4) BC-6000 Software System - Service Training

This document provides a 3-page summary of service training for the BC-6000 Series hematology analyzer. It includes sections on startup and calibration procedures, key features like reproducibility and special rack setup, how to perform upgrades, export data, backup and recovery procedures, and an overview of the software system layout and various setup options.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPTX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
57 views49 pages

4) BC-6000 Software System - Service Training

This document provides a 3-page summary of service training for the BC-6000 Series hematology analyzer. It includes sections on startup and calibration procedures, key features like reproducibility and special rack setup, how to perform upgrades, export data, backup and recovery procedures, and an overview of the software system layout and various setup options.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPTX, PDF, TXT or read online on Scribd
You are on page 1/ 49

BC-6000 Series Service Training

IVD Technical Support


October, 2017
Version : V1.0
Startup---Skip the initialization
 Click the 4 corners in sequence when the green bar appears.
 After the basic initialization(within 1 min), input service password to skip
subsequent initialization steps.

Click Click

Click Click
Calibrator --- Calibration
 Open Vial Model: AL WB mode + OV PD mode + OV Micro WB mode
 Close Tube Model: CT WB mode + CT PD mode + CT Micro WB mode
一、常用功能 - 校准物校准
PARAMETER CALIBRATION PRINCIPLE

UC2: Only can manually edit


UC2:只能手动修改
Target
用户靶值 Value WB全血CDR
CDR Mode PD CDR Mode
预稀释CDR 用户校准
User Calibrate
WBC UC1- W WBC UC2-W WBC
HGB U C1- H HGB UC2-H HGB
RBC U C1- R RBC UC2-R RBC
MCV UC1- M MCV UC2-M MCV
PLT UC1-P PLT UC2-P PLT

厂家全血靶值
WB Target Value WB全血CDR
CDR Mode PD CDR Mode
预稀释CDR 厂家校准
Factory
WBC MC1- W WBC TC3-W WBC Calibrate
HGB MC1-H HGB TC3-H HGB
RBC MC1-R RBC TC3-R RBC
MCV MC1- M MCV TC3-M MCV
PLT MC1- P PLT TC3-P PLT
RBC-O MC1-RO RBC-O TC3-RO RBC-O
PLT-O MC1-PO PLT-O TC3-PO PLT-O
WBC-D WBC-D
TC1 TC4
Micro Blood
Micro Blood
Target Value
厂家微量全血靶值 微量全血CDR
CDR mode
WBC MC2- W WBC
HGB MC2-H HGB
RBC MC2-R RBC
MCV MC2- M MCV
PLT MC2- P PLT
RBC-O MC2-RO RBC-O
PLT-O MC2-PO PLT-O
WBC-D
TC2
一、常用功能 - 重复性
PERFORMANCE---REPRODUCIBILITY
PERFORMANCE---BACKGROUND
SAMPLE INFO ACQUIRING MODE --- COMMUNICATION SETUP
 Select LIS or LabXpert in Data channel menu
 Select Bi-directional LIS if needed
 If the customer use LabXpert, we can choose Run as per worklist
SAMPLE INFO ACQUIRING MODE --- MODE

 Run as Bi-directional LIS or Worklist


---End users can use Custom Mode in Open Vial Mode
---End users can select different modes
---End users can choose manually input or auto increase while typing in the
sample ID
 If you choose auto scan barcode:
---End user should manually input the initial ID. During the test, the sample
ID will auto increase.
---Auto scan barcode option will be selected while using bi-directional LIS.
---Auto scan barcode option will not be selected while running as per
worklist.
一、常用功能 - 复检
KEY FEATURES--- RE-EXAM
 If connected to LabXpert
--- Machine will use the re-exam rules in LabXpert.
 If not connected to LabXpert
--- Machine will use 3 built-in Re-exam rules
--- Re-exam will happen during autoloading procedure. After piercing, the rack
will feed one step forward and wait for the result.
--- If the result triggers re-exam rules, the rack will be moved one step
backward. The triggered sample will be mixed again and test again.
KEY FEATURES--- SPECIAL RACK SETUP

