Measurements & Instuments
Measurements & Instuments
Ubaidullah
1
Measurement and
Instrumentation
Lecture#1
Engr. M.Ubaidullah
2
Course Outline:
1. Basics Concepts
2. Essentials of Analog Instruments
3. Galvanometers, Ammeters & Voltmeters
4. Watt meters & Energy meters
5. Measurement of Resistance
6. Measurement of Inductance and Capacitance (AC
Bridges)
7. Oscilloscopes
8. Transducers
9. Misc. Electrical Instruments
10. Electronics Instruments
3
Advance Outline:
Signal Generation By: Helfrick Chapter: 8
Frequency Counters and Time interval
Measurements By: Helfrick Chapter: 10
4
Reference Books:
Berlin, Electronic Instruments &
Measurements
Modern Electronic Instrumentation &
Measurement Techniques – A.D. Hell Frick
and W.D Cooper.
Electronic Instrumentation & Measurements -
David A Bell
A course in Electronic and electrical
measurements and Instrumentation by J.B.
Gupta
5
Objective of the Course:
6
Grading Scheme:
Assignments 3-4 10%
Quizzes 3-4 10%
Midterm 30%
Final Term 50%
8
Electronics:
9
Measurement:
10
By measurement of a quantity,
we mean “an act of comparison”
11
Instrumentation:
12
1.1 Measurement Methods
Direct Comparison method:
• Mostly used method
• Unknown quantities are measured with comparison of known
standards, e.g. length, mass etc
• This is not very accurate method, as human errors are
involved.
Indirect Comparison method:
• More accurate than Direct method
• More sensitive
• Quantity is converted in analog form
• Analog signal is processed and fed to the measuring
instrument
• The device which converts the quantity into suitable form is
called transducer.
13
1.1 Measurement Methods
14
1.1 Measurement Methods
Indirect Comparison method:
• Amplification of the weak signal before giving it to the
measurement.
15
1.2 Type of Measurement
Primary Measurements:
Direct measurement with comparison of respective standard.
Needs no conversion in the measuring data.
E.g. length, time, mass etc
Secondary Measurements:
Needs one conversion.
E.g. to measure pressure, 1st it will be converted into suitable
form by a transducer then fed into the instrument.
16
1.2 Type of Measurement
Tertiary Measurements:
Needs two conversions.
E.g. temperature measurement by thermocouples, involves
1st temperature conversion into voltage and then 2 nd
conversion into some readable form through a meter.
17
1.2 Type of Measurement
Tertiary Measurements:
Needs two conversions.
E.g. temperature measurement by thermocouples, involves
1st temperature conversion into voltage and then 2 nd
conversion into some readable form through a meter.
18
1.3 Instruments
19
1.4 Classification of Instrumentations.
(A)
Mechanical instruments.
Used to measure static
and stable quantities.
E.g. force, pressure etc
They cant measure
dynamic and variable
quantities.
E.g. are pressure gauge,
speedometer, water meter
Tire pressure gauge
20
1.4 Classification of Instrumentations.
(A)
Electrical Instruments.
Faster than mechanical instruments
However, their measuring part is also mechanical like
needle, springs etc
They are more accurate and sensitive
They can measure dynamic quantities like current (varying
at the rate of 50Hz/sec) measuring by ammeter.
E.g. are ammeters, voltmeters etc
21
1.4 Classification of Instrumentations.
(A)
Electronic Instruments.
Modern than the previous 2 types.
Uses semiconductor devices like diodes, transistors etc
which have excellent time response and have no noise
pollution.
E.g. are CRO, Power supplies etc.
22
1.4 Classification of Instrumentations.
(B)
Instruments can also be classified as:
Absolute instruments.
Secondary Instruments.
Analog instruments
Digital Instruments
23
1.4 Classification of Instrumentations.
(B)
Absolute instruments.
Which gives measurement quantity in terms of constant of
the instrument.
Used in the laboratory for calibration and standardization.
E.g. Tangent galvanometer
Secondary Instruments.
Gives the measurement of the quantity directly on the
instrument.
E.g. frequency meter, ohm meter, volt meter etc
24
1.4 Classification of Instrumentations.
(B)
Secondary Instruments.
Analog instruments.
Measure analog signals.
E.g. temperature
Digital Instruments.
Gives the digital measurement like DMM
25
1.4 Classification of Instrumentations.
Errors in analog instruments:
Parallax error
While taking reading, eye and pointer should be in
the same line, otherwise error occurs, called parallax
error.
This error cant be removed completely.
Human error
Reading wrongly or recorded wrongly.
Parallax is also a type of human error.
26
1.4 Classification of Instrumentations.
Instrumentations can also be classified as:
Indicating Instruments
Indicating instruments indicate, generally the
quantity to be measured by means of a pointer which
moves on a scale. Examples are ammeter, voltmeter,
wattmeter etc.
Recording Instruments
These instruments record continuously the variation
of any electrical quantity with respect to time. In
principle, these are indicating instruments but so
arranged that a permanent continuous record of the
indication is made on a chart or dial.
27
1.4 Classification of Instrumentations.
Integrating instruments.
These instruments record the consumption of the total
quantity of electricity, energy etc., during a particular
period of time.
That is, these instruments totalize events over a specified
period of time. No indication of the rate or variation or the
amount at a particular instant are available from them.
Some widely used integrating instruments are: Ampere-
hour meter: kilowatt-hour (kWh) meter, kilovolt-ampere-
hour
(kVARh) meter.
28
1.5 Terms related to measurement
Instrument.
Device to measure the value
True Value.
The true or expected value of a quantity to be measured may be
defined as the average of an infinite number of measured values
when the average deviation due to the various contributing factors
tends to zero.
It also refers to a value of the quantity under consideration that
would be obtained by a method (known as exemplar method)
agreed upon by experts.
In other words, it is the most probable value that calculations
indicate and one should expect to measure. Note that the value of
the unknown obtained by making use of primary standards and
measuring instruments is considered to be its true value.
29
1.5 Terms related to measurement
Error.
it is the deviation of the measured (or indicated) value from
the true (or expected) value of a quantity.
In other words, error is the difference between the measured
value and the true value of the unknown quantity.
It is also called absolute error or maximum possible error.
Then error of measurements is given by
∈A = Am – At ---------(1)
where Am = measured value of the quantity
At = true value of the quantity
30
1.5 Terms related to measurement
Accuracy.
Closeness of readings to the true value
It is expressed as %age of true value.
Precision.
It is measure of the consistency of reproducibility (repeatability) of
the measurement (i.e., the successive reading do not differ).
For a given fixed value of an input variable, precision is a measure
of the degree to which successive measurement differ from one
another.
31
1.5 Terms related to measurement
Sensitivity.
It is defined by the change in the output or response of the
instrument for a unit change of input or measured variable.
Efficiency.
Ratio of the measured quantity at full scale to the input
power taken by the instrument.
Resolution or Discrimination.
Resolution is the smallest change in a measured variable (or
measurand) to which the instrument will respond.
32
Numerical
33
Numerical
34
1.5 Terms related to measurement
Fidelity.
If the instrument produces the o/p exactly the replica of the input,
process is said to 100% fidelity.
E.g. if a sine wave is fed to a CRO, the output of CRO should be
an exact sine wave without any distortion.
Torque weight Ratio: (T/W)
Ratio of the deflecting torque produced on the moving system and
weight of the moving system.
For accuracy this ratio should be less.
Response speed and response time:
Rapid response of the instrument to the applied input.
Time taken by the instrument to respond to the changes in the i/p.
35
1.6 Errors in measurement
1. Gross error
2. Systematic error
3. Random error
36
1.6 Errors in measurement
Gross error
The class of error mainly covers human mistake in reading
instruments recording and calculating result.
These can be avoided by two means :-
Great care should be taken with reading & recording the
data.
2, 3 or more reading should be taken for the quantity
under measurement.
some gross error can be detected & some others can not.
complete elimination is not possible.
37
1.6 Errors in measurement
Systematic error
Instrumental
Environmental
Observational
38
1.6 Errors in measurement
Systematic error
These are divided as
Instrumental error:
Due to inherent short comings in the instrument.
Due to misuse of instrument.
Due to loading effect of instrument.
Cab be reduced by:
Substitution method or calibration against
standard may be used for the purpose.
Correction factors should be applied after
determining instrumental errors.
Instrument may be recalibrated carefully.
39
1.6 Errors in measurement
Systematic error
These are divided as
Environmental
due to surroundings E.g. Noise from
electrical machine. Magnetic field,
temperature
Observational
may be due to misreading of instruments.
40
1.6 Errors in measurement
Random Error:
Errors due to unknown Causes.
Magnitude and direction not known
Expressed as average deviation of probable errors or
standard deviation
Can be reduced by:
Increasing the no. of readings and using
statically method.
41
Statistical Analysis:
Arithmetic Mean
42
Statistical Analysis:
Deviation from Mean
d1=x1- A.M.
d2=x2- A.M.
43
Statistical Analysis:
Average Deviation
D= |di| / n
44
Statistical Analysis:
Standard Deviation
= sqrt ( |d i| 2/ n)
For finite observations
= sqrt ( |d i| 2/ (n-1))
45
Statistical Analysis:
Variance
v= 2
46
Statistical Analysis:
Probable error
= +/- 0.6745
47
Statistical Analysis:
Limiting error
specified by manufactures
48
Summary:
1.1 Measurement Methods
1.2 Type of Measurement
1.3 Instruments
1.4 Classification of Instrumentations.
1.5 Terms related to measurement
1.6 Errors in measurement
1.7 Types of errors
Statistical Analysis
49
Numerical Examples:
50
Course Outline:
1. Basics Concepts
2. Essentials of Analog Instruments
3. Galvanometers, Ammeters & Voltmeters
4. Watt meters & Energy meters
5. Measurement of Resistance
6. Measurement of Inductance and Capacitance (AC
Bridges)
7. Misc. Electrical Instruments
8. Electronics Instruments
9. Transducers
51
2.Essentials of Analog Instruments
2.1 classification of analog instruments
2.2 effects of current utilized by analog
instruments
2.3 Essential forces of indicating instruments
2.4 Production of deflecting torque
2.5 Production of controlling torque
2.6 Production of damping torque
52
Instruments in which the output is a continuous
function of the time and bears a constant relation
to the input-Analog Instrument.
