Draeger
Draeger
Draeger
9000/9500
Subtitle
Location. date. first name last name
No environmental influences
Measuring results with
such as temperature and
laboratory quality due to gas
humidity when in Analysis
chromatography technology
measurement mode
1 – Gas sample inlet 3 – Pump 1 5 – Analysis PID 7 – Valve 9 - Carrier gas inlet
2 – Seeker PID 4 – GC bundle of 6 – Pump 2 8 – Activated carbon 10 – Gas sample outlet
columns filter
© Drägerwerk AG & Co. KGaA, 2020 7
Dräger X-pid® 9000/9500
Target compounds, LOQ & upper range
Target compounds CAS number Retention time, s LOD1, ppm LOQ2, ppm UR3, ppm
Acrolein 107-02-8 7.8 0.33 1.00 100
Benzene 71-43-2 19.3 0.02 0.05 25.0
Butadiene, 1,3- 106-99-0 6.4 0.07 0.20 25.0
Butanone, 2- 78-93-3 12.9 1.00 3.00 300
Butyl acrylate 141-32-2 125.5 - - -
Chlorobenzene 108-90-7 75.6 1.00 3.00 200
Dichloroethene, 1,1- 75-35-4 8.9 0.07 0.20 50.0
Dichloroethene, cis-1,2- 156-59-2 13.4 0.07 0.20 50.0
Dichloroethene, trans-1,2- 156-60-5 10.9 0.07 0.20 50.0
Epichlorohydrin 106-89-8 27.3 0.67 2.00 200
Ethyl acetate 141-78-6 14.6 1.00 3.00 300
Ethyl acrylate 140-88-5 24.9 1.00 3.00 200
Ethylbenzene 100-41-4 88.7 1.00 3.00 300
Ethylene oxide 75-21-8 6.8 0.33 1.00 100
Hexane, n- 110-54-3 13.7 0.33 1.00 100
Isobutylene 115-11-7 6.3 0.07 0.20 100
Methyl acrylate 96-33-3 14.4 0.67 2.00 200
Methyl bromide 74-83-9 6.8 0.17 0.50 100
Phosphine 7803-51-2 5.3 0.67 2.00 100
Propylene oxide 75-56-9 8.2 0.17 0.50 25.0
Styrene 100-42-5 111.3 1.00 3.00 300
Tetrachloroethylene 127-18-4 58.9 0.67 2.00 150
Tetrahydrofuran 109-99-9 16.5 1.00 3.00 200
Toluene 108-88-3 41.6 0.33 1.00 100
Trichloroethylene 79-01-6 24.9 0.33 1.00 100
Vinyl chloride 75-01-4 6.3 0.33 1.00 100
Xylene, m- 108-38-3 95.7 1.00 3.00 300
Xylene, o- 95-47-6 114.5 1.00 3.00 300
Xylene, p- 106-42-3 96.6 1.00 3.00 300
01 02 03 04
Pre-testing and localisation of Selection and start of an analysis Visualisation of the measuring signal Display of concentration per target
measuring points program with associated target in real-time to check plausibility compound and alarm when threshold
compounds is exceeded
6.45 ppm Duration of analysis
Total VOC Analysis program for 0.15 ppm butadiene
30 seconds
butadiene and benzene
2.58 ppm benzene
1.44 ppm other VOC
Accessories
Test and Case with Sampling Upgrade X-pid Connect
calibration gas Inlay for X- Accessories Mobile App to annual licence
isobutene & pid 9000/9500 (probes and Version 9500
toluene hoses)
3 registered patents
for X-pid 9000/9500
years of experience in gas
>70 detection technology
01 Clearance Measurement
02 Exposure Measurement
03 Emission Measurement
Solution:
04 Exploration Measurement X-pid 9000
or X-pid 9500
© Drägerwerk AG & Co. KGaA, 2020 13
Dräger X-pid® 9000/9500
Application Scenarios Challenge:
Low Threshold Limit Values
01 Clearance Measurement
02 Exposure Measurement
03 Emission Measurement
Solution:
X-pid® 9000 04 Exploration Measurement
or X-pid® 9500
© Drägerwerk AG & Co. KGaA, 2020 14
Dräger X-pid® 9000/9500
Challenge: Application Scenarios
Unknown danger
01 Clearance Measurement
02 Exposure Measurement
03 Emission Measurement
01 Clearance Measurement
02 Exposure Measurement
03 Emission Measurement
Performance
Requirements differ by the planned
number of measurements per year
and the performance requirements
for measuring systems.
Purchase Price
The X-pid 9000/9500 is
suitable due to the short measuring Tubes X-am 8000 X-pid 9000 / 9500
time and the absence of consumables
Regular user with low Multiple use of Demanding users with
for demanding users with a high
measurement volume clearance measurement high measurement
measurement volume. Investment costs device volume
are amortised starting at around 200
measurements per year.
© Drägerwerk AG & Co. KGaA, 2020 17
Dräger X-pid® 9000/9500
Technical Data
Environmental Temperature -10 to +35 °C Robust behavior – even at low temperatures and high
conditions Pressure 700 to 1.300 mbar humidity
Relative humidity 10% to 95% RH
Operating times typically 6 h,
reduces with lower ambient temperatures
Dimensions & ca. 132 x 281 x 56 mm (W × H × D) Small and light device instead of complex laboratory
weight sensor unit ca. 880 g (2 lb) analytics
Sensors* 10.6 eV PID with separation by a gas chromatograph
Limit of 0.05 ppm for benzene
quantification* 0.20 ppm for 1,3-butadiene
Upper range* 25 ppm for benzene
25 ppm for 1,3-butadiene
Selectivity* For benzene no cross sensitivity to toluene, ethyl benzene, High selectivity for valid results and
xylene, n-hexane and other VOCs with differing volatility reduced false alarms
Measurement 30 seconds for benzene and 1,3-butadiene simultaneously Short measuring cycles save time and staff at high
duration* in one analysis measurement volume
*Analysis Mode
© Drägerwerk AG & Co. KGaA, 2020 18
Dräger X-pid® 9000/9500
Markets & User
Target markets
User groups
Many thanks