Analytical Method Validation
Analytical Method Validation
Analytical Method Validation
Background
NDA and ANDA must include the analytical procedures necessary
to ensure:
Identity, Strength, Quality, Purity, and Potency of the Drug Substances
and Drug product [21CFR 314.50(d)(l) and 314.94(a)(9)(i)]
Data to establish and reliability [21CFR 211.169(e) and 211.194(a)
(2)]
Sample Preparation
Analysis
Optimization of Validation
Considerations Prior to
Method Validation
Suitability of Instrument
Status of Qualification and Calibration
Suitability of Materials
Status of Reference Standards, Reagents, Placebo Lots
Suitability of Analyst
Status of Training and Qualification Records
Suitability of Documentation
Written analytical procedure and proper approved protocol with
pre-established acceptance criteria
Analytical Method Validation
Validation Step
Qualification of material.
Perform pre-validation experiments.
Adjust method parameters and/or
acceptance criteria if necessary.
Perform full validation experiments.
Develop SOP for executing the method in routine analysis.
Document validation experiments and results in the
validation report.
Analytical Method Validation
System Suitability
Validation
Calibration
Pump
Injector
Detector
Data System
Analyst
Method
Sample
Analytical Method Validation
10
11
ICH Guidelines
Q2A, Text on Validation of Analytical procedures
(March 1995)
Q2B, Validation of Analytical Procedures: Methodology (May 1997)
12
13
14
Precision
Specificity
Linearity
Range
Accuracy
Repeatability
Intermediate Precision
Limit of Detection
Limit of Quantitation
Robustness
15
Assay Category 2
Assay Category 1
Quantitative
Limit Tests
Assay
Category 3
Accuracy
Yes
Yes
Precision
Yes
Yes
No
Yes
Specificity
Yes
Yes
Yes
LOD
No
No
Yes
LOQ
No
Yes
No
Linearity
Yes
Yes
No
Range
Yes
Yes
Ruggedness
Yes
Yes
Yes
Yes
16
Type of Analytical
Procedure
Impurity testing
Identification
Assay
Quantitative
Limit Tests
No
Yes
No
Yes
Repeatability
No
Yes
No
Yes
Interm. Prec.
No
Yes
No
Yes
Specificity
Yes
Yes
Yes
Yes
LOD
No
No
Yes
No
LOQ
No
Yes
No
No
Linearity
No
Yes
No
Yes
Range
No
Yes
No
Yes
Accuracy
Precision
17
Specificity
Ability of an
analytical method to
measure the analyte
free from
interference due to
other components.
Selectivity
Bias
18
Specificity: ICH/USP
An investigation of specificity should be conducted
during the validation of an identification test, an
impurities assay, and a potency assay.
Procedures used will depend on the intended
objective of the analytical procedure.
If a method can not completely discriminate, two of
more procedures are recommended.
19
Specificity: Identification
Should be able to discriminate between compounds
closely related in structure.
Confirmed by obtaining negative results for samples
with spiked related compounds and positive results
for samples with analyte.
Choice of potential interfering substances should be
based on sensible scientific judgment considering
substances that could likely occur.
Analytical Method Validation
20
Specificity: Impurities/Assay
Chromatographic Methods
Demonstrate Resolution
Impurities/Degradants Available
Spike with impurities/degradants
Show resolution and a lack of interference
21
22
Placebo
Drug Substance Degradants
Drug Product Degradants
Related Substances
Packaging Extractables
23
Humidity
Acid Hydrolysis
Base Hydrolysis
Oxidation
Light
24
Specificity Study
Condition
DEG
#1
DEG
#2
DEG
#3
DEG
#4
Active
ingredients
2 N HCl
17.23
4.71
95.17%
0.1 N
NaOH
100%
30%
H2O2
1.12
1.41
1.65
1.2
99.98%
UV/Vis
4.68
94.67%
25
Linearity
Ability of an assay
to elicit a direct and
proportional
response to
changes in analyte
concentration.
Analytical Method Validation
26
27
Linearity Example
R square = 0.999
Slope = 0.97
y-intercept = 0.233
Line Eq.: Y = 0.97X + 0.233
Std. Error = 1.319
Std. Deviation of Slopes = 0.0079
Analytical Method Validation
28
Range
The interval between the
upper and lower
concentrations of analyte
in the sample that have
been demonstrate to have
a suitable level of
precision, accuracy, and
linearity.
Analytical Method Validation
29
Range
30
31
Accuracy
32
Accuracy
Should be established across specified range of analytical
procedure.
Should be assessed using a minimum of 3 concentration levels,
each in triplicate (total of 9 determinations)
Should be reported as:
Percent recovery of known amount added (reference material) or
The difference between the mean assay result and the accepted
value
Analytical Method Validation
33
Amount
Added (mg)
Amount
Found
(mg)
Percent
Recovery
0.0
0.0
---
50.2
50.4
100.5
79.6
80.1
100.6
99.9
100.7
100.8
% Recovery
% Recovery
99.2
98.9
99.3
99.3
99.4
99.7
99.3
99.3
Std.dev.