53
2.1 Classification of analog instruments
55
Active (Real or True) Power:
Apparent Power:
56
2.1 Classification of analog instruments
57
2.1 Classification of analog instruments
58
2.1 Classification of analog instruments
59
2.1 Classification of analog instruments
60
2.1 Classification of analog instruments
61
2.2 Effects of current utilized by analog instruments
1)Magnetic effect:
When electric current flows through a wire, it
behaves like a magnet. This is called magnetic
effect of electric current.
A coil of wire with an electric current flowing through
it creates a magnetic field.
Putting iron inside a current-carrying coil increases
the strength of the electromagnet.
A changing magnetic field induces an electric current
in a conductor.
63
Oersted showed that magnetic effects could be
produced by moving electrical charges;
64
2.2 Effects of current utilized by analog instruments
Magnetic effect:
65
2.2 Effects of current utilized by analog instruments
66
2.2 Effects of current utilized by analog instruments
3) Thermal effect:
When electric current flows through a wire, the wire
gets heated. This is called the heating effect of
electric current.
Uses of heating effects of electric current :-
The heating effect of electric current is used in
electrical appliances like electric heater, electric iron,
electric room heater, electric kettle, hair dryer etc.
All these appliances have a coil of wire called an
element. When electric current flows through the
element it becomes hot and gives out heat.
The amount of heat produced in a wire depends upon
its material, length and thickness.
67
2.2 Effects of current utilized by analog instruments
Thermal effect:
68
2.2 Effects of current utilized by analog instruments
4) Electrostatic effect:
A force exist between two charged plates.
This force can be utilized to move one of the two
plates.
Electrostatic voltmeters are used for
measurement.
69
2.2 Effects of current utilized by analog instruments
5) Induction effect:
The finding that electric current can produce
magnetic fields led to the idea that magnetic
fields could produce electric currents.
The production of emfs and currents by the
changing magnetic field through a
conducting loop is called induction.
Electric currents induced in conductors when
exposed to a changing magnetic field-
70
2.2 Effects of current utilized by analog instruments
6) Chemical effect:
71
2.2 Effects of current utilized by analog instruments
7) Hall effect:
If an electric current flows
through a conductor in a
magnetic field, the magnetic
field exerts a
transverse (at right angle )fo
rce
on the moving
charge carriers which tends
to push them to one side of
the conductor.
The Hall effect can be
used to measure the
average drift velocity of
the charge carriers 72
2.3 Essential forces of indicating instruments
73
2.3 Essential forces of indicating instruments
74
2.3 Essential forces of indicating instruments
75
2.3 Essential forces of indicating instruments
Degree of Damping:
Under Damped.
When the pointer oscillates about the final position and takes same
time to come to rest.
Over damped.
When the pointer moves slowly to its final position in an lethargic
way. In this case it is very difficult to read the instrument.
Critical Damping / Dead Beat.
When the pointer moves quickly and stops immediately to its final
position of rest.
This situation is ideal and called critical damping.
76
2.3 Essential forces of indicating instruments
77
2.4 Production of deflecting torque
Damping Methods:
78
2.5 Production of controlling torque
Two methods are used to provide the
controlling torque:
1) Spring Control:
2) Gravity Control
79
2.5 Production of controlling torque
1) Spring Control:
80
2.5 Production of controlling torque
1) Spring Control:
81
2.5 Production of controlling torque
1) Spring Control:
Advantages:
82
2.5 Production of controlling torque
1) Spring Control:
Disadvantages:
83
2.5 Production of controlling torque
1) Spring Control:
Design of spring:
Usually flat springs are used as the space required by
theses springs is less.
For flat spring the controlling torque is:
85
2.5 Production of controlling torque
2) Gravity Control:
86
2.5 Production of controlling torque
2) Gravity Control:
87
2.5 Production of controlling torque
2) Gravity Control:
Advantages:
88
2.5 Production of controlling torque
2) Gravity Control:
Disadvantages:
90
2.5 Production of controlling torque
Numerical Example: 2.1
91
2.6 Production of damping torque
92
2.6 Production of damping torque
93
2.6 Production of damping torque
The piston moves in a fix air chamber, close
to the one end.
Piston moves to and fro when there are
oscillations.
The oscillations of the pointer are controlled
by the to and fro motion of the piston.
This helps in settling down the pointer in its
final position.
94
2.6 Production of damping torque
2) Fluid friction Damping:
This damping is similar to air friction damping,
except that air is replaced by high viscosity oil.
95
2.6 Production of damping torque
2) Fluid friction Damping:
This damping is similar to air friction damping,
except that air is replaced by high viscosity oil.
Viscosity is a measure of a
fluid's resistance to flow.
96
2.6 Production of damping torque
2) Fluid friction Damping:
This damping is similar to air
friction damping, except that air is
replaced by high viscosity oil.
The friction b/w fluid and disc is
used for opposing the motion.
97
2.6 Production of damping torque
2) Fluid friction Damping:
98
2.6 Production of damping torque
2) Fluid friction Damping:
Disadvantages:
99
2.6 Production of damping torque
2) Fluid friction Damping:
100
2.6 Production of damping torque
3)Eddy current or Electromagnetic Damping:
101
2.6 Production of damping torque
3)Eddy current or Electromagnetic Damping:
102
2.6 Production of damping torque
Numerical Examples: 2.2-2.3-2.4-
103
2.Essentials of Analog Instruments
Summary:
2.1 classification of analog instruments
2.2 effects of current utilized by analog
instruments
2.3 Essential forces of indicating instruments
2.4 Production of deflecting torque
2.5 Production of controlling torque
2.6 Production of damping torque
104
Course Outline:
1. Basics Concepts
2. Essentials of Analog Instruments
3. Galvanometers, Ammeters & Voltmeters
4. Watt meters & Energy meters
5. Measurement of Resistance
6. Measurement of Inductance and Capacitance (AC
Bridges)
7. Misc. Electrical Instruments
8. Electronics Instruments
9. Transducers
105
3. Galvanometers, Ammeters & Voltmeters
Summary
106
3.1 Types of Galvanometers.
Galvanometer is an electromechanical
instrument which is used for the detection of
electric currents through electric circuits.
we can measure very small currents by using
galvanometer but the primary purpose of
galvanometer is the detection of electric current.
107
3.1 Types of Galvanometers.
108
3.1 Types of Galvanometers.
1. D'arsonval Galvanometer
It is very simple and mostly used galvanometer.
2. Ballastic Galvanometer
It is used to measure the quantity of charge passed
through it.
Basically it is employed to measure the small changes
such as obtained in magnetic flux measurement.
The “flux meter” is a special type of Ballastic
galvanometer.
3. Vibration Galvanometer
Most widely used tuned detectors particularly as null point
detectors in ac bridges.
They are manufactured for a frequency range of 5Hz to
1KHz
109
3.1 Types of Galvanometers.
1. Duddels Galvanometer
Duddels oscillograph is basically a vibration
galvanometer, having a low period of
vibration.
The natural frequency is about 12 KHz.
110
3.1 Types of Galvanometers.
1. D'arsonval
Galvanometer
Also called
Permanent-Magnet
Moving Coil (PMMC).
Based on the moving-
coil galvanometer
constructed by Jacques d’ Arsonval
Jacques d’ Arsonval in
1881.
111
3.1 Types of Galvanometers.
1. D'arsonval Galvanometer
112
3.1 Types of Galvanometers.
1. D'arsonval Galvanometer
113
3.1 Types of Galvanometers.
1. D'arsonval
Galvanometer
114
3.1 Types of Galvanometers.
1. D'arsonval
Galvanometer
115
3.1 Types of Galvanometers.
1.D'arsonval Galvanometer
WORKING PRINCIPLE
116
3.1 Types of Galvanometers.
1. D'arsonval Galvanometer
ESSENTIAL PARTS OF GALVANOMETER
117
3.1 Types of Galvanometers.
1. D'arsonval Galvanometer
118
3.1 Types of Galvanometers.
1. D'arsonval Galvanometer
119
What is
Full Scale Deflection?
120
Moving the indicator all the way up
What is
without hitting the stop pin.
The more sensitive the instrument,
Full Scale Deflection?
the less input it takes to reach
full scale deflection.
121
3.3 Conversion of D’arsonval Galvanometer into
Ammeter and voltmeter.
122
3.3 Conversion of D’arsonval Galvanometer into
Ammeter and voltmeter.
+
Rsh
Rm
_
D’Arsonval
_ Movement
123
3.3 Conversion of D’arsonval Galvanometer into
Ammeter and voltmeter.
124
3.3 Conversion of D’arsonval Galvanometer into
Ammeter and voltmeter.
I sh Rsh I m Rm +
Ish
Im
I +
I sh I I m Rsh
Rm
_
I m Rm D’Arsonval
Rsh _ Movement
I Im
125
3.3 Conversion of D’arsonval Galvanometer into
Ammeter and voltmeter.
Example:
126
3.3 Conversion of D’arsonval Galvanometer into
Ammeter and voltmeter.
127
3.3 Conversion of D’arsonval Galvanometer into
Ammeter and voltmeter.
_
Basic DC Voltmeter
128
3.3 Conversion of D’arsonval Galvanometer into
Ammeter and voltmeter.
129
3.3 Conversion of D’arsonval Galvanometer into
Ammeter and voltmeter.
V I m ( Rs Rm ) Rs
+
V I m Rm V Multiplie Im
Rs Rm V
r
Rm
Im Im
_
V
Rs Rm Basic DC Voltmeter
Im
130
3.3 Conversion of D’arsonval Galvanometer into
Ammeter and voltmeter.
Example:
A basic D’ Arsonval movement with a full-
scale deflection of 50 uA and internal
resistance of 500Ω is used as a DC voltmeter.
Determine the value of the multiplier resistance
needed to measure a voltage range of 0-10V.
131
3.3 Conversion of D’arsonval Galvanometer into
Ammeter and voltmeter.
