Lecture 7 Attenuation of X-Rays

This document provides an overview of attenuation of x-rays. It defines key terms like attenuation, linear attenuation coefficient, and mass attenuation coefficient. It explains that x-ray beams lose intensity as they pass through matter due to interactions with atoms. The lecture objectives are to understand how x-ray quantity, quality, and intensity are reduced based on the material's properties. Formulas are provided to calculate attenuation for monochromatic and polychromatic x-ray beams.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPT, PDF, TXT or read online on Scribd
0% found this document useful (1 vote)
186 views51 pages

Lecture 7 Attenuation of X-Rays

This document provides an overview of attenuation of x-rays. It defines key terms like attenuation, linear attenuation coefficient, and mass attenuation coefficient. It explains that x-ray beams lose intensity as they pass through matter due to interactions with atoms. The lecture objectives are to understand how x-ray quantity, quality, and intensity are reduced based on the material's properties. Formulas are provided to calculate attenuation for monochromatic and polychromatic x-ray beams.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPT, PDF, TXT or read online on Scribd
You are on page 1/ 51

Resident Physics Lectures

Attenuation of X-rays
Prof. J.K Tonui, PhD

School of Medicine,
Department of Radiology & Imaging
Learning Objectives
 At the end of this lecture, the student is expected to:

 Define attenuation;

 Know the characteristics of x-ray beams;

 Understand monochromatic and polychromatic x-ray beams;

 Define linear and mass attenuation coefficients;

 Define Half Value Layer (HVL) and

 Calculate various aspects of attenuation of x-rays when passing

through matter.
2 RIB 810, 811, 812 Imaging Physics 8/21/2019
Introduction
 Attenuation

 Is defined as the gradual loss in intensity of any kind of

energy flux when it travels thro’ a medium,


 For example, sunlight is attenuated by dark (sun) glasses,

x-rays are attenuated by lead or human body etc, and


 Hence, when x-rays pass thro’ the pt.’s body, the intensity of

exiting radiation is less than that of incident radiation.

3 RIB 810, 811, 812 Imaging Physics 8/21/2019


Radiation interaction with matter
 When x-rays interact with body tissues,

 They are attenuated as they pass thro’ the pt.’s body, and

 The amount of attenuation increases as the X-ray beam passes

thro’ longer paths in the pt.’s body and


 Basically, each interaction of an X-ray photon with an atom of

the tissue removes an X-ray photon from the beam, which


results in decrease in x-ray beam intensity.

4 RIB 810, 811, 812 Imaging Physics 8/21/2019


Beam Characteristics
 Quantity
 Refers to the number of photons in the x-ray beam

1, 2, 3, ...
~
~
~ ~
~
5 RIB 810, 811, 812 Imaging Physics 8/21/2019
Beam Characteristics
 Quality
 Refers to the energy distribution of photons in the x-ray beam

1 @ 27 keV, 2 @ 32 Energy Spectrum


keV, 2 at 39 keV, ...

~ ~
~ ~ 10 20 30 40 50
Energy
60 70 80

~
~ ~
~
6 RIB 810, 811, 812 Imaging Physics 8/21/2019
Beam Characteristics
 Intensity
 Weighted product of number and energy of photons and
 Equal to the product of quantity and quality.

324 mR

~
~ ~
~ ~ ~
~ ~
7 RIB 810, 811, 812 Imaging Physics 8/21/2019
What is a Roentgen?
 Roentgen

 Is a unit of measurement

for amount of ionizing


radiation that produces 2.58
x 10-4 Coulomb/kg of air
@ STP
 1 C ~ 6.241509324×1018

electrons

8 RIB 810, 811, 812 Imaging Physics 8/21/2019


Beam Intensity
 Beam Intensity

 Can be measured in terms of # of ions created in air by

beam and
 Is valid for monochromatic as well as polychromatic beams.

324 mR
-
~
+

9 RIB 810, 811, 812 Imaging Physics 8/21/2019


Monochromatic Radiation
 Monochromatic radiation

 Are produced by radioisotope but not x-ray beam and

 All photons in a beam have same energy.

 Attenuation results in

 Change in beam quantity but

 No change in beam quality, that is

 The # of photons & total energy of beam changes by

same fraction.

