Test and Inspection Data Checklist: Service Actions Not Requiring Verification

The document is a Test and Inspection Data Checklist for incident report number 3004464965 regarding the ACL ELITE PRO system. It outlines service actions performed, including performance verification and mandatory actions taken, as well as quality verification responsibilities for the customer. The customer waived control verification, and detailed instructions are provided in the service manual.

Uploaded by

pranshu126545
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
3 views1 page

Test and Inspection Data Checklist: Service Actions Not Requiring Verification

The document is a Test and Inspection Data Checklist for incident report number 3004464965 regarding the ACL ELITE PRO system. It outlines service actions performed, including performance verification and mandatory actions taken, as well as quality verification responsibilities for the customer. The customer waived control verification, and detailed instructions are provided in the service manual.

Uploaded by

pranshu126545
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

Test and Inspection Data Checklist

Incident Report Number: 3004464965 Serial Number: 09051214R Date: Jul 9, 2025

System Activity: Performance Verification Model: ACL ELITE PRO

Service Actions Not Requiring Verification


Non-instrument service action performed (e.g. printer, UPS, etc.). See Incident Report (IR) for further detail
System not repaired. See Incident Report (IR) for further detail
Service Mandatory Actions
✔ Reviewed File Error History, observations will be captured in IR code

Perform Acquisition adjustment. Coagulation channel range is 2.3vdc to 2.9vdc using 200ul of Wash-
Reference Emulsion in Rotor position #20. The Chromogenic channel range is 8.65vdc to 8.75vdc with no
Rotor
✔ Verify proper drainage of Rinse / Waste reservoir and decontaminate as necessary
✔ Verify the Needles alignment at the Rotor position
✔ Verify the Software Identification is correct and current software diskettes are readily available
Create a backup of the acl8000.ini file, if necessary
✔ Confirm that customer has a current database backup and it is accessible for use at all times
Quality Verification Actions
It is the responsibility of the customer, based on local Regulatory requirements, to select which QC tests are to
be completed in order to determine functional performance to their satisfaction.
Request customer perform laboratory established levels of control and verify QC results are within
established ranges
Customer was not available to perform QC verification following completion of Service activity
Customer waived control verification

Detailed instructions are contained in the Service Manual.


Form# QMPS-40001-04-C, Rev 06
Authorization Group HMNZ

Page 1 of 1