IEC62353
IEC62353
IEC62353
Introduction
Specially formulated for in house and routine electrical safety testing of medical equipment at hospitals Introduced by mid of 2007
IEC 62353
Historical Perspective
1960-1970: Increase in adoption and use of biomedical instrumentation in hospitals
IEC 62353
Cardiac arrest
Ventricular fibrillation Burns Electrolysis
IEC 62353
IEC 62353
Class I
Equipment with exposed conductive parts connected to the protective earth in the fixed wiring of the installation
Class II
Internally Powered
Equipment with additional protection to basic insulation, but having no direct connections to protective earth
Equipment is powered from battery
IEC 62353
F- Type
Applied part which is electrically isolated from Earth and other parts of the medical equipment. i.e. floating
B- Type
BF- Type CF- Type
Applied parts, which are usually Earth referenced. Not suitable for direct cardiac application. e.g. Dental chair
Applied part complying with a higher degree of protection against electric shock than Type B applied parts. Not suitable for direct cardiac application. e.g. SPO2 Applied Part complying with the highest degree of protection against electric shock. Suitable for direct cardiac application. e.g. ECG monitor
IEC 62353
Tests Conducted
Insulation Measurements
Leakage Measurements
IEC 62353
Earthbond Testing
Tests the integrity of the low resistance connection between the earth conductor and any metal conductive part (Class I Equipment) Earth bond resistance should be less than 0.3 ohm Test voltage to be applied - less than 24V Test current to be applied - +/- 200 mA constant
IEC 62353
This test is applicable to both CLASS I (earthed) and CLASS II (unearthed) equipments. During the measurement all switches of the MAINS PART shall be in operating position (ON). Insulation resistance measurements shall be performed with 500V (D.C.). Specifications Measurement : 1. EUT to Earth / Ground 2. AP to Ground Test voltage : 500V DC @1mA Safe limit : >100K
IEC 62353
IEC 62353
Leakage Tests
Equipment leakage Current flowing from the mains part to earth via the protective earth conductor, accessible conductive parts of the enclosure and applied parts Measured by three different methods Direct method, Alternative method and Differential method Applied part leakage Current flowing from main parts and the accessible conductive parts of the enclosure to the applied parts Measured by two different methods - Direct method and Alternative method
IEC 62353
Leakage Tests
a. Direct method
Measures true leakage that would occur whilst the medical equipment is in typical use, this will give highest accuracy on low leakage current measurement. The DUT will be powered ON from mains through the analyzer
Specifications Test Voltage Measuring Device : 230V, 50Hz : As per IEC 60601-1 requirements
b. Differential method
Measures the total leakage current and DUT does not need to be isolated during measurement. The DUT will be powered ON from mains through the analyzer and the measurement is taken as the difference between line and neutral currents
Specifications Test Voltage Measuring Device : 230V, 50Hz : As per IEC 60601-1 (1KOhm)
c. Alternative method
This test provide highest safety for person doing the test. Here the supply inputs of the DUT will be shoted and an equivalent voltage generated by the analyzer will be connected as prescribed in the standard
Specifications Test Voltage Measuring Device : 250V, 50Hz : As per IEC 60601-1 (1KOhm)
IEC 62353
Leakage Tests
Equipment leakage a a Applied part leakage
Internally Powered
IEC 62353
www.cdac.in
IEC 62353
IEC 60601
Manufacturers type-test standard
Need special laboratory condition Stringent measurement parameters
IEC 62353
Routine and in house testing of equipment
No need of special laboratory condition Nominal measurement parameters suitable for routine testings
IEC 62353
CDAC-T and SCTIMST jointly developed Indias 1st indigenous medical electrical safety analyzer