HPLC
HPLC
HPLC
Solvent Sample
Pump
Reservoirs Injector
Data
Column(s) Detector
System
Mobile Phase Reservoirs
Inert container with inert lines leading to the pump are required.
- Vacuum filtration
- Sparge with inert gas
Degassed solvent (N2 or He)
- Ultrasonic under vacuum
Constant pressure
pumps
Reciprocating Pumps
• One, two, or three pump heads
- more heads, less pulse
• Small head volumes (50 to 250 mL)
• Short piston stroke
• Inert pistons (generally sapphire)
• Continuous use (no refill time)
• Pulse dampeners
Constant flow rate pump
Non-pulsating flow
Syringe
Low flow rates (1 to 100 mL/min)
Pumps
Isocratic flow only
Low cost
Sample Introduction
• Valve-type injectors
- Six port fixed volume Rheodyne
reproducible injection volumes
variable loop size
easy to use, reliable
- Six port variable volume Waters
variable injection volumes without loop change increased
maintenance, operator skill required more expensive
Continuous injections operator free
A guard column is introduced before the analytical column to increase the life
of the analytical column by removing not only particulate matter and
contaminants from the solvents but also sample components that bind
irreversibly to the stationary phase. The guard column serves to saturate
the mobile phase with the stationary phase so that losses of this solvent
from the analytical column are minimized. The composition of the guard-
column packing is similar to that of the analytical column; the particle size
is usually larger. When the guard column has become contaminated, it is
repacked or discarded and replaced with a new one.
Detectors: Unlike gas chromatography, liquid chromatography has
no detectors that are as universally applicable and as reliable as
the flame ionization and thermal conductivity detectors. A major
challenge in the development of liquid chromatography has been
in detector improvement.
• Types of Detectors: Liquid chromatographic detectors are of two
basic types. Bulk property detectors respond to a mobile-phase
bulk property, such as refractive index, dielectric constant, or
density.
• In contrast, solute property detectors respond to some property
of solutes, such as UV absorbance, fluorescence, or diffusion
current, that is not possessed by the mobile phase.
• Absorbance Detectors: Is a Z-shaped, flow-through cell for
absorbance measurements on eluents from a chromatographic
column. Many absorbance detectors are double-beam devices in
which one beam passes through the eluent cell and the other
through a filter to reduce its intensity.
• Ultraviolet Absorbance Detectors with Filters: The simplest UV
absorption detectors are filter photometers with a mercury lamp
as the source. Most commonly the intense line at 254 nm is
isolated by filters. Deuterium or tungsten filament sources with
interference filters also provide a simple means of detecting
absorbing species.
• UV Absorbance Detector with Monochromator: There are
detectors that consist of a scanning spectrophotometer with
grating optics. Some are limited to uv radiation; others
encompass both uv and visible radiation. The most powerful uv
spectrophotometric detectors are diode-array instruments.
• Infrared Absorbance Detectors: Two types of infrared detectors
are offered commercially. The range of the first instrument is
from 2.5 to 14.5 m or 4000 to 690 cm-1. The second type of
infrared detector is based upon Fourier transform instruments.
• Fluorescence Detectors: Fluorescence is observed by a
photoelectric detector located at 90 deg to the excitation beam.
The simplest detectors employ a mercury excitation source and
one or more filters to isolate a band of emitted radiation. More
sophisticated instruments are based upon a Xenon source and
employ a grating monochromator to isolate the fluorescent
radiation. An inherent advantage of fluorescence methods is
their high sensitivity, which is typically greater by more than an
order of magnitude than most absorbance procedures.
• Refractive-Index Detectors: Refractive-index detectors have the
significant advantage of responding to nearly all solutes. That is
they are general detectors analogous to flame or thermal
conductivity detectors in gas chromatography. In addition, they are
reliable and unaffected by flow rate. They are, however, highly
temperature sensitive and must be maintained at a constant
temperature to a few thousandths of a degree centigrade.
Furthermore, they are not as sensitive as most other types of
detectors.
• Electrochemical Detectors: These devices are based upon
amperometry, polarography, coulometry, and conductometry.
They appear to offer advantages, in many instances, of high
sensitivity, simplicity, convenience, and widespread applicability.
• Mass Spectrometric Detectors: A problem in coupling liquid chromatography with
mass spectrometry is the enormous mismatch between the relatively large solvent
volumes and the vacuum requirements. Several interfaces have been developed for
solving this problem. In a first type the eluent from the column is split, with only a
tiny fraction being introduced directly into the mass spectrometer. In a second type
of interface the effluent is deposited on a continuous, moving-belt or moving-wire
that transports the solvent and analyte to a heated chamber for removal of the
former by volatilization.
Partition Chromatography
Partition chromatography can be subdivided into
(i) liquid-liquid chromatography and
(ii) bonded-phase chromatography.
M+E- + A- M+A- + E-
IEC Stationary Phases
• Silica based materials
- Pellicular particles
• Organic materials
- porous beads
- styrene/divinylbenzene crosslinked co-
polymers
- methacrylic acid/divinylbenzene
crosslinked co-polymers
• Inorganic materials
IEC Stationary Phases
• Ce3+ > Al3+ >> Ba2+ > Pb2+ > Ca2+ >
Ni2+ > Cu2+ > Mg2+ > UO22+ >> Tl+ >
Ag+ > K+ > NH4+ > H+
Anion M+A- Bond Strengths
• Citrate > SCN- > CrO42- > SO42- > NO3- >
Br- > CN- > Cl- > HCO3- > HCOO- > OH-