0% found this document useful (0 votes)
76 views46 pages

CH 502 Experimental & Analytical Methods Supplement - 1

This document provides an overview of various spectroscopic methods and instrumentation. It discusses topics like glass membrane electrodes, the electromagnetic spectrum, types of interactions between electromagnetic radiation and matter, sources of radiation, wavelength selectors, UV absorption spectroscopy, infrared spectroscopy using Michelson interferometers, NMR spectroscopy concepts involving spin states and magnetic fields, and an example of proton NMR spectroscopy of water. The document contains diagrams to illustrate spectroscopy concepts and instrumentation.

Uploaded by

aasthakatara
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPT, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
76 views46 pages

CH 502 Experimental & Analytical Methods Supplement - 1

This document provides an overview of various spectroscopic methods and instrumentation. It discusses topics like glass membrane electrodes, the electromagnetic spectrum, types of interactions between electromagnetic radiation and matter, sources of radiation, wavelength selectors, UV absorption spectroscopy, infrared spectroscopy using Michelson interferometers, NMR spectroscopy concepts involving spin states and magnetic fields, and an example of proton NMR spectroscopy of water. The document contains diagrams to illustrate spectroscopy concepts and instrumentation.

Uploaded by

aasthakatara
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPT, PDF, TXT or read online on Scribd
You are on page 1/ 46

CH 502 Experimental & Analytical

Methods

Supplement - 1
Glass Membrane Electrodes
• pH electrode • Ion Selective Field Effect
Transistor (ISFET) electrode

csrg.ch.pw.edu.pl/tutorials/isfet/
www.ph-meter.info
Spectroscopic Methods

• Based on the interaction between electromagnetic radiation


(EMR) and matter
• EM spectrum ranges from Gamma rays to Radio waves
• The energy associated with the rays is related to the
wavelength by the Planck’s law equation
– E = hc / λ
• h – Plancks Constant c – velocity of light λ - wavelength
• Gamma rays have the lowest wavelength and hence the
highest energy
Electromagnetic Spectrum
• Gamma Rays: < 0.01 nm
• Hard X-rays: 0.01 – 0.1 nm
• Soft X-rays: 0.1 – 10 nm
• Vac. UV: 10 – 180 nm
• Visible: 380 – 700 nm
• Near UV: 180 – 380 nm
• Near IR: 700 – 1400 nm
• Mid IR: 1400 – 15000 nm
• Far IR: 15 – 1000 μm
• Microwave: 1 – 10 mm
• Radio waves: > 10 mm

http://nationalatlas.gov/articles/mapping/IMAGES/
Types of Interaction between EMR and Matter
• NMR (Radio frequency):
– Change of Spin – absorption on RF energy
• Microwave
– Change of Orientation
• Infrared
– Change of Configuration – rotational/vibrational
• UV/Vis
– Change in electronic structure
• X-rays
– Change in electronic structure
• Gamma Rays
– Change in nuclear configuration/mutation/isotopes
Instrumentation for Spectroscopy
Absorption Spectroscopy

Source Sample
Wavelength Detector Data
Selector Readout

Emission Spectroscopy
Sample

EMISSION Wavelength
Selector

EXCITATION

Wavelength Detector Data


Selector Readout

Source
Optical Materials

• Transmittance is the main property for selecting material for


optical elements
– LiF – transmittance between 150 – 7000 nm
– Borosilicate (normal glass) absorbs below ~380 nm
– Quartz or Fused Silica is ideal for UV applications and have very high
transmittance
– IR optical elements – halide salts (AgCl, KBr, NaCl), polymers have high
transmittance
Sources

• Continuous Sources
– Xenon Arc Lamp – 250 – 600 nm
• Molecular Fluorescence
– Hydrogen or Deuterium Lamp – 180 – 380 nm
• UV Molecular Absorbance
– Tungsten/Halogen – 240 – 2500 nm
• UV/Vis/NIR – Molecular Absorbance
– Tungsten – 350 – 2200 nm
• Vis/NIR
– Nernst Glower – 400-20,000 nm
• IR – molecular absorbance
– Nichrome – 750 – 20,000 nm
– Globar – 1200 – 40,000 nm
– Tunable Dye Lasers
Sources

• Continuous Source Lamp Spectra & Irradiance

From “Oriel Corporation”: (www.lot-oriel.com)


