Media Fills and Environment Atl Monitoring 26 July 2011 Presentation One
Media Fills and Environment Atl Monitoring 26 July 2011 Presentation One
Media Fills and Environment Atl Monitoring 26 July 2011 Presentation One
26/7/2011
Documents Applied
ISO 13408-1:2008 Aseptic processing of health care products Part 1: General requirements (parts 2-8 also deal with aseptic processing)
PIC/S Guide to Good Manufacturing Practice for Medicinal Products Annex 1 Manufacture of Sterile Medicinal Products PIC/S Recommendation on the Validation of Aseptic Processes January 2011 PDA Technical Report No. 28 Process Simulation Testing for Sterile Bulk Pharmaceutical Chemicals
27/07/2011
ISO
OLSS Chief Microbiologist is the Australian representative on the ISO Technical Committee 198 (which deals with sterilisation standards) through Standards Australia she represents Australia, not the TGA. OLSS Microbiology is heavily involved in the ISO standards writing and review process. Review process is open to everyone. We are currently involved in the drafting of ISO 13408-7 Aseptic processing of healthcare products Part 7 Alternative processes for atypical medical devices and combination products and ISO 13408-8 Aseptic processing of health care products Part 8 Cell-based health care products ISO 13408-7 draws heavily on detail already included in ISO 13408-1, but includes an excessive amount of information concerning application of risk assessment is likely to undergo modification before entering final draft stage
Listed documents are applied during evaluation of pre-market submissions, during audit and during assessment of post-market problems What do we expect to see?? In an ideal world: -Every sterile bulk hold period would be simulated for the maximum hold period -This would immediately be followed by simulation of the maximum permitted filling time -Every possible intervention, stoppage, process, procedure or worst case situation would be simulated -Every possible container/closure combination would be tested
-Zero contamination would be detected -Results of environmental monitoring would be zero -All would be repeated every 6 months thereafter Reality is somewhat different can be considerable variation in performance of media fills but this is not always bad!
Simulation of procedures, interventions, maximum number of personnel Frequency of simulations Simulation of container/closure combinations (bracketing is accepted provided worst case combinations are tested) Incubation period and conditions for filled units
Problem areas
-Simulation of prolonged holding periods directly followed by filling -Simulation of the maximum permitted filling time -Use of alternative media that may not be appropriate -Attitudes to media fills as a whole -Investigation of failures -Campaign filling
Filling times
Maximum permitted filling time is not simulated, or is simulated in frequently
ISO 13408-1:2008 states that Process simulation runs shall be conducted over the maximum permitted filling time. Where this is not possible, simulation runs shall be of sufficient duration to cover representative manipulations, interventions and shift changes performed in actual processing ( Clause 10.3.8) PIC/S Recommendation on the Validation of Aseptic Processes states that where filling takes place over extended periods ie longer than 24 hours, the process simulation test should extend over the whole of the standard filling period (Clause 4.1.2)
Filling times
-The premise is that media fills will follow manufacturing conditions as closely as possible
- Several manufacturers do perform media fill simulations to cover the maximum permitted filling time this is easier when the fill time is less than 24 hours - However, significant proportion do not, particularly if they are US based manufacturers
Anaerobic media use obviously acceptable if a media fill needs to be performed to detect anaerobes - Sterile bulk APIs can be a challenge to validate usually a wet phase and a dry phase - can use media to validate, but more often see combinations of media and dried substances like lactose or mannitol - Have seen saline washed through the system with the whole lot filtered and filter placed on agar not ideal
Failures
Typical scenario is that a submission for a new company will contain media fill data, but it may be several years old not uncommon for us to received data generated in 2004 or sometimes earlier.
Always will ask for more recent data Sometimes comes out at this point that there has been a major or catastrophic failure in media fill validation
What happens then?? Failure investigation is examined checking to see what has been done to rectify the problem, - does not mean that regulatory disaster will follow!
Campaign Filling
-
Usually several batches of sterile API or product manufactured and filled over several days, sometimes weeks Can be problematic to validate Have accepted piggy back media fills where a fill is performed after a campaign, before CIP and SIP If CIP/SIP occurs during campaign, have accepted simulation of period between these events - if no CIP/SIP occur during campaign, expectation is that the duration of the campaign will be simulated
Non intact units are not expected to be incubated What about units that are intact but would normally be rejected??
Views on this vary below is a compilation of views from OLSS Microbiology All intact units should be incubated and counted in media fill If units would be routinely discarded during normal production process, then should be discarded and not incubated Units discarded should be set aside and incubated separately, and considered in terms of results