Detectors

Download as pptx, pdf, or txt
Download as pptx, pdf, or txt
You are on page 1of 9

• GC DETECTORS


As solutes elute from the column, they interact with the detector. The detector converts this
interaction into an electronic signal that is sent to the data system. The magnitude of the signal is
plotted versus time (from the time of injection) and a chromatogram is generated.
Some detectors respond to any solute eluting from the column while others respond only to solutes
with specific structures, functional groups or atoms
Detectors that exhibit enhanced response to specific types of solutes are called selective detectors.
Most detectors require one or more gases to function properly. There are combustion, reagent,
auxiliary and makeup gases. In some cases, one gas may serve multiple purposes. The type of
detector gas is dependent on the specific detector and is fairly universal between GC
manufacturers. The flow rates for each type of detector varies between GC manufacturers. It is
important to follow the recommended flow rates to obtain the optimal sensitivity, selectivity and
linear range for a detector

FLAME IONIZATION DETECTOR (FID):
• Mechanism: Compounds are burned in a hydrogen-air flame. Carbon
containing compounds produce ions that are attracted to the collector.
The number of ions hitting the collector is measured and a signal is
generated.
Selectivity: Compounds with C-H bonds. A poor response for some non-
hydrogen containing organics (e.g., hexachlorobenzene).
Sensitivity: 0.1-10 ng
Linear range: 105-107
Gases: Combustion - hydrogen and air; Makeup - helium or nitrogen
Temperature: 250-300°C,and 400-450°C for high temperature analyses.
• NITROGEN PHOSPHORUS DETECTOR (NPD):
• Mechanism: Compounds are burned in a plasma surrounding a
rubidium bead supplied with hydrogen and air. Nitrogen and
phosphorous containing compounds produce ions that are attracted
to the collector. The number of ions hitting the collector is
measured and a signal is generated.
Selectivity: Nitrogen and phosphorous containing compounds
Sensitivity: 1-10 pg
Linear range: 104-10-6
Gases: Combustion - hydrogen and air; Makeup - helium
Temperature: 250-300°C
• ELECTRON CAPTURE DETECTOR (ECD):
• Mechanism: Electrons are supplied from a 63Ni foil lining
the detector cell. A current is generated in the cell.
Electronegative compounds capture electrons resulting in a
reduction in the current. The amount of current loss is
indirectly measured and a signal is generated.
Selectivity: Halogens, nitrates and conjugated carbonyls
Sensitivity: 0.1-10 pg (halogenated compounds); 1-100 pg
(nitrates); 0.1-1 ng (carbonyls)
Linear range: 103-104
Gases: Nitrogen or argon/methane
Temperature: 300-400°C
• THERMAL CONDUCTIVITY DETECTOR (TCD):
• Mechanism: A detector cell contains a heated filament with an
applied current. As carrier gas containing solutes passes through
the cell, a change in the filament current occurs. The current
change is compared against the current in a reference cell. The
difference is measured and a signal is generated.
Selectivity: All compounds except for the carrier gas
Sensitivity: 5-20 ng
Linear range: 105-106
Gases: Makeup - same as the carrier gas
Temperature: 150-250°C
• FLAME PHOTOMETRIC DETECTOR (FPD):
• Mechanism: Compounds are burned in a hydrogen-air flame. Sulfur
and phosphorous containing compounds produce light emitting
species (sulfur at 394 nm and phosphorous at 526 nm). A
monochromatic filter allows only one of the wavelengths to pass. A
photomultiplier tube is used to measure the amount of light and a
signal is generated. A different filter is required for each detection
mode.
Selectivity: Sulfur or phosphorous containing compounds. Only one
at a time.
Sensitivity: 10-100 pg (sulfur); 1-10 pg (phosphorous)
Linear range: Non-linear (sulfur); 103-105 (phosphorous)
Gases: Combustion - hydrogen and air; Makeup - nitrogen
Temperature: 250-300°C
PHOTOIONIZATION DETECTOR (PID):
• Mechanism: Compounds eluting into a cell are bombarded with
high energy photons emitted from a lamp. Compounds with
ionization potentials below the photon energy are ionized. The
resulting ions are attracted to an electrode, measured, and a signal
is generated.
Selectivity: Depends on lamp energy. Usually used for aromatics
and olefins (10 eV lamp).
Sensitivity: 25-50 pg (aromatics); 50-200 pg (olefins)
Linear range: 105-106
Gases: Makeup - same as the carrier gas
Temperature: 200°C
• ELECTROLYTIC CONDUCTIVITY DETECTOR (ELCD):
• Mechanism: Compounds are mixed with a reaction gas and passed
through a high temperature reaction tube. Specific reaction products are
created which mix with a solvent and pass through an electrolytic
conductivity cell. The change in the electrolytic conductivity of the solvent
is measured and a signal is generated. Reaction tube temperature and
solvent determine which types of compounds are detected.
Selectivity: Halogens, sulfur or nitrogen containing compounds. Only one
at a time.
Sensitivity: 5-10 pg (halogens); 10-20 pg (S); 10-20 pg (N)
Linear range: 105-106 (halogens); 104-105 (N); 103.5-104(S)
Gases: Hydrogen (halogens and nitrogen); air (sulfur)
Temperature: 800-1000°C (halogens), 850-925°C (N), 750-825°C (S)
• MASS SPECTROMETER (MS):
• Mechanism: The detector is maintained under vacuum. Compounds are bombarded with electrons
(EI) or gas molecules (CI). Compounds fragment into characteristic charged ions or fragments. The
resulting ions are focused and accelerated into a mass filter. The mass filter selectively allows all
ions of a specific mass to pass through to the electron multiplier. All of the ions of the specific mass
are detected. The mass filter then allows the next mass to pass through while excluding all others.
The mass filter scans stepwise through the designated range of masses several times per second.
The total number of ions are counted for each scan. The abundance or number of ions per scan is
plotted versus time to obtain the chromatogram (called the TIC). A mass spectrum is obtained for
each scan which plots the various ion masses versus their abundance or number.
Selectivity: Any compound that produces fragments within the selected mass range. May be an
inclusive range of masses (full scan) or only select ions (SIM).
Sensitivity: 1-10 ng (full scan); 1-10 pg (SIM)
Linear range: 105-106
Gases: None
Temperature: 250-300°C (transfer line), 150-250°C (source)