10 RIB 810, 811, 812 Imaging Physics 8/21/2019


Attenuation of Photons in Matter
 As a result of interactions between photons and
tissues,
 The intensity of the beam (i.e. the number of

photons remaining in the beam), decreases as the


beam passes thro’ body as shown in this Fig.

 This loss of photons

 Is called attenuation, and

 The matter through which the beam passes is Concept of


Attenuation
called attenuator.

11 RIB 810, 811, 812 Imaging Physics 8/21/2019


Attenuation Coefficient
 Attenuation coefficient

 Is a parameter indicating the fraction of radiation attenuated by

a given absorber thickness, and

Is a function of:

o Absorber, and

o Photon energy.

12 RIB 810, 811, 812 Imaging Physics 8/21/2019


Factors Affecting Attenuation

13 RIB 810, 811, 812 Imaging Physics 8/21/2019


Linear Attenuation Coefficient
 Why called linear?

 distance expressed in linear dimension “x”

 Formula

N = No e –mx
where

N = number of incident photons


o N N
N = number of transmitted photons o
e = base of natural logarithm (2.718…)

m = linear attenuation coefficient (1/cm); property of

energy material x
x = absorber thickness (cm)

14 RIB 810, 811, 812 Imaging Physics 8/21/2019


Linear Attenuation Coefficient
 Larger attenuation coefficient  more attenuation

 Units:
N = No e - m x
 1 / cm ( or 1 / distance)

 Properties

 Reciprocal of absorber thickness that reduces beam intensity by e (~2.718…)


o ~63% reduction

o 37% of original intensity remaining

 As photon beam energy increases


o penetration increases or attenuation decreases, i.e. attenuating distance increases and

o linear attenuation coefficient decreases

 Note: Same equation as used for radioactive decay


15 RIB 810, 811, 812 Imaging Physics 8/21/2019
Mass Attenuation Coefficient
 Mass attenuation coefficient

 Is defined as linear attenuation coefficient divided by density

o i.e. normalized with density

 It expresses attenuation of a material regardless (independent)

of it physical state.

 Note that

 Most references often give mass attenuation coefficient but

 The linear attenuation coefficient is be more useful in radiology.


16 RIB 810, 811, 812 Imaging Physics 8/21/2019
Linear & Mass Attenuation Coefficient
 Linear attenuation • Mass attenuation coefficient
coefficient (m) (mm)
 units: 1 / cm  Define as:
 absorber thickness(x) o mm= m /,

 linear o = is density of tissue

 units: cm  Units: cm 2 / g

o i.e. linear attenuation


coefficient divided density
17 RIB 810, 811, 812 Imaging Physics 8/21/2019
Worked Example
 Question:

 A 2 mm thickness of tissue transmits 25 % of monochromatic

beam of radiation. Calculate the HVL of the beam.

 Solution:

 Start with attenuation equation and substitute the known

values, i.e. Iout/Iin = 25% = 0.25, x = 2 mm = 0. 2 cm hence


determine m, from which find HVL.
 mx
I out = I ine
18 RIB 810, 811, 812 Imaging Physics 8/21/2019
Monochromatic Radiation
 Fig. below shows

 A graph of attenuation of a monochromatic x-ray beam as they

pass through an attenuator thickness, where


 We see that the no. penetrating reduces.

19 RIB 810, 811, 812 Imaging Physics 8/21/2019


Monochromatic Radiation
 A plot of no. remaining as a function of attenuator thickness

 Yields straight line on semi-log graph as shown below

Fraction .1
(also fraction of
energy) .01
Remaining or
Transmitted
.001

1 2 3 4 5
Attenuator Thickness
20 RIB 810, 811, 812 Imaging Physics 8/21/2019
Polychromatic Radiation
 X-Ray beam contains spectrum of photon energies

 Highest energy = peak kilovoltage (kVp) applied to tube and

 The mean energy is usually between 1/3 - 1/2 of peak, but

o Depends also on filtration

21 RIB 810, 811, 812 Imaging Physics 8/21/2019


X-Ray Beam Attenuation
 The reduction in x-ray beam intensity is caused by Lower Higher
Energy Energy
 absorption (photoelectric)

 deflection (scattering)

 Attenuation alters beam

 quantity

 quality
o higher fraction of low energy photons removed (see fig.)
and results
o Beam Hardening

22 RIB 810, 811, 812 Imaging Physics 8/21/2019


Half Value Layer (HVL)
 HVL

 Is defined as absorber thickness that reduces beam intensity by

exactly a half, and


 Has units of thickness and

 Is the value of “x” which makes N equal to No / 2

N = No e -mx HVL=0.693/m

23 RIB 810, 811, 812 Imaging Physics 8/21/2019


Half Value Layer (HVL)
 Valid concept for all beam types

 Mono-energetic, and

 Poly-energetic.

