Gauges
Gauges
Gauges
Vacuum Gauges and Control Instruments 10-12 - 2000 mbar (10-12 Torr) Calibration Service
179.01.02 Excerpt from the Oerlikon Leybold Vacuum Full Line Catalog Product Section C16 Edition 2010
Contents
General Applications for Vacuum Gauges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Select the Combination of Sensor and Gauges which is ideal for your Application . . . . . . . . . . . . . . . . . Basic Terms of Vacuum Metrology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Connection Accessories for Small Flanges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Products Mechanical Gauges (Vacuum Gauges Analog Readout) Bourdon Vacuum Gauges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C16.09 Capsule Vacuum Gauges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C16.10 Diaphragm Vacuum Gauge DIAVAC DV 1000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C16.12 Handheld Measurement Instruments THERMOVAC Sensor TM 90 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C16.14 THERMOVAC Sensor TM 100 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C16.16 Active Sensors / Transmitters CERAVAC Transmitters CTR 100/CTR 91 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C16.18 THERMOVAC Transmitters TTR 91/TTR 91S/TTR 96S . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C16.22 THERMOVAC Transmitters with Fieldbus Interface TTR 211 D/PB / TTR 216 D/PB . . . . . . . . . . . . . . . . C16.26 THERMOVAC Transmitters TTR 100/TTR 100 S2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C16.30 PENNINGVAC Transmitters PTR 225/PTR 225 S/PTR 237 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C16.34 PENNINGVAC Transmitter PTR 90 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C16.38 IONIVAC-Transmitter ITR 90 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C16.42 IONIVAC-Transmitter ITR 200 S . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C16.46 Spiral Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C16.50 Connection Cables for Active Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C16.52 Operating Units for Active Sensors DISPLAY ONE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . DISPLAY TWO / THREE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CENTER ONE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CENTER TWO / THREE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Operating Units for Passive Sensors PIEZOVAC PV 20 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C16.62 Ultra High Vacuum Gauges IONIVAC IM 540 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C16.64 COMBIVAC CM 51 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C16.68 Additional Sensors IONIVAC Sensors IE 414/IE 514 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . THERMOVAC Sensors TR 211/TR 211 NPT/TR 212/TR 216 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . PENNINGVAC Sensors PR 25/PR 26/PR 27/PR 28 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Linear Pressure Sensors DI 200/DI 201/DI 2000/DI 2001/DI 2001 rel . . . . . . . . . . . . . . . . . . . . . . . . . . . Spare Sensors for Older Operating Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Pressures Switches and Control Instruments Pressure Regulation Valves Move . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Low Pressure Safety Switch PS 113 A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Pressure Switches PS 115 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Switching Amplifier SV 110 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Diaphragm Pressure Regulators MR 16/MR 50 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Miscellaneous Oerlikon Leybold Vacuum Calibration Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C16.98
C16 02
Oerlikon Leybold Vacuum Full Line Catalog
C16.72 C16.74 C16.76 C16.78 C16.80 C16.82 C16.90 C16.92 C16.94 C16.96
Mechanical engineering
Car industry; filling of brake and air conditioning systems Vacuum conveying technology Packaging technology Isolation vacuum
G au ge s
Chemical processes
Absolute pressure measurements in gas mixtures Drying and degassing processes Solvent recovery Vacuum pressure control in existing central vacuum supply systems
Electrics/electronics/optics
Evaporation and coating systems Monitoring and controlling of sputter systems Semiconductor technology (CVD, plasma etching etc.) Ion implantation Lamp production
Research
Measurement of ultimate pressure in UHV systems Application in MBE systems Particle accelerators Beam guidance systems, cyclotron Fusion experiments Space simulation chamber
Calibration
Calibration of vacuum gauges and mass spectrometers Reference instruments for the determination of the physical properties of gases Precision measurements of low pressures also in the presence of corrosive or reactive gases
Miscellaneous
Vacuum annealing, melting, soldering and hardening furnaces Cooling and air conditioning technology Electron beam welding Metallurgy
03 C16
Capacitance Diaphragm CERAVAC Transmitters (x = 2, 3, 4 or 5) 1) CTR 91 / CTR 101 CTR 91 / CTR 101 CTR 91 / CTR 101 CTR 91 / CTR 101 CTR 91 / CTR 101 CTR 100 CTR 100 CTR 100 CTR 100 Thermal Conductivity (according to Pirani) THERMOVAC Transmitters 1) TTR 100 (Pirani combined with capacitance diaphragm) TTR 91 TTR 96 S TTR 211 Profibus/DeviceNet TTR 216 Profibus/DeviceNet Handheld Measurement Instruments THERMOVAC Sensors TM 90 + 100 Cold Cathode Ionization (according to Penning) PENNINGVAC Transmitters PTR 225, DN 25 KF PTR 237, DN 40 CF PTR 90, DN 40 CF PTR 90, DN 40 KF PTR 90, DN 25 CF Hot Cathode Ionization IONIVAC Transmitters (Bayard-Alpert combined with Pirani) 1) ITR 90 / ITR 200 ITR 90 / ITR 200 with Display IONIVAC Sensors IE 414 (Bayard-Alpert) IE 514 (Extractor) Linear Diaphragm Sensors 2) DI 200 DI 201 DI 2000 DI 2001 DI 2001 rel.
