Imaging With Xrays
Imaging With Xrays
RAYS
By Dr. Prajwal Lakkol
Under the guidance of
Dr. B.P Venkatesh
IMAGING QUALITY
Is the term used to describe overall appearance
of the image and its fitness for purpose. There
are two basic description,
1. Contrast is the ability to distinguish between
adjacent area in the image.
2. Spatial resolution is the ability to detect fine
details.
CONTRAST
• The information content of the invisible x-ray image is
“decoded” by the x-ray film in to a pattern of variation
optical density, known as “radiographic contrast.”
• It is the density difference between image area in the
radiograph.
• It is depends on subject contrast and film contrast.
• Subject contrast depends on the differential attenuation of
x-ray beam as it passes through the patient.
• Subject contrast was seen to be affected by the thickness,
density, atomic differences of the subject, radiation energy ,
contrast material and scatter radiation.
• The information content of the x-
ray image is the pattern of varying
intensity of the x-ray beam caused
by differential attenuation of x-ray
by subject.
• Film contrast depends on,
1.characteristic curve of the film
2. film density
3. screen or direct x-ray exposure
4. film processing.
SHAPE OF THE CHARECTERISTC CURVE
SCATTER
GRID CONSTRUCTION
• The grid is constructed using lead strips usually
about 0.05-0.07mm wide. The number of strips
per cm, the line density, is generally in range of
30-80, with 40 strips per cm being typical. In a
grid of this specification repetition interval is
0.25mm, with gap between strips being 0.19mm
with 0.06 mm wide lead strips. The gap is filled
with inter space material that may be aluminium
but is preferably a lower attenuation material
such as carbon fibers.
GRID CONSTRUCTION
GRID RATIO
• Is the ratio of depth of the inter space channel divided
by its width and is typically 8:1 .
• Larger the grid ratio the smaller is the angle of
acceptance and the more efficient is the grid at
absorbing scattered radiation, thus the greater is the
contrast in the image.
• With very large fields, especially at high kV , more
scatter is produced and a high ratio grid is preferable.
• A grid is generally not used with thin parts of the body
such as extremities or children, because less scatter is
generated.
GRIDS:
GRIDS
GRIDS
GRIDS
P-B DIAPHRAGM
• Grid cut off:
• Within certain tolerance ,the focused grid
must be used at a specified distance from
anode, the tube must be accurately centered
over grid, and grid must not be tilted about an
axis parallel to the lead strips. Otherwise , cut-
off of primary rays will occur.
• The tolerance reduce with higher grid ratio.
GRID ERRORS
MOVING GRID
• It was invented by Dr. Hollis E. Potter in 1920 .
• Grids are moved to blur out the shadows cast
by lead strips.
• They are advantageous because they
eliminate grid line from film.
• Disadvantages are costly, subject to failure
may vibrate x-ray table and increases patients
radiation dose.
MAGNIFICATION AND DISTORTION
MAGNIFICATION is reduced by using a longer FFD ‘F’ or by
decreasing the object film distance ‘h’. when positioning
the patient, the film is therefore usually placed close to
the structure of interest. If the tissues were compressed
, this would also reduce patient dose.
DISTORTION this refers to a difference between the shape
of a structure in the image and in the subject. It may
caused by foreshortening of the shadow of tilted object.
It may also caused by differential magnification of the
parts of structure nearer to and further away from the
imaging device. It can be reduce by using a longer FFD.
HEAT RATING
• The heat loading of an x-ray tube is calculated in
joules and is equal to kV × mAs for a constant
potential or three phase generator. For single phase
generators, the heat loading is 0.7 × kVp × mAs.
• Single radiographic exposure;
In order to freeze and display a moving structure
individual exposure should be as short as possible
and is limited by the heating of x-ray tube. A
combination of kV , mA and exposure time used
should be such that, at the end of the exposure, the
temperature of anode does not exceeds its safe
value.
HEAT UNIT
• Is the original unit of energy input to target. It
was defined in the days when single-phase
generators were standard.
• The number of heat unit is simply kVp × mAs.
COLLIMATION
• X-ray set have a collimator system
so that the beam can be adjusted to
the required size by the operator. In
radiographic equipment, this
Collimators
comprises two sets of parallel
blades made of high attenuation
material that can be driven in to the
beam to define required area. The
collimator incorporates a light
source to the side of the x-ray beam
and mirror in the beam.
COLLIMATORS
Collimator in SPS Top view open View along beam End view
Carbon-carbon jaw
ring installed collimator path assembled jaw
PERFORMANCE TEST ON THE X-RAY
TUBE AND GENERATOR
• LIGHT BEAM DIAPHARAGM AND BUCKY
CENTERING;
Regular checks should be made to ensure that
the beam and x-ray beam match and that the
centre of the field lies at the centre of the film
when the cassette is positioned in the bucky
assembly behind the patient. Any misalignment
would lead to repeat examination or use of
unnecessarily large field x-ray
• KILOVOLTAGE AND OUTPUT;
The kV is generally measured indirectly using a penetrameter.
The instrument usually has two radiation detectors placed
behind metal filters of different thickness.
The output of the tube can be measured using a dosimeter,
often an ionization chamber.
STEP WEDGE;
A simple alternative to the use of a kV meter and ionization
chamber is to use a step wedge. This is a block material,
generally aluminium, which is machined so as to have
successive steps of differing thickness.
Penetrameter
SUMMARY
• Contrast in the image depends on thickness, density and
atomic number of structure with in the body.
• Contrast media use high atomic number materials with suitable
k-edge.
• For a particular imaging device, patient dose is affected by kV ,
FFD and filtration
• Scattered radiation reduces image contrast.
• The grid is generally focused to minimize cut off and moved
during exposure to blur out the grid lines.
• Heat rating of an x-ray tube indicates the maximum allowable
mA for the specified kV and exposure time and the maximum
sustainable continuous mA without damage to tube.
• X-Ray Film