Dosimetry Commissioning of The Lns-Infn: Proton Therapy Facility
Dosimetry Commissioning of The Lns-Infn: Proton Therapy Facility
Dosimetry Commissioning of The Lns-Infn: Proton Therapy Facility
INFN-LNS
INFN-LNS
INFN-LNS
110
BRAGG PEAK
100
90
80
70
60
50
40
30
PLATEAU REGION
20
10
INFN - LNS
0
0
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
CENTRE
MAXIMUM
RANGE
EQUIVALENT
ENERGY
(mm)
(MeV)
PEAK
PLATEAU
RATIO
F.W.H.M.
(mm)
Distal-dose
falloff
d90%-10%
(mm)
CATANA
30.60
60.2
4.68
3.29
0.81
CCO^
31.0
60.4
4.85
3.65
0.80
PSI^
30.0
60.0
4.47
1.10
INFN-LNS
INFN-LNS
100
MarkusLNS
R elative D ose (% )
90
ScanditronixLNSDiode
80
70
60
50
40
30
20
10
0
0
10
15
20
25
30
35
DETECTOR
PEAKPLATEAU
RATIO
F.W.H.M
(mm)
Distal-dose
falloff(1)
d90%-10%
(mm)
Distal-dose
falloff(2)
d80%-20%
(mm
PRACTICAL
RANGE
(d10%,ICRU59)
PEAK
DEPTH
(mm)
MARKUSPTW
30.14
4.68
3.19
0.71
0.50
31.15
DIODE
SCANDITRON
IX
30.07
4.89
3.07
0.80
0.60
31.06
INFN-LNS
110
100
20 mm SOBP
RS 1.8 mm
90
80
14.5 mm SOBP
RS 14 mm
70
60
50
11 mm SOBP
RS 10 mm
40
30
20
10
0
0
10
15
20
25
30
35
MODULATOR
RANGE
SHIFTER
MODULATION
(SOBP)
DISTAL-DOSE
FALLOFF(1)
d90%-10%
(mm eye-tissue)
DISTAL-DOSE
FALLOFF(2)
d80%-20%
MAXIMUM
DOSE
INSOBP
%
BEAMRANGE
(90%DISTAL)
(mm eye-tissue)
MOD: 000/00
RS: 14 mm
14.5
0.91
0.65
101.6
14.38
MOD: 010/02
RS: 10 mm
10.5
0.90
0.60
104
18.27
MOD: 009/02
RS: 1.8 mm
20.30
0.80
0.55
103.9
27.04
CCO
17
0.90
0.75
<102
INFN-LNS
RS = 10 mm
80
RS = 2 mm
RS = 14 mm
70
60
50
40
30
20
10
0
0
10
15
20
25
30
35
INFN-LNS
TOLERANCES
1) [(MEASURED RANGE) / (REQUIRED RANGE)]: 0.2 mm
2) [(MEASURED MODULATION) / (REQUIRED MODULATION)]: [ -0.1 mm
// +0.9 mm ]
INFN-LNS
1) LATERAL PENUMBRA:
H
2) Field ratio:
d80%-20%
90% fieldsize
50% fieldsize
3) L95%
4) SIMMETRY (AREA RATIO):
5) FLATNESS:
RT %
Sr
Pmax Pmin
100%
Pmax Pmin
110
100
90
80
N .O .D .%
70
60
50
40
30
20
10
-20
1.3 -10
mm
0
0
10
20
X (mm)
INFN-LNS
Isocenter
Nozzle
PMMA Phantom
Beam axis
PMMA Fantom
Final collimator
Exposition point
1 cm
120
R
e
la
tiv
eIo
n
iza
tio
n%
100
80
60
40
20
2.5 cm
6 cm
0
0
10
15
20
25
30
35
Depth in w ater (m m )
2
1.8
y=-4E-05x 2+0,0207x-0,0095
R2=0,9989
1.6
N
.O
.D
.
1.4
1.2
1
1.2
1
0.8
y=0,0193x-0,0047
R2=0,998
0.8
N
.O
.D
.
0.6
0.6
0.4
0.4
0.2
0.2
0
0
10
20
30
40
50
60
Dose (Gy)
0
0
20
40
60
Dose (Gy)
80
100
120
INFN - LNS
PROFILE MEASUREMENTS
MD - 55 RADIOCHROMIC FILM
120
100
=25mm
Relative dose
80
60
40
20
0
-20
-15
-10
-5
10
15
20
EXPERIMENTAL
SET-UP
(=25mm)
SOBP
(WIDTH)
MOD: 000/00
RS: 14 mm
14mm
zirr.=7mm
MOD: 010/02
RS: 10 mm
Lateral
penumbra (mm)
L95%
(mm)
Sr%
Rt%
(simmetry)
(flatness)
0.92
23
3%
3%
1.40
10mm
zirr.=12mm
0.92
23
2%
3%
1.30
MOD: 009/02
RS:1.8 mm
20mm
zirr.=14mm
0.93
23
3%
3%
1.15
18
zirr.=SOBP
3%
1.10
CCO
CAL
18
zirr.=SOBP
d80%-20%
1.50
INFN-LNS
KODAK XV FILMS AND SCANDITRONIX DIODE WERE USED IN COMMISSIONING
MOD 010/02
RS:10 MM
Lateral penumbra
(mm)
H
L95%
Sr%
(mm)
(simmetry)
zirr.=12 mm
d80%-20%
KODAK XV
1.30
0.92
22.8
0.50
MD-55-2
1.27
0.92
22.7
0.90
120
Diodo
MD55
100
X-OMATV
N.O.D. %
80
60
40
20
0
-25
-20
-15
-10
-5
X Axis (Water m m )
10
15
20
25
INFN-LNS
BEFORE EACH TREATMENT
PLANNED TOLERANCES
1)
3)
160
140
120
25
1.2
0.1
1.3
R90/50:
0.89922
Omogeneit95% [mm]:
21.7
Simmetria [%]:
0.54297
Rt [%]:
2.7259
Path:
Z:\Misure\Trattamenti Marzo2002\20mar2002\preliminari\profilo11.prf
Dose [a.u.]