2 x 10-11 0.01 0.0001 5 x 10-10 5x 10-10 1000 1000 1 x 10-9 1x 10-9 0.01 0.01 1000 1000 1000 0.0005 0.0005 0.0005 0.0005 0.0005 1500 1000 1000 1000 1000 0.00013 0.013 0.0013 1.3 13 0.013 0.0013 0.00013 0.000013 0.13 0.13 1330 133 1.3 13 0.13 1330 133
1 x 10-12 1 10-12
mbar
1) 2)
10-11 10-10 10-9 10-8 10-7 10-6 10-5 10-4 10-3 10-2 10-1
10
100 1000
Different Part Numbers depend on the vacuum connection Differences of the sensors in the same operating range caused by the materials in contact with the medium
C16 04
1 1 1 1 1 1 1 1 1
1, 2 1, 2 1, 2 1, 2 1, 2 1, 2 1, 2 1, 2 1, 2
1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3
3, 4 3, 4 3, 4 3, 4 3, 4 3, 4 3, 4 3, 4 3, 4
1 1 1 1 1
1, 2 1, 2 1, 2 1, 2 1, 2
1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3
1 1 1 1 1
1, 2 1, 2 1, 2 1, 2 1, 2
1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3
3, 4 3, 4 3, 4 3, 4 3, 4
1, 2 1, 2 1 1 1 1, 2 1, 2 1, 2
1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3
1 1 1 1 1
1, 2 1, 2 1, 2 1, 2 1, 2
1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3 1, 2, 3
1, 2
1, 2, 3
1, 2 1, 2
05 C16
Vacuum Gauges where the Pressure Readings are Independent of the Type of Gas (Mechanical Vacuum Gauges) BOURDON Vacuum Gauge
The inside of a tube which is bent into a circular arc (the so-called Bourdon tube) is connected to the vacuum system. Due to the effect of the external atmospheric pressure, the end of the tube bends more or less during the evacuation process. This actuates the pointer arrangement which is attached to this point. The corresponding pressure can be read off on a linear scale. With Bourdon gauges it is possible to roughly determine pressures between 10 mbar (7.5 Torr) and atmospheric pressure.
sing evacuation, the difference between the pressure which is to be measured and the pressure within the reference chamber becomes less, causing the diaphragm flex. This flexure may be transferred by mechanical means like a lever, for example, to a pointer and scale, or electrically by means of a strain gauge or a bending bar for conversion into an electrical measurement signal. The measurement range of such diaphragm vacuum gauges extends from 1 mbar (0.75 Torr) to over 2000 mbar (1500 Torr).
C16 06
Vacuum Gauges where the Pressure Readings depend of the Type of Gas Thermal Conductivity Gauge (Pirani)
This measurement principle utilizes the thermal conductivity of gases for the purpose of pressure measurements in the range from 10-4 mbar/Torr to atmospheric pressure. Today, only the principle of the controlled Pirani gauge is used by Oerlikon Leybold Vacuum in order to attain a quick response. The filament within the gauge head forms one arm of a Wheatstone bridge. The heating voltage which is applied to the bridge is controlled in such a way, that the filament resistance and thus the temperature of the filament remains constant regardless of the quantity of heat given off by the filament. Since the heat transfer from the filament to the gas increases with increasing pressures, the voltage across the bridge is a measure of the pressure. Improvements with regard to temperature compensation have resulted in stable pressure readings also in the face of large temperature changes, in particular when measuring low pressures.
Other electrode arrangements permit access to a higher range of pressures from 10-1 mbar/Torr down to 10-10 mbar/Torr. For the measurement of pressures below 10-10 mbar/Torr so-called extractor ionization sensors after Redhead are employed. In extractor ionization gauges the created ions are focused onto a very thin and short ion detector. Due to the geometrical arrangement of this system, interfering influences such as X-ray effects and ion desorption can be almost completely eliminated. The extractor ionization gauge permits pressure measurements in the range from 10-4 to 10-12 mbar/Torr.
07 C16
DN 10 KF
Part No. 183 53 Part No. 883 95
DN 16 KF
Part No. 183 53 Part No. 883 96
DN 25 KF
Part No. 183 54 Part No. 883 97
Ordering Information
Outer centering ring with O-ring Aluminum / FPM (FKM) ((Viton)) Fine filter on centering ring with O-ring Stainless steel / FPM (FKM) ((Viton)) Baffle with centering ring (FPM (FKM)) Connection accessories for metal seals or degassing room up to 150 C Ultra sealing ring, aluminum (Set of 3) Outer support ring Clamping ring Connection accessories for CF connections Copper seals, (set of 10 pieces) Screw (set of 25 pieces) Replacement sinter filter with centering ring
DN 40 KF
Part No. 183 55
DN 16 CF
DN 40 CF
C16 08
80 105 64
30
Typical Applications
- Vacuum distillation
140
101
50
Dimensional drawing for the BOURDONVAC A (top) and the BOURDONVAC C (bottom)
Technical Data
Measurement range Measurement uncertainty Class 1 (EN 837) Overload range (abs. briefly) Storage temperature range Nominal temperature range Flange connection Length of scale Diameter Overall weight Weight Leak tightness mbar % FS % FS bar C C DN mm mm mm kg (lbs) mbar x l/s
BOURDONVAC A
1 to 1020 1 1 1.5 -25 to +60 10 to 60 16 KF 207 80 105 0.25 (0.55) 1 x 10-8 nickel plated standard steel, bronze, soft solder
BOURDONVAC C
1 to 1020 1 1 1.3 -25 to +60 10 to 100 (max.) 16 KF 188 101 140 0.5 (1.10) 1 x 10-8 stainless steel 1.4404
Ordering Information
Bourdon vacuum gauge
BOURDONVAC A
Part No. 160 40
BOURDONVAC C
Part No. 161 20
09 C16
Typical Applications
- Measurement of absolute pressures (for inert gases only) - Vacuum conveying systems - Operation monitoring - Packaging
110
10 100 40 53 3,5 53
101+1
92
4,5 120 13
Dimensional drawings and panel cut-outs for the capsule vacuum gauges Part Nos. 160 63/64 (left) and Part No. 160 68 (right)