 HVL

 Indicates the beam quality, where

o Higher value means more penetrating beam and

o Lower attenuation coefficient.

24 RIB 810, 811, 812 Imaging Physics 8/21/2019


Factors Affecting Attenuation
 Attenuation

 Depends on both x-ray beam and matter (body tissue), where

 Energy of radiation or beam quality – higher energy means

more penetration and hence less attenuation.


 Body tissue -

o Density ()

o atomic number (Z)

o electrons per gram (e/g)

o Where higher density, atomic number, or electrons per gram increases

attenuation.
25 RIB 810, 811, 812 Imaging Physics 8/21/2019
Polychromatic Attenuation
 A plot of penetrating beam
 Yields curved line on semi-log graph as shown below;
 The line straightens with increasing attenuation, but
 The slope approaches that of monochromatic beam at the peak
energy (see figure).
 The mean energy increases with attenuation
 beam hardening
1

Fraction .1
Transmitted Polychromatic
.01

Monochromatic
.001

26 RIB 810, 811, 812 Imaging Physics 8/21/2019


Attenuator Thickness
Photoelectric vs. Compton
 Fractional contribution of each is determined
by
The photon energy and
The atomic number of absorber.
 The Equation describing attenuation w.r.t to each
interactions is:
m = mcoherent + mPE + mCompton
Small

27 RIB 810, 811, 812 Imaging Physics 8/21/2019


Photoelectric vs. Compton

 As photon energy m = mcoherent + mPE + mCompton


increases
 Both PE & Compton
decrease, but Interaction
 PE decreases faster Probability
o Fraction of m that is
Compton increases
o Fraction of m that is PE
Compton
decreases

Photoelectric

28 RIB 810, 811, 812 Imaging Physics Photon Energy 8/21/2019


Photoelectric vs. Compton

m = mcoherent + mPE + mCompton

 As atomic # increases
 Fraction of m that is PE increases and
 Fraction of m that is Compton decreases

29 RIB 810, 811, 812 Imaging Physics 8/21/2019


Interaction Probability
Atomic
Number of
Absorber
Photoelectric

Pair
Production
Compton

Photon Energy

• PE dominates for very low energies


30 RIB 810, 811, 812 Imaging Physics 8/21/2019
Interaction Probability

Atomic
Number of
Absorber Pair
Photoelectric Production

Compton

Photon Energy

• For lower atomic numbers


– Compton dominates for high energies
31 RIB 810, 811, 812 Imaging Physics 8/21/2019
Interaction Probability
Atomic Photoelectric
Number of
Absorber Pair
Production

Compton

Photon Energy
• For high atomic # absorbers
– PE dominates throughout diagnostic energy range
32 RIB 810, 811, 812 Imaging Physics 8/21/2019
Attenuation & Density
 Attenuation proportional to density, hence

 The difference in tissue densities accounts for much of optical

density (OD) difference seen in radiographs (x-ray films)

 # of Compton interactions depends on electrons/unit path,

which in turn depends on:


 Electrons per gram and

 Density

33 RIB 810, 811, 812 Imaging Physics 8/21/2019


Relationships
 Density generally

 Increases with atomic # and

 Different states have got different density, e.g.

 The 3-states of water (ice, water, steam) have got different densities

 Note that

 There is no relationship between density and electrons per gram and

 Atomic # vs. electrons / gram

o hydrogen ~ 2X electrons / gram as most other substances and

o as atomic # increases, electrons / gram decreases slightly

34 RIB 810, 811, 812 Imaging Physics 8/21/2019


Applications
 As photon energy increases

 Subject (and image) contrast decreases

 differential absorption decreases


100
90
o at 20 keV bone’s linear attenuation 80
70
60
coefficient 6 X water’s 50 Bone
40 Water
30
o at 100 keV bone’s linear attenuation 20
10
coefficient 1.4 X water’s 0
20 keV 100 ke