Example:
A basic D’ Arsonval
Solution: movement with a full-
scale deflection of 50 uA and internal
resistance ofV 500Ω is10used
V as a DC voltmeter.
Rs Rm 500 199.5k
Determine theI m value50 ofuAthe multiplier resistance
needed to measure a voltage range of 0-10V.
132
3.4 Extension of range of Ammeter
and Voltmeter.
The range of the dc ammeter is extended by a number
of shunts, selected by a range switch.
The resistors are placed in parallel to give different
current ranges.
Switch S (multi position switch) protects the meter
movement from being damage during range changing.
Increase cost of the meter.
133
3.5 Extension of range of Ammeter
by shunt.
MULTI-RANGE AMMETER
+
+
R1 R2 R3 R4
Rm
_
D’Arsonv
al
S Moveme
nt
_
Im
V2
V1 V3
Rm
+
V4
_ 135
Examples: 3.1-3.10
136
3.8 Ammeters & Voltmeters.
I I1 I 2 I 3
138
139
3.8 Ammeters & Voltmeters.
Dividing equation yields:
Im R1
Ie R1 Rm
140
3.8 Ammeters & Voltmeters.
Ammeters
Ammeters measure current flowing in a circuit.
They are galvanometers that are placed in parallel with a
resistor.
The resistor in this capacity is referred to as a shunt. Its value is
chosen to be very small, in order to limit the amount of internal
resistance of the galvanometer.
Galvanometers as ammeters can directly measure DC current
but to measure AC, a diode is placed inside the device.
The purpose of this diode is to limit the current flow to one
direction. Without it, the needle would fluctuate back and forth or
not move at all, and it would not be possible to take a reading.
141
3.8 Ammeters & Voltmeters.
Voltmeter
Voltmeters measure the potential difference or
voltage in a circuit, and are galvanometers that are
placed in series with a resistor.
Ideal galvanometers used this way must have an
infinite resistance, and this is approximated having
the value of the resistor be very large.
Once again, a diode must be placed inside in order
for it to measure AC currents.
142
3.8 Ammeters & Voltmeters.
143
3.9 Types of instruments using
Ammeter.
The main types of instruments used as ammeters
and voltmeters are;
144
3.9 Types of instruments.
1. Moving iron type
145
3.10 Moving Iron (M.I) Instruments.
1. Moving iron type
146
3.11 Repulsion Type
1. Repulsion type
147
3.11 Repulsion Type
1. Repulsion type
In repulsion type moving–iron instrument consists of two
cylindrical soft iron vanes mounted within a fixed current-
carrying coil.
One iron vane is held fixed to the coil frame and other is free
to rotate, carrying with it the pointer shaft.
Two irons lie in the magnetic field produced by the coil that
consists of only few turns if the instrument is an ammeter or of
many turns if the instrument is a voltmeter.
Current in the coil induces both vanes to become magnetized
and repulsion between the similarly magnetized vanes
produces a proportional rotation.
148
3.12 Deflecting torque equations.
149
Moving Iron (M.I) Instruments.
Measurement of Electric Voltage and
Current
Moving iron instruments are used as Voltmeter
and Ammeter only.
Both can work on AC as well as on DC.
150
Moving Iron (M.I) Instruments.
Measurement of Electric Voltage and Current
Ammeter
Instrument used to measure current in the circuit.
Always connected in series with the circuit and carries the current
to be measured.
This current flowing through the coil produces the desired
deflecting torque.
It should have low resistance as it is to be connected in series.
Voltmeter
Instrument used to measure voltage between two points in a
circuit.
Always connected in parallel.
Current flowing through the operating coil of the meter produces
deflecting torque.
It should have high resistance. Thus a high resistance of order of
kilo ohms is connected in series with the coil of the instrument.
151
Moving Iron (M.I) Instruments.
Advantages
The instruments are suitable for use in AC and
DC circuits.
The instruments are robust, owing to the
simple construction of the moving parts.
The stationary parts of the instruments are
also simple.
Instrument is low cost compared to moving
coil instrument.
152
Moving Iron (M.I) Instruments.
Errors
153
3.13 Attraction Type
2. Attraction Type
154
3.13 Attraction Type
2. Attraction Type
155
Numerical Examples:
156
3.17 Moving Coil (M.C) Instruments
157
3.17 Moving Coil (M.C) Instruments
Deflecting force:
F= NIBL Sinα
N= No. of turns in the coil
I=Current flowing through the coil
B=Flux density
L=Length of the coil
α= Angle between the coil and the field
158
3.17 Moving Coil (M.C) Instruments
159
3.17 Moving Coil (M.C) Instruments
Advantages:
These instruments have low power
consumption.
They have uniform scale.
They are free from hysteresis losses.
They provide current damping by eddy
current.
160
If the magnetic field applied to a magnetic
material is increased and then decreased
back to its original value, the magnetic field
inside the material does not return to its
original value. The internal field 'lags' behind
the external field. This behavior results in a
loss of energy, called the hysteresis loss.
161
3.17 Moving Coil (M.C) Instruments
Disadvantages:
They cant be used for ac measurements.
They are costly than moving iron instruments.
Ageing of control spring may cause error.
162
163
164
165
166
Chapter-4:
Watt meters & Energy meters
ELECTRODYNAMOMETER-TYPE METERS
Types of Wattmeter
Dynamometer Wattmeter
Types of Dynamometer Wattmeter
Wattmeter connections
Poly phase Wattmeter
Energy meter
Single phase induction type energy meter
Lag adjustment in Energy meters
Errors in induction type energy meter
Adjustments in an energy meter
Calibration of 1 phase energy meter
summary
167
ELECTRODYNAMOMETER-TYPE METERS
168
ELECTRODYNAMOMETER-TYPE METERS
170
ELECTRODYNAMOMETER-TYPE METERS
METER ACCURACY
171
ELECTRODYNAMOMETER-TYPE METERS
METER MOVEMENT
METER MOVEMENT
METER MOVEMENT
178
Watt Meter
Wattmeter Connection
179
Watt Meter
Wattmeter Errors:
180
Watt Meter
Wattmeter Errors:
181
Watt Meter
Wattmeter Errors:
184
Watt Meter
Wattmeter Overloads:
185
Types of Wattmeter
1. Dynamometer wattmeter
2. Induction wattmeter
3. Electrostatic wattmeter
186
1. Dynamometer wattmeter
Principle:
When a current carrying
moving coil is placed in a
magnetic field produced by the
current carrying fixed coil, a
mechanical force is exerted on
the coil sides of the moving
coil and deflection takes place.
In other words, when the field
produced by the current
carrying moving coil (Fr) tries
to come in line with the field
produced by the current
carrying fixed coil (Fm), a
deflecting torque is exerted on
the moving system.
187
1. Dynamometer wattmeter
When the instrument of this type is used as a wattmeter,
the fixed coil, which is divided into two equal portions in
order to provide uniform field, is employed as current coil
and the moving coil is used as pressure coil i.e. the fixed
coil carries the current flowing through the circuit and the
moving coil carries the current proportional to the voltage
across the circuit.
A high non-inductive resistance is connected in series with
the moving coil in order to limit the current.
The magnetic fields of the fixed and moving coils react on
one another causing the moving coil to turn about its axis.
The movement is controlled by hair-springs which also
lead the current into and out of moving element.
188
1. Dynamometer wattmeter
189
1. Dynamometer wattmeter
Damping is provided by light aluminum vanes moving
in an air dash pot.
The pointer is fixed to the moving – coil spindle, and
moves over a suitably calibrated scale.
Dynamometer type wattmeter has the advantages of
(i) uniform scale and (ii) high degree of accuracy and
disadvantages of (i) errors due to stray field acting
on the moving coil and (ii) serious error due to
inductance of pressure coil (Potential coil) unless
special precautions are taken to reduce this effect.
190
1. Dynamometer wattmeter
The inductance error
can further be reduced
by connecting an
additional winding
called “compensating
winding” in series with
the pressure (voltage)
coil and so placed that it
produces a field in the
opposite direction to
that of the current coil.
191
2. Induction wattmeter
The operation of all
induction instruments
depends upon the
production of the torque
due to flux1 and eddy
current produced because
of flux2.
192
193
2. Induction wattmeter
194
Poly phase Wattmeter
Such an instrument is used to measure
power in a polyphase (3-phase)
system.
Error:
In order to avoid mutual interference
between the two units, these are
magnetically shielded from each other
by placing a nickel iron plate between
them.
196
Poly phase Wattmeter
Connection:
The electrical connection of a poly-phase
wattmeter for measurement of power in a
3-Ø system is identical with those of two
wattmeter method of measuring power in
3-Ø system.
199
Induction type Single Phase Energy
Meter:
Construction:
A single phase induction type energy meter
consists of driving system, moving system,
breaking system and registering system. Each of
the systems is briefly explained below.
200
Induction type Single Phase Energy
Meter:
201
202
Induction type Single Phase Energy
Meter:
Driving system:
203
Induction type Single Phase Energy
Meter:
A coil having large number of
turns of fine wire is wound on
the middle limb of the shunt
magnet. This coil is known as
“pressure or voltage” coil and is
connected across the supply
mains.
An adjustable copper
shading rings are
provided on the central
limb of the shunt
magnet to make the
phase angle
displacement between
magnetic field set up by
shunt magnet and
supply voltage is
approximately 90
degree.
205
Induction type Single Phase Energy
Meter:
The copper shading bands
are also called the power
factor compensator or
compensating loop.
The series electromagnet is
energized by a coil, known
as “current” coil which is
connected in series with the
load so that it carry the load
current. The flux produced
by this magnet is
proportional to, and in phase
with the load current.
206
Induction type Single Phase Energy
Meter:
Moving system:
208
Induction type Single Phase Energy
Meter:
Moving system:
209
Induction type Single Phase Energy
Meter:
Braking system:
211
Induction type Single Phase Energy
Meter:
Registering or
Counting system:
The registering or
counting system
essentially consists of
gear train, driven either
by worm or pinion gear
on the disc shaft, which
turns pointers that
indicate on dials the
number of times the
disc has turned.