0.1
0.4
95%C.I
99.30.25
99.30.99
Mean
34
Analyte ratio
Unit
100
100 %
98-102
10
10-1
10 %
98-102
10-2
1%
97-103
0.1
10-3
0.1%
95-105
0.01
10-4
100 ppm
90-107
0.001
10-5
10 ppm
80-110
0.0001
10-6
1 ppm
80-110
0.00001
10-7
100 ppb
80-110
0.000001
10-8
10 ppb
60-115
0.0000001
10-9
1 ppb
40-120
35
Precision
The closeness of agreement
Ball Ball
Ball
Ball Ball Ball
Ball Ball
Strik Strike
Strike
e Strike
Strike
Strike
36
Repeatability
Intermediate Precision
Reproducibility
37
Repeatability
Express the precision
Should be assessed
using minimum of 9
operating conditions
determinations
(3 concentrations/ 3
of time.
replicates) or
Also referred to as
Minimum of 6
Intra-assay precision
determinations at the
100% level.
Analytical Method Validation
38
Intermediate Precision
Express within-laboratory variations.
Depends on the
to be used.
USP
equipment, etc.
39
Day 2
100.6
99.5
100.8
99.9
100.1
98.9
100.3
99.2
100.5
99.7
100.4
Mean = 100.5
RSD = 0.24%
CI = 100.5 0.24
99.6
Mean = 99.5
RSD = 0.36%
CI = 99.5 0.36
Grand
Mean = 100.0
RSD = 0.59%
40
Reproducibility
41
Reproducibility Study
Lab 1
Lab 2
Lab 3
Day 1
Day 2
Day 1
Day 2
Day 1
Day 2
Analyst
1
Analyst
2
Analyst
1
Analyst
2
Analyst
1
Analyst
2
3 Preps
3 Preps
3 Preps
3 Preps
3 Preps
3 Preps
42
Analyte ratio
Unit
RSD (%)
100
100 %
1.3
10
10-1
10 %
2.7
10-2
1%
2.8
0.1
10-3
0.1%
3.7
0.01
10-4
100 ppm
5.3
0.001
10-5
10 ppm
7.3
0.0001
10-6
1 ppm
11
0.00001
10-7
100 ppb
15
0.000001
10-8
10 ppb
21
0.0000001
10-9
1 ppb
30
43
precision.
Estimated by Signal to Noise
Ratio of 10:1.
44
45
46
DL =
3.3s
S
QL =
10s
S
47
Robustness
Definition: Capacity to remain unaffected by small
but deliberate variations in method parameters
Determination: Comparison results under differing
conditions with precision under normal conditions
Variations may include: stability of analytical solution,
variation of pH in a mobile phase, different column
(lot/supplier), temperature, flow rate.
48
Repeatability
Reproducibility
Ruggedness
Intermediate
Precision
Robustness
49
Precision Terms
Instrument Precision
- 10 Std. Injections
Repeatability
Intermediate Precision
- Two Analyses
Reproducibility
- Two Lab.
Ruggedness
- Many Variables
Robustness
- Intentional Changes
Analytical Method Validation
50
Robustness Variations
All Assays
HPLC Assays
GC Assays
-Different Columns
-Temperature
-Flow Rate
Analytical Method Validation
51
MeOH/
Water
Retention
Time 1
Retention
Time 2
Resolution
75:25
11.94
16.41
7.39
80:20
8.47
11.17
6.17
85:15
7.81
10.18
5.93
52
USP 23 <621>
Definition: evaluation of
equipment, electronic,
analytical operations and
samples as a whole
Determination: repeatability,
tailing factor (T), capacity
Recommendations
In general k 2.0
T2
Repeatability
RSD 2.0% (n 5)
53
Guidance on Re-Validation
When sponsors make changes in the analytical
procedure, drug substance, drug product, the
changes, may necessitate revalidation of the
analytical procedures.
The degree of revalidation depends on the nature of
the change.
FDA intends to provide guidance in the future on
post-approval changes in analytical procedures.
Analytical Method Validation
54
References
Analytical Methods Validation for FDA Compliance The Center for
Professional Advancement 2003. 3.12-14.
Guideline for submitting samples and analytical data for kethods validation (Feb.
1987)
ICH Q2A
ICH Q2B
21 Code of Federal Registrations Part 210 and 211
Michael E. Swatrz and Ira S. Krull, Analytical method development and validation.
Mrcel Dekker, Inc. New York, 1997.
USP 23 <1225>
http://www.waters.com
Ludwig Huber, Validation and Qualification in Analytical Laboratories, Interpharm
Press Inc. Buffalo Grove, Illinois, 1999.
Analytical Method Validation
55