C16 10
Technical Data
Measurement range Measurement uncertainty Overload range (abs. briefly) Storage temperature range Nominal temperature range Length of scale Dead volume, approx. Diameter Weight Vacuum Connection mbar % FS bar C C mm cm3 mm kg (lbs) DN 1 to 100 1.0 1.5 -25 to +60 10 to 50 205 235 132 0.7 (1.54) 16 KF
45 brass standard steel nickel plated glass NBR aluminum copper beryllium soft and hard solder resin
45 brass standard steel nickel plated glass NBR aluminum copper beryllium soft and hard solder resin
Ordering Information
Capsule vacuum gauge Part No. 160 63
11 C16
Typical Applications
- Chemical processes - Vacuum destillation - Absolute pressure measurements for gas mixtures - For use in explosion hazard rated areas - Drying processes - Lamp manufacture
180
166
166
C16 12
Technical Data
Measurement range Measurement uncertainty 01 - 10 mbar (1.0 - 7.5 Torr) 10 - 1000 mbar (7.5 - 750 Torr) Storage temperature range Nominal temperature range Permissible overload (abs.) Lenght of scale / dead volume Vacuum Connection Dimension (W x H x D) Weight C C bar mm DN mm kg (lbs) mbar (Torr)
DIAVAC DV 1000
1 to 1000 (1 to 750)
1 mbar (Torr) 10% of meas. value -25 to +60 0 to 60 3 270 / 130 40 KF 180 x 166 x 100 2.7 (5.95) stainless steel 1.4301, 1.4310 (diaphragm), FPM (FKM)
Ordering Information
DIAVAC DV 1000 mbar readout Torr readout DKD calibration Replacement sintered filter with DN 40 KF centering ring Replacement housing, complete
1)
DIAVAC DV 1000
13 C16
Typical Applications
- Pressure measurements on refrigerating and air-conditioning systems before filling in the refrigerant - Pressure measurements during servicing assignments
29,3
172,6
23,7
30
C16 14
Technical Data
Power source Battery life Connection Controls Indicated units of measurement Measurement range h DN
THERMOVAC Sensor TM 90
9 Volt battery (type 6 LR61) <5 16 KF or 25 KF On/Off/Unit selector mbar, microns, Pa atm. to 1 x 10-2 mbar (atm. to 1 Pa, atm. to 1 micron) after 15 minutes; last read out is displayed again after restarting mbar microns < 2 5%, > 2 10%, < 20.000 5%, > 20.000 microns 10% microns: in steps down to 1 micron Pascal: in steps down to 1 Pa mbar: in steps down to 0.01 mbar C C kg (lbs) bar abs. 0 to 50 -10 to +60 0.22 (0.49) 10
Resolution
Ordering Information
TM 90, DN 16 KF TM 90, DN 25 KF
THERMOVAC Sensor TM 90
Part No. 230 076 Part No. 230 077
15 C16
Typical Applications
- Mobile pressure measurements of all kinds - Pressure measurements on refrigerating and air-conditioning systems before filling in the refrigerant - Pressure measurements during servicing of vacuum pumps - Pressure measurements during the production of gas filled pressurised lamps, respectively refilling
60
25
36
120
30
44
C16 16
Technical Data
Measurement principle Indicated units of measurement Measurement range Maximum overload mbar (Torr) bar abs.
Measurement uncertainty with reference to full-scale value (at 1200 - 10 mbar) % measured value (at 10 to 0.01 mbar) % Materials in contact with the vacuum Measurement cycle Settling time Operating temperature Storage temperature Supply voltage s ms C C
0.4 (full-scale) 10 (full-scale) Stainless steel, gold, tungsten, nickel, glass, Viton 1.6 400 +5 to +50 -20 to +60 Rechargeable 9 V battery (type 6LR61) or 12 V AC adaptor (miniature jack, + terminal at the tip)
Power consumption, approx. Operating duration Li battery 6LR61 alkaline Display Connection Dimensions (without flange) Protection class Weight (including battery)
mW
110 (pulse)
h h
DN mm IP kg (lbs)
Ordering Information
TM 100, DN 16 KF
17 C16
Typical Applications
- General pressure measurements in the fine and rough vacuum range, also for corrosive process gases - Chemical process engineering - Semiconductor production processes - Suited as a reference sensor for monitoring test instruments in accordance with DIN/ISO 9000
C16 18
Technical Data
Full Scale (FS) / Measurement range
Pressure Units: 1 Torr = 1.33 mbar = 133 Pascal Materials exposed to gases Max. overrange pressure
ceramic (Al2O3), stainless steel 316, Vacon 70 ceramic (Al2O3), stainless steel 316, Vacon 70 2000 Torr for 1/10/100 Torr sensors, 3000 Torr for 1000 Torr sensors 0.2% of reading temperature effect 0.003% of FS for 0.1/1 Torr sensors 2000 Torr for 1/10/100 Torr sensors, 3000 Torr for 1000 Torr sensors 0.2% of reading temperature effect 0.003% of FS for 10/100/1000 Torr sensors
Measurement uncertainty Resolution Temperature effects Zero coefficient Span coefficient Response time Nominal temperatur range Supply voltage Power consumption Signal Output Interface Weight approx. Dead Volume Connection cable kg (lbs) cm3
%/C %/C ms C V DC W V
0.015 of Full scale 0.01 of reading 30 +5 to +50 +14 to +30 1 0 - 10; linear 0.26 (0.57) RS 232 C 6 see section Connection Cable for active Sensors
0.005 of Full scale 0.01 of reading 30 +5 to +50 +14 to +30 1 0 - 10; linear 0.26 (0.57) RS 232 C 6 see section Connection Cable for active Sensors
Calibration
see section Miscellaneous, para. see section Miscellaneous, para. Oerlikon Leybold Vacuum Calibration Service Oerlikon Leybold Vacuum Calibration Service
1.1xF.S. 1.0xF.S. 0.9xF.S. 0.8xF.S. Pressure p 0.7xF.S. 0.6xF.S. 0.5xF.S. 0.4xF.S. 0.3xF.S. 0.2xF.S. 0.1xF.S. 0.0xF.S. 0 1 2 3 4 5 6 7 8 9 10 Measurement signal Uout [V]
Characteristic of the CERAVAC Transmitter CTR 100 Oerlikon Leybold Vacuum Full Line Catalog
19 C16
Technical Data
Pressure Units: 1 Torr = 1.33 mbar = 133 Pascal Materials exposed to gases Max. overrange pressure