Path: Setup --- System Setup --- Rack Setup


UPGRADE
 Menu --- Status --- Version Info. --- Details --- Update

 Tips:
 Software package name : update.tar.gz . DO NOT unzip the file.
 Flash disk should be in FAT32 format.
 Update.tar.gz is for BC-6000 series. Update_cal.tar.gz is for CAL6000 series.
DO NOT mix them up.
DATA EXPORT
Sample result data export
Table review , next page , click EXPORT

 Tips:
1. Use the filter to select the data you want.
2. Select inf file and other information
3. 30,000 data needs 50G storage space.
DATA EXPORT
 One key export
Service Advanced toolbox Debug data export
DATA EXPORT

 Notice
The folder name will be SN+EXPORT DATE
BACKUP & RECOVERY
 During the deep sleep status and shut down process, data will be backup
into flash chip in the mainboard

 After replace SSD, engineer should log in using the service account. If there
are backup file in the mainboard, the machine will give a prompt which will
allow you to do data recovery.
 If you only replace mainboard, after logging in using service account. The
machine will ask you to do data backup from SSD into flash chip in the new
mainboard.
 If you only replace other modules/power driver board/autoloader board, it
will not affect the user and factory parameters. Data Backup
 SSD can upgrade from lower version to higher version.
SSD FLASH

Data Recovery
Software System --- Software Layout Description

4
Software System---Graph Review
Software System---Graph Review---Other Parameter
Software System---Graph Review---Special Info
Software System --- Graph Review --- Trace Info
The screen will display the reagent and QC information of the sample you select
Software System --- Graph Review --- Edit Result
Click the parameter and edit the result.
There will be an ‘E’ behind the parameter you edit in table review and graph review page
Software System --- Communication Setup

Path: Click ‘Menu’- ‘Setup’- ‘System Setup’- ‘Communication’


Software System --- Barcode Setup

Longest Length: 20 Digits

Select the corresponding


digit length

Path:
Click ‘Menu’- ‘Setup’- ‘System Setup’- ‘Barcode’
Software System --- Get Sample Info Setup

Path : Click ‘Menu’- ‘Setup’- ‘Auxiliary Setup’- ‘Get Sample Information’


Software System --- Auto-loading

Path: Click ‘Menu’- ‘Setup’- ‘Auto-loading’


Software --- Re-exam Rules Setup
Software System --- Reference Range Setup
Software System --- Microscopic Parameter Setup
Software System --- Auto Startup/Shutdown Setup

Path: Click ‘Menu’- ‘Setup’- ‘Auto Startup/Shutdown Setup’


Software System --- Maintenance

Path : Click ‘Menu’- ‘Setup’- ‘Maintenance’


Software System --- Gain Setup
Software System --- Flag Alarm Sensitivity
Software System --- Other Setup

Path: Click – ‘Menu’– ‘Setup’- ‘Auxiliary Setup’- ‘Other Setup’


Software System --- CV Calculation

1. Click Menu – ‘Table Review’


2. Select the samples which want to calculate the CV
3. Click CV button
4. The screen will display Mean Value/ SD Value/ CV Value
Software System --- QC file setup
Software System --- Manual Calibration
Software System --- Calibrator Calibration
Software System --- Fresh Blood Calibration
Before using fresh blood to do the calibration, engineers should make sure:
1. WBC/RBC/PLT have good morphology under the microscope.
2. The result of the fresh blood are within reference range
3. The volume is enough for the whole calibration test
Software System --- Print Setup
Software System --- Reagent Setup
Software System --- Reagent Setup
Software System --- Service – Maintenance - Reagent
Software System --- Service – Maintenance - Fluidic
Software System --- Service - Log
Software System --- Export Log
Software System --- Status