Chapter 1 Introduction in Biomedical Sensors-1
Chapter 1 Introduction in Biomedical Sensors-1
Note –These lectures taken from text book and different internet resources :
Modified by Dr. Noman AL Naggar
UST-Sana’a 2016
And modified by Dr. Nasr Al Awdi
UST-Aden 2023
BBM322- Biomedical instrumentation II
3-Slide
Introduction to sensors
:Introduction
A sensor (also called detector) is a converter, that measures a physical •
quantity and converts it into a signal which can be read by an observer
.or by an electronic instrument
Sensor/Transducer -Integrated with other parts to “read” out the signal (electrically, •
optically, chemically)
Some are used in vivo to perform continuous, “invasive or non- •
invasive” monitoring of critical physiological variables
pressure, flow, concentration of gas –
Some are used in vitro to help clinicians in various diagnostic •
procedures
electrolytes, enzymes, metabolites in blood –
:Concepts
in vivo: inside a living body (human or animal) •
ex vivo: outside the living body •
in vitro: in a test tube •
in situ: right in the place where reactions happen (could be in the cells and • 4
Transducers
Transducer
a device that converts a primary form of energy into a-
corresponding signal with a different energy form
Primary Energy Forms: mechanical, thermal, electromagnetic, optical,
.chemical, etc
take form of a sensor or an actuator-
Sensor (e.g., thermometer)
a device that detects/measures a physical energy-
”acquires information from the “real world-
Actuator (e.g., heater)
a device that generates a signal or stimulus-
sensor intelligent
real
feedback
world
actuator system
sensor
photodiode
light I
Sensors
Classification of Transducers
:The transducers can be classified in many ways, such as
By the process used to convert the energy into an electrical )i(
:signal
Active Transducers .1
a transducer that converts one form of energy directly into another. For _
example: photovoltaic cell in which light energy is converted into electrical
.energy
Passive Transducers .2
a transducer that requires energy to be translate changes due to the measured ._
For example: a variable resistance placed in a bridge in which the voltage at the
.output of the circuit reflects the physical variable
Temperature measurement •
Thermocouples: Thermoelectric effect (Discovered by Seebeck in 1821), Peltier –
,effect, Thomson effect
Thermistors –
Radiation thermometry –
Fiber-optic temperature sensor –
12-Slide
Measurements systems
Optical Measurements are used in biomedical sensors
Optical Measurement
Optical systems are widely used in medical diagnosis, especially in clinical-chemistry –
.lab
Application example: Blood or tissue sample analysis, oxygen saturation of hemoglobin, –
.cardiac output
Radiation sources: Tungsten lamp, Arc discharges, LEDs, LASERS –
: Figure
a) General block diagram of an optical
.instrument
b) Highest efficiency is obtained by
using an intense lamp, lenses to gather
and focus the light on the sample in the
.cuvette, and a sensitive detector
c) Solid _ state lamps and detectors
.may simplify the system
13-Slide
Measurements systems
Optical Measurements are used in biomedical sensors
14
Signal Acquisition
Physiological output Signals
Three types of output signal
:Self-generating (active) transducers-1
The electrical signal output of –
transducer is generated from another
.form of input energy
:Tandem transducers-3
The original input signal energy is –
converted to a final output of electrical
energy through two or three effects or
.conversions in tandem
Signal Acquisition
Signal Acquisition
Medical Instrumentation typically entails monitoring a signal off
the body which is analog, converting it to an electrical signal,
.and digitizing it to be analyzed by the computer
Connecting Sensors to Microcontrollers
sensor µC keypad
signal timing
sensor memory display
:Analog instrument
Output Output
Input Input
Input
Sensitivity Error
The pressure transducer discussed above may have an actual sensitivity
.of 7.8 mV/V/mm Hg instead of 10 mV/V/mm Hg
:Sensor Terminology
Range = Maximum and Minimum values of applied
.parameter that can be measured
If an instrument can read up to 200
mmHg and the actual reading is 250
mmHg then you have exceeded the
.range of the instrument
:Sensor Terminology
Dynamic Range: total range of sensor for minimum to
maximum. if your instrument can measure from -10V to
+10 V your dynamic range is 20V
Precision = Degree of reproducibility denoted as the
range of one standard deviation σ
Resolution = smallest detectable incremental change of
input parameter that can be detected
What’s requirements to Sensors
Accuracy -2 Sensor Terminology
Xi Xo
What’s requirements to Sensors
Offset Error -3 Sensor Terminology
Output Output
Input Input
Offset Error
Zero offset error
The difference between the actual output value and the specified
.output value under some particular set of conditions
What’s requirements to Sensors
Linearity -4 Sensor Terminology
Linearity = Extent to which actual measure curved or
.calibration curve departs from ideal curve
Nonlinearity (%) = (Din(Max) / INfs) * 100%
Nonlinearity is percentage of nonlinearity
Din(max) = maximum input deviation
INfs = maximum full-scale input
Full Scale Input eal r e
Id as u
M e
Output Din(Max)
Input
What’s requirements to Sensors
Hysteresis -5 Sensor Terminology
Hysteresis = measurement of how sensor changes
.with input parameter based on direction of change
Output = F(x)
P
F2
F1 Input = x
B
Q
What’s requirements to Sensors
Response Time -6 Sensor Terminology
Ton Time
Dynamic Linearity-7 Sensor Terminology
Measure of a sensor’s ability to follow rapid changes in the input parameters.
Difference between solid and dashed curves is the non- linearity as depicted by
.the higher order x terms
Where F(X) is the output signal, while the X terms represent the input parameter
.and its harmonics, and K is an offset constant
K K
Input X Input X
Frequency Response-8
Ideal System
:In Ideal
When you look at the frequency response of an instrument, ideally
.you want a wideband flat frequency response
Av Av = Vo/Vi
1.0
Av Av = Vo/Vi
1.0
0.707
FL FH
Frequency (w) radians per second
:Frequency Response-8
Examples of Filters
Ideal Filter has sharp cutoffs and a flat pass
band
Most filters attenuate upper and lower
frequencies
Other filters attenuate upper and lower
frequencies and are not flat in the pass
band
Sensor Calibration
Sensors can exhibit non-ideal effects
offset: nominal output ≠ nominal parameter value
nonlinearity: output not linear with parameter changes
cross parameter sensitivity: secondary output variation with other
.parameter, e.g., temperature
digital calibration
6.000
5.000
,T = a + bV +cV
1001
2
1010
Frequency (MHz)
4.000
Compensation
1.000
39-Slide