ceramic (Al2O3), stainless steel 316, Vacon 70 ceramic (Al2O3), stainless steel 316, Vacon 70 1000 Torr for 0.1 Torr sensors 2000 Torr for 1/10/100 Torr sensors, 3000 Torr for 1000 Torr sensors 0.15% of reading temperature effect 0.0025% of FS for 1 Torr sensors 0.0015% of FS for 10/100/1000 Torr sensors
Temperature effects Zero coefficient %/C Span coefficient Reaction time Nominal temperatur range Supply voltage Power consumption Signal Output Weight approx. Dead Volume Connection cable %/C ms C V DC W V kg (lbs) cm3 0.005 of Full scale 0.01 of reading 30 +15 to +40 either 15 or +24 19 0 - 10; linear 0.6 (1.32) 6 see section Connection Cable for active Sensors
1/10/100/1000 Torr sensor 0.0025 of Full scale 0.01 of reading 30 +15 to +40 either 15 or +24 19 0 - 10; linear 0.6 (1.32) 7 see section Connection Cable for active Sensors
Calibration
see section Miscellaneous, para. see section Miscellaneous, para. Oerlikon Leybold Vacuum Calibration Service Oerlikon Leybold Vacuum Calibration Service
1.1xF.S. 1.0xF.S. 0.9xF.S. 0.8xF.S. Pressure p 0.7xF.S. 0.6xF.S. 0.5xF.S. 0.4xF.S. 0.3xF.S. 0.2xF.S. 0.1xF.S. 0.0xF.S. 0 1 2 3 4 5 6 7 8 9 10 Measurement signal Uout [V]
C16 20
35 36 37 37 39
45 46 47 48 49
55 56 57 58 59
The installation of a spiral tube is recommended in connection with applications involving contamination (oil vapors or dusts)
The installation of a spiral tube is recommended in connection with applications involving contamination (oil vapors or dusts)
A
3,8 89 117 113
View
CTR91
(26)
1/2" Tube DN 16 KF
44
DN 16 CF-R 8-VCR
21 C16
Typical Applications
- Analytical engineering - Safety circuits in vacuum systems - Controlling ionization gauges - General pressure measurement and control on systems in the fine and rough vacuum range pressure value (atmospheric, zero, reference pressure) by pressing a button. Integration of the transmitters in programmable control systems is facilitated by the linear characteristic, which can be defined by entering a simple equation into the computer. The transmitters TTR 91 / TTR 91 S / TTR 96 S are compatible to older TTR models like TTR 211, TTR 216 S or TTR 90. Built-in relays allow switching functions to be performed directly by the transmitter, without the need of a programmable control.
Sensor
The THERMOVAC transmitter has a Pirani sensing cell equipped with a tungsten (TTR 91/TTR 91 S) or nickel (TTR 96 S) filament. If required, the sensing cells can easily be aligned to any precisely known
50
65 22.2
70.4
10 9 8 7 6 5 4 3 2 1
1/8" NPT
33
15
DN 16 ISO-KF DN 16 CF-R
14 29
U(a) / Volt
115
0 2 10-4
4 68
10-3
10-2
10-1
100
101
Pressure
DN 16 CF-R
C16 22
Technical Data
Measurement range Measurement uncertainty Principle of measurement Supply voltage mbar (Torr)
Power consumption Storage temperature range Nominal temperature range Max. rel. humidity Protection class Weight, approx., for DN 16 KF Sensor Bake out temperature, max. Dead volume, approx. Materials in contact with the medium - filament Over-pressure rating, abs. Signal output (Ra 10 k) Measurement signal Status signal
kg (lbs)
C cm3
80 (250 with long tube) 2 (10 with long tube) Stainless steel, Ni, glass, NiFe Tungsten (TTR 91, TTR 91 S) or nickel (TTR 96 S)
bar
10 0 to 10.3 V DC 1.9 to 10 V DC, corresp. 5 x 10-4 to 1 x 103 mbar 1.286 V/decade, logarithmic Error: 0.5 V Normally open relay contact 2 x 10-3 to 500 (1.5 x 10-3 to 375) 10% 30 V, 0.5 A DC, floating Relay contact open Trigger (active): Green LED FCC-68/RJ45 socket, 8 way with shield
Trigger (only TTR 91 S / TTR 96 S) Adjustment range mbar (Torr) Hysteresis Rating Error status Status indicators (only TTR 91 S / TTR 96 S) Electrical connection Cable length, max. m
100
23 C16
Ordering Information
Without switching threshold TTR 91, DN 16 KF TTR 91, 1/8" NPT TTR 91, DN 16 CF TTR 91, 1/2" Tube, DN 16 CF bakeable With switching threshold TTR 91 S, DN 16 KF TTR 91 S, 1/8" NPT TTR 91 S, DN 16 CF TTR 91 S, 1/2" Tube, DN 16 CF bakeable TTR 96 S, DN 16 KF TTR 96 S, 1/8" NPT TTR 96 S, DN 16 CF TTR 96 S, 1/2" Tube, DN 16 CF bakeable Replacement sensing cell for TTR 91 / TTR 91 S DN 16 KF 1/8" NPT DN 16 CF 1/2" Tube, DN 16 CF bakeable for TTR 96 S DN 16 KF 1/8" NPT DN 16 CF 1/2" Tube, DN 16 CF bakeable Calibration Connection cable, FCC 68 on both ends, 8 way, shielded 005 m 010 m 015 m 020 m 030 m 040 m 050 m 075 m 100 m Accessories
Part No. 230 040 Part No. 230 043 Part No. 230 041 Part Part Part Part No. No. No. No. 230 230 230 230 042 045 048 046
Part No. 230 050 Part No. 230 053 Part No. 230 051 Part No. 230 052 Part No. 230 055 Part No. 230 058 Part No. 230 056 Part No. 230 057 see section Miscellaneous, para. Oerlikon Leybold Vacuum Calibration Service
Type A Part No. 124 26 Part No. 230 012 Part No. 124 27 Part No. 124 28 Part No. 124 29 Part No. 124 30 Part No. 124 31 Part No. 124 32 Part No. 124 33 The installation of a spiral tube is recommended in connection with applications involving contamination (oil vapors or dusts)
Note
- Measurement errors caused by the increased conductance of the component need to be taken into account - Low vibration mounting must be ensured - The sensor must be connected at the upper end
C16 24
Notes
25 C16
THERMOVAC Transmitter with Field Bus Interface Profibus or DeviceNet TTR 211 PB/D / TTR 216 PB/D
The THERMOVAC transmitters have been developed especially for integration into vacuum systems. Being active sensors (pressure to voltage converters) with a well-proven Pirani sensing cell and new operating and processing electronics these units offer a measurement range which spans 5 x 10-4 to 1000 mbar (3.75 x 10-4 to 750 Torr).
