100 um
SOLUTION: HR Mode
• ALTA RESOLUCION
• Cortes más delgados.
• Ángulo (40°), Tiempo (9 seg)
• Ideal desagregar distorsiones mamarias.
• Visualizar espiculaciones y distribución 3D de
las microcalcificaciones
50 um
SOLUTION: ST/HR Mode
2D image ST mode image HR mode image
SYNTHETIC IMAGE / S-VIEW+
SOLUTION: S-VIEW
• Imagen 2D reconstruída a partir de DBT
• Reduce la dosis generada en la exposición 2D
SOLUTION: S-VIEW
Raw Irradiation
Adipose Tissue
Glandular Tissue
Skinline
14
Advantage (1/3)
With DV2m all breast tissue is visible with better contrast and with no saturation.
Advantage (2/3)
With DV2m, calcifications are better visualized through dense breast tissue.
Advantage (3/3) 16
With DV2m lesions are better visualized through dense breast tissue.
CEDM – CONTRAST ENHANCEMENT DIGITAL MAMMO
Contrast Enhancement Digital Mammo
Energy Subtraction
Purpose - New blood vessels (angionesis) in breast lesions are confirmed by dense
accumulation of iodine contrast medium.
Future - It is expected to be one of the methods to evaluate the lesion malignancy degree,
similar as performed by MRI examination but simply and inexpensively
Exposure method
1. Injection of iodine contrast agent
2. Compress the breast.
3. Perform 2 exposures. Low Energy and High Energy. AWS perfoms the Energy Subtraction
LE HE ES
B B
A A
Contrast agent
Fat Breast
Tissue Tissue
Breast Tissue
The mammary gland structure disappears,
33kV while contrast agents are enhanced. It
becomes conspicuous image quality
29kV 49kV
CEDM - Workflow 1 回のポジショニング 2 連続撮影
Intravenous
ES
injection LE HE
of contrast
medium
BELLUS 2
(v1.2)
PACS
LE and ES
W+Rh W+Cu
Auto AEC 49KV
2 ~ 3ml/sec
20
ANODE MATERIAL – TUNGSTEN O MOLIBDEN?
PROBLEM: Mo or W?
Mo/Mo26kV. W/Rh28 k V
23
Visualización de calcificaciones
SOLUTION: ISC Image-based Spectrum Conversion
ISC hace posible tener la penetración y reducción de dosis del Tungsteno (W) resolución de contraste lograda
con Molibdeno (Mo)
Anode: W Anode: Mo
Corrección
Automática
Pre-Exposición Exposición
Mama densa?, grasa? implante? La dosis se determina acorde a la morfología y composión
mamaria
Análisis de imagen
PATIENT CARE
PATIENT CARE: Fit Sweet Paddle
la tracción dérmica
Compresor articula
hacia el borde del pezón
Opiniones de Radiólogos
Algunos doctores reportaron que las imágenes tomadas con FIT SWEET PADDLE fueron mejores que las
estandars porque la glándula mamaria es comprimida en forma más efectiva.
BIOPSY
3M o 5M B/W
monitor
( for BIOPSY )
Touch panel
de Operación
Posicionador
Compresor
2M Color monitor
Soft Pad ( for control of AMULET )
Lámina de protección
descartable
del detector
Workflow 5. Other useful functions -
Spacers
Spacer for positioning (Vertical approach
only)
120mm
Lower Part
120mm
B 15mm 50mm
*Two steps use
50mm
30mm
Upper Part
It is used to secure the thickness of the breast in case of a thin breast.
When the thickness of the breast is insufficient and the inspection executes it at the inspection with needle (CNB,VAB) of the stroke
structure difficult.
This product is composed of 2 layers. Please adjust according to the thickness of the patient’s breast.
A: To make the height of the spacer 5-10mm, only the top layer is necessary.
A+B: To make the height of the spacer 10-20mm, please use the 2 layers.
* Layer is not to be used independently.
Workflow
MAMMO QC
No estaré generando Cómo sé que lo que adquiero en
diagnósticos falsos negativos? Mamo Digital es realmente lo que
veo en el negatoscopio o en los
monitores?
Cómo sé que mi
El sistema digital, se calibra?
mamógrafo funciona
Funciona siempre bien? Y…
correctamente? Cómo se
puede esconderme algo?
calibra?
Quién establece los criterios de calidad?
IEC (International Electrotechnical Commission)
Articulación entre EUREF y FDA lograda.
Display QC
Herramientas
• Fantomas conocidos (ACR, CDMAM, etc)
Fibra: 0,75 mm
Microcalcificaciones:
CDMAM PHANTOM 0,32 mm
Massa: 0,75 mm
Microcalcificaciones: 0,25 mm
Masa: 4,0/2,0 mm
Bajo contraste: 1,3%
Fibra: 0.70 mm
Resolución : 12 pl/mm
Programa Mamo QC
Diario/Semanal Semianual Anual
Tejido toráxico perdido Imagen Residual
Ruido - Contrast Noise Ratio
Evaluación Compresor Evaluación Tubo RX
Sensibilidad
Distorsión Medición Negatoscopio Medición kVp
Mammography QC software
Fantoma Multiparámetros
Dectabilidad Bajo CNR
Contraste
Distorsión
Uniformidad Uniformidad
Range Dinámico
Sensibilidad Constancia de
Uniformidad Linealidad
Uniformidad
Resolución Espacial
Colimación Pared
Torax
Summary