Calibration
Calibration
What is calibration
As per ISO/IEC Guide 99 ; International vocabulary of metrology — Basic and general concepts and associated terms(VIM) and
or as per JCGM 200:2012*
Calibration consists of operation that, under specified conditions, in a first step, establishes a relation between the quantity values with
measurement uncertainties provided by measurement standards and corresponding indications with associated measurement
uncertainties and, in a second step, uses this information to establish a relation for obtaining a measurement result from an indication
Note 1 : A calibration may be expressed by a statement, calibration function, calibration diagram, calibration curve, or calibration table.
In some cases, it may consist of an additive or multiplicative correction of the indication with associated measurement uncertainty
Note 2 :Calibration should not be confused with adjustment of a measuring system, often mistakenly called “self-calibration”, nor with
verification of calibration.
Wrong concept
Calibration refers to the act of evaluating and adjusting the precision and accuracy of measurement
equipment. Instrument calibration is intended to eliminate or reduce bias in an instrument's readings over
a range for all continuous values.
Why calibration is necessary ?
• It is impossible to make perfect hardware that would not need any form of error correction. Even
making the hardware good enough to eliminate the need for error correction for most devices would
be extremely expensive.
• The best balance is to make the hardware as good as practically possible, balancing performance and
cost. Calibration is then a very useful tool to improve measurement accuracy.
• Calibration of a measuring equipment or working standard enables the user to take note of the
deviations and errors from the nominal values so that corrections can be made to minimize the errors
during measurement.
• A well-calibrated instrument provides the confidence to the user as well as the consumer that the
service or the product meets the required specifications and ensures the acceptance in international
market.
Steps of calibration
1. Determine which procedure to be used during calibration.
2. Compare to a reference standard (full-range or as per customer test point).
3. Find out the error.
4. Adjust the equipment if capable. This step is not included in calibration. Still some organizations does it.
5. Check the traceability .The traceability, the unbroken chain of comparison linking to the highest level (At least to national level).
6. Estimate measurement uncertainty.
7. Record results to a calibration certificate.
A calibration report usually provides a set of results that show the performance of the instrument being
calibrated in comparison to a reference instrument or standard. Measurements are recorded for each
function, range and points tested, and the uncertainty of each measurement is reported.
Calibration certificate
A calibration certificate usually states the compliance of an instrument with the manufacturer's specification.
Measurements made by the instrument being calibrated are compared to the manufacturer's specification. If
the instrument is within specification, a statement that the instrument meets the manufacturer's specification
at the points tested is made in the conformance certificate.
Sample
calibration
certificate
Thank you
Verification
What is verification ?
As per ISO/IEC Guide 99 ; International vocabulary of metrology — Basic and general concepts and
associated terms(VIM) and or as per JCGM 200:2012*
• Provision of objective evidence that a given itemFulfils specified requirements
Example 1
Confirmation that a given reference material as claimed is homogeneous for the quantity
value and measurement procedure concerned, down to a measurement portion having a mass of 10 mg.
Example 2
Confirmation that performance properties or legal requirements of a measuring system are achieved.
What is verification : Let us make it simple
• Compliance to the Requirements of ISO Standards – under ISO 17025:2017 and ISO 9001:2015 Standards which controls and
requires calibration interval analysis
Let us check ISO 9001:2015 and ISO/IEC 17025 :2017 clause requirement
ISO 9001:2015 , Clause no 7.1.5.2 : Measurement traceability
I. When measurement traceability is a requirement or is considered by the organization to be an essential part of providing
confidence in the validity of measurement results, measuring equipment shall be: a) calibrated or verified, or both, at specified
intervals, or prior to use, against measurement standards traceable to international or national measurement standards;
II. The organization shall determine if the validity of previous measurement results has been adversely affected when measuring
equipment is found to be unfit for its intended purpose, and shall take appropriate action as necessary
Let us check ISO 9001:2015 and ISO/IEC 17025 :2017 clause requirement
ISO/IEC 17025:2017 Cl no 6.4.7 , 6.4.13, 7.1.5.2,7.7.1
I. When measurement traceability is a requirement or is considered by the organization to be an essential part of providing
confidence in the validity of measurement results, measuring equipment shall be: a) calibrated or verified, or both, at specified
intervals, or prior to use, against measurement standards traceable to international or national measurement standards;
II. The organization shall determine if the validity of previous measurement results has been adversely affected when measuring
equipment is found to be unfit for its intended purpose, and shall take appropriate action as necessary.
