Pacemaker
Pacemaker
ANAESTHETIC IMPLICATIONS
What is a pacemaker ?
It is a medical device which generates electrical impulses
and delivers by electrodes contracting the heart muscles
to regulate the beating of heart.
COMPONENTS OF PACEMAKER
Pulse Generator - Containing batteries and electrical circuit.
Electrodes - Exposed metal connected to heart ( may be endo /epi/myocardial ).
- Electrodes are either unipolar or bipolar.
UNIPOLAR BIPOLAR
Cathode electrode inside the Cathode is distal within the
heart heart
Anode on the case of pulse Anode is shortly proximal on
generator the same lead in the heart
Larger pathway exposes the Smaller pathway ….less EMI
deice to more EMI
PACING THRESHOLD
- Minimum amount of energy required to consistently cause depolarization and therefore
contraction of the heart.
- Pacing threshold is measured in terms of both amplitude and duration for which it is applied
to the myocardium.
FACTORS AFFECTING PACING THRESHOLD
INCREASES DECREASES
THRESHOLD THRESHOLD
Hyperkalemia Stress, Anxiety
Hypothyroidism Hyperthyroidism
Hypothermia Sympathomimetic Drugs
Severe Hypoxia Anticholinergics
MI Hypermetabolic States
Anti Arrythemic Drugs Glucocorticoids
1-4 weeks after insertion
R WAVE SENSITIVITY
- Measure of minimal voltage of intrinsic R wave necessary to activate the sensing circuit of
the pulse generator and thus inhibit or trigger the pacing circuit.
TYPES OF PACING
1. Temporary
2. Permanent
TEMPORARY PACING
- External, battery powered, pulse generators with exteriorized electrode produce
electrical cardiac stimulation to treat a brady or tachy arrhythmia until it resolves or until
long term therapy can be initiated.
- Used for less than 3 days.
1. Trans venous
2. Trans thoracic
3. Trans cutaneous
4. Trans esophageal
PERMANENT PACING
- Implantable pulse generators with endo or myocardial electrodes for permanent use.
- Implanted with long batteries of mercuric zinc (3 years) or lithium (10 years)
TECHNIQUE OF PERMANENT PACING
- Leads placed through - subclavian or cephalic vein.
- Leads positioned in the right atrial appendage - for atrial pacing
in right ventricular apex - for ventricular pacing.
- The pulse generator lies in the subcutaneous pocket below the clavicle.
- Epicardial lead placement is used when no trans venous or if the chest is open.
- To understand the language of pacemakers coding system was developed originally by the
international conference on heart disease.
I II III IV V
CHAMBER CHAMBER MODE OF PROGRAMMABILITY ANTICARDIAC
PACED SENSED RESPONSE FUNCTIONS
- Determining that a CIED is present and defining the functionality of the device(e.g
pacemaker or ICD).
- Whether significant EMI will be present during the planned procedure that might
affect the programmed behavior of the CIED.
- Whether the patient is dependant on pacing and whether or not reprogramming of
- Identification of the manufacturer , type and mode of CIED.
- Have it interrogated by a specialist with a documented written report.
- Determination of patients underlying rhythm/ rate for back up pacing
support.
- If present all rate and anti tachycardia response should be turned off.
- Increasing pacemaker rate to optimize oxygen delivery in major cases.
- Correction of any electrolyte abnormality prior to elective surgery.
- Emergency drugs should be readily available.
- Confirmation of magnet response if it is planned.
IMPORTANT INFORMATION GIVEN TO CIED TEAM
- Intended surgical procedure
- Location of the pulse generator
- Patient position during the procedure
- Types of electro cautery to be used
- Other sources of EMI likely to be present
- Whether cardio version or defibrillation will be necessary
INTRA OPERATIVE
- Ensuring the availibility of a back up source of pacing , defibrillation or
both.
- Maintaining vigilance and monitoring in accordance with ASA
standards so as to rapidly detect any hemodynamic compromise as a
result of interference with CIED function.
- Management of EMI.
- Rapid implementation of the back up source of pacing , defibrillation
or both as required.
MONITORING
- Frequent palpation of patients pulse or arterial waveform in monitor.
- ECG.
- CVC and PAC ; better avoided if CIED recently inserted <2 weeks as they cause lead
dislodgement and arrhythmia during placemant ; safe after 6 week as fibrosis around leads
makes it more stable.
- ETCO2, NIBP, temperature, pulse oximeter.
CONDUCT OF ANAESTHEISA
RA:- Considered safe
- anticoagulants used, coagulation profile to be checked and guidelines to be followed.
GA:- Etomidate and ketamine can cause myopotential interference.
- Succinylcholine is better avoided ( cause myopotential interference)
- A complete array of drugs and equipment must be immediately
available for cardiopulmonary resuscitation.
MAINTENANCE
- Drugs and situation that can increase pacing threshold should be avoided.
- Drugs avoided - Dexmedetomidine (supresses SA and AV node )
- N2O is avoided in a patient with newly implanted pacemaker as it causes an expansion of
gas in the pocket which leads to loss of anodal contact and pacing system malfunction.
MAGNETIC RESPONSE
- Application of magnet is not an advisable practice in all CIED.
- Each CIED has specific response to magnet: - Asynchronous mode
- Turn off transiently
- Turn off permanently
- Reprogramming
FACTORS ASSOCIATED WITH EMI GENERATION
1. Electrocautery
2. Nerve stimulator
3. Evoked potential monitor
4. Fasciculation
5. Shivering
6. Large tidal volumes
7. Radiofrequency ablation
8. ESWL
9. ECT