TEC5500 Tech Seminar 2008
TEC5500 Tech Seminar 2008
TEC5500 Tech Seminar 2008
Contents
Features Board/Unit location and Block Diagram Screens for Maintenance Troubleshooting Case Study
Nihon Kohden 2008
Features
AC and battery power (Ni-MH) operation 5.7 inch color TFT-LCD Synchronized cardioversion AED (Automated External Defibrillation) 50 mm width thermal array recorder External pacing (TEC-5531) External paddles, internal paddles disposable pads Options: 5 lead-ECG, SpO2, etCO2 or AUX out, TEC-5500: Voice prompt/Sound recording in SD card
2/36
Board/Unit Location
3/36
4/36
CPU board
Mother board
Key board
5/36
6/36
ECG signals from the electrodes are input through the LEAD ECG floating circuit. ECG signals are selected in the ECG MPX as lead signals. The lead signal (Lead I, II, III, aVR, aVL, aVF to V) is selected by the user. This lead information is sent from the Main CPU to the Sub CPU.
ECG MPX
Selected Lead ECG signal
Differential amplifiers - This circuit amplifies differential (biological) signals and reduces common mode signals (AC interference)
High-cut and Low-cut filter - The frequency response is from 0.05 to 150 Hz for input through ECG connection cable 0.05 to 30 Hz when High-cut filter (System Maintenance screen Configuration) and AC filter are set to ON Notch filter - To check if noise is caused by AC interference Signal generator To generate test and calibration waveforms
ii)
iii)
iv)
SUB CPU
The Sub-CPU receives the signals from the ECG Amplifier. The Sub-CPU converts the analog ECG data to digital data and sends it to Main CPU as parallel data through the Main FPGA. The QRS detection, Heart rate count, VF analysis for AED are performed by the Main CPU.
7/36
ECG signals from the paddles are input through the Paddle ECG floating circuit.
The lead signal in this case is fixed to the paddle lead (Apex to Sternum) . ECG MPX
Paddle Lead ECG
Differential amplifiers This circuit only amplifies potential difference between two electrode. High-cut and Low-cut filter The frequency response is from 0.5 to 20Hz for input through external paddle Notch filter To check if noise is caused by AC interference Signal generator To generate test and calibration waveforms
SUB CPU
The Sub-CPU receives the signals from the ECG Amplifier. The Sub-CPU converts the analog ECG data to digital data and sends it to Main CPU as parallel data through the Main FPGA. The QRS detection, Heart rate count, VF analysis for AED are performed by the Main CPU.
8/36
SpO2
SpO2 signal from patient
probe/adapter
The Red and Infrared signals are detected through the detector and are sent to the DSI Interface Unit/DSI/AUX Out Interface unit as serial data
Serial data
Sends the serial data (SpO2 value, PR (Pulse Rate) and pulse waveform data) to the FPGA.
FPGA
Converts the received serial data to parallel data and sends it Main CPU.
MAIN CPU
9/36
Sensor kit
Serial data
The Infrared signals are detected through the detector and are sent to the DSI Interface Unit/DSI/AUX Out Interface Unit as serial data
Sends the serial data (etCO2 value, PR and CO2 waveform data) to the FPGA.
FPGA
Converts the received serial data to parallel data and sends it to the MAIN CPU.
MAIN CPU
10/36
11/36
Lead ECG
floating circuit
ECG signals from patient are fed through the ECG electrodes to the Lead ECG floating circuit
This circuit continuously detects if any ECG leads have come off
The ECG leads status are sent to the Main Input port
12/36
Paddle ECG
floating circuit
ECG signals from patient are fed through the Paddles electrode to the Lead ECG floating circuit
Paddle contact
Detector When both Discharge buttons are not pressed, the paddles electrodes pick up the patients physiological signal. From this picked up signal, Main Input the electrode-to-skin contact resistance can be determined
The paddle contact status is sent from Main Output port to the LED status on the External paddle
LED status on
External Paddle
13/36
14/36
Main CPU
FPGA
LCD Unit
LCD Inverter
15/36
FPGA
Voice CPU
Sound Amplifier
Amplifies the voice signal or sound signal and sends it to the Speaker.
Speaker
16/36
FPGA
Converts the waveform data and measurement data to the recording data and sends to the thermal head. Generates the motor driving signal (pulse) for the Motor Driver.