212
Induction type Single Phase Energy
Meter:
Registering or Counting
system:
213
Induction type Single Phase Energy
Meter:
Basic operation
Induction instruments operate in alternating-current
circuits and they are useful only when the frequency
and the supply voltage are approximately constant.
The most commonly used technique is the shaded pole
induction watt-hour meter, shown in fig.
214
Induction type Single Phase Energy
Meter: Basic operation
The rotating element is an aluminum disc, and the
torque is produced by the interaction of eddy currents
generated in the disc with the imposed magnetic fields
that are produced by the voltage and current coils of the
energy meter.
215
Induction type Single Phase Energy
Meter: Basic operation
216
Induction type Single Phase Energy
Meter: Basic operation
217
Induction type Single Phase Energy
Meter: Basic operation
So in all induction instruments we have two
fluxes produce by currents flowing in the
windings of the instrument.
These fluxes are alternating in nature and so
they induce emfs in a aluminum disc or a
drum provided for the purpose.
These emfs in turn circulate eddy currents in
the disc.
218
Induction type Single Phase Energy
Meter: Basic operation
219
Induction type Single Phase Energy
Meter: Basic operation
220
T1 and T2 are opposite in directions.
221
222
Induction type Single Phase Energy
Meter: Basic operation
223
Induction type Single Phase Energy
Meter: Basic operation
224
Induction type Single Phase Energy
Meter: Basic operation
225
Induction type Single Phase Energy
Meter: Basic operation
226
Induction type Single Phase Energy
Meter: Basic operation
The flux generated by the current coil is in phase with the
current and flux generated by the voltage coil is adjusted to be
exactly in quadrature with the applied voltage by means of the
copper shading ring on the voltage or shunt magnet.
The average torque acting upon the disc
227
228
Induction type Single Phase Energy
Meter: Basic operation
229
Induction type Single Phase Energy
Meter: Opposing or Brake Torque:
230
Induction type Single Phase Energy
Meter: Opposing or Brake Torque:
231
Induction type Single Phase Energy
Meter: Opposing or Brake Torque:
232
Induction type Single Phase Energy
Meter: Opposing or Brake Torque:
233
234
Lag Adjustment.
235
Induction type Single Phase Energy
Meter: Opposing or Brake Torque:
For a given disc and brake magnet, the braking torque varies
with the distance of the poles from the center of the disc.
The maximum braking torque occurs when the distance of the
center of the pole faces from the center of the disc is equal to
83% of the radius of the disc.
A movement of the poles of brake magnet towards the center
of the disc reducing the braking torque (as the distance of brake
magnet reduces from the center of the disc), and vise versa.
236
Induction type Single Phase Energy
Meter: Errors in the energy meter
Assuming the supply voltage and frequency constant, the
induction type energy may have the following errors:
1) Speed error:
237
Unity power factor has a value of 1.0. This
means the current and voltage waveforms
are in phase. This is only possible if the net
load is non-reactive (resistive). If the load is
either capacitively or inductively reactive, the
power factor will be other than unity.
238
Induction type Single Phase Energy
Meter: Errors in the energy meter
Continued…
239
Induction type Single Phase Energy
Meter: Errors in the energy meter
2) Meter phase error:
241
Induction type Single Phase Energy
Meter: Errors in the energy meter
4) Creep:
243
Induction type Single Phase Energy
Meter: Errors in the energy meter
4) Creep:
244
245
Induction type Single Phase Energy
Meter: Errors in the energy meter
5) Temperature effect:
247
Induction type Single Phase Energy Meter:
Extension of Instrument Range:
We have seen earlier M.C. instrument’s range can be
extended by properly designed non inductive shunts and
multipliers in cases of ammeter and voltmeter respectively.
Similarly for MI instruments shunts and multipliers can be
designed for extension of range.
Sometimes transformers are used in ac systems for the
measurement of the basic quantities such as current,
voltage and power.
The transformers used in connection with the instruments
for measurement purpose are referred to as Instrument
Transformers.
248
Induction type Single Phase Energy Meter:
Extension of Instrument Range: Continued..
They are classified as Current Transformer (C.T.) used for
current measurement and potential Transformer (P.T.)
used for voltage measurement.
These transformers are used not only for extension of the
range of the instrument, but also for isolating the
instrument from a high current or voltage line.
The advantages of these transformers are
• Single range instrument can be used to cover a wide range.
• Indicating instrument can be located at some distance from
the circuit. This is a great advantage particularly
for high voltage situation.
249
Chapter 4 Numerical Examples
4.1
A dynamometer wattmeter with its potential
coil connected across the load side reads
230W. If the load voltage is 200V, what power
is being taken by the load? The potential coil
has resistance of 2000 ohm.
250
Chapter 4
Numerical Examples.
4.1 Power consumed by the potential coil=
P=V2
/R=(200) 2
/2000=20W
A dynamometer wattmeter with its potential
coil As
connected
the potentialacross the loadacross
coil is connected side the
reads
load,
230W. If power
so the the load voltage
consumed is wattmeter
by the 200V, what power
is also
is being taken
included in theby thereading.
meter load? The potential coil
has resistance of 2000 ohm.
Power taken by the load=230-20=210W
251
Chapter 4 Numerical Examples
4.2
A 5A, 200V meter on full load, unity power
factor test makes 50 revolutions in 360
seconds. If normal disc speed is 500
rev/KWh, what is the %age error?
Ans. No error
as the consume is same.
252
Chapter 4 Numerical Examples
4.3
For a 5A, 230V energy meter the number of
revolutions per Kwh is 480. if in a test of full
load , unity power factor, the disk makes 6
revolutions in 32 seconds. Calculate the error
if any?
Ans. 22.5%
meter runs faster
253
Chapter 4 Numerical Examples
4.4
The constant of an energy meter is 1400
rev/Kwh. On connecting a load of 5 lamps of
100W each and 6 fans of 80W each, the disc
makes 1200 revolutions in one hour. Find
error if any.
Ans. 9.77%
The meter is slow.
254
Chapter 4 Numerical Examples
4.5
The power of a single phase 6.6KV line
drawing a current of 60A is required to be
measured by a wattmeter marked as 220V,
6A. find suitable rating of PT and CT for this
purpose.
Ans.
PT Rating: 30
CT Rating: 10
255
Chapter 4 Numerical Examples
4.6
A meter of constant of 500 makes 5
revolutions in 20 seconds. Calculate the load
in Kw.
Ans.
1.818 KW
256
Chapter 4 Numerical Examples
4.7
The disc of a metal makes 600 rev/KwH.
When a load of 1 Kw is connected for 12 Hrs,
the disc rotates at a rate of 11 r.p.m. Find the
errors if any?
Ans.
Error= 1.2KwH
257
Electronic Energy Meter (EEM)
Electronic Energy meter is based on digital micro
technology (DMT).
It uses no moving parts. So EEM is known as “Static
Energy Meter”
Functioning is controlled by a specially designed IC
called ASIC (Application Specified Integrated Circuit)
ASIC are constructed only for specific applications.
Similar ASIC’s are used in Washing Machines, Air
Conditioners, Automobiles, Digital Camera etc. 258
Electronic Energy Meter (EEM)
259
Chapter 5:
Measurement of Resistance
Measurement of low/medium resistance
Voltmeter ammeter method
Wheatstone bridge
Errors in Wheatstone bridge
Ohm meter
Earth tester
Measurement of high resistance
Method of measurement of high resistance
Meggar
Summary
260
Measurement of low/medium
resistance
Low Resistance:
All resistances of 1Ὼ value or below are classified as low
resistance. such as wires, switches, current-sense resistors,
fuses, relays etc.
Medium Resistance:
The resistance of the value between 1Ὼ and 100k Ὼ are
classified as medium resistance. Most of the electrical
apparatus have medium resistance.
High Resistance:
The resistance of the value above 100 k Ὼ are classified as
high resistance. E.g Heating elements, but the level of
resistance depends on the temperature.
261
Measurement of low/medium
resistance
The important methods of measuring low/medium
resistance are as given below.
262
Measurement of low/medium
resistance
1. Ammeter Voltmeter
Method
Voltmeter is connected in
parallel and ammeter in
series with the resistance.
R= V/I
Measurement of
This method is very simple, resistance.
but it doesn't give accurate
results.
263
Measurement of low/medium
resistance
2. Wheat stone bridge
264
Measurement of low/medium
resistance
2. Wheat stone bridge
Working Principle:
A Wheatstone bridge is an electrical circuit
used to measure an unknown
electrical resistance by balancing two legs of
a bridge circuit, one leg of which includes the
unknown component.
265
Measurement of low/medium
resistance
2. Wheat stone bridge
In the figure, Rx is the unknown
resistance to be measured.
R1,R2,R3 are resistors of known
resistance and the resistance of
R2 is adjustable.
If the ratio of the two resistances
in the known leg R2/R1 is equal
to the ratio of the two in the
unknown leg Rx/R3 , then the
voltage between the two
midpoints (B and D) will be zero
and no current will flow through
the galvanometer .
266
Measurement of low/medium
resistance
2. Wheat stone bridge
If the bridge is
unbalanced, the
direction of the current
indicates whether R2
is too high or too low.
R2 is varied until there
is no current through
the galvanometer,
which then reads zero.
267
Measurement of low/medium
resistance
2. Wheat stone bridge
Voltage drop in R1=voltage drop in R2
Voltage drop in R3=voltage drop in Rx
268
Measurement of low/medium
resistance
2. Wheat stone bridge
269
Measurement of low/medium
resistance
270
Wheat stone bridge
271
Wheat stone bridge
272
Errors in Wheat stone bridge
Human error such as finding a balance point,
taking reading and making calculations.
Thermal junctions in the bridge may induce
error by producing thermo electric emf.
There is a change of value of the resistance
due to self heating.
If the sensitivity of the galvanometer is not
sufficient, the balance point will not be exact.
The error due to resistance of the leads and
contacts.
273
Wheatstone bridge
A Wheatstone bridge is to be used to
measure a high resistance. The ratio arms of
the bridge are 1000Ὼ and 10Ὼ. The
adjustable arm S has a maximum value of
10000Ὼ. What is the maximum resistance
which can be measured?