35 RIB 810, 811, 812 Imaging Physics 8/21/2019


Applications Photo-
electric Pair
Production

Compton

 At low x-ray energies


 attenuation differences between bone & soft tissue primarily caused
by photoelectric effect
o related to atomic number & density

 At high x-ray energies


 attenuation differences between bone & soft tissue primarily caused
by Compton scatter
o related entirely to density

36 RIB 810, 811, 812 Imaging Physics 8/21/2019


Applications
 Difference between water & fat only visible at low energies

 effective atomic # of water slightly higher

o yields photoelectric difference

 electrons / cm almost equal

o No Compton difference

 Photoelectric dominates at low energy

37 RIB 810, 811, 812 Imaging Physics 8/21/2019


K-Edge
 Each electron shell has threshold for PE effect

 Photon energy must be >= binding energy of electron shell

 For photon energy > K-shell binding energy

 k-shell electrons are candidates for PE

 PE interactions increase as photon energy exceeds k-shell

binding energy

38 RIB 810, 811, 812 Imaging Physics 8/21/2019


K-Edge
 step increase in attenuation at K-edge energy

 K-shell electrons become available for interaction

 exception to rule of decreasing attenuation with increasing

energy

Linear
Attenuation K-edge
Coefficient

Energy

39 RIB 810, 811, 812 Imaging Physics 8/21/2019


K-Edge Significance
 K-edge energy

 insignificantly low for low Z materials and

 Is in diagnostic range for high Z materials

 higher attenuation above k-edge useful in


 contrast agents

 rare earth screens

 Mammography beam filters

40 RIB 810, 811, 812 Imaging Physics 8/21/2019


Scatter Radiation
 NO Socially Redeeming Qualities

 no useful information on image

 detracts from film quality

 exposes personnel, public

 represents 50-90% of photons exiting patient

41 RIB 810, 811, 812 Imaging Physics 8/21/2019


Abdominal Photons
 ~1% of incident photons on adult abdomen reach film

 fate of the other 99%

 mostly scatter

o most do not reach film

 absorption

42 RIB 810, 811, 812 Imaging Physics 8/21/2019


Scatter Factors
 Factors affecting scatter

 field size

 thickness of body part

 kVp

An increase in any of above increases scatter.

43 RIB 810, 811, 812 Imaging Physics 8/21/2019


Scatter & Field Size
 Reducing field size causes significant reduction in scatter
radiation

II II
Tube Tube

X-Ray X-Ray
Tube Tube

44 RIB 810, 811, 812 Imaging Physics 8/21/2019


Field Size & Scatter
 Field Size & thickness determine volume of irradiated

tissue

 Scatter increase with increasing field size

 initially large increase in scatter with increasing field size

 saturation reached (at ~ 12 X 12 inch field)

 further field size increase does not increase scatter reaching film

 scatter shielded within patient

45 RIB 810, 811, 812 Imaging Physics 8/21/2019


Thickness & Scatter

 Increasing patient thickness

 Leads to increased scatter, but

 Saturation point reached, where

 Scatter photons produced far from film and

 shielded within body

46 RIB 810, 811, 812 Imaging Physics 8/21/2019


kVp & Scatter
 kVp has less effect on scatter than
 field size

 thickness

 Increasing kVp
 increases scatter

 more photons scatter in forward direction

47 RIB 810, 811, 812 Imaging Physics 8/21/2019


Scatter Management
 Reduce scatter by minimizing

 field size

o within limits of exam

 thickness

o mammography compression

 kVp

o low kVp increases patient dose

o in practice we maximize kVp

48 RIB 810, 811, 812 Imaging Physics 8/21/2019


Scatter Control Techniques: Grid
 Scatter is controlled by using grid

 directional filter for photons

 Increases patient dose

49 RIB 810, 811, 812 Imaging Physics 8/21/2019


Angle of Escape
 angle over which scattered radiation misses primary
field
 escape angle larger for
 small fields
Larger Angle of Escape
 larger distances from film

X
X

Film
Film
50 RIB 810, 811, 812 Imaging Physics 8/21/2019
Scatter Control Techniques: Air Gap
 Gap intentionally

 left between patient & image

receptor

 Natural result

 of magnification radiography

 Grid not used

51 RIB 810, 811, 812 Imaging Physics 8/21/2019