Motor Driver
Stepping Motor
Fees the recording paper. A stepping motor rotates by an exactly angle with a pulse from the motor driver IC.
Thermal head
17/36
18/36
HV Circuit Charging/Discharging
External relay Opens and closes the connection between the Paddle ECG input circuit and pacing output circuit, and the HV charging circuit. When the Charge button is pressed, the set of relay switches closes to the connection between the DC/DC converter and HV capacitor for charging. When the Discharge buttons are pressed simultaneously, the set of relay switches closes the connection between the HV capacitor and Patient
Step-up transformer - DC/DC converter that provides the voltage charge the HV capacitor. This block receives the start charging signal from the Sub CPU, and outputs the control signal to the switching FET that switches on or off the HV stepup transformer
Internal relay Used to disarm the capacitor when a fault is detected, the capacitor is in charged condition for more than 40 seconds (default setting, can be changed in Charge hold time), or when the rotary switch is set to Disarm. In this situation, the internal relay closes the circuit to discharge the stored energy of the HV capacitor
19/36
4. Calculates TTR (Transthoracic Resistance) from integration of discharge current change - The change of magnetic flux in the primary winding induces the voltage in the
pickup coil (secondary winding). - The voltage at the secondary winding is integrated. Sub CPU calculates the TTR and delivered energy according to the integrated value.
20/36
High-voltage Capacitor
Protection from sudden failure of capacitor by segmentation of dielectrics and thin lead connection of cell capacitors Dielectrics
Thin Lead
Electrode (metal)
Basic Checks
Setup
Report Setup - Defibrillation Report/Event List Report Battery Test HV Capacitor Test Recorder Test Report History (Instrumnt Report) Check Hardware
22/36
System Setup
System Maintenance
Press the Reset key to delete the previous battery test history when the battery is replaced.
Test result Description Action
Battery OK XXX% (Capacity XXX%:100%, 80% 70%) Battery WEAK XXX% Capacity XXX%: 60%, 50%)
Capacity is enough.
Battery operation is available anytime. When the defibrillator is not used, always do trickle charging. Battery operation is available only when the battery is fully charged. The time to replace the battery is coming soon. It is recommended that the battery be replaced with a new one. Only AC operation is available. Immediately replace the battery with a new one.
23/36
Troubleshooting
The following repots are useful for troubleshooting.
Defibrillation report (Setup screen Report Setup Defib) Setting and waveforms when discharge operation is performed. Event list report (Setup screen Report Setup Event List) Instrument operation history Instrument report (System Setup screen Report History Instrument report) Error code, battery test and HV capacitor test result We can know what error occurred and what the operator was doing from the reports.
24/36
Case Study
Power on problem
Problem: The defibrillator does not turn on when the Energy/Mode Select control (rotary knob) is set to the position other than the OFF position.
25/36
Power on problem
Cause: When the AC power cord is connected to AC outlet on the wall, if the AC power lamp and battery charge lamp are lit, the power unit is operative. Check if you can hear the click sound by pressing any of the key switches on the front panel. There is the click sound: LCD or LCD inverter is faulty. No click sound: Faulty power circuit on the mother board. Solution: Replace the faulty board or unit.
26/36
Normal example
27/36
Cause: The battery temperature exceeds 60C. When the battery temperature exceeds 60C, the battery test stops and the battery check result appears on the screen. Then, the message appears to prompt you to fully charge the battery by tuning the Energy/Mode select control to OFF position. This is normal operation.
28/36
A512 HV Error
Problem: A512 error is displayed and energy cannot be charged. The sub CPU monitors the output of the HV unit. When the output of HV unit does not match the control signal, this error is displayed. Block diagram Error code
29/36
A512 HV Error
Possible Causes:
Connection cable between Biphasic HV unit and HV capacitor. Connection cable between Biphasic HV unit and CPU board. HV capacitor Biphasic HV unit
30/36
Problem 1: When discharge test is performed in the paddle holder, the Test fail message appears. Problem 2: When the paddle is placed in the paddle holder, paddle contact LED turns red.
31/36
Cause:
Resistance of the resistor in the Test Load Board is changed or the resistor is damaged due to test discharge.
Solution:
Replace resistors. Test Load Photo
32/36
Cause: Test electrode plate is rusted. Solution: Clean or replace the plate.
33/36
Problem:
The Out of paper message appears even when paper is loaded normally and magazine door is tightly closed.
34/36
35/36