Asnc Guidelines
Asnc Guidelines
Resolution :
• Test consist of imaging flood
source intrinsically through a
spatial resolution phantom.
• Source can be point/ sheet
source(count rate-10-20 kcps).
• Phantoms commonly used: parallel
line equal space bar phantom and
4- quadrant phantoms
CENTER OF ROTATION
• COR is the center point of projections recorded by
detector at all angle.
• Center of electronic matrix should always align with the
mechanical COR.
• Error results in doughnut/tuning fork artefact
High count flood field uniformity
correction
• Anger cameras use stored high count flood field correction
maps to compensate for variation in sensitivity across the
detector FOV before reconstruction.
• To correct for uniformity 60-100 million flood image is
acquired for each detector and stored.
CT QC procedure
• CT number calibration:
• The quantitative CT values are transformed into the attenuation coefficients or the SPECT
radiotracer.
• Any errors in CT numbers will be propagated as errors in the estimated attenuation coef ficients
for the radiotracer.
• The CT calibration is then checked daily with a water-filled cylinder, usually 24 cm in diameter
provided by the manufacturer. The manufacturer will specify the acceptable range of HU values
measured for water. In practice, if the error is greater than 5 to 10 HU (i.e., different than the
anticipated value of 0 HU), the CT system is considered to be out of calibration.
• An air calibration (or blank-scan acquisition) can be performed to determine if this corrects the
error. If it does not, the manufacturer’s field service engineer must be called. Calibration must
be checked by a qualified person on an annual basis, or after any major repair or calibration.
• Field uniformity: Using water filled phantom
Quarterly and annual QC tests
• Collimation hole angulation:
• It is the geometric relationship of the actual collimator holes and
septa to the crystal face of the planar/SPECT scintillation camera
or to the axis of rotation of SPECT scintillation camera.
• Ideally all edges and septa in the collimators has to be exactly
perpendicular to the crystal and the axis of rotation. Any
difference in the angle is known as collimator angulation error.
• Tilt angle check:
• Refers to the detector plane
and the axis of rotation,
measured along the axis of
rotation.
• Should be 0 degree at the
beginning of tomographic
acquisition
• It can be determined from
summed projection images
over 360 degree of a point
source placed off the
Acquisition
• Activity:
• Depends on several parameters:
• Patient body habitus
• Study protocol
• Scanner type
• Software type
• Radiotracer used
Protocol
• Acquisition type:
• Step and shoot: camera acquires a projection, then stops
recording data when moving to the next angle; this results in a
small amount of dead time because the camera is not acquiring
data while it is moving.
• Continuous’ mode: camera moves continuously and acquires each
projection over an angular increment. This eliminates dead time
and thus increases image counts at the expense of a small amount
of blurring due to the motion of the camera head while acquiring.
• Gating :
• Requires stable and consistent heart rhythm.
• 8 or 16 frames per cycle.
• Attenuation correction:
Shallow tidal breahing and tube current of 10-20 mA and tube
voltage of 80-140 kvp, scan duration-10-30 seconds. But lower
currents can also be used.
PROCESSING
• FILTERING:
• Quality of an image is determined by signal to noise ratio.
• Higher at lower spatial frequencies and lower at higher spatial
frequency.
• Types :
• Low pass(hanning and butterworth)
• High pass(ramp filter)
• Reconstruction :
• Filter backprojection (analytical reconstruction):
• Iterative reconstruction:
• Algorithm :
• Expectation maximization
• MLEM
• MAP
• Newer trend:
• RESOLUTION RECOVERY
Attenuation correcetion for SPECT
system
• TYPES:
• RADIONUCLIDE TRANSMISSION COMPUTED TOMOGRAPHY
• CT SCANNER
• Reorientation and reslicing:
• Quantitation:
• Perfusion parameters, i.e. defect extent, severity values with polar
map display, reversibility stress-rest, percent hypo perfused
myocardium, percent ischaemic myocardium, categorical summed
scores (rest, stress and difference), and even total perfusion
deficit.
• Left ventricle (LV) global functional parameters, i.e. left ventricle
ejection fraction (LVEF), end-diastolic volume (EDV) and end-
systolic volume (ESV), and the LV regional parameters, i.e.
regional wall motion and thickening, in conjunction with phase
analysis and 3D gated surface display,
THANK YOU