LC-2010 Road Show
LC-2010 Road Show
LC-2010 Road Show
1
PRODUCT OVERVIEW
Fast injection System integrity
- 15 sec / 10 uL injection - Pre-inspected system performance
High sample capacity Automatic Validation
- 350 samples ( 1mL), MTP : 4 - Modules and system
SIL
Oven
Pump UV
Automatic operation Intelligent LC analyzer
- Automatic system preparation - Method transfer with FD
Timer
2 operation Simple maintenance
- Automatic warm-up/ shutdown - All front access
DESIGN OVERVIEW
Auto injector Others
- 350 ( 1mL), 210(1.5mL), 100(4mL) - Column management device
- 100uL (std), Max. 2mL (op) - FD in MS-DOS format
- Automatic rack recognition
Pump UV detector
- Built-in active inlet valve (AIV) - Cell with temperature control
-3Automatic drain - Automatic cell recognition
- Automatic seal wash - Built-in Hg lamp for validation
PRINCIPAL SOLVENT DELIVERY UNIT SPECIFICATIONS
5
LC-2010 CONCEPT :~ HAVE ~
H : High Throughput
•High Speed Injection
– 15 seconds (when injecting 10L)
•Multiple Sample Processing
– 350x Samples (1mL vial)
– 4x Microtiter Plates
6
LC-2010 CONCEPT :~ H AVE ~
A : Automation
•Automated Analytical Operation
– Auto Start Up
–Auto Purge
–Auto Baseline Check
–Auto Shutdown
7
LC-2010 CONCEPT :~ HA VE ~
V : Validation
•Auto Validation
•Guarantee of system performance
•Column Management Device
8
LC-2010 CONCEPT :~ HAV E
E : Ease of Use
•Graphical Operation System
–GUI Capability
–Wizard Function
•Front access for maintenance
9
SOLVENT DELIVERY UNIT FEATURES
10
PRINCIPAL DETECTOR SPECIFICATIONS
・ Light Source: Deuterium lamp
・ Wavelength Range: 190 to 600nm
・ Spectral Bandwidth: 8 nm
・ Wavelength Accuracy: below +/- 1 nm
・ Wavelength Reproducibility:
below +/- 0.1 nm
・ Cell Optical Path: 10 mm
・ Cell Volume: 8L
・ Drift: below 1×10-4AU/hr (250 nm , air in cell )
・ Noise Level: below +/- 0.25 x10-5AU ( as above )
・ Cell Temperature Control System: 3 switching steps - OFF,
High, & Low
11
DETECTOR FEATURES
・ Adoption of a temperature controlled cell unit
- Maximizes the reproducibility of quantitative values
・ Built-in validation device (upgraded validation)
- Built-in mercury lamp to check wavelength accuracy
- Built-in filter to check absorbance accuracy
・ Automatic cell recognition
capability
12
EFFECT OF THE TEMPERATURE CONTROLLED CELL
The temperature controlled cell unit maximizes stability
・ It eliminates variations in the quantitative value due to thermal expansion
of the mobile phase or temperature fluctuations in the sample spectrum.
Quantitative values are kept constant irrespective of the measurement time
(morning / evening) or season (summer / winter). → Reproducibility of
quantitative values
・ Stable analysis even in the short wavelength region where there are large
variations in the absorption spectrum
Shortening of stabilization times → Improved productivity
Spectrum at low temperature
Measurement Wavelength
13
PRINCIPAL AUTOSAMPLER SPECIFICATIONS
15
PRINCIPAL COLUMN OVEN SPECIFICATIONS
・ Block heating system (with preheating)
・ Temperature Settings Range: 4 to 60deg. C (1deg. C step)
・ Temperature Control Range:
(room temperature‐15deg. C) to 60deg. C
・ Temperature Control Precision: +/- 0.1deg. C
・ Size within oven: 2x 250 mm columns
16
VALIDATION SUPPORT (1)
Auto Validation
- The advantages of an all-in-one system are put to effective use
- Fully automatic
- Approximately 1.5 hours with the Isocratic mode
- Approximately 3 hours even if the Gradient Check is included
17
VALIDATION SUPPORT (2)
・ Performance Check
- Validation support through the Wizard
・ System Check
- Validation report
・ IQ/OQ documents are attached as standard
18
SUMMARY
LC-2010CHT
HAVE
• Configured for High Throughput (H)
• Automation (A)
• Validation (V)
• Ease of Use (E)
Improved Basic Performance
• Dynamic Inlet Valve
• Temperature controlled cell unit