DEPARTMENT OF TECHNICAL EDUCATION
ANDHRA PRADESH
Name : D. Simhachalam
Designation : Lecturer
Branch : A.E.I.E
Institute : K.D.R.G.P.T, Wanaparthy
Semester : III
Subject : Process Instrumentation-I
Subject code : AEI-305
Topic : Conductivity, Weight and Humidity
measurement
Duration : 100 minutes
Subtopic : Conductivity meters Specifications
Teaching Aids : PPTS, animations, Diagrams
AEI305.73 T0 74 1
Objectives
Upon completion of this topic, you will be able
to know about
• Specifications of components of conductivity
meter
• Specifications of conductivity meters
AEI305.73 T0 74 2
Specifications of components
Material for the cells :
• Epoxy
• Glass
• Quartz
• Pyrex Glass
AEI305.73 T0 74 3
Specifications of components
• The electrodes are made up of
• Platinum plates
• Distance between electrodes may be large
or small
AEI305.73 T0 74 4
Specifications of components
• For low conductivity measurement
• Range : 0 to 2µS/cm, 0 to 20µS/cm,
0 to 200µS/cm, 0 to 40MΩ·cm
• Cell Model : SC72SN-11(0.75m)
• Cell constant :0.05cm-1
AEI305.73 T0 74 5
Specifications of components
• For high conductivity measurement
• Range :0 to 2 mS/cm, 0 to 20 mS/cm,
0 to 200 mS/cm, 0 to 2 S/cm
• Cell Model : SC72SN-41(0.75m)
• Cell constant : 50cm - 1
AEI305.73 T0 74 6
Specifications of conductivity meters
AEI305.73 T0 74 7
Features
• Before seeing the specifications, let us see the
features of conductivity meters
• Waterproof, sealed case and connector cover to keep
out moisture
• Auto ranging allows 0 to 2000 mS/cm capability
• Large, easy-to-read display
AEI305.73 T0 74 8
Features
• Automatic Temperature Compensation (NaCl curve
or User-defined coefficient)
• Self-Diagnostics warn of fault conditions
• Auto power off extends battery life
AEI305.73 T0 74 9
Features
• 300 Points of Data Memory (DM)
• DM stores conductivity value, temperature, date, and
time.
• Audio Alarm Function
AEI305.73 T0 74 10
General specifications
Measuring Ranges
• Conductivity : 0 to 2000 mS/cm
(dependent on sensor selected)
• Resistivity : 0-40.0 MΩ.cm
(high purity sensor only)
• Temperature : 0 to 80°C
AEI305.73 T0 74 11
General specifications
Resolution
• Conductivity : 0.001 µS/cm
(for 0 to 2 µS/cm range)
• Resistivity : 0.1 MΩ.cm
• Temperature : 0.1°C
AEI305.73 T0 74 12
General specifications
• Display : Digital LCD
• Repeatability : ±2%
• Process Temperature : 0 to 80°C
• Ambient Temperature : 0 to 50°C
• Power Supply : Two AA size Alkaline
Batteries
AEI305.73 T0 74 13
Model SC72
Personal Conductivity Meter
AEI305.73 T0 74 14
AEI305.73 T0 74 15
Specifications of Personal Conductivity
Meter
• Measurement range
• Conductivity : 0 to 2 µS/cm through 0 to 2 S/cm
(depends on sensor to be
used)
• Resistivity : 0 to 40.0 MΩ·cm
• Temperature : 0 to 80°C
AEI305.73 T0 74 16
Specifications of PCM
• Resolution
• Conductivity : 0.05% of full scale
• Resistivity : 0.1 MΩ·cm
• Temperature : 0.1°C
AEI305.73 T0 74 17
Specifications of PCM
Repeatability
• Conductivity : ±2% (±5% when general-purpose type
sensor is used in the range of 0 to 2 ms/cm)
• Display : Digital LCD
AEI305.73 T0 74 18
Specifications of PCM
Indication
• Conductivity or resistivity, solution temperature and
temperature coefficient (simultaneously), various
messages
• Range switching : Automatic/manual
AEI305.73 T0 74 19
Specifications of PCM
Temperature compensation
• Temperature coefficient : (0 to 9.99%/°C) or NaCl
coefficient, reference temperature at 25°C
• Ambient temperature : 0 to 50°C
• Construction :IP67 (JIS C0920)
AEI305.73 T0 74 20
Specifications of PCM
• Weight :Approx. 220 g (without sensor)
• Power source : 2 size AA dry cell batteries
• Battery life : Approx. 200 hours*3,
Auto Power Off function
• Functions : Data memory (300 points),
alarm clock
AEI305.73 T0 74 21
Summary
We have discussed about
• Specifications of components of conductivity meter
• Specifications of conductivity meters.
AEI305.73 T0 74 22
QUIZ
1. The conductivity cells are made up of
a) Epoxy
b) Quartz
c) Glass
d) All the above
AEI305.73 T0 74 23
• The conductivity electrodes are made up of
a) Platinum
b) Pyrex Glass
c) Quartz
d) None
AEI305.73 T0 74 24
1. The following is the general measuring range
specification for conductivity
a) 0 to 2000 mS/cm
c) 0 to 200 mS/cm
e) 0 to 3000 mS/cm
g) 0 to 300 mS/cm
AEI305.73 T0 74 25
1. The following is the general measuring range
specification for Resistivity
a) 0-20.0 MΩ.cm
c) 0-30.0 MΩ.cm
e) 0-40.0 MΩ.cm
g) 0-50.0 MΩ.cm
AEI305.73 T0 74 26
Frequently asked questions
1. List the specifications of components of the
conductivity meter.
3. List the general specifications of conductivity meters.