CEms Uras 26
CEms Uras 26
ANALYZER
ABB EL3020(URAS26)
Gases and their sensing wavelengths
This change in pressure evokes an Electrical signal via a membrane capacitor. The
correlation between the detector gas and the sample gas provides an extremely
high selective sensitivity with regard to gases such as CO,CO2, SO2, NO, CH4 and
NO – to name but a few.
Flow Diagram of Sample Gases
Calibration
Calibration can be performed automatically or manually. Automatic calibration
for all sample components together – is normally started on a cyclically time-
controlled basis; it can also be started by an external control signal or via the
Modbus as well as manually on the display and operator control unit of the gas
analyzer.
Normally, simplified calibration methods with the built-in calibration cells or
the so-called single-point calibration are used for automatic calibration.
If calibration with test gases is required the control of solenoid valves for
switching on test gases, zero gas and sample gas via digital outputs can be
configured.
Calibration Cells
The use of calibration cells (option) enables the end-point calibration
of the Uras26 to be carried out without test gas cylinders.
A calibration cell is installed in each optical beam path of the analyzer
as per the order. Each calibration cell is filled with a test gas, whose
composition and concentration has been adapted to the sample
components and measuring ranges which have been installed in the
respective optical beam path.
The calibration cells are moved into the optical beam path for the end-
point calibration.
Calibration Cells