Ans. 1 MῺ
274
Ohm meter
275
Ohm meter
The purpose of an ohmmeter, of course, is to
measure the resistance placed between its leads.
This resistance reading is indicated through a
mechanical meter movement which operates on
electric current.
The ohmmeter must then have an internal source of
voltage to create the necessary current to operate the
movement, and also have appropriate ranging
resistors to allow just the right amount of current
through the movement at any given resistance.
276
Ohm meter
Starting with a simple movement and battery circuit,
let's see how it would function as an ohmmeter:
277
Ohm meter
When there is infinite resistance
(no continuity between test
leads), there is zero current
through the meter movement,
and the needle points toward the
far left of the scale.
In this regard, the ohmmeter
indication is "backwards"
because maximum indication
(infinity) is on the left of the
scale, while voltage and current
meters have zero at the left of
their scales.
278
Ohm meter
If the test leads of this ohmmeter
are directly shorted together
(measuring zero Ω), the meter
movement will have a maximum
amount of current through it,
limited only by the battery
voltage and the movement's
internal resistance:
279
Ohm meter
With 9 volts of battery potential
and only 500 Ω of movement
resistance, our circuit current will
be 18 mA, which is far beyond
the full-scale rating of the
movement. Such an excess of
current will likely damage the
meter.
280
Ohm meter
If full left-of-scale on the meter face represents an
infinite amount of resistance, then full right-of-scale
should represent zero
We need a way to make it so that the movement just
registers full-scale when the test leads are shorted
together. This is accomplished by adding a series
resistance to the meter's circuit:
281
Ohm meter
To determine the proper
value for R, we
calculate the total circuit
resistance needed to
limit current to 1 mA
(full-scale deflection on
the movement) with 9
volts of potential from
the battery, then
subtract the
movement's internal
resistance from that
figure:
282
Ohm meter Range:
In the left side of the
scale we have "infinity"
and on the right side we
have zero.
this scale is strange
because it goes from
nothing to everything,
rather than from nothing
to a finite value (such
as 10 volts, 1 amp, etc.)
283
Ohm meter Range:
What value of resistance
between the test leads will
cause exactly 1/2 scale
deflection of the needle?
If we know that the
movement has a full-scale
rating of 1 mA, then 0.5 mA
(500 µA) must be the value
needed for half-scale
deflection.
Following our design with the
9 volt battery as a source we
get:
284
Ohm meter Range:
With an internal
movement resistance of
500 Ω and a series
range resistor of 8.5 kΩ,
this leaves 9 kΩ for an
external (lead-to-lead)
test resistance at 1/2
scale.
285
Ohm meter Range:
Using Ohm's Law a few
more times, we can
determine the test
resistance value for 1/4
and 3/4 scale deflection
as well:
1/4 scale deflection
(0.25 mA of meter
current):
286
Ohm meter Range:
3/4 scale deflection
(0.75 mA of meter
current):
287
Ohm meter Range:
So, the scale for this
ohmmeter looks
something like this:
288
Ohm meter Drawback:
Ohmmeters contain internal sources of voltage to
supply power in taking resistance measurements.
An analog ohmmeter scale is "backwards" from that
of a voltmeter or ammeter, the movement needle
reading zero resistance at full-scale and infinite
resistance at rest.
Analog ohmmeters also have nonlinear scales.
Analog ohmmeters are not precision instruments.
Ohmmeters should never be connected to an
energized circuit (that is, a circuit with its own source
of voltage). Any voltage applied to the test leads of
an ohmmeter will invalidate its reading.
Reference: http://www.allaboutcircuits.com
289
Megger or Ohmmeter:
Megger is a portable instrument which is used to
measure insulation resistance of the electrical
machinery or system.
It can be battery operated or mechanically operated
(hand dc generator) and gives direct reading in ohms.
For this reason it is also called as ohm meter.
Onboard ship, different systems are present with
large voltage ratings and therefore Megger comes in
the range of 100V to 5000V.
290
Construction :
The important construction features of Megger consist of
following parts:
291
Construction :
292
Construction :
293
294
Working:
The voltage for testing is supplied by a hand
generator incorporated in the instrument or by battery
or electronic voltage charger. It is usually 250V or
500V and is smaller in size.
- A test volt of 500V D.C is suitable for testing ship’s
equipment operating at 440V A.C. Test voltage of
1000V to 5000V is used onboard for high voltage
system onboard.
- The current carrying coil (deflecting coil) is
connected in series and carries the current taken by
the circuit under test. The pressure coil (control coil)
is connected across the circuit.
295
Working:
Current limiting resistors are connected in series with
pressure and current coil to prevent damage in case of
low resistance in external source.
296
Working:
When the external circuit is open, torque due
to voltage coil will be maximum and the
pointer will read “infinity”.
When there is short circuit the pointer will
read “0”.
297
Earth Tester
Earth Resistance Testers are used for
measuring earth electrode resistance, soil
resistance, earth continuity, resistance of
various components etc.
They are used in Railways, Defense, Utilities,
Power Plants, Geological Surveys and other
Industries.
298
Earth Resistance
“Earth resistance” is the resistance
of the soil to the passage of electric
current.
299
Reasons for Earth Resistance
To provide zero reference for electrical
service.
To provide a low resistance path to protect
against electric faults.
To protect equipment for personal safety.
300
Measurement of Earth Resistance:
301
Use of Earth Resistivity.
302
Measurement of Earth resistance by
Voltmeter and Ammeter:
303
Measurement of Earth resistance by
Voltmeter and Ammeter:
304
Measurement of Earth resistance by
Voltmeter and Ammeter:
305
Earth Tester
306
Earth Tester
307
Earth Tester
308
Earth Tester
Working of Earth Tester
309
Current can produce or speed
up chemical change, this ability
of current is called chemical
effect (shown by dc not by ac).
Ref:http://www.transtutors.com
310
Earth Tester
311
Summary Chapter-5
Measurement of low/medium resistance
Voltmeter ammeter method
Wheatstone bridge
Errors in Wheatstone bridge
Ohm meter
Earth tester
Measurement of high resistance
Method of measurement of high resistance
Meggar
312
Chapter: 6
Measurement of Inductance and Capacitance
1. Measurement of inductance and capacitance
2. Ammeter and voltmeter method to measure L and C
3. AC bridges
4. Maxwell’s bridge
5. Wein bridge
6. Desauty’s Bridge
7. Shering Bridge
8. LCR Bridge
9. LCR Meter
10. Errors in the Bridge circuits
11. Precautions to reduce errors in bridge circuits
12. Summary
313
1.Measurement of inductance and capacitance
314
Resistance
315
Inductive Reactance
Inductive reactance is the opposition
to current flow in an inductive circuit.
This causes the current in a circuit to
lag behind the applied voltage. This
can be calculated by using the
formula:
X L 2fL
316
Capacitive reactance
Capacitive reactance is the
opposition to current flow in a
capacitive circuit. It will cause the
current in the circuit to lead ahead of
the voltage. It can be calculated
from:
1
XC
2fC
317
1.Measurement of inductance and
capacitance
The various methods are used for
measuring the value of L and C in a circuit.
A. Voltmeter and Ammeter method
B. Potentiometers
C. AC Bridges
D. Analog and Digital LCR Bridges.
318
1.Measurement of inductance and
capacitance
Voltmeter and Ammeter method
This is an approximate method of finding the
value of L and C.
We find the value of their reactance either
capacitive or inductive.
319
1.Measurement of inductance and
capacitance
Voltmeter and Ammeter method
Measurement of L (Inductance):
Let we have V and A,
XL=V/A
But X L 2fL
So, V/A= X L 2fL
So L= V/2πfA
320
1.Measurement of inductance and
capacitance
Voltmeter and Ammeter method
Measurement of C (Capacitance):
Let we have V and A,
XC=V/A
1
XC
But 2fC
So, V/A=1/2 πfC
So C= A/2πfV
321
3.AC bridges:
322
4.Maxwell’s Bridge:
The bridge circuit is used for
medium inductance and can be
arranged to yield results of
considerable precision.
In the two arms, there are two
pure resistances so that for
balance relations, the phase
balance depends on the
remaining two arms.
323
324
Maxwell’s Bridge:
325
Example
326
Advantages:
327
Disadvantages:
328
Wein bridge
Because the phase shifts of inductors and
capacitors are exactly opposite each other.
A capacitive impedance can balance out an
inductive impedance if they are located in
opposite legs of a bridge.
In the simplest implementation, the standard
capacitor (C) and the resistor in parallel with it
are made variable, and both must be
adjusted to achieve balance.
329
Wein bridge
331
Wein bridge
332
Wein bridge
Z3 = R3 + jωL3 and Z4 = R4
Balance condition is Z1Z3 = Z2Z4
L3 = CR2R4
333
5. Maxwell’s Wein bridge
334
Wein bridge
Advantages:
The equations are independent of frequency.
The bridge is very useful for measurement of wide
range of inductance at audio frequency (20Hz to
20KHz)
This bridge is also used as a feedback circuit in wein
bridge oscillator.
Disadvantages:
This requires a variable standard capacitor, which is
very expensive for getting a good accuracy.
The bridge is suited to measure inductance of coils
having medium Q factor
The bridge is not suited for high Q values.
335
6.Desauty’s Bridge
De Sauty Bridge measures an unknown
capacitance in term of a standard
capacitance i.e. comparing
two capacitance’s.
Two ratio arm of this bridge consist pure
resistor and two consist capacitor where one
is of known value and another is standard
capacitor.
336
6.Desauty’s Bridge
337
6.Desauty’s Bridge
338
6.Desauty’s Bridge
Advantages:
It is very simple circuit
The calculations are also very easy
Disadvantages:
It is not possible to obtain zero sound in the headphones
because of phase difference in arms due to capacitor.
It is impossible to get the balance of the bridge if the
capacitors are not free from the dielectric losses.
This method is therefore used for air capacitors, which are
more or less loss free.
The capacitance of a capacitor filled with a dielectric is
greater than it would be in a vacuum.
