Module 5
Interlocks
• C-Arm Movement • Mirror and Filter
– Vertical motion and rotation
disabled when 13lb or
Interlock
greater of compression is – Prevents X-Ray when
displayed mirror or filter are not
• Automatic Compression positioned correctly
Release
– Installing a localization
• Smart Paddle System
paddle disables the Position Interlock
automatic compression – X-Ray is prevented if
release
paddle does not
• Early Release
match the view.
– Immediate termination of
exposure if X-Ray switches
are released prior to the
end of exposure
Safety
• No rotation on loss of power
• Compression release is disabled
when localization paddle is
installed
• Compression force limited to 45lbs
• 3 Emergency Stop switches
– Turns off gantry in Emergency
– Not for normal use.
Power Up and Down
• Pre-Start Checks • Daily
– Check Physical – Re-Start the
Integrity system
• Mechanical Controls
• Weekly
– Emergency Off
– Complete power
switches
down
• Turn on system
– Press Power On
– Login to the OS
– Login to the
Application
System Startup process
Icons on CDE appear,
Power Applied Computer Login As Processes start Application
to Sun performs POST “tech” Digital Detector Launches
Computer Checks temp bar
Gantry
Brick Receives Brick booting Brick booted Starts
Power Array
Off Blinking Off
Off On Off Off Blinking On
Brick ~4 mins
Clicking Heard Array
HV Checks Running
Array ~2 mins
Logins & Passwords
• Logins for Operating
System Level (OSL)
access: – reboot Allows the
– tech Access allows the user to initiate a
technologist to capture reboot from the
images. After accessing the Solaris log in screen.
system, the technologist
then signs in to the – shutdown Allows the
application with a unique user to power down
identifier. the Acquisition Station
– mgr Access allows a Lead computer.
Tech to create reports, delete
patients from the database,
and edit procedures.
– apps Has the most
privileges, with access to
administration and service
functions.
Usernames and Passwords are case sensitive;
type lower case letters
Log On to the Operating System
Common Desktop Environment
Application Main Screen
Automatic Array Shutdown and Startup
• Configure the system to
automatically shutdown the array at
night (TB-00213.)
• In the morning, the array can be
automatically started before anyone
arrives so that the array has time to
warm up to temperature.
• This features is enabled or disabled
by Hologic Field Service
Acquiring Images
Choosing a patient
Patient
Add New
Patient
Procedur Procedure
Local dB e
Search for View
existing
View
patient
manually or View
using BCR Modality View
Worklist
Choosing a Patient Using the Modality Worklist Query
• Modality Worklist
– that allows you to retrieve patient information
from a Radiology Information System (RIS).
– This feature follows the DICOM Modality
Worklist standard, and requires your facility to
have a DICOM Modality Worklist provider.
• If this feature is configured on your
system, the Enter or search for patient
window contains an enhanced user
interface that includes two tabs:
– Worklist tab.
– Local dB tab
Choosing a Patient Using the Modality
Worklist Query (cont.)
• Worklist tab. This tab includes two
subtabs.
– The Patient Query tab
• allows you to search for a
patient in the Modality Worklist
provider using several
different search criteria.
– The Todays Worklist tab
• allows you to retrieve a daily
worklist of patients and visits
from the RIS.
• Local dB tab
– allows you to manually enter
patient information, or search the
local database for a patient.
Local database
Acquiring Images
Making X-Ray
• Press both X-Ray buttons
• Generator Status
– Boost
– Rotor
– X-Ray On
– X-Ray Off
• Hold the X-Ray buttons until the
high pitched tone stops
Preview Image
Options
• Technique Used
• Comments
– Not printed on film
• Markers
– View Left and Right
– Standard Markers
Left and Right
– Click and Drag and
Drop
Options
• Magnification
• Measure
• Region of Interest
(ROI)
• Orientation
Patient Orientation and Image Laterality
Image Laterality Patient Orientation Column
Laterality is the localization of a The value in the patient orientation
particular side of the body. For column represents the image orientation
example, is the image the left hand in the Y direction as viewed in the Image
or the right hand. There are four Preview window. The values for this
Image Laterality values the user can option are:
select: A for Anterior L for Left
R for Right P for Posterior H for Head
L for Left R for Right F for Feet
U for Unpaired
B for Both (Left and Right)