Moisture Analyzer

ANGGER SEMBODO
LNG ACADEMY 6
ELECTRICAL & INSTRUMENTATION
6516010047

2018
SHE – Q Moment

2
AGENDA
SEC ELEC INST WHSC

LAB TOP
Introduction of Moisture Analyzer
 Definition and Function of Moisture Analyzer
 The Role of Moisture Anlayzer in LNG Process
 Moisture Analyzer in Badak LNG (24)

Moisture Analyzer

5
Moisture Analyzer Parts : Field Unit

Electronic
Solenoid
Control PC
valve
Board
Power
Moisture
Generator

Sample Gas Reference Gas


Control Flow Control Flow

Back Pressure
Moisture Regulator
Generator Flow Control
Control
Back Pressure
Flow Meter Gauge

6
7
How Does Moisture Analyzer Work

8
Maintenance & Troubleshooting

In AMETEK 5000, the routine maintenances are :

1. Weekly
• Operate internal generator for calibration
• Adjust span if required

2. Monthly
• Inspect system and check for leakage
• Replace filter

3. Yearly
• Replace or refill molecular sieve on dryer
• Replace or refill activated carbon on contaminant trap
• Replace super-activated dryer
• Refill moisture generator with distillated and deionized water

9
Special Activity
 Preparation
 Work Permit
 TRA
 Personal Protective Equipment
 PM Moisture Analyzer ( calibration )
Every Moisture Generator has different value of ppm
 Press Calib on the Keypad E2-AT-4. Then Choose
Flow Moisture Gen and press enter. Access Code :
.3-18
0.0662 ppm Vol H2O 0.0662 ppm Vol H2O
030-smpl XX:XX-Cal 015-ref XX:XX-Cal
Start Auto Cal. ENTER ACCESS CODE :
>Flow Moisture Gen<

10
 After entering the code, the display shown as below
0.001 Current H2O
2.4 Moist Gen ppm
Moisture Gen Starting

0.009 Current H2O


2.4 Moist Gen ppm
WAIT FOR INSTRUMENT
TO STABILIZE THEN

11
 Wait until the value of Current H2O ± 0.1
 Then press enter on Keypad and wait a minute until
the measurement back to the normal process

0.01 ppm Vol H2O


015-smpl XX:XX-Cal
123.4 hz Delta freq
SYSTEM CALIBRATING

 Do the same for E2-AT-5 and E2-AT-6

12
Calibration Result

Moisture As Found As Left


Tag Number Generator Mark
(ppm) Process M.G. Process M.G.
E2-AT-6 3,2 0,1084 2,34 - - Uncalibrated
E2-AT-5 2,6 0,1532 2,55 0,1196 2,6042 Calibrated
E2-AT-4 2,4 0,0662 2,34 0,077 2,4043 Calibrated

Why ???

13
Troubleshooting E2-AT-6

 Leakage of Moisture Generator


 Flow of moisture generator is not appropriate

 Leakage of Cell/Crystal

14