Unit Iii
Unit Iii
Unit Iii
• The validation parameters that will be evaluated will depend on the type of
method to be validated.
• Analytical methods that are commonly validated can be classified into
three main categories:
– Identification
– Testing for impurities
– Assay
A normal validation protocol should contain the following minimum
contents:
• Objective of the protocol
• Validation parameters that will be evaluated
• Acceptance criteria for all the validation parameters evaluated
• Details of the experiments to be performed
• Draft analytical procedure.
1. Selectivity/specificity
6. Stability
Many analytes readily decompose prior to chromatography
investigations, for example during the preparation of the sample solutions,
during extraction, clean-up, phase transfer, and during storage of prepared
vials. Under these circumstances, method development should investigate the
stability of the analyte. Accuracy test takes care of stability.
It is required to mention in the method how long a sample after extraction can
be stored before final analysis, based on the duration taken for accuracy test.
7. Limit of detection(LOD) and limit of quantitation(LOQ)
The term LOD is defined as the lowest concentration at which the instrument
is able to detect but not quantify and the noise to signal ratio for LOD should be 1:3.
The term LOQ is defined as the lowest concentration at which the instrument is able
to detect and quantify. The noise to signal ratio for LOQ should be 1:10.
8. Robustness (or ruggedness)
It is the ability of the procedure to provide analytical results
of acceptable accuracy and precision under a variety of conditions.
The results from separate samples are influenced by changes in
the operational or environmental conditions.
Robustness should be considered during the development phase,
and should show the reliability of an analysis when deliberate
variations are made in the method parameters.