Pacing - CCN Class
Pacing - CCN Class
Pulse generator
Pacing lead
Healthy myocardium
Pulse Generator
Pacing pulse generator
The pulse generator is designed to generate
an electrical current that travels through
pacing lead and exits through electrode that
is in direct contact with the heart. This
electrical current initiates a myocardial
depolarization.
Pacing Lead System
Can be bipolar or unipolar
The negative
electrode is attached
to the negative
terminal, and the
positive electrode is
attached to the
positive terminal of
the pulse generator,
either directly or via a
bridging cable
Unipolar Pacing System
A unipolar pacing
system has only one
electrode (the
negative electrode)
making contact with
the heart.
Pacing Routes
Transcutaneous
pacing
Transthoracic
pacing
Epicardial pacing
Transvenous pacing
Transvenous pacing
It involves threading an electrode catheter
through a vein into the right atrium or right
ventricle. Five different veins can be used.
Antecubital approach
Femoral approach
Subclavian
Internal jugular
External jugular
• Transvenous pacemaker: Leads are threaded
transvenously to the chambers and attached to
the power source
• Epicardial pacing: The pacing leads are attached
to the epicardium during heart surgery
• Transcutaneous pacemaker: Noninvasive, power
source is attached to large electrodes placed
over the anterior and posterior chest
Types of Pacemaker
Permanent and temporary
pacemakers
Permanent pacemaker
It is implanted totally with in the body.
The power source is implanted subcutaneously
usually over the pectoral muscle on the patients
non dominant side.
The pacing leads are threaded transvenously to
the chamber to be paced
Permanent Pacemakers
Indications
• Acquired AV block
• Second- and third-degree block
• Cardiomyopathy
• BBB
• SA Node dysfunction
• Tachydysrhythmias
Temporary
pacemaker
• It has the power source
outside the body
• Transvenous,Epicardial
pacing and Transcutaneous
pacemakers
Classification based the heart chamber
being paced
• Single chamber pacing: right atrium or the right
ventricle is paced. Only one pacing lead is used.
• Dual chamber pacing, both the right atrium and
right ventricle are paced. Requires two pacing leads.
• One lead is placed in the right atrium, and the other
lead is placed in the right ventricle.
• Note that the designation “MR conditional” means
that the implanted pacemaker pose no known
hazards in a specified MRI environment
Classification according to type of
pacing
• Atrial pacing, ventricular pacing, and AV pacing.
• Atrial pacing: SA node failure. – Symptomatic
sinus bradycardia or sinus arrest.
• Ventricular pacing - complete av block
• AV pacing - SA node failure and a complete AV
block.
Classification according to
pacemaker modes
Asynchronous vs synchronous.
• Asynchronous mode/fixed rate
A pacemaker set to to asynchronous mode will fire at a
consistent, constant rate regardless of the heart’s
electrical activity.
• Synchronous mode/demand type
A pacemaker set to synchronous mode will fire only
when the heart’s intrinsic rate falls below a preset
number. Another name for this is a demand pacemaker.
Classification according to Pacemaker
nomenclature
• Pacemaker nomenclature is most often written as a series of three
letters (i.e., VVI) but can be as much as five (DDDRO
• In order, the 3 or 5 letter nomenclature stands for:
1. Chamber Paced: A (atrium), V (ventricle), D (dual)
2. Chamber Sensed: A (atrium), V (ventricle), D (dual)
3. Response to Sensed Beat: I (inhibit), T (trigger), D (dual), O (none)
4. Rate Responsiveness: R (rate responsive), O (none)
• Automatically increases the pacing (and cardiac output) to meet
an increase in exertion
5. Multisite Pacing: O (none), A (atrium), V (ventricle), D (dual)
n.b. only the first three characters matter.
Indications
Bradycardia
Sinus bradycardia and arrest
Sick sinus syndrome
Heart blocks
Tachydysrhythmias
Supraventricular
Ventricular
Permanent pacemaker failure
Support cardiac output after cardiac surgery
Diagnostic studies
• Electrophysiological studies
• Atrial electrogram
Pacemaker Settings
1. RATE
2. OUTPUT
3. SENSITIVITY
RATE
It regulates the number of impulse
that can be delivered to the heart per
minute.
The rate setting depend on the
physiological needs of the patient,
but it is generally maintained
between 60-80 beats/ min.
If the pacemaker is operating in dual-
chamber mode, the ventricular
control rate also regulate the atrial
rate.
OUTPUT
It is the amount of electrical current
(measured in milliamperes [mA]) that is
delivered to the heart to initiate
depolarization.
The point at which depolarization occurs
is termed threshold and is indicated by a
myocardial response to the pacing
stimulus (capture)
SENSITIVITY
Troubleshooti
ng
Failure to Fire
Failure of the pacemaker deliver the pacing
stimulus results in the disappearance of the
pacing artifact, even though the patient’s
intrinsic rate is less than the set rate on the
pacer.
This can occur either intermittently or
continuously and can be attributed to failure of
the pulse generator or its battery
Failure to fire
•CAUSES
• Dead battery
• Replace battery
• Catheter dislodged
• Loose connections
Failure to Capture
•INTEREVENTIONS
• Reposition catheter
• Replace battery