Typical Applications
- Analytical engineering - Safety circuits in vacuum systems - Controlling ionization gauges - Vacuum furnaces - General pressure measurement and control on systems in the fine and rough vacuum range which have the following requirements: - Immediate data transfer to a programmable control/computer via analog interface - Coverage of greater distances between the point of the measurement and processing location - Several locations which are to be monitored continuously - Low voltage supply - Simple, cost and space saving installation - Increased reliability - Simple operation - Increased requirements concerning electromagnetic compatibility (EMI)
C16 26
Sensor
The well-proven Pirani sensing cells with tungsten filament and a DN 16 KF flange are built into the TTR 211 PB/D THERMOVAC transmitters. The TTR 216 PB/D transmitter is equipped with a stainless steel sensing cell, with a platinum filament and an Al2O3 ceramics current feedthrough for use in connection with corrosive media or where much water vapor is present. The sensing cells can be easily exchanged on all transmitters. If required, the sensing cell can easily be aligned at atmospheric pressure and Zero pressure through two potentiometers. Integration of the transmitter in programmable control systems is facilitated by the straight characteristic which may be defined by entering a simple equation into the computer. Through the built-in relays it is possible to perform important switching functions directly through the transmitter without the need of a programmable control.
10 9 8 7 6 5 4 3 2 1 0 2 10-4 U(a)/Volt
56
73
0 Trigger
7 13 DN 16 KF 1/8" NPT
16
4 68
10
-3
10-2
101
Dimensional drawing for the TTR 211 PB/D and TTR 216 PB/D
27 C16
Technical Data
Display range Measurement uncertainty mbar (Torr)
Power consumption Storage / nominal temperature range Max. rel. humidity (climatic class F) Protection class Dimensions (H x W x D) Weight, approx. Inflammability Sensor Filament Vacuum connection Bake out temperature, max Dead volume, max. Materials in contact with the medium
Over-pressure rating, abs. Signal output (Ra 10 k) Measurement signal Status signal Trigger Adjustment range Hysteresis Reaction time Rating Error status Status indicators Monitor output (Ra 100 k) Electrical connection Cable length, max. Interface TTR 211 D / 216 D TTR 211 PB / 216 PB
mbar (Torr) ms
DeviceNet Profibus DP
C16 28
Ordering Information
TTR 211 D, DN 16 KF Tungsten filament (1 trigger) DeviceNet interface Replacement sensing cell TTR 211 PB, DN 16 KF Tungsten filament (1 trigger) Profibus DP interface Replacement sensing cell TTR 216 D, DN 16 KF Platinum filament (1 trigger) DeviceNet interface Replacement sensing cell TTR 216 PB, DN 16 KF Platinum filament (1 trigger) Profibus PB interface Replacement sensing cell Calibration Connecting cable, FCC 68 on both ends, 8 way with shield 005 m 010 m 015 m 020 m 030 m 040 m 050 m 075 m 100 m Accessories
Part No. 896 52 Part No. E 157 77 see section Miscellaneous, para. LEYBOLD Calibration Service
Type A Part No. 124 26 Part No. 230 012 Part No. 124 27 Part No. 124 28 Part No. 124 29 Part No. 124 30 Part No. 124 31 Part No. 124 32 Part No. 124 33 The installation of a spiral tube is recommended in connection with applications involving contamination (oil vapors or dusts)
Note
- Measurement errors caused by the increased conductance of the component need to be taken into account - Low vibration mounting must be ensured - The sensor must be connected at the upper end
29 C16
Typical Applications
- Loadlock control - Forevacuum pressure monitoring - Safety circuits in vacuum systems - General measurement and control in the medium and rough vacuum range - Control of high vacuum ionization gauges
Option
Dust and other particles may cause increasing measurement errors and reduced lifetime. Therefore we recommend the installation of a fine filter in critical applications. Fine filters are listed in section General, para. Connection Accessories for Small Flanges.
