partners and ability for binding, for instance in drug discovery and
biopharmaceutical development.
• To determine the interaction capabilities, for example epitope mapping with monoclonal antibodies
• To measure the interaction kinetics and affinity, made possible by realtime detection of interaction events
. Biacore Processing unit contains following components:
Figure 2
i) Two pump
ii) Syringe Needles
iii) Integrated µ-Fluidic Cartridge (IFC),
iv) Optical detector for measuring SPR response.
v) Autosampler
vi) Microprocessor
vii) Two racks for samples for holding samples in vials
viii) Sensor chips holder
Pump: One pump (Eluent) transport running buffer over sensor chip for maintain constant flow
of liquid and another pump (Autosampler) for injecting samples in the auto sampler
One needle is connected to the pump for constant flow of
Types of biosensors
• Piezoelectric biosensors
• Cantilever biosensors
Piezoelectric effect
• Aluminum phosphate (berlinite)
• Aluminum nitride
• Zinc Oxide
• Gallium orthophosphate
but also other properties like viscosity of ambient solution has impact on the
frequency shift.
• Optimization of desired characteristics of biosensors
• sensitivity,
• selectivity,
• stability,
• detection limit,
• reliability,
• safety,
• simplicity
01/23/2025 Department of Biotechnology, AIT, Biosensor and Applications- 21BT742 168
• cost, and parameters like operating conditions,
calibration, positive and negative controls.
Sensitivity
The minimum amount of analyte that can be detected by a biosensor defines its
limit of detection (LOD) or sensitivity.
Click to Edit
Click to Edit • A solution (usually a buffer containing salts) containing the antigen is
then exposed to the transducer where antibodies interact only with the
antigens.
These disturbances can cause a drift in the output signals of a biosensor under
measurement.
This can cause an error in the measured concentration and can affect the precision and
accuracy of the biosensor.
Stability is the most crucial feature in applications where a biosensor requires long
incubation steps or continuous monitoring.
Click to Edit The response of transducers and electronics can be temperature-sensitive, which may
influence the stability of a biosensor.
Another factor that can influence the stability is the affinity of the bioreceptor, which is
the degree to which the analyte binds to the bioreceptor.
01/23/2025 Department of Biotechnology, AIT, B-107 Plant Physiology an 173
d Phytohormones-BBT306D
Bioreceptors with high affinities encourage either strong electrostatic
bonding or covalent linkage of the analyte that fortifies the stability of
a biosensor.
Click to Edit
Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ) are terms used to
describe the smallest concentration of a measurand that can be reliably measured by an analytical
procedure.
LoB is the highest apparent analyte concentration expected to be found when replicates of a blank
sample containing no analyte are tested.
LoD is the lowest analyte concentration likely to be reliably distinguished from the LoB and at which
detection is feasible. LoD is determined by utilising both the measured LoB and test replicates of a
Click to Edit sample known to contain a low concentration of analyte.
LoQ is the lowest concentration at which the analyte can not only be reliably detected but at which
some predefined goals for bias and imprecision are met. The LoQ may be equivalent to the LoD or it
could be at a much higher concentration.
Sensor reliability is the likelihood that a sensor will function as intended without failure or
degradation over a specific time period and under certain conditions. It's influenced by
many factors, including:
Sensor design: The design of the sensor should consider the required strength.
Manufacturing quality: The quality of manufacturing can impact the reliability of a sensor.
Operational stress: The operational stress on the sensor can impact its reliability.
Response time is the time taken by the sensor device to respond to a step
concentration change from initial to a definite value of concentration.
Sensor type
The type of sensor affects its response time. For example, a pressure sensor's response
time depends on the media being measured, such as air, water, or oil.
Environmental conditions
The environmental conditions, such as still air, motion, still liquid, or type of liquid, affect the
Click to Edit
sensor's response time.
Precision is the ability of the sensor to provide alike results every time a sample is
measured and accuracy indicates the sensor's capacity to provide a mean value close
to the true value when a sample is measured more than once.
Click to Edit
Reproducible signals provide high reliability and robustness to the inference made on
the response of a biosensor.
Linearity is the attribute that shows the accuracy of the measured response (for a set
of measurements with different concentrations of analyte) to a straight line,
mathematically represented as y=mc, where c is the concentration of the analyte, y is
the output signal, and m is the sensitivity of the biosensor.
Linearity of the biosensor can be associated with the resolution of the biosensor and
range of analyte concentrations under test.
The resolution of the biosensor is defined as the smallest change in the concentration
of an analyte that is required to bring a change in the response of the biosensor.
Click to Edit
Depending on the application, a good resolution is required as most biosensor
applications require not only analyte detection but also measurement of
concentrations of analyte over a wide working range.
Another term associated with linearity is linear range, which is defined as the range of
analyte concentrations for which the biosensor response changes linearly with the
concentration.
Environmental factors:
Temperature,
Reaction nature
Bacterial Interaction
pH
Microbial growth
Click to Edit
• Easy to fabricate
• Direct measurement
Calibration methods
Physical references: For some sensors, like rangefinders, physical references like
rulers and meter sticks can be used for calibration. For temperature sensors, boiling
Click to Edit water and the triple point of pure water can be used.
Intrinsic and extrinsic calibration: In the automotive industry, intrinsic calibration is used
to find parameters for interpreting raw sensor data, while extrinsic calibration is used to
find the offset between a sensor's coordinate system and another coordinate system.
Negative controls
Do not produce a result and are used to confirm that only the antigen of interest is
causing a reaction. This helps to rule out non-specific binding or other
Click to Edit
• Nucleic acids have become the ultimate tools in the recognition and monitoring of
many important compounds.
• DNA biosensor technologies are under intense investigation owing to their great
promise for rapid and low-cost detection of specific DNA sequences in human, viral
and bacterial nucleic acids.
• As the sequencing of the human genome continues, the mutations responsible for
numerous inherited human disorders are now mapped.
Click to Edit • Pathogens responsible for disease states, bacteria and viruses are also detectable via
their unique nucleic acid sequences and interest in their detection continues to grow.
• The analysis of nucleic acids has gained broad acceptance in diagnostic testing,
pharmacological research, and numerous other fields including animal husbandry and
detection of transgenes.
A biosensor is made from a biological sensing element attached to a signal transducer. The
sensing element can be enzymes, antibodies (as in immunosensors), DNA, or
microorganisms; and the transducer may be electrochemical, optical, or acoustic in nature.
Click to Edit A signal transducer is an essential component of a biosensor. It converts the recognition
event into a measurable signal.
The transducer can take many forms depending upon the parameters being measured.
(A) The two half-matching oligos hybridize with the MB to form a nick and may open
the stem slightly. (B) The DNA ligase binds to the nick and catalyzes the ligation of two
short oligos to form a longer oligo. The ligation product then hybridizes with the MB,
which restores the quenched fluorescence of the MB.
Click to Edit
DNA dendrimers---increase sensitivity
Click to Edit
Click to Edit
Concept of DNA microarray
Click to Edit
Figure 2
DNA microarray
Click to Edit
DNA Microarray (cont’s)
Click to Edit
Click to Edit
The light directed probe array synthesis process used for the preparation of
Affymetrix’s Gene Chip
DNA biosensor not limited to DNA detection, but
more…
Click to Edit