Determination of Linear Attenuation Coefficient of Al, Fe and PB Applying Varying Channel Method
Determination of Linear Attenuation Coefficient of Al, Fe and PB Applying Varying Channel Method
Determination of Linear Attenuation Coefficient of Al, Fe and PB Applying Varying Channel Method
Al, Fe and Pb
Applying Varying Channel Method
Prelim Project
Group I
Supervisor
Ma Lwin Lwin Soe (Prelim-Yapa-6)
Dr Win Zaw
Ma Nway Nway Aung (Prelim-Yapa-7)
Lecturer
Department of Physics Ma San Yin Phoo Maung (Prelim-Yapa-10)
University of Mandalay
Ma Moe Moe Aye (Prelim-Yapa-13)
3
Introduction
When 𝛾-ray beam passes through an absorber, it is attenuated.
The degree of attenuation depends on the scattering and various absorption
processes.
The absorption coefficient 𝜇 can be derived from the Lambert-Beer law
x 1 I
I I 0e ln( 0 )
x I
where 𝐼0 = the incident 𝛾-ray intensity when measured without sample
𝐼 = the transmitted 𝛾-rays intensity through the sample
x = the sample thickness (cm)
The experimental mass-absorption coefficient (𝜇/𝜌) of elements is given by
1 I0
ln
x I
where 𝜌 = the material density (g/cm3).
4
Experimental Setup and Procedure
Source
Fiber Case
Absorber
Detector
Lead Shielding
PMT
PreAmp
HV Amp MCA PC
I0
I
x 5
CT – 600s
CG – 20
FG – 0.688
6
Energy Spectrum after passing through the absorber
30 channel 30 channel
35 channel 35 channel
40 channel 40 channel
7
Result
In this project, NaI(Tl) scintillation detector, standard calibration
sources
Cs-137 and Co-60, standard absorbers Aluminium and Lead are used.
But there is no standard ion absorber, normal cast iron sheets are used
instead of standard.
The spectrum analyzing was performed at the observed spectra with
various channel, 30 channel left and right from centroid, 35 channel
left and right from the centroid and 40 channel left and right from the
centroid.
8
Result
The sample analyzed data are shown in Table (1) and the graph
plotting with analyzed data is shown in Figure(1).
The linear attenuation data was taken from the slope of the graph.
Comparison of reference values and measured values are shown in
Table(2).
The typical attenuation chart is shown in Figure(2).
9
Table (1) Analysed data of observed spectrum (Al, 1173keV)
1 0 18104 18104 0
0.1
0.06 AlCo(1173keV)
Linear (AlCo(1173keV))
0.04
0.02
0
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7
Thickness(cm)
Figure (1) The graph of thickness vs ln(I0/I)
11
Table (2) Comparison of reference value and measured data with discrepancy
Ch_30 Al 0.20131 0.15314 0.14358 0.2048 0.1553 0.1414 -1.7336 -1.4370 1.5183
Ch_35 Al 0.20131 0.15314 0.14358 0.1991 0.1529 0.1457 1.0978 0.1567 -1.4765
Ch_40 Al 0.20131 0.15314 0.14358 0.1981 0.1576 0.1484 1.5946 -2.9124 -3.3570
Ch_30 Fe 0.83100 0.43514 0.40807 0.5863 0.4376 0.4202 -1.3816 -0.5653 -2.9725
Ch_35 Fe 0.83100 0.43514 0.40807 0.5868 0.4298 0.4195 -1.4681 1.2272 -2.8010
Ch_40 Fe 0.83100 0.43514 0.40807 0.5857 0.4427 0.4308 -1.2779 -1.7374 -5.5701
Ch_30 Pb 1.26000 0.70355 0.64001 1.171 0.7577 0.5736 7.0635 -7.6967 10.3764
Ch_35 Pb 1.26000 0.70355 0.64001 1.182 0.6714 0.652 6.1905 4.5697 -1.8734
Ch_40 Pb 1.26000 0.70355 0.64001 1.1919 0.6866 0.6258 5.4048 2.4092 2.2203
12
1.4
Linear Attenuation Coefficient
1.2
0.8
Pb
Fe
Al
0.6
0.4
0.2
0
660 760 860 960 1060 1160 1260 1360 1460
Energy (keV)
Figure (2) Typical Attenuation Chart for various energies
13
Discussion
According to table 2, the reference linear attenuation value and
measured values are agreed within the discrepancy of 0.1567 to
10.3764.
Actually , the measured value and reference value should be
matched almost 100% .
The greatest tolerance is 10%. It is due to the weak intensity of
sources, lack of collimation and less counting time.
14
THANK YOU
15