~65 28
10 9 8 7 6 5 4 3 2 1 0 2 10-4
4 68
68.3
U(a) / Volt
45
26.3 10
-3
10
-2
10
-1
10
10
10 mbar 10
DN 16 ISO-KF
45
1/8" NPT
Pressure
Characteristic of the THERMOVAC Transmitters
C16 30
Technical Data
Measurement principle Measurement range (air, O2, CO, N2) Accuracy 1 x 10-3 to 50 mbar 50 to 950 mbar ATM (atmospheric pressure) Repeatability Trigger (only TTR 100 S2) Setting range with potentiometer Relay contacts closed open Hysteresis Contact rating Relay status Output signal analog Measurement range Voltage vs. pressure Output impedance Minimum load impedance Response time Power supply Voltage (ripple 1 Vpp) Consumption, max. Fuse to be connected Electrical connection Cable length, max. Materials exposed to vacuum (process media) Vacuum connection Pirani filament Capacitance sensor cell Feedthrough Other materials Internal volume DN 16 ISO-KF 1/8" NPT Over-pressure rating, abs. Temperature Operation (ambient) Storage Bakeout at flange, max. Filament temperature Relative humidity Mounting orientation any Use Protection class Weight DN 16 ISO-KF 1/8" NPT IP kg (lbs) kg (lbs) cm3 cm3 bar C C C C m mbar (Torr)
stainless steel tungsten AI2O3 glass Ni, Cu, NiFe, SnAg, AgPd 6 8 5 +10 to +50 -20 to +65 +80 < 160 < 80% at temperatures < +31 C, decreasing to 50% at +40 C any Indoors only, altitudes up to 2000 m NN 40 0.09 (0.20) 0.09 (0.20)
31 C16
Ordering Information
Without switching threshold TTR 100, DN 16 KF TTR 100, 1/8" NPT With switching threshold TTR 100 S2, DN 16 KF TTR 100 S2, 1/8" NPT Calibration Connection cable, FCC 68 on both ends, 8 way, shielded 005 m 010 m 015 m 020 m 030 m 040 m 050 m 075 m 100 m Accessories
Part No. 230 027 Part No. 230 029 see section Miscellaneous,para. Oerlikon Leybold Vacuum Calibration Service
Type A Part No. 124 26 Part No. 230 012 Part No. 124 27 Part No. 124 28 Part No. 124 29 Part No. 124 30 Part No. 124 31 Part No. 124 32 Part No. 124 33 The installation of a spiral tube is recommended in connection with applications involving contamination (oil vapors or dusts)
Note
- Measurement errors caused by the increased conductance of the component need to be taken into account - Low vibration mounting must be ensured - The sensor must be connected at the upper end
C16 32
Notes
33 C16
Typical Applications
- Evaporation and sputtering systems - Analytical engineering - Vacuum furnaces - High vacuum systems - General pressure measurement and control on systems in the fine and rough vacuum range which have the following requirements: - Immediate data transfer to a programmable control/computer via analog interface - Coverage of greater distances between the point of the measurement and processing location - Several locations which are to be monitored continuously - Low voltage supply - Simple, cost and space saving installation - Increased reliability, also in argon processes (sputtering) - Simple operation - Increased requirements concerning electromagnetic compatibility (EMI)
Option
For protection of the sensors PTR 225 against contamination, radiation and other disturbing factors the installation of a baffle is recommmended.
C16 34
Sensor
Cold cathode sensors - based on the well-proven principle of the inverted Penning - are built into the PENNINGVAC transmitters PTR 225/ 225 S/237 which have a DN 25 KF or DN 40 CF flange. The housing of the transmitter, including its electronics, as well as the magnet can easily be removed for degassing of the all-metal sensor with Al2O3 current feedthrough. The design of the Penning sensors with its closed magnetic field causes a negligible stray field. Thus the PTR 225/225 S/237 may also be installed close to sensitive parts within a system. The anode ring and the titanium cathode plates can be exchanged easily for quick maintenance of the sensors in case of contamination. The shape of the cathode plates is such that they also act as a baffle for the sensors.
73 80
POWER POWER READY
Integration of the transmitter in programmable control systems is facilitated by the linear characteristic which can be defined by entering a simple equation into the computer. Built-in relays (PTR 225 S) allow switching functions to be performed directly by the transmitter, without the need of a programmable control.
73
80
READY
114 130
114 144
17
30
10 9 8 7 6 5 4 3 2 1 0 2 10-9 U(a)/Volt
4 68
10
-8
10-7
10-4
35 C16
Technical Data
Display range Measurement uncertainty Principle of measurement Supply voltage Power consumption Storage temperature range Nominal temperature range Max. rel. humidity (climatic class F) Protection class Dimensions (H x W x D) Weight, approx. Inflammability Sensor Vacuum connection Degassing temperature, max. Dead volume, max. Materials in contact with the medium Over-pressure rating (abs.) Signal output (Ra 10 k) Measurement signal bar DN C cm3 VA C C % n.c. IP mm kg (lbs) mbar (Torr)
Trigger (PTR 225 S) Adjustment range Hysteresis Rating Error status High voltage control input Status output Ready to measure Error (no ignition) Status indicators
Monitor output (Ra 100 k) Electrical connection Cable length, max. Interface PTR 225 PB PTR 237 D
C16 36
Ordering Information
PTR 225, DN 25 KF PTR 225 S, DN 25 KF PTR 225 PB, DN 25 KF Profibus interface PTR 237, DN 40 CF PTR 237 D, DN 40 CF DeviceNet interface Baffle, DN 25 KF, with centering ring Replacement cathode plates, titanium (set of 5 pieces) Replacement anode ring Calibration Connection cable, FCC 68 on both ends, 8 way, shielded 005 m 010 m 015 m 020 m 030 m 040 m 050 m 075 m 100 m
Part No. EK 162 91 Part No. 200 28 711 see section Miscellaneous, para. Oerlikon Leybold Vacuum Calibration Service
Type A Part No. 124 26 Part No. 230 012 Part No. 124 27 Part No. 124 28 Part No. 124 29 Part No. 124 30 Part No. 124 31 Part No. 124 32 Part No. 124 33
37 C16
75 60 60
75
55 116 63,5
28
18 DN 25 ISO-KF
18,5
24
DN 40 ISO-KF DN 40 CF-F
Typical Applications
- Sputtering and coating technology - Analytical technology - Vacuum furnaces - Multipurpose pressure measurement and control in the medium and high vacuum range
C16 38
Option
For protection of the sensor ITR 90 against contamination, radiation and other disturbing factors the installation of a baffle is recommmended. Two types of baffles are available: A build-in version for CF connections is mounted in the sensor; the baffle for KF connections is integrated in a centering ring.
