DR800 Service and Maintenance

Como realizar el mantenimiento de un colorimetro DR/890 de la marca Hach, incluye los metodos que puedes realizar los componentes que debes tener en cuneta al momento de realisar un matenimiento completo
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPT, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
387 views21 pages

DR800 Service and Maintenance

Como realizar el mantenimiento de un colorimetro DR/890 de la marca Hach, incluye los metodos que puedes realizar los componentes que debes tener en cuneta al momento de realisar un matenimiento completo
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPT, PDF, TXT or read online on Scribd
You are on page 1/ 21

Hach DR800 Colorimeter

Service and Maintenance

Hach DR800 Colorimeter


Specifications
Wavelength Ranges
DR890 - 420, 520, 560 and 610 nm
DR850 520 and 610 nm
DR820 520 nm

Wavelength Accuracy +/-1 nm


Light Source LED
Detector Silicon Photodiode
LCD Display

Hach DR800 Colorimeter


Specifications
5 User entered programs
External Output Infrared to RS-232
Serial using Data Transfer Adapter
Battery Power 4 AA alkaline batteries
(battery life ~6 months)
Photometric Range - 0-2 A
Stray Light - <1.0% at 400nm
Environmental Rating IP67 Standard

Hach DR800 Colorimeter


Basic Instrument Configuration
4844000 DR820
4845000 DR850
4847000 DR890

Hach DR800 Colorimeter


Instrument Components
Main Components
4943100 Cell Assembly Cap
4943200 Battery Cover
4843400 Battery Holder Assy
4845300 Bottom Housing
Top Housing
4943301 DR/820 ASSEMBLY
4943302 DR/850 ASSEMBLY
4943304 DR/890 ASSEMBLY

Hach DR800 Colorimeter


Instrument Components
Circuit Board Assemblies
48448-01 DR/820 COLORIMETER
48448-02 DR/850 COLORIMETER
48448-04 DR/890 COLORIMETER
*if circuit board is changed the serial number will not
be on the board

Hach DR800 Colorimeter


Instrument Components
Instrument Optics (DR890)
4844100 Dichroic filter
Interference Filters

4844200 420 nm
4844300 520 nm
4844400 560 nm
4844500 610 nm

LEDs
4848200 LED, white (D2 and
D3)
4942100 LED, blue (D1)

5262800 Detector

Hach DR800 Colorimeter


Instrument Components
Due to the discontinuation of the photo-amplifier
4849400, changes were made in January of 2007
resulting in modifications to the main board and
optics.
The following components are available for replacement

4844100 filter, Dichroic


4844200 420 nm interference filter
4844300 520 nm interference filter
4844400 560 nm interference filter
4844500 610 nm interference filter
4848200 LED, white
4942100 LED, blue

Hach DR800 Colorimeter


Instrument Components
The following components are no longer
available for replacement
4844600 Housing, Top Optics
4844700 Housing, Bottom Optics
4849400 Photoamplifier OPT101

The following optics assemblies are now


obsolete
4943401 Optics assembly, DR820
4943402 Optics assembly, DR850
4943404 Optics assembly, DR890

Hach DR800 Colorimeter

Test and Method Number at 420nm

Boron
Chlorine Dioxide, MR
COD, LR
COD, Manganic
Color*
Copper, porphyrin
Hydrazine
Hydrazine, AV
Molybdenum, Molybdate, HR
Nitrate, TNT, CAF
Nitrogen, TKN
Nitrogen, TN, TNT, CAF
Nitrogen, total, HR, TNT

4
7
16
18
19
22
31
32
44
57
65
58
69

Phosphorous, Molybdovanadate
Phosphorous, Molybdovanadate, AV
Phosphorous, Phosver 3
Phosphorous, Phosver 3, AV
Phosphorous, Reactive HR, TNT
Phosphorous, Total, HR, TNT
Silica, HR
Silica, UHR
Triazole, Benzotriazole
Triazole, Tolytriazole