In order to use the circuit for the loss capacitors, modified
“Desauty's” bridge is used in which extra resisters are used
in series with C1 and C2.
339
7.Shering Bridge
A Schering Bridge is a bridge circuit used for
measuring an unknown electrical capacitance
and its dissipation factor.
The dissipation factor of a capacitor is the
ratio of its resistance to its capacitive
reactance.
The Schering Bridge is basically a four-arm
alternating-current (AC) bridge circuit whose
measurement depends on balancing the
loads on its arms.
340
7.Shering Bridge
Figure 1 below shows a diagram of the Schering
Bridge.
341
7.Shering Bridge
In the Schering Bridge above, the resistance values
of resistors R1 and R2 are known, while the
resistance value of resistor R3 is unknown.
The capacitance values of C1 and C2 are also
known, while the capacitance of C3 is the value being
measured.
To measure R3 and C3, the values of C2 and R2 are
fixed, while the values of R1 and C1 are adjusted
until the current through the ammeter between points
A and B becomes zero.
This happens when the voltages at points A and B
are equal, in which case the bridge is said to be
'balanced'.
342
7.Shering Bridge
When the bridge is
balanced, Z1/C2 = R2/Z3,
where Z1 is the impedance
of R1 in parallel with C1 and
Z3 is the impedance of R3 in
series with C3.
In an AC circuit that has a
capacitor, the capacitor
contributes a capacitive
reactance to the impedance.
The capacitive reactance of
a capacitor C is 1/2πfC.
343
7.Shering Bridge
As such,
Z1 = R1/[2πfC1((1/2πfC1) + R1)]
= R1/(1 + 2πfC1R1)
Z3 = 1/2πfC3 + R3.
Thus, when the bridge is balanced:
344
7.Shering Bridge
When the bridge is balanced, the negative and positive
reactive components are equal and cancel out, so
2πfC2R3 = 2πfC1R2 or
R3 = C1R2 / C2.
Similarly, when the bridge is balanced, the purely resistive
components are equal, so
C2/C3 = R2/R1 or
C3 = R1C2 / R2.
Note that the balancing of a Schering Bridge is
independent of frequency.
Reference: http://www.ecelab.com/schering-bridge.htm
345
Hay’s Bridge:
A Hay Bridge is an AC bridge circuit used for
measuring an unknown inductance by
balancing the loads of its four arms.
one of which contains the unknown
inductance.
One of the arms of a Hay Bridge has a
capacitor of known characteristics, which is
the principal component used for determining
the unknown inductance value.
346
Hay’s Bridge:
Figure 1 below shows a diagram of the Hay Bridge.
347
Hay’s Bridge:
As shown in Figure 1, one arm of the
Hay bridge consists of a capacitor in
series with a resistor (C1 and R2)
and another arm consists of an inductor
L1 in series with a resistor (L1 and R4).
The other two arms simply contain a
resistor each (R1 and R3).
The values of R1and R3 are known,
and R2 and C1 are both adjustable.
The unknown values are those of L1
and R4.
348
Hay’s Bridge:
When the Hay Bridge is
balanced, it follows that
Z1/R1 = R3/Z2
wherein Z1 is the impedance of
the arm containing C1 and R2
while Z2 is the impedance of the
arm containing L1 and R4.
Thus,
Z1 = R2 + 1/(2πfC) while Z2 =
R4 + 2πfL1.
349
Hay’s Bridge:
Mathematically, when the bridge is
balanced,
[R2 + 1/(2πfC1)] / R1 = R3 / [R4 + 2πfL1];
or
[R4 + 2πfL1] = R3R1 / [R2 + 1/(2πfC1)]; or
R3R1 = R2R4 + 2πfL1R2 + R4/2πfC1 +
L1/C1.
When the bridge is balanced, the reactive
components are equal, so
2πfL1R2 = R4/2πfC1, or R4 =
(2πf)2L1R2C1.
350
Hay’s Bridge:
Substituting R4, one comes up with the
following equation:
R3R1 = (R2+1/2πfC1)((2πf)2L1R2C1) +
2πfL1R2 + L1/C1; or
L1 = R3R1C1 / (2πf)2R22C12 + 4πfC1R2 + 1);
or
L1 = R3R1C1 / [1 + (2πfR2C1)2] after dropping
the reactive components of the equation since
the bridge is balanced.
Thus, the equations for L1 and R4 for the Hay
Bridge in Figure 1 when it is balanced are:
L1 = R3R1C1 / [1 + (2πfR2C1)2]; and
R4 = (2πfC1)2R2R3R1 / [1 + (2πfR2C1)2]
Note that the balancing of a Hay Bridge is
frequency-dependent.
351
8. LCR Bridge
A simple circuit for the measurement of
resistance, capacitance and inductance.
It permits the operator to set up a number of
different bridge circuits simply by plugging
standard and unknown units into the proper
terminals.
352
8. LCR Bridge Contd…
354
8. LCR Bridge Contd…
355
8. LCR Bridge Contd…
356
8. LCR Bridge Contd…
357
Errors in the Bridge Circuits
When analysis the ac bridges, it is observed that the effect of
low frequencies may be neglected but higher frequencies
changes the behavior of the bridge considerably. the various
errors in ac bridges are:
Errors due to the stray fields.
The errors due to the stray field may be reduced by
providing the electrostatic shield to the circuit.
Leakage errors.
It is because of the insulation failure of different components,
and leakage of current may occur between the arms. To
eliminate this the variable component may be mounted on
insulating stands.
358
Errors in the Bridge Circuits
(Contd…)
Eddy current error
Eddy current may be induced in the resisters and inductors, for this
conducting materials are avoided near the bridge circuits.
Residual Error
The resisters are assumed as non inductive and inductors are
assumed as non resistive, but practically they contains some
inherent values of inductance and resistance respectively. They
will cause error. The error may be reduced by using high quality
components.
Frequency and waveform error
The frequency of the supply effects the of the inductance and
capacitance of the bridge. Similarly presence of harmonic in the
waveform is also important for the same reason. Moreover the
bridge can not be simultaneously balanced for the fundamental
frequency and harmonics present in the wave form.
359
Short Questions:
1. What are the various applications of AC bridges?
2. Explain the basic bridge. Show how resistance can be measured.?
3. With the help of the sketch, explain the Maxwell bridge?
4. Explain Desauty's bridge for measurement of capacitance?
5. Name any two bridges for measurement of inductance and
capacitance?
6. What are the precautions to be observed while handling the ac
bridges?
7. What are the detectors and supply sourced used in ac bridges?
8. How a Wein bridge can be used to determine the frequency of the
supply.
9. What do you mean by Q factor of a coil?
10. The calculations of ac bridge is different than dc bridge. Explain
how?
360
Chapter: 6 Measurement of Inductance and Capacitance
Summary
364
1.Introduction
Analog & Digital
365
1.Introduction
Analog & Digital
In contrast, a digital oscilloscope samples the waveform
and uses an analog-to-digital converter (or ADC) to
convert the voltage being measured into digital
information.
It then uses this digital information to reconstruct the
waveform on the screen.
For many applications either an analog or digital
oscilloscope will do.
However, each type does possess some unique
characteristics making it more or less suitable for specific
tasks.
366
1.Introduction
Analog & Digital
People often prefer analog oscilloscopes when it is
important to display rapidly varying signals in "real time"
(or as they occur).
Digital oscilloscopes allow you to capture and view events
that may happen only once.
They can process the digital waveform data or send the
data to a computer for processing.
Also, they can store the digital waveform data for later
viewing and printing.
367
2. How Does an Oscilloscope Work?
368
2. How Does an Oscilloscope Work?
Analog Oscilloscopes
1. When you connect an oscilloscope probe to a circuit, the
voltage signal travels through the probe to the vertical
system of the oscilloscope.
2. Following Figure is a simple block diagram that shows
how an analog oscilloscope displays a measured signal.
369
2. How Does an Oscilloscope Work?
373
2. How Does an Oscilloscope Work?
375
2. How Does an Oscilloscope Work?
Analog Oscilloscopes (Contd….)
1. The signal also travels to the trigger system to start or
trigger a "horizontal sweep."
2. Horizontal sweep is a term referring to the action of the
horizontal system causing the glowing dot to move
across the screen.
3. Triggering the horizontal system causes the horizontal
time base to move the glowing dot across the screen from
left to right within a specific time interval.
4. Many sweeps in rapid sequence cause the movement of
the glowing dot to blend into a solid line.
5. At higher speeds, the dot may sweep across the screen up
to 500,000 times each second.
376
2. How Does an Oscilloscope Work?
377
2. How Does an Oscilloscope Work?
Analog Oscilloscopes (Contd….)
In conclusion, to use an analog oscilloscope, you need to adjust three
basic settings to accommodate an incoming signal:
The attenuation or amplification of the signal. Use the
volts/div control to adjust the amplitude of the signal before it is
applied to the vertical deflection plates.
The time base. Use the sec/div control to set the amount of time
per division represented horizontally across the screen.
The triggering of the oscilloscope. Use the trigger level to
stabilize a repeating signal, as well as triggering on a single
event.
Also, adjusting the focus and intensity controls enables you to create
a sharp, visible display.
378
2. How Does an Oscilloscope Work?
Digital Oscilloscopes
Some of the systems that make up digital oscilloscopes
are the same as those in analog oscilloscopes; however,
digital oscilloscopes contain additional data processing
systems.
With the added systems, the digital oscilloscope collects
data for the entire waveform and then displays it.
When you attach a digital oscilloscope probe to a circuit,
the vertical system adjusts the amplitude of the signal,
just as in the analog oscilloscope.
379
2. How Does an Oscilloscope Work?
Digital Oscilloscopes (Contd….)
Next, the analog-to-digital converter (ADC) in
the acquisition system samples the signal at
discrete points in time and converts the signal's
voltage at these points to digital values called
sample points.
The horizontal system's sample clock determines
how often the ADC takes a sample.
The rate at which the clock "ticks" is called the
sample rate and is measured in samples per
second.
