UT Lesson7 08 12
UT Lesson7 08 12
UT Lesson7 08 12
Introduction
Introduction
• Equipment calibration is the process of repeatedly
verifying that the ultrasonic equipment is
performing as intended.
• Calibration is carried out at the transducer level,
the instrument level and the integrated system
level.
• The frequency of calibration is based on practical
field experience and is often mandated through
consensus codes and standards.
Equipment Calibration
Introduction
Before tests are begun,
transducers are usually
checked for their general
condition and conformance
to specified performance
criteria, such as beam
angle, depth resolution
and absence of excessive reverberation noise.
Horizontal and vertical screen linearity, as well as
proper dB control function, are also checked.
Equipment Calibration
Amplitude Linearity
• In analog instruments, measurements of signal
strength (pulse height) and transit time (related to
distance from the sending transducer) are taken
directly from the display screen.
• It is important in these instruments that the visual
A-scan axes be directly proportional to incoming
signal strength (vertical axis) and expended time
(horizontal axis).
Equipment Calibration
Amplitude Linearity
• Onsite checks of amplitude linearity (vertical axes)
can be performed by observing how pairs of
pulses, which differ in amplitude by some fixed
amount, maintain their relative amplitude difference
while changing the instrument’s amplification.
• If the amplifier is linear, the ratio of the two pulses
will remain the same as the gain of the instrument
is changed over its operational range.
• ASTM E-317 provides instructions for how this
procedure can be performed.
Equipment Calibration
Amplitude Linearity
International Institute of
Welding (IIW) Block
IIW Block
IIW Block
This block provides the operator with many options for
straight beam calibrations:
• For straight beam calibration, the block can be
scanned from the side (2.5 cm [1 in.] thickness) to
set up screen widths in 2.5 cm (1 in.) multiples.
• Or it can be scanned from the edge through the
10 cm (4 in.) width for wider screen widths.
• In addition, the 0.2 cm (0.08 in.) notch can be viewed
from the opposite side of the block to check
transducer resolution.
Equipment Calibration
IIW Block
Performance of a
straight-beam inspection
of the base material
adjacent to a weld,
looking for laminations or
small laminar reflectors,
is required by AWS code
prior to shear wave
inspection.
Equipment Calibration
IIW Block
AWS Calibration
The indication rating is calculated using:
A–B–C=D
where A is the indication level in decibels (dB), B is
the reference level in decibels (dB), C is the
attenuation factor and D is the indication rating.
Equipment Calibration
AWS Calibration
• The formula is simple to use and the results are
objective, leaving no room for error.
• Due to the ease of defining the discontinuity value,
the AWS code presents a defined method for
deciding whether a discontinuity is acceptable.
• A personal study of section 6 of the AWS code will
provide you, the shear wave technician, with a
wealth of information for the proper conduct of
shear wave weld inspection.
Equipment Calibration
Indication Evaluation
(dB Rating Chart Example)
Equipment Calibration
AWS Calibration
The indication rating by itself is just a number
(without units). To determine whether or not the
indication is rejectable, the governing code or
specification has to give ranges of values for
rejection. A typical set of ranges might be as follows:
1. If the indication rating is less than +5, the
indication is rejectable regardless of length.
2. If the indication rating is from +6 through +9,
the indication might be rejectable if the length
is greater than 1.9 cm (0.75 in.).
Equipment Calibration
AWS Calibration
3. If the indication rating is from +10 through
+12, the indication might be rejectable if the
length is greater than 5 cm (2 in.).
4. If the indication rating is greater than +12, the
indication might be acceptable regardless of
length.
The above ranges are only examples of how a code
or specification might define indication ratings. For
actual values, the governing specification or code
should be consulted.
Equipment Calibration
DSC Block
A third commonly used
calibration block is the
distance sensitivity
calibration (DSC)
block. This type of
block has two radii of
2.5 and 7.6 cm
(1 and 3 in.).
Equipment Calibration
DSC Block