39 C16
Technical Data
Display range Measrement uncertainty mbar (Torr) %
Principle of measurement
Reproducibility
Output signal (measurement signal) Voltage range Measurement range Relationship voltage-pressure Error signal
0 - 10.5 1.828.6 Logarithmic, 0.6 V/decade < 0.5 V no power supply > 9.5 V Pirani sensor is defective (broken filament) V DC 15 - 30 FCC 68 socket, 8 way C C +5 to +55 -40 to +65 Stainless steel, ceramics, Mo, Ni, Au, W bar IP 10 40
Power supply Electrical connection Operating temperature Storage temperature Materials in contact with the medium Overpressure resistance (absolute) Protection class
Ordering Information
PTR 90 DN 25 KF DN 40 KF DN 40 CF Baffle DN 25 KF DN 40 KF Calibration Connection cable, FCC 68 on both ends, 8 way, shielded 005 m 010 m 015 m 020 m 030 m 040 m 050 m 075 m 100 m
PENNINGVAC Transmitter
Part No. 230 070 Part No. 230 071 Part No. 230 072
Part No. 230 078 Part No. 230 079 see section Miscellaneous, para. Oerlikon Leybold Vacuum Calibration Service
Type A Part No. 124 26 Part No. 230 012 Part No. 124 27 Part No. 124 28 Part No. 124 29 Part No. 124 30 Part No. 124 31 Part No. 124 32 Part No. 124 33
C16 40
Notes
41 C16
Typical Applications
- Analytical - Evaporation and coating - Vacuum furnaces - General purpose pressure measurements in the fine and high vacuum ranges
C16 42
Sensor
The sensor of the ITR 90 contains a dual filament Pirani system as well as a Bayard-Alpert measurement system. When using the bake out extension, measurements will be possible also at flange temperatures up to 150 C.
10 Volt 8 Equation 7 p = 10(u-7,75)/ 0,75 6 5 4 3 2 1 0 2 468 10-10 10-9 10-8 10-7 10-6 10-5 10-4 10-3 10-2 10-1 U(a) Pressure
100
10 mbar 1000
58 (2.28 )
67 (2.64 )
Top View
153 (6 ) ~105
48 (1.89 )
42 (1.65 )
DN 40 CF-R
DN 25 ISO-KF
40
Dimensional drawing for the ITR 90; dimensions in brackets ( ) are in inch
43 C16
Technical Data
Display range mbar (Torr)
Measurement uncertainty, 10-8 - 10-2 mbar Reproducibility, 10-8 - 10-2 mbar Principles of measurement
15% of the meas. value 5% of the meas. value Hot cathode ionization according to Bayard-Alpert combined with thermal conductivity according to Pirani Electron bombardment 3 minutes, max. 20 to 28 V DC, typ. 24 V DC
Degas Supply voltage Power consumption, max. Storage / nominal temperature range Protection class Weight, approx. ITR 90, DN 25 KF ITR 90, DN 40 CF Sensor Bake out temperature, max. Dead volume, max. C cm3 W C
kg (lbs) kg (lbs)
0.285 (0.64) 0.550 (1.24) Fully sealed, exchangeable 150 1) 24 at DN 25 KF 34 at DN 40 CF Cu, W, Glas, NiFe, Mo, stainless steel, Aluminum, Iridium, Yttrium, NiCr,
Materials in contact with the medium Over-pressure rating (abs.) Signal output (Ra 10 k) Measurement signal Error signal Interface (standard / optional) Electrical connection Cable length, max. m bar
0 - 10 V, 0.774 - 10 V, 0.75 V pro decade < 0,5 V RS 232 C / ProfiBus 15 way Sub-D male connector / pin contacts 100 / 30 at RS 232 C
1)
C16 44
Ordering Information
ITR 90, DN 25 KF ITR 90, DN 25 KF Profibus interface ITR 90, DN 40 CF-R, rotatable CF flange ITR 90, DN 40 CF-R, rotatable CF flange Profibus interface Options Power supply for IONIVAC transmitter 100 V - 240 V AC / 24 V DC incl. 5 m connection cable and 5 m RS 232 C cable Bake out extension (100 mm, approx.) Baffle, DN 25 KF, with Installation baffle for CF/KF variant Replacement sensor IE 90, DN 25 KF 1) IE 90, DN 40 CF-R 1) Calibration
Connection cable
1)
45 C16
Typical Applications
- Analytical engineering - Sputtering and coating technology - Vacuum furnaces - Multipurpose pressure measurement in the medium and high vacuum range
Sensor
The sensor of the ITR 200 S includes besides the Pirani system a dual cathode measurement system according to Bayard-Alpert. If one of the hot cathodes should burn out during operation, then the second cathode is automatically switched on. Moreover, each sensor contains a memory chip with the relevant system data. Thus after having exchanged a sensor, an automatic alignment is performed between sensor and operating electronics (plug and play).
Options
Upon request, the ITR 200 S can be supplied with an integrated display, a Profibus or a DeviceNet interface.