The sample points from the ADC are stored in
memory as waveform points. 380
2. How Does an Oscilloscope Work?
382
2. How Does an Oscilloscope Work?
383
Sampling Methods
386
4. Oscilloscope Terminology
Learning a new skill often involves learning a new
vocabulary. This idea holds true for learning how to use an
oscilloscope. This section describes some useful
measurement and oscilloscope performance terms.
387
4. Oscilloscope Terminology
Measurement Terms
The generic term for a pattern that repeats over time is a wave –
An oscilloscope measures voltage waves. One cycle of a wave is
the portion of the wave that repeats.
A waveform is a graphic representation of a wave. A voltage
waveform shows time on the horizontal axis and voltage on the
vertical axis.
Waveform shapes tell you a great deal about a signal. Any time
you see a change in the height of the waveform, you know the
voltage has changed.
Straight diagonal lines mean a linear change - rise or fall of
voltage at a steady rate.
Sharp angles on a waveform mean sudden change.
388
4. Oscilloscope Terminology
Measurement Terms (Contd…)
Figure shows below common waveforms.
389
4. Oscilloscope Terminology
Waveform Measurements in Oscilloscope
You use many terms to describe the types of
measurements that you take with your oscilloscope. This
section describes some of the most common measurements
and terms.
1.Frequency and Period
If a signal repeats, it has a frequency. The frequency is
measured in Hertz (Hz) and equals the number of times the
signal repeats itself in one second (the cycles per second). A
repeating signal also has a period - this is the amount of time
it takes the signal to complete one cycle. Period and
frequency are reciprocals of each other.
390
4. Oscilloscope Terminology
Waveform Measurements in Oscilloscope
Frequency and Period
391
4. Oscilloscope Terminology
Waveform Measurements in Oscilloscope
2.Voltage
Voltage is the amount of electric potential (a kind of signal
strength) between two points in a circuit.
Usually one of these points is ground (zero volts) but not
always - you may want to measure the voltage from the
maximum peak to the minimum peak of a waveform,
referred to at the peak-to-peak voltage.
The word amplitude commonly refers to the maximum
voltage of a signal measured from ground or zero volts.
392
4. Oscilloscope Terminology
Waveform Measurements in Oscilloscope
Voltage
The waveform shown in Figure 8 has an amplitude of one
volt and a peak-to-peak voltage of two volts.
393
4. Oscilloscope Terminology
Waveform Measurements in Oscilloscope
3.Phase
Phase is best explained by looking at a sine wave.
Sine waves are based on circular motion and a circle has 360
degrees.
Using degrees, you can refer to the phase angle of a sine
wave when you want to describe how much of the period has
elapsed.
Phase shift describes the difference in timing between two
otherwise similar signals.
394
4. Oscilloscope Terminology
Waveform Measurements in Oscilloscope
Phase
In following figure, the waveform labeled "current" is said to be
90 degree out of phase with the waveform labeled "voltage,"
since the waves reach similar points in their cycles exactly 1/4 of a
cycle apart (360 degrees/4 = 90 degrees). Phase shifts are common
in electronics.
395
4. Oscilloscope Terminology
Performance Terms
The terms described in this section may come up in your
discussions about oscilloscope performance. Understanding these
terms will help you evaluate and compare your oscilloscope with
other models.
1.Bandwidth
The bandwidth specification tells you the frequency range the
oscilloscope accurately measures.
2.Rise Time
Rise time is another way of describing the useful frequency range
of an oscilloscope. Rise time may be a more appropriate
performance consideration when you expect to measure pulses and
steps. An oscilloscope cannot accurately display pulses with rise
times faster than the specified rise time of the oscilloscope.
396
4. Oscilloscope Terminology
Performance Terms (Contd…)
3.Vertical Sensitivity
The vertical sensitivity indicates how much the vertical
amplifier can amplify a weak signal. Vertical sensitivity is
usually given in milli volts (mV) per division.
The smallest voltage a general purpose oscilloscope can
detect is typically about 2 mV per vertical screen division.
4.Sweep Speed
For analog oscilloscopes, this specification indicates how
fast the trace can sweep across the screen, allowing you to
see fine details. The fastest sweep speed of an oscilloscope is
usually given in nanoseconds/div.
397
4. Oscilloscope Terminology
Performance Terms (Contd…)
5.Gain Accuracy
The gain accuracy indicates how accurately the vertical
system attenuates or amplifies a signal. This is usually listed
as a percentage error.
6.Time Base or Horizontal Accuracy
The time base or horizontal accuracy indicates how
accurately the horizontal system displays the timing of a
signal. This is usually listed as a percentage error.
398
4. Oscilloscope Terminology
Performance Terms (Contd…)
7.Sample Rate
On digital oscilloscopes, the sampling rate indicates how many
samples per second the ADC (and therefore the oscilloscope) can
acquire. Maximum sample rates are usually given in mega samples
per second (MS/s). Typically, the sample rate changes with
changes made to the sec/div control to maintain a constant number
of waveform points in the waveform record.
8.ADC Resolution (Or Vertical Resolution)
The resolution, in bits, of the ADC (and therefore the digital
oscilloscope) indicates how precisely it can turn input voltages
into digital values. Calculation techniques can improve the
effective resolution.
399
4. Oscilloscope Terminology
Performance Terms (Contd…)
9.Record Length
The record length of a digital oscilloscope indicates how
many waveform points the oscilloscope is able to acquire for
one waveform record. Some digital oscilloscopes let you
adjust the record length. The maximum record length
depends on the amount of memory in your oscilloscope.
400
5. Setting Up
Setting the controls of Oscilloscope
After plugging in the oscilloscope, take a look at the front
panel. It is divided into three main sections labeled
Vertical, Horizontal, and Trigger.
Your oscilloscope may have other sections, depending on
the model and type (analog or digital).
Notice the input connectors on your oscilloscope. This is
where you attach probes.
Most oscilloscopes have at least two input channels and
each channel can display a waveform on the screen.
Multiple channels are handy for comparing waveforms.
401
5. Setting Up (Contd…)
Setting the controls of Oscilloscope (Contd…)
402
5. Setting Up (Contd…)
Setting the controls of Oscilloscope (Contd…)
Some oscilloscopes have an AUTOSET or PRESET button that sets
up the controls in one step to accommodate a signal. If your
oscilloscope does not have this feature, it is helpful to set the controls
to standard positions before taking measurements.
Standard positions include the following:
Set the oscilloscope to display channel 1
Set the volts/division scale to a mid-range position
Turn off the variable volts/division
Turn off all magnification settings
Set the channel 1 input coupling to DC
Set the trigger mode to auto
Set the trigger source to channel 1
Turn trigger holdoff to minimum or off
Set the intensity control to a nominal viewing level
Adjust the focus control for a sharp display
403
6. The Controls
Display Controls
Display systems vary between analog and digital
oscilloscopes. Common controls include:
An intensity control to adjust the brightness of the
waveform. As you increase the sweep speed of an analog
oscilloscope, you need to increase the intensity level.
A focus control to adjust the sharpness of the waveform.
Digital oscilloscopes may not have a focus control.
A trace rotation control to align the waveform trace with
the screen's horizontal axis. The position of your
oscilloscope in the earth's magnetic field affects waveform
alignment. Digital oscilloscopes may not have a trace
rotation control.
404
6. The Controls
Vertical Controls
Use the vertical controls to position and scale the waveform vertically.
Position and Volts per Division
The vertical position control lets you move the waveform up or
down to exactly where you want it on the screen.
Input Coupling
Coupling means the method used to connect an electrical signal
from one circuit to another. In this case, the input coupling is the
connection from your test circuit to the oscilloscope. The coupling
can be set to DC, AC, or ground. DC coupling shows all of an
input signal. AC coupling blocks the DC component of a signal so
that you see the waveform centered at zero volts.
405
6. The Controls
Input Coupling
406
6. The Controls
Input Coupling
The ground setting disconnects the input signal from the
vertical system, which lets you see where zero volts is on
the screen. With grounded input coupling and auto trigger
mode, you see a horizontal line on the screen that
represents zero volts.
Bandwidth Limit
Most oscilloscopes have a circuit that limits the bandwidth
of the oscilloscope. By limiting the bandwidth, you reduce
the noise that sometimes appears on the displayed
waveform, providing you with a more defined signal
display.
407
6. The Controls
Vertical Controls (Contd…)
Channel Invert
Most oscilloscopes have an invert function that allows you to
display a signal "upside-down." That is, with low voltage at the
top of the screen and high voltage at the bottom.
Math Operations
Your oscilloscope may also have operations to allow you to add
waveforms together, creating a new waveform display. Analog
oscilloscopes combine the signals while digital oscilloscopes
mathematically create new waveforms. Subtracting waveforms is
another math operation. Subtraction with analog oscilloscopes is
possible by using the channel invert function on one signal and
then use the add operation. Digital oscilloscopes typically have a
subtraction operation available. Following Figure illustrates a third
waveform created by adding two different signals together.
408
6. The Controls
Horizontal Controls
Use the horizontal controls to position and scale the waveform
horizontally.
Position and Seconds per Division
The horizontal position control moves the waveform from left and
right to exactly where you want it on the screen.
Time Base Selections
To see events more clearly or even see events not visible with just
the main time base sweep.
Trigger Position
It actually represents "the horizontal position of the trigger in
the waveform record." Horizontal trigger position control is
only available on digital oscilloscopes. Varying the
horizontal trigger position allows you to capture what a
signal did before a trigger event (called pre trigger viewing).
409
6. The Controls
Horizontal Controls (Contd…)
Magnification
Oscilloscope may have special horizontal magnification
settings that let you display a magnified section of the
waveform on-screen.
XY Mode
Most oscilloscopes have the capability of displaying a
second channel signal along the X-axis (instead of time).
This is called XY mode.
410
6. The Controls
Trigger Controls
The trigger controls let you stabilize repeating waveforms
and capture single-shot waveforms
412
CRT (Contd…)
413
CRT (Contd…)
A CRT works by sweeping an electron beam of varying intensity
across a phosphor-coated screen. The basic components of the CRT
are described below:
Electron Gun -- The electron gun, which consists of the
cathode, choke, accelerator, and lensing region, is the
device which generates and focuses the electron beam used
to project an image on the phosphor screen.