Sources
• Line Sources
– Narrow Discrete Spectra

From “Oriel Corporation”: (www.lot-oriel.com)


Sources

• Pulsed Sources
– Lasers (Light Amplification by Stimulated Emission of Radiation)
• Permits light of low divergence
• Well defined wavelength
• Low Energy and High Energy
• Nd: YAG Lasers
Wavelength Selectors

• Radiation is measured in ‘bands’


– Selectivity – narrow bands
– Sensitivity – broader bands – allow more energy
• Devices to select wavelength
• Absorption/Emission Wavelengths
– Monochromator
– Band Filters
– Interference Filters
Wavelength Selectors

• Monochromator
– Czerny-Turner
– A – incident beam
– B – Entrance Slit
– C – Mirror 1
– D – Diffraction
Grating (Rotating)
• Blazed Grating
• Prism
• Concave
• Holographic
– E – Mirror 2
– F- Exit Slit

Image:
http://www.cem.msu.edu/~reusch/VirtualText/Spectrpy/spectro.htm#contnt
Wavelength Selectors

• Blazed Diffraction Grating

From www.edmundoptics.com
UV Absorption Spectroscopy
• Transmittance, T = I / I0
• Absorbance, A = -log T
• Beer’s Law
– A = ε.L.C
– ε = absorption coefficient
– L = path length
– C = concentration of analyte
• Example Absorption
Spectra shown here

Image:
http://www.cem.msu.edu/~reusch/VirtualText/Spectrpy/spectro.htm#contnt
UV Absorption Spectroscopy

• Double Beam Spectrophotometer (figure above)


– Analysis of Reference and Sample
Simultaneously
• Single Beam
– Only one detector – reference and sample
cuvettes move to come in line to the path of
light.
Infra-red Spectroscopy

Structural Features:
• Stretching or Bending Vibrations
http://www.cem.msu.edu/~reusch/VirtualText/Spectrpy/spectro.htm#contnt
• Organic or Inorganic Bonds
IR Vibrational Spectroscopy
Formaldehyde – Gas Phase IR spectra

Some Common Structural Features

http://www.cem.msu.edu/~reusch/VirtualText/Spectrpy/spectro.htm#contnt
IR Vibrational Spectroscopy
Typical Example of Infrared Absorption Frequencies
Stretching Vibrations Bending Vibrations
Intensit
Functional Class Range (cm-1) Intensity Assignment Range (cm-1) Assignment
y
Alkanes 2850-3000 str CH3, CH2 & CH 1350-1470 med CH2 & CH3 deformation
2 or 3 bands 1370-1390 med CH3 deformation
720-725 wk CH2 rocking

Alkenes 3020-3100 med =C-H & =CH2 (usually sharp) 880-995 str =C-H & =CH2
1630-1680 var C=C (symmetry reduces 780-850 med (out-of-plane bending)
intensity) 675-730 med cis-RCH=CHR
1900-2000 str
C=C asymmetric stretch

• Examples of Functional classes that exhibit IR absorbance


– S compounds
• S-H, S-OR, S-S, C=S, S=O
– P compounds
• P-H, (O=)PO-H, P-OR, P=O
– Si Compounds
• Si-H, Si-OR, Si-CH3
– Oxygen-Nitrogen
• NOH, N-O, N=O

http://www.cem.msu.edu/~reusch/VirtualText/Spectrpy/spectro.htm#contnt
Michelson Interferometer

http://www.biophysik.uni-freiburg.de/Spectroscopy/Time-Resolved/spectroscopy.html
IR Spectrophotometer

www.thermonicolet.com
Fourier Transform

FTIR

Processor

Absorption Spectra
Interferogram

• The interferogram is the signal that is received by the detector (which is a


thermal detector).
• This interferogram is processed using a fourier transform to obtain spectral
information (or an absorption spectrum) shown on the right side.
• The transmission is the usual representation of IT spectra – absorption can
be converted to transmission easily.

http://www.biophysik.uni-freiburg.de/Spectroscopy/Time-Resolved/spectroscopy.html
NMR Spectroscopy
• Nuclei of elements have a characteristic spin – I
– Integral: e.g. 1,2,3… (# protons and # nuetron are odd)
– Half integral: e.g. ½, 3/2, 5/2…(#protons + # nuetrons are odd)
– Zero (#protons and #nuetrons are even)
• Spinning charge generates a small magnetic field resulting in a
magnetic moment
• In the presence of an external magnetic field, B0 , two spin
states exist, +1/2 and -1/2.
• The magnetic moment of the lower (+1/2) energy spin state is
aligned with the external magnetic field and that of the higher
(-1/2) is aligned in the opposite direction to B0.
Spin states (two energy) in external magnetic
field
No Magnetic
Field