77
78
79
79
86
87
89
88
3
3

Hach DR800 Colorimeter

Test Method and Number at 520nm

Aluminum, Aluminon
Bromine
Bromine AV
Chlorine, free, AV
Chlorine, free, HR
Chlorine, free, PP
Chlorine, free, TNT
Chlorine, total, AV
Chlorine, total, HR
Chlorine, total, PP
Chlorine, total, TNT
Cyanuric acid
Hardness, calcium
Hardness, magnesium
Iron, Ferrous
Iron, Ferrous, AV
Iron, total, Ferrover

1
5
6
11
8
9
10
11
8
9
10
24
29
30
33
33
33

Iron, total, Ferrover, AV


Manganese, HR
Nitrate (Cd reduction) HR AV
Nitrate (Cd reduction) HR PP
Nitrate (Cd reduction) MR AV
Nitrate (Cd reduction) MR PP
Nitrate, LR
Nitrite,LR
Nitrite,LR, AV
Nitrite,TNT
Oxygen,dissolved, HR, AV
pH
Phosphorous, amino acid
Sulfate
Sulfate, AV
Turbidity
Volatile Acids

33
41
50
51
53
54
55
60
62
63
70
75
85
91
92
95
96

Hach DR800 Colorimeter

Test Method and Number at 560nm

Chromium, Hexavalent
Chromium, Hexavalent, AV
Chromium, total
Copper, Bichinchoninate
Copper, Bichinchoninate, AV
DEHA^
Iron, total, Ferrozine
Manganese, LR
Nickel, PAN
Nitrite, HR

13
14
15
20
20
25
37
43
48
59

Hach DR800 Colorimeter

Test Method and Number at 610nm

COD-HR, HR+
Cyanide
Detergents^ (downsize)
Fluoride, SPADNS
Fluoride, SPADNS, AV
Iron, total, FerroMo
Iron, total, TPTZ*
Iron, total, TPTZ, AV
Molybdenum, ternary complex
Nitrogen, Ammonia, HR, TNT
Nitrogen, Ammonia, LR, TNT
Nitrogen, Ammonia, Salicylate
Nitrogen, monochloramine &
free ammonia, Salicylate, AV

17
23
26
27
28
38
39
39
47
67
66
64
49

Nitrogen, monochloramine &


free ammonia, Salicylate
Nitrogen, Total Inorganic TNT
Oxygen, dissolved,LR, AV
Ozone, HR, AV
Ozone, LR, AV
Ozone, MR, AV
Phosphorous, acid hydrolyzable,
Phosver 3, TNT
Phosphorous, Phosver 3, TNT
Phosphorous, total, Phosver 3, TNT
Silica, LR
Sulfide
Suspended Solids
Zinc

49
68
71
74
72
73
82
82
82
90
93
94
97

Hach DR800 Colorimeter


Maintenance
1. Clean or replace sample compartment
2. Replace batteries
3. Replace interference filters as needed

Hach DR800 Colorimeter


Troubleshooting
Replace Main Board

E-1
E-2
E-3
E-6

unconfigured instrument
could not read program/data
could not write program/data
measurement offset error

E-4 measurement battery error replace batteries


E-5 measurement A/D error replace main board or
detector
E-7 measurement low light error replace interference
filter, LED or detector
E-8 measurement over-range error check samples replace main board

Hach DR800 Colorimeter


Troubleshooting
1. Keypad spills or missing keys, older
version may short out. Replace top housing.
2. Interference filters cause accuracy issues
3. Dirty Sample Compartment can get
scratched or stained causing accuracy issues
4. Circuit board wont turn on, software
issues

Hach DR800 Colorimeter


Service
Check instrument accuracy by using Absorbance
Spec Check Standards - Catalog Number 27639-00
Includes a zero standard and 3 absorbance standards
Example Certificate

Hach DR800 Colorimeter


Service

Hach DR800 Colorimeter


Service

Hach DR800 Colorimeter


Service

Hach DR800 Colorimeter


Questions & Answers