Cathode -- The cathode is a grounded metal plate that is
super-heated so that electrons are literally jumping off the
surface.
Accelerator Plate -- This metal ring is held at a large,
positive voltage and is used to "grab" loose electrons from
the cathode and hurl them forwards into the lensing
chamber (towards the right in the diagram).
414
CRT (Contd…)
Choke -- This metal ring is located between the cathode and
accelerator plate and held at a slightly negative charge. The
electric fields from the choke help columnate the electrons; they
also can be used to quickly modulate the number of electrons in
the beam and, thus, the brightness or intensity of the picture.
Lensing Region -- The lensing region consists of two
adjacent metal tubes that are located just after the accelerator.
The electrons that have jumped off the cathode begin to focus.
Ideally, the focal point will occur at the point when the beam
strikes the display, thereby providing pinpoint resolution on the
screen. The last metal tube of the lensing chamber is held at the
highest potential of all the electron gun components so that
exiting electrons have a very high forward velocity.
415
CRT (Contd…)
Phosphor Screen –
If all works well, a pinpoint electron beam strikes
the screen with the appropriate intensity and
causes the phosphor to fluoresce.
The intensity modulation is synchronized with the
horizontal and vertical retraces so that one frame
of video is displayed. The process repeats itself
rapidly (24 frames/second for analog television)
so that the moving scene appears seamless.
416
8.Delay Line
The delay line is used to delay the incoming signal vertical
information and sync it with the horizontal.
In the vertical signal path, before the delay line, there is typically a
trigger pick-off which supplies an undelayed copy of the vertical
signal to the trigger and sweep circuitry.
Trigger and sweep circuitry need about 60ns to react when presented
with the trigger event.
Without a delay line, the trigger event would already have come and
gone before the scope can trigger and sweep.
By sending the input signal through a delay line, the scope will have
triggered and begun sweeping by the time the trigger event emerges
from the delay line.
Thus, the trigger event is drawn on the screen where the operator can
see it, photograph it, or record it by other means.
417
Chapter:7 Oscilloscopes
Summary
1. Introduction
2. How Does an Oscilloscope Work?
3. Oscilloscope Block Diagram
4. Oscilloscope Terminology
5. Setting Up
6. The Controls
7. Cathode Ray Tube
8. Delay Line
418
Chapter: 8
Measurement of Non-Electrical Quantities
1. Classifications of Transducers
2. Advantages of electrical transducers
3. Requirement of transducers
4. Classification of electrical transducers
5. Resistance(wire) strain gauges
6. Semiconductor strain gauges
7. Capacitive transducers
8. Linear variable differential transformer
9. Hall effect transducer
10. Resistance thermometer
11. Thermocouple transducer
12. Piezoelectric transducers
419
Measurement of Non-Electrical Quantities
Introduction:
A transducer is a device that converts one form of energy
to another.
Energy types include (but are not limited to) electrical,
mechanical, electromagnetic (including light), chemical,
acoustic or thermal energy.
While the term transducer commonly implies the use of a
sensor/detector, any device which converts energy can be
considered a transducer.
Transducers are widely used in measuring instruments.
They are used to measure non electrical quantities like
pressure, force, velocity, ph value, temperature, light
intensity etc. 420
Measurement of Non-Electrical Quantities
Classification/Applications of
Transducers:
Electromagnetic Transducers:
Antenna – converts propagating electromagnetic waves to
and from conducted electrical signals
Magnetic cartridge – converts relative physical motion to
and from electrical signals
Tape head, Disk read-and-write head - converts magnetic
fields on a magnetic medium to and from electrical signals
Hall effect sensor – converts a magnetic field level into an
electrical signal
Electrochemical Transducers :
pH probes
Hydrogen sensor
421
Measurement of Non-Electrical Quantities
Classification/Applications of
Transducers:
Electromechanical (electromechanical output devices are
generically called actuators):
Galvanometer
Rotary motor, linear motor
Vibration powered generator
Potentiometer
Accelerometer
Strain gauge
Air flow sensor
422
Measurement of Non-Electrical Quantities
Classification/Applications of
Transducers:
Electroacoustic Transducers:
Loudspeaker, earphone – converts electrical signals into sound
(amplified signal → magnetic field → motion → air pressure)
Microphone – converts sound into an electrical signal (air pressure
→ motion of conductor/coil → magnetic field → electrical signal)
Pickup (music technology) – converts motion of metal strings into
an electrical signal (magnetism → electrical signal)
Tactile transducer – converts electrical signal into vibration
( electrical signal → vibration)
Piezoelectric crystal – converts deformations of solid-state
crystals (vibrations) to and from electrical signals
423
Measurement of Non-Electrical Quantities
Classification/Applications of
Transducers:
Electro-optical (Photoelectric):
Fluorescent lamp – converts electrical power into incoherent light
Incandescent lamp – converts electrical power into incoherent
light
Light-Emitting Diode – converts electrical power into incoherent
light
Laser Diode – converts electrical power into coherent light
• Thermoelectric:
• Resistance temperature detector (RTD) - converts temperature into
an electrical resistance signal
• Thermocouple - converts relative temperatures of metallic
junctions to electrical voltage
424
Reference:wikipedia.org
Measurement of Non-Electrical Quantities
Advantages of Electrical Transducers:
The following are the advantages of Electrical Transducer Elements
425
Describe elements and requirements of
transducers?
426
Measurement of Non-Electrical Quantities
Requirement of Transducers:
A transducer may be thought of consisting of two
closely related elements (Fig. 1).
427
Measurement of Non-Electrical Quantities
Requirement of Transducers:
1. Sensing or Detecting Element. This element senses
detects the physical quantity under measurement.
2. Transduction Element. This element transfers the
output of the sensing element into, an electrical output.
428
Transducer’s Transfer Efficiency:
Transfer Efficiency:
429
Measurement of Non-Electrical Quantities
Requirement of Transducers:
(b) Requirements. The transducers should have the
following requirements:
1. Mechanical Ruggedness. The transducer should be
mechanically rugged. It should be able to withstand the
working conditions and mechanical strains.
2. Linear Output / Input Characteristic. The transducers
should have linear output/input characteristic.
3. Reliability. The results of the transducers should be
reliable. The result should also be stable and should not
change with temperature and other changes.
430
Measurement of Non-Electrical Quantities
Requirement of Transducers:
4. Good Response. If the input to the transducer varies
with time (in few cases, it does) the transducers should
have good response to the variations.
5. Low Noise. The noise / signal ratio of the transducers
should be low so that its output can be directly measured
with sufficient accuracy.
431
Measurement of Non-Electrical Quantities
Classifications of Electrical Transducers:
They can be classified as:
A) Primary Transducers
B) Secondary Transducers
The primary transducer basically senses and sends the signal without
any modification, like a displacement as a result of measurement of
displacement, force, pressure, temperature, etc. This displacement is
amplified in the second stage and used to indicate the reading in the
third stage.
A secondary transducer, works subsequent to a primary transducer
receiving the input from it. For example, ????????
432
Measurement of Non-Electrical Quantities
Classifications of Electrical Transducers:
Analog:
convert I/p quantity into an analog o/p
Analog o/p- a continuous fn. Of time
E.g. Strain gauge, thermocouple.
433
Measurement of Non-Electrical Quantities
Classifications of Electrical Transducers:
434
Measurement of Non-Electrical Quantities
Classifications of Electrical Transducers:
PASSIVE TRANSDUCERS:
435
Measurement of Non-Electrical Quantities
Classifications of Electrical Transducers:
436
Measurement of Non-Electrical Quantities
Resistance(wire) strain gauges
437
Measurement of Non-Electrical Quantities
Resistance(wire) strain gauges
For a uniform distribution of internal resisting forces, stress can be
calculated (Figure 2-1) by dividing the force (F) applied by the unit
area (A):
438
Measurement of Non-Electrical Quantities
Resistance(wire) strain gauges
440
Measurement of Non-Electrical Quantities
Resistance(wire) strain gauges
441
Measurement of Non-Electrical Quantities
Resistance(wire) strain gauges
442
Measurement of Non-Electrical Quantities
Types of Resistance(wire) strain gauges
443
444
A roll of Aluminum Foil
Measurement of Non-Electrical Quantities
Types of Resistance(wire) strain gauges
445
Measurement of Non-Electrical Quantities
Types of Resistance(wire) strain gauges
447
Measurement of Non-Electrical Quantities
Types of Resistance(wire) strain gauges
448
Measurement of Non-Electrical Quantities
Types of Resistance(wire) strain gauges
449
Measurement of Non-Electrical Quantities
Types of Resistance(wire) strain gauges
450
Measurement of Non-Electrical Quantities
Types of Resistance(wire) strain gauges
451
Measurement of Non-Electrical Quantities
Types of Resistance(wire) strain gauges
452
Measurement of Non-Electrical Quantities
Semiconductor Strain Gauges
453
Measurement of Non-Electrical Quantities
Semiconductor Strain Gauges
455
Measurement of Non-Electrical Quantities
Semiconductor Strain Gauges
457
Measurement of Non-Electrical Quantities
Semiconductor Strain Gauges (Example)
458
Measurement of Non-Electrical Quantities
Semiconductor Strain Gauges (Example)
459
What are the desirable characteristics for resistance wire
strain gauge?
460
Measurement of Non-Electrical Quantities
Capacitive Transducers
463
Measurement of Non-Electrical Quantities
Working of Capacitive Transducers
464
Measurement of Non-Electrical Quantities
Working of Capacitive Transducers
Advantages
It produces an accurate frequency response to both static and
dynamic measurements.
Its power requirement is very small.
Disadvantages
An increase or decrease in temperature to a high level will
change the accuracy of the device.
As the lead is lengthy it can cause errors or distortion in
signals.
466
How capacitive transducers are superior
to other transducers?
467
Chapter:9
AC Power Analysis
Know the difference between instantaneous power
and average power
Learn the AC version of maximum power transfer
theorem
Learn about the concepts of effective or rms value
Learn about the complex power, apparent power
and power factor
Understand the principle of conservation of AC
power
Learn about power factor correction
468