Calibration of MCII

1) The document provides steps for calibrating the MC-II flow analyzer meter. 2) It involves obtaining the meter's calibration factor, calculating the divisor and rate multiplier, setting jumpers, and entering the divisor by pressing buttons on the analyzer's interface. 3) Correctly entering the divisor of 2328 is key to calibrating the MC-II to measure volumes in barrels based on the specific meter's calibration factor.

Uploaded by

lucaz273747
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPT, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
117 views18 pages

Calibration of MCII

1) The document provides steps for calibrating the MC-II flow analyzer meter. 2) It involves obtaining the meter's calibration factor, calculating the divisor and rate multiplier, setting jumpers, and entering the divisor by pressing buttons on the analyzer's interface. 3) Correctly entering the divisor of 2328 is key to calibrating the MC-II to measure volumes in barrels based on the specific meter's calibration factor.

Uploaded by

lucaz273747
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPT, PDF, TXT or read online on Scribd
You are on page 1/ 18

MC-II

Entering the Calibration


Step #1
Obtain The Meter Calibration
Factor
• Locations Of the Calibration Factor:
• Tag on Meter Collar
• End Of The Shipping Box
• Call The Factory With The Meter Serial Number.

Note: We only have


the original calibration
factor of the meter. If
the kit has been
changed, This number
will not be valid.
Step #2
Calculate The Divisor &
Rate Multiplier
Meter Size: 2”
Calibration Factor: 55.43
Volume Units: Barrels
Rate Units: BPD
Divisor: 2328
Rate Multiplier: ONC

Divisor:
55.43 P/Gal X 42 Gal/Bbl = 2328.06 P/Bbl

Rate Multiplier:
Since This Is a ‘Per Day’ Rate, We Will Use ONC.
Step #3
Set The Jumpers
• To Open The MC-II, Loosen
the Eight Screws Around
the Back of the Housing.
• These Are Captive Screws
and Will Not Come All the
Way Out.
• When They Are All Loose,
the Face Plate Will Fall
Open.
• Plastic Straps Attach the
Face Plate to the Housing So
It Does Not Fall off.
• The Jumpers Are Located
on the Bottom of the
Circuit Card.
Step #3
Set The Jumpers

ONC Jumper

ONC Enable

RST Calibrate
Jumper
CAL Enable

Enable the ONC, Place the Jumper Next to ONC


Enable the Calibration. Place the Jumper Next to CAL
Note: The RST Jumper Allows the totals to be reset from the front panel.
It unnecessary to change its position for calibration.
Step #4
Enter The Divisor
MODEL MC-II
• Enter Calibration FLOW ANALYZER
Mode 0
• Press The Enter and TOTAL (BBLS)
Access Buttons
simultaneously and
release. RATE (BPD)
CALIBRATION

ACCESS DEC. INCRE- ENTER VIEW RESET


POINT MENT DIV.
STEP

FLOW MEASUREMENT SYSTEMS


Step #4
Enter The Divisor
MODEL MC-II
• Enter Calibration FLOW ANALYZER
Mode 0
• Press The Enter and TOTAL (BBLS)
Access Buttons
simultaneously and
release. RATE (BPD)

• A single ‘0’ should CALIBRATION

appear in the top ACCESS DEC.


POINT
INCRE-
MENT
ENTER VIEW
DIV.
RESET

display. The bottom STEP

display should be
blank.
FLOW MEASUREMENT SYSTEMS
Step #4
Enter The Divisor
MODEL MC-II
• Enter The Divisor FLOW ANALYZER
From Right To Left 0
TOTAL (BBLS)

2 3 28
Left To Right RATE (BPD)
CALIBRATION

ACCESS DEC. INCRE- ENTER VIEW RESET


POINT MENT DIV.
STEP

FLOW MEASUREMENT SYSTEMS

Divisor = 2328
Mul. Fac. = ONC
Step #4
Enter The Divisor
MODEL MC-II
• Enter The Divisor FLOW ANALYZER
From Right To Left 07
4
0
3
2
8
5
1
6
• Press The Increment TOTAL (BBLS)
Button Repeatedly
Until the Display
Shows the Proper
RATE (BPD)
Number in That Place.
CALIBRATION
• Press The Enter DEC. INCRE- ENTER VIEW
ACCESS RESET
Button Once To Set POINT MENT
STEP
DIV.

