Permanent Pacemaker: Supervisor: DR Pipin Ardhianto SPJP (K), Fiha
Permanent Pacemaker: Supervisor: DR Pipin Ardhianto SPJP (K), Fiha
Permanent Pacemaker: Supervisor: DR Pipin Ardhianto SPJP (K), Fiha
Supervisor:
dr Pipin Ardhianto SpJP(K), FIHA
PACEMAKER
Devices that can be placed in your body, usually by surgery, to support the electrical
system in your heart.
When intrinsic cardiac automaticity or conduction integrity fails, the electrical
excitability of cardiac tissue allows a small, external electrical stimulus to drive
myocytes to threshold Pacemakers provide that external stimulus
1. Mulpuru SK, Madhavan M, et al. Cardiac Pacemakers: Function, Troubleshooting, and Management: Part 1 of a 2-Part Series. J Am Coll Cardiol. 2017;69(2):189–210.
1. Mulpuru SK, Madhavan M, et al. Cardiac Pacemakers: Function, Troubleshooting, and Management: Part 1 of a 2-Part Series. J Am Coll Cardiol. 2017;69(2):189–210.
IMPLANTABLE PACEMAKER CIRCUIT
Leads
Implantable pulse generator IPG
(IPG):
Battery
Circuitry
Connector(s)
Leads or wires
Cathode (negative
electrode)
Anode (positive electrode)
Body tissue
Anode
Cathode
THE PULSE GENERATOR
Common battery compositions include:
Lithium-Iodine
Lithium silver vanadium oxide
with carbon monoflouride
Starting battery voltage will vary
Circuitry
depending on composition
Longevity
Dependent on impedance and output
Commonly ranges from 6-12 years Battery
PACEMAKER
Two leads
One lead implanted in the
atrium
One lead implanted in
the ventricle
Provides AV synchrony and
pacing support in both atrium
and ventricle if needed
DDD Pacemaker
TRIPLE CHAMBER SYSTEM
Three Leads:
Right Atrium
Right Ventricle
Left Ventricle (via the Coronary
Sinus vein)
Most commonly called a Bi-
Ventricular Pacemaker but also
called Cardiac Resynchronization
Therapy (CRT–P)
Paces both ventricles together to
“resynchronize” the beat
VVI EXAMPLE (60 BPM)
V V V V
P P S P
GENERAL TERMINOLOGY
The ability of the CIED to detect intrinsic myocardial depolarization (i.E., The
potential difference between the lead’s cathodal and anodal electrodes)
CAUSES :
• generally lead failure (lead fracture, lead
displacement)
• generator failure
• battery failure
• crosstalk inhibition
• oversensing
OUTPUT FAILURE
CAPTURE FAILURE
Capture failure occurs when the generated pacing stimulus does not initiate
myocardial depolarization.
On the surface ECG, pacing spikes are present, but they are not followed by a
QRS complex in the event of ventricular noncapture or by the lack of P waves
in the event of atrial noncapture
CAUSES :
• lead dislodgment
• low output
• lead maturation (inflammation or fibrosis at the electrode tip)
• lead or pacer failure (lead fracture, low pacing voltage, or elevated
myocardial pacing thresholds)
CAPTURE FAILURE
UNDERSENSING
Undersensing occurs when the pacemaker fails to detect spontaneous myocardial
depolarization, which results in asynchronous pacing
This typically results in the appearance of too many pacing spikes, as seen on ECG
Causes :
• pacemaker programming problems (improper sensing threshold)
• insufficient myocardial voltage signal
• lead or pacer failure (fibrosis, fracture, etc.)
• an electrolyte abnormality.
UNDERSENSING
UNDERSENSING
Initial management
• Consider placement of a magnet to enable
asynchronous pacing
Treatment of the underlying cause
• Correct electrolyte abnormalities, treat
ischaemia, and remove non-essential
medications
Programming changes
• Increase the sensitivity
Device changes
• Lead revision and/or insertion (if a lead
problem)
OVERSENSING
• Oversensing occurs when the pacemaker senses electrical signals that it
should not normally encounter, which results in inappropriate inhibition of
the pacing stimulus
• In addition to the native cardiac depolarization signals (P or R waves), any
electrical signal with sufficient amplitude and frequent occurrence can be
sensed and can inhibit the pacemaker when pacing is needed
Causes :
• Physiologic signals like t waves
• By myopotential (and nonphysiologic) signals like electromagnetic
interference
• A lead failure (an insulation break or a lead fracture)
OVERSENSING
OVERSENSING
Initial management
• If the patient is not pacemaker dependent, then monitoring may be
appropriate pending elucidation of the underlying cause and provision of a
definitive treatment
• If the patient is pacemaker dependent, then IV isoproterenol may be
attempted prior to placement of a transvenous temporary pacemaker lead
Programming changes
• Lower the sensitivity
• Lengthening of the blanking or refractory period
Device changes
• Generator/battery change (if a generator/battery problem)
• Lead revision and/or insertion (if a lead problem)
PSEUDOMALFUNCTION
• Ventricular safety pacing typically results in the appearance of 2 very closely atrial
and ventricular paced events on ECG
PACEMAKER MEDIATED TACHYCARDIA (PMT)
• Pacemaker mediated tachycardia is an endless-loop tachycardia,
sustained, in part, by the presence of the pacemaker
RUNAWAY PACEMAKER
• A runaway pacemaker is a rare medical emergency
• It is due to the total failure of a pacemaker generator, resulting in rapid tachycardia
• A magnet application is usually ineffective. Often the device must be manually disabled manual (disconnection of
the lead from the generator)
TWIDDLER SYNDROME
• It occurs due to unintentional or deliberate manipulation of the pacemaker pulse generator within its skin pocket by
the patient. This causes coiling of the lead and its dislodgement, resulting in failure of ventricular pacing.
• More commonly reported among elderly females with impaired cognition, the phenomenon usually occurs in the first
year following pacemaker implantation.
• Treatment involves repositioning of the dislodged leads and suture fixation of the lead and pulse generator within
its pocket.
TERIMA KASIH
MOHON BIMBINGAN
Troubleshooting PPM
Cause