If you read these requirements, there are no specific guides or standards that are required to follow for calibration interval analysis or
method, therefore, any methods that work for you are ok.
But it is better to have published documents or guides as a reference like the ILAC G24
Objectives of calibration interval analysis
1. To designate Initial Calibration Interval – this is the starting calibration interval based on experience and recommendations.
2. Use to extend interval of calibration – this means that a given calibration interval will be extended for a specific period until a fixed
interval is reached. This is also applicable like for example during this pandemic where calibration is difficult to access and we want
to extend a little more temporarily.
3. Use to reduce interval of calibration – this is applicable when we encounter out of tolerance where calibration is reduced as per
implementation rule. See below example presentation
4. Use to determine the fixed calibration interval (final interval) – this is the main calibration interval that we need to achieve based
on actual use which should be adequately justified.
Types of interval
1. To designate Initial Calibration Interval – this is the starting calibration interval based on experience and recommendations.
2. Use to extend interval of calibration – this means that a given calibration interval will be extended for a specific period until a fixed
interval is reached. This is also applicable like for example during this pandemic where calibration is difficult to access and we want
to extend a little more temporarily.
3. Use to reduce interval of calibration – this is applicable when we encounter out of tolerance where calibration is reduced as per
implementation rule. See below example presentation.
4. Use to determine the fixed calibration interval (final interval) – this is the main calibration interval that we need to achieve based
on actual use which should be adequately justified.
Initial Calibration Interval
1. Initial Calibration Interval means an interval that we used initially as per the decision of the expert, this should not be the final
calibration frequency to be used, this is just our starting point. Since we do not have yet data to justify this interval (in most cases
for start up), it is known as ‘Engineering Intuition’.
2. ILAC G-24 : “The so-called “engineering intuition” which fixed the initial calibration intervals, and a system which maintains fixed
intervals without review, are not considered as being sufficiently reliable and are therefore not recommended. “
Initial Calibration Interval determination helplines
The decision on where to base the initial calibration interval depends solely on you as the user. These could be based on below criteria:
1. Manufacturer Requirements – recommended by manufacturer.
2. On the frequency of use – the more it is used, the shorter the calibration interval.
3. Required by the regulatory bodies (example: required by the government).
4. Past experience of the user with the same type of instrument.
5. Based on the criticality of use. – more critical instruments have higher accuracy or very strict tolerance, therefore shorter calibration
interval.
6. Customer Requirements.
7. Conditions of the environment where it is being used.
8. Published Documents
Initial interval
Initial calibration intervals in some cases could become the ‘fixed/final calibration interval’, considering that we already have evidence
to justify why we decide this calibration interval for a specific instrument
Determination of fixed interval : Floating interval method
2 Initial interval 6
251.0 3 Initial interval 6
250.8 4 Initial interval 6
Standard value (MPa)
250.6
UCL at 100% Tol 4 Initial interval 6
250.4 UCL at 80% Tol
5 New adjusted 12( 6+6X100%)
interval(Additional 100%)
250.2 UUC value
6 Fixed interval 12
250.0
Standard value 7 New adjusted 24(12+12X100
interval(Additional 100%) %)
249.8
8 Final interval 24
249.6 LCL at 80% Tol
LCL at 100% Tol 9 Fixed interval 24
249.4
10 Fixed interval 24
249.2 12-12X50%
249.0
6 6 6 6 12 12 24 24 24 24
Interval (Month)
Thank you