B0 – Magnetic Field Strength

http://www.cem.msu.edu/~reusch/VirtualText/Spectrpy/nmr/nmr1.htm
Magnetic Fields
• Earth’s Magnetic field – 10-4 T (Tesla)
• Modern NMR instruments use magnetic fields 1 – 20 T.
• Even at high B0, the energy difference observed between the two spin
states is less than 0.1 cal/mole
– IR (1-10 cal/mole), Electronic transitions are 100 times higher
• Energy difference is expressed in terms of frequency in MHz
– Ranging from 20 – 900 MHz depending on the field and the molecule
• Irradiation of the sample with Radio Frequency energy corresponding to
this energy difference between the two spin states causes absorption and
excitation from the +1/2 to the -1/2 state.
• For ½ spin nuclei, the energy difference is proportional to their magnetic
moments at a given B
• Common nuclei used in NMR spectroscopy are
– 1H, 13C, 19F, 31P,
– Most common is the 1H NMR also referred to as Proton NMR.
Proton NMR Spectroscopy

http://www.cem.msu.edu/~reusch/VirtualText/Spectrpy/spectro.htm#contnt
Proton NMR Spectroscopy
• Consider a sample of water
– External magnetic field B = 2.3487 T
– 100 MHz - radiation
• If B is increased slowly to 2.3488 T
– H nuclei will absorb RF energy and a resonance signal will
appear

http://www.cem.msu.edu/~reusch/VirtualText/Spectrpy/spectro.htm#contnt
Chemical Shift
• Different H nuclei have different signals at different frequencies
– Electron density around a nucleus produce a magnetic field that opposes the
much stronger external field.
– This secondary field shields the nucleus from the external field
– To compensate for this shielding B must be increased to achieve resonance
– Scanning of B at a RF frequency produces a SHIFT
– Compounds that have higher electron density have the resonance at higher
energy frequencies
Chemical Shift
• The characteristic frequencies are dependent on the magnitude of B and the radio
frequency
• No two magnets have the same field and resonance frequencies vary accordingly
– Reference proton: TMS (Trimethylsilane)
– Shift is measured as shown below in terms of ppm
• Solvent – CDCl3
Chemical Shift
NMR Spectroscopy – examples

Red peaks indicate different characteristic groups of the chemical,


which exhibit a certain chemical shift with respect to a reference such
as TMS.
Chemical Shift
• Signal Strength
– Number of Hydrogen atoms
– Concentration
• OH proton signal
– Aliphatic or aromatic
• Π-electrons
• Solvent effects
• Spin-Spin Interactions
– J Coupling
J-Coupling
• Signal Splitting
– Doublet
– Triplet
– Quartet
– Quintet
– Concentration
13
C NMR Spectroscopy
• For Molecules that lack significant C-H bonds – one
cannot use proton (or 1H NMR spectroscopy)
– Polychlorinated compounds
– Triple bond compounds
– Polycarbonyl ass
• 13 C isotope is a viable alternative
– Less abundant  less sensitive signal splitting due to
interaction with H
• Solid State NMR
– anisotropy
Raman Spectroscopy
• Used in chemistry,
– Vibrational information is very specific
for the chemical bonds in molecules.
– Provides a fingerprint by which the
molecule can be identified. The
fingerprint region of organic molecules
is in the range 500-2000 cm-1.
– Study changes in chemical bonding, e.g.
when a substrate is added to an enzyme.
– Raman gas analyzers have many
practical applications, for instance they
are used in medicine for real-time
monitoring of anaesthetic and
respiratory gas mixtures during surgery.
– Characterize materials – measure T
– Crystallographic Orientation of a sample
• Inelastic Scattering or Raman scattering
of monochromatic light usually from a
laser in the visible, NIR or near UV
range.
• Holographic diffraction gratings and
multiple dispersion stages to separate
the weak inelastic Raman scattering and
the strong Rayleigh scattering
• Spontaneous and stimulated RS
Atomic Spectroscopy
• Qualitative/Quantitative determination of elements
– Analysis of metals in water, soils, solids, tissue
• Ionization/Atomization of sample
– Flame Atomization
– Plasma
• Analysis
– Atomic Optical Spectrometry
– Atomic Mass Spectrometry
• Generic Instruments
– Atomic Absorption (AA) - Flame
– ICP-AES/OES - Plasma
– ICP-MS - Plasma