The Number and Move


to the Next Place to
The Left
FLOW MEASUREMENT SYSTEMS

Divisor = 2328
Mul. Fac. = ONC
Step #4
Enter The Divisor
MODEL MC-II
• Enter The Divisor FLOW ANALYZER
From Right To Left 00
2
18
• Press The Increment TOTAL (BBLS)
Button Repeatedly
Until the Display
Shows the Proper
RATE (BPD)
Number in That Place.
CALIBRATION
• Press The Enter DEC. INCRE- ENTER VIEW
ACCESS RESET
Button Once To Set POINT MENT
STEP
DIV.

The Number and Move


to the Next Place to
The Left
FLOW MEASUREMENT SYSTEMS

Divisor = 2328
Mul. Fac. = ONC
Step #4
Enter The Divisor
MODEL MC-II
• Enter The Divisor FLOW ANALYZER
From Right To Left 02
0
3
128
• Press The Increment TOTAL (BBLS)
Button Repeatedly
Until the Display
Shows the Proper
RATE (BPD)
Number in That Place.
CALIBRATION
• Press The Enter DEC. INCRE- ENTER VIEW
ACCESS RESET
Button Once To Set POINT MENT
STEP
DIV.

The Number and Move


to the Next Place to
The Left
FLOW MEASUREMENT SYSTEMS

Divisor = 2328
Mul. Fac. = ONC
Step #4
Enter The Divisor
MODEL MC-II
• Enter The Divisor FLOW ANALYZER
From Right To Left 02
0
13 28
• Press The Increment TOTAL (BBLS)
Button Repeatedly
Until the Display
Shows the Proper
RATE (BPD)
Number in That Place.
CALIBRATION
• Press The Enter DEC. INCRE- ENTER VIEW
ACCESS RESET
Button Once To Set POINT MENT
STEP
DIV.

The Number and Move


to the Next Place to
The Left
FLOW MEASUREMENT SYSTEMS

Divisor = 2328
Mul. Fac. = ONC
Step #4
Enter The Divisor
• Press The Enter Button MODEL MC-II
Again. FLOW ANALYZER
• Notice the ‘L’. This Is 0232
L80
Asking If the Divisor
TOTAL (BBLS)
Represents Full Barrels or
Represents 1/10, 1/100, or 0.0
1/1000 of a Barrel.
RATE (BPD)
• Since Our Divisor Is CALIBRATION
Pulses Per Full Barrel, We
DEC. INCRE- ENTER VIEW
Pass This up by Pressing ACCESS
POINT MENT
STEP
DIV.
RESET

Enter Again

FLOW MEASUREMENT SYSTEMS

Divisor = 2328
Mul. Fac. = ONC
Step #5
Verify The Divisor
MODEL MC-II
• Press And Hold The
FLOW ANALYZER
View Divisor Button
• After a Short Delay, the 2 3 2 80
Numbers Will Change. TOTAL (BBLS)

• The Top Display Shows


the Divisor 0.0
3 7.2
• Notice That It Matches RATE (BPD)
CALIBRATION
Our Divisor.
• This Means We Correctly ACCESS DEC. INCRE- ENTER VIEW RESET
POINT MENT DIV.
STEP
Entered the Divisor.

FLOW MEASUREMENT SYSTEMS

Divisor = 2328
Mul. Fac. = ONC
Step #5
Verify The Divisor
MODEL MC-II
• Press And Hold The
FLOW ANALYZER
View Divisor Button
• The Bottom Display 2328
Shows the Rate Multiplier TOTAL (BBLS)

• Notice That A Rate


Multiplier was Calculated 3 7.2
By the MC-II. It Is The RATE (BPD)
Rate Multiplier For CALIBRATION
Barrels Per Day. ACCESS DEC. INCRE- ENTER VIEW RESET
POINT MENT DIV.

• This Is the result of STEP

enabling the ONC


Jumper.
FLOW MEASUREMENT SYSTEMS

Divisor = 2328
Mul. Fac. = ONC
Step #5
Verify The Divisor
MODEL MC-II
• Release The View
FLOW ANALYZER
Divisor Button
• The Display Returns to Its 2 3 2 80
Previous Reading. TOTAL (BBLS)

• If The MC-II Were On A


Meter, It Would Continue 0.0
3 7.2
To Count While View RATE (BPD)
Divisor Was Pressed. CALIBRATION

ACCESS DEC. INCRE- ENTER VIEW RESET


POINT MENT DIV.
STEP

• We Aren’t Done Yet! We


Have One More Task.
FLOW MEASUREMENT SYSTEMS
Close the MC-II
• Place the Faceplate
on the housing. Inserts
• Be sure no wires
overlap the gasket
area.
• Tighten the eight
screw. Gasket
• Theses should be
snug but do not over
tighten.
• Overtightening can
break the Inserts out
of the face plate.
You Have Successfully
Calibrated The MC-II.