100
80
60
40
20
Experimental data
Interpolated data
0
-30
-20
-10
0
Distance from axes [mm]
10
20
30
INFN-LNS
COMMISSIONING TESTS
1) PRECISION
1a) SHORT TERM PRECISION: (CV) OF THE RATIO R OF DOSE MONITOR UNITS TO
DOSIMETER SCALE READING FOR n=5 CONSECUTIVE IRRADIATIONS OF 15 Gy.
MEASURED CV = 0.1%
1b) LONG TERM PRECISION (WEEKLY STABILITY):
2) LINEARITY
THE RATIO V BETWEEN THE MEASURED PROTON DOSE AND DOSE MONITOR UNITS
SHALL BE WITHIN 1% AT ALL MONITOR SCALE READINGS, i.e. IN THE CLINICAL
RANGE UP TO 1520 Gy.
MEASURED RATIO
Vmax
1.005
Vmin
INFN - LNS
2400
2200
2000
1800
1600
1400
1200
1000
800
600
400
200
0
R2=1
(Doserate=51.10Gy/min.)
0
50000
100000
150000
U.M. SENT
200000
250000
cGy/U.M.
102
101
100
99
98
97
0
3
tim e (days)
INFN-LNS
DOSE RATE MONITORING (BEAM INTENSITY MONITOR)
THE PROTON BEAM RATE IS MONITORED BY THE VOLTAGE SIGNAL (VSF) PROVIDED BY THE
FIRST SCATTERING FOIL, LOCATED IN VACUUM IMMEDIATELY UPSTREAM OF THE EXIT
WINDOW.
(R2=0.9999)
(VSF) IS USED BY THE C.S. TO STOP THE BEAM IF THE RATE XCEEDS A PRESET LIMIT.
1.05
1.04
U.M. =K
1.03
1.02
Alarm current
1.01
1
Zmarkus = 1 mm
Full energy
beam
0.99
Normal
current
0.98
0.97
0
10
INFN-LNS
M Q M ( C ) kT ,P k pol ks
(sw,air)q
(nC)
pp
D , air ,Q0
ref
cGy
ref
N Dref,air Pcav
Pcelref
cGy/nC
pp
A RADIOACTIVE CHECK DEVICE (90Sr) IS USED FOR OPERATIONAL AND CONSTANCY CHECKS
OF DOSIMETRY SYSTEM (1%, NCRP).
INFN-LNS
AN EXRADIN T1 THIMBLE ION CHAMBER [A-150 TE, 0.05 cm3, NK(60Co)]
CAN BE USED AT LNS IN THE WATER PHANTOM, ACCORDING TO ICRU 59
WORK IN PROGRESS
NOW AVAILABLE BY PTW ND,w,60Co FOR T1 EXRADIN AND MARKUS LNS CHAMBERS
(Nk=1%)
EXRADIN T1
2.6%
2.1%
PTW MARKUS
3.1%
2.5%
INFN-LNS
) INTERCOMPARISON (2) AT PSI (PHYSICA MEDICA VOL.XVII, S.3, PTCOG XXX 1999)
% = 1.2%
INFN-LNS
-2
Standard deviation of the mean = 1.13%
Difference between maximum and minimum dose =
3.3%
-4
-6
-8
MARKUS EXRADIN FARMER MARKUS FWT 683
LNS
T1 LNS
LNS
CCO
CCO
INFN-LNS
110
100
90
Relative output
80
GAF
Markus
Diode1
Diode2
TLD2
TLD1
70
60
50
40
30
MODULATED BEAM
Z = 21 mm WED
PMMA PHANTOM
20
10
0
0
10
12
14
16
18
20
22
24
26
28
INFN-LNS
110
100
90
80
GAF
Markus
Diode1
Diode2
TLD2
TLD1
Relative output
70
60
50
40
30
MODULATED BEAM
Z = 21 mm WED
PMMA PHANTOM
20
10
0
0
10
12
14
16
18
20
22
24
26
28
INFN-LNS
PATIENT DOSES (STRAY RADIATION)
(PERSONAL MONITORING, TWO PATIENTS TESTED)
0.20 mSv
15 Sv/h
(10)
1.5 Sv/h