SAMPLE Sample Spray Flame or Analyzer


Processin /Vapor Plasma
g (Nebulizer
or
Vaporizer
Atomic Absorption Spectroscopy - AA
• Source – Hollow Cathode Lamp
– Tungsten anode/Cylindrical Cathode in quartz with inert gas
– Cathode is made of the analyte metal or has a coating of the metal
– Line spectra (narrow)
– Individual lamps for different metals – about 70 elements
• Sample is aspirated into the atomizing zone as a spray –
– Nebulizer
• Ionization/Atomization of sample
– Flame – Burner – 1700 – 3500 C
• Fuel + air (natural gas, Hydrogen, acetylene)
• Flow rate is high – efficiency of atomization is not good
• Higher T increases sensitivity
– Graphite Furnace
• Graphite tube with electric current
• Thermal energy causes electronic transition
– Light Absorption
• Monochromator
• Detector
– Photometers
– Spectrophotometer – UV/Vis monochromator
Atomic Absorption Spectroscopy - AA
• Source of Error
– Flame Emission
– Modulation – Chopper/Reflected/Sample
– Alternating signal from the lamp and continuous signal from the flame
– Correction is made at the data analysis.
• Background Correction
– Hollow cathode lines are narrow so very little chance of interferences
– Molecular species in the range of the wavelength absorbed can cause interference
– AT = AA + AB
– Continuous Source Background Correction
• Part of the flame used for quantitative measurement
– Primary combustion zone (very close to nozzle) - hottest
– Interzonal region
– Secondary combustion zone (outer ring) – oxidation may occur
– Flame position must be adjusted since the response of each metal varies with it.
• Beer’s Law
• Internal Standards
• Hg – Cold Vapor AAS
• Detection limits
– 0.2 – 30 ppb (g/L) (for different elements)
Plasma For Atomization
• Plasma tube
– 3 concentric quartz tubes
– Argon Gas – as ionizing gas
and for cooling
– RF Coils – 2KW energy at
27/40 MHz
– Temperature – 6000 – 8000 K
– Ionization efficiency is high
– Tangential Gas flow allows for
thermal separation of plasma
zone from the region before it
ICP-AES
ICP-AES
• Interference
– Blank interference
• MilliQ water – relative
• Blank – MDL and IDL
– Spectral interference
– Analyte interference
• Physical/chemical effects
• Operating conditions
• Detection limits
ICP-AES
• Wavelength selection
• Monochromator
– Sequential analysis/scan of wavelengths and
a PMT
• Spectrograph
– Large aperture for a band and detection
using a CCD or array detector
• Polychromator
– Grating + Simultaneous detection using a
CCD or array detector
ICP-MS

icpms.ucdavis.edu/InstrumentationQuad.html
ICP-MS
• Operates under Vacuum – 10-6 torr
• After ionization sample is introduced to the
analyzer zone through two orifices
– Skimmer cone
– Sampler cone
• Ion Optics
• Mass Analyzer
• Selection on the basis of Isotope (or
Mass/Charge ratio)
– Capable of isolating “isotopes”
• Very low detection limits.
ICP-MS – Mass Analyzer
• Ion Optics
– Focus the ions in a
barrow beam towards
the Mass Analyzer
• Mass Analyzer
– Quadrupole
– Allows only a certain
M/Z set to pass through
to the detector • Other Types of analysers
• Detector – Ion Trap
– PMT/Dynode – Time of Flight(TOF) MS
– Interferences
• Ar+, ArO+ , ArH+, H2O+,
Ar2+ or combination
• Same isotope
• Matrix effects
ICP-MS
• During analysis, the use of blank is important
since it accounts for contamination
– in the instrument (previous sample memory effects)
– the surroundings (cleanliness of the lab).
– For very sensitive (or low concentration) work, a clean
room (with very low particulate contamination) may be
required.
• Internal standard is recommended in all samples
(same amount in all samples, standards and
blanks)
– The internal standard should be an element that is not
present in the sample
– Selection of internal standard depends on the nature
of the analysis matrix

You might also like