C16 46
58 (2.28 )
67 (2.64 )
Top View
153 (6 )
107.8
48 (1.89 )
42 (1.65 )
DN 40 CF-R
DN 25 ISO-KF
Bakeout area
Dimensional drawing for the ITR 200; dimensions in brackets ( ) are in inch
10 Volt 8 Equation 7 p = 10(u-7,75)/ 0,75 6 5 4 3 2 1 0 2 468 10-10 10-9 10-8 10-7 10-6 10-5 10-4 10-3 10-2 10-1 U(a) Pressure
Characteristic of the ITR 200
100
10 mbar 1000
47 C16
Technical Data
Display range mbar (Torr)
Measurement uncertainty, 10-8 - 10-2 mbar Reproducibility, 10-8 - 10-2 mbar Principles of measurement
Degas Supply voltage Power consumption, max. Storage / nominal temperature range Protection class Weight, approx. ITR 90, DN 25 KF ITR 90, DN 40 CF Sensor Bake out temperature, max. Dead volume, max. C cm3
kg (lbs) kg (lbs)
0.285 (0.64) 0.550 (1.24) Fully sealed, exchangeable 150 1) 24 at DN 25 KF 34 at DN 40 CF Cu, W, Glas, NiFe, Mo, stainless steel, Aluminum, Iridium, Yttrium, NiCr,
Materials in contact with the medium Over-pressure rating (abs.) Signal output (Ra 10 k) Measurement signal Error signal Interface (standard / optional) Switching function Standard Profibus or DeviceNet Electrical connection Cable length, max. m bar
1 normally open contact 2 normally open contacts 15 way Sub-D male connector / pin contacts 100 / 30 at RS 232 C
1)
C16 48
Ordering Information
ITR 200 S, DN 25 KF 1 switching function ITR 200 SP, DN 25 KF Profibus interface, 2 switching functions ITR 200 SD, DN 25 KF DeviceNet interface, 2 switching functions ITR 200 S, DN 40 CF-R, rotatable CF flange 1 switching function ITR 200 SP, DN 40 CF-R, rotatable CF flange Profibus interface, 2 switching functions ITR 200 SD, DN 40 CF DeviceNet interface, 2 switching functions DN 40 CF-R, ITR 200 SL rotatable CF flange, 1 switching function, 150 C bakeable Options Power supply for IONIVAC transmitter 100 V - 240 V AC / 24 V DC incl. 5 m connection cable and 5 m RS 232 C cable Baffle, DN 25 KF, with Installation baffle for CF/KF variant Replacement sensor IE 200, DN 25 KF 2) IE 200, DN 40 CF-R 2) IE 200 SL 1), DN 40 CF-R 2) Calibration
1),
Part No. 240 020 Part No. 240 021 Part No. 240 022
see section Miscellaneous, para. Oerlikon Leybold Vacuum Calibration Service see section Connection Cable for active Sensors
Connection cable
1) 2)
SL = long version (bake out extension) including hex. socket screw key
49 C16
Spiral Tube
Installation position of the sensor DN
65
51
Dimensional drawing for the spiral tube
Operating Principle
Through the specially developed geometry which provides a constant slope, possibly occurring contamination is removed.
Note
- Measurement errors caused by the increased conductance of the component need to be taken into account - Low vibration mounting must be ensured - The sensor must be connected at the upper end
Maintenance
Depending on the type of application regular maintenance on the spiral tube is recommended.
Technical Data
Materials
Spiral Tube
Stainless steel
Ordering Information
Spiral Tube DN 16 ISO-KF DN 25 ISO-KF DN 40 ISO-KF
Spiral Tube
Part No. 230 082 Part No. 230 083 Part No. 230 084
C16 50
Notes
51 C16
THERMOVAC Transmitter TTR 91, TTR 91 S, TTR 96 S a. o. TTR 211 PB/D, TTR 216 PB/D PENNINGVAC Transmitter PTR 90, PTR 225, PTR 225 S, PTR 237 CERAVAC Transmitter CTR 100, CTR 101 (digital signal) IONIVAC Transmitter ITR 90, ITR 200 S
Type A Type A
(only PTR 90)
Type A Type A
CMove
Type A
Type B
Type B Type C
Type E
C16 52
Technical Data
Cables Type Type Type Type
Connection Cable
A B C E
FCC 68 (RJ45) on both ends, 8 way, shielded Sub-D 15 way female to FCC 68 (RJ45), 8 way, shielded Sub-D 15 way female to Sub-D 15 way male, shielded Sub-D 15 way female to bare wire ends, shielded
Connection Cable
Ordering Information
Cable length 005 m 010 m 015 m 020 m 030 m 040 m
1)
Type C
Part No. 124 55 Part No. 230 022 Part No. 124 56 Part No. 124 57 Part No. 124 58
1)
Type E
Part Part Part Part Part Part No. No. No. No. No. No. 124 163 124 124 124 124 63 69 64 65 66 67
Longer cable runs are not specified because of the RS 232 C connection
53 C16
Connectable Sensors
THERMOVAC - TTR 100 - TTR 211 - TTR 216 S - TTR 90 - TTR 90 S - TTR 91 - TTR 91 S - TTR 96 S PENNINGVAC - PTR 90
66 78
85
91,5 105,5
106,5
C16 54
Technical Data
Number of measurement channels Display for measured values Display range Unit of measurement (selectable) Switching thresholds Chart recorder output (Ra 2.5 k) Main connection EURO version US version V / Hz V / Hz mbar (Torr)
DISPLAY ONE
1 digital, 7 segment LED 5 x 10-9 to 2000 (3.8 x 10-9 to 1500) mbar, Torr, Pascal from the transmitter are run to a terminal strip 0 - 10 Volt, characteristic corresponds to the connected transmitter
Ordering Information
EURO version, including mains cord US version, including mains cord THERMOVAC transmitter PENNINGVAC transmitter PTR 90 Connection cable, FCC 68 on both ends, 8 way, shielded 005 m 010 m 015 m 020 m 030 m 040 m 050 m 075 m 100 m Adapter panel for installation in a 3 HU, 19 in. rack
DISPLAY ONE
Part No. 230 001 Part No. 235 001
Type A Part No. 124 26 Part No. 230 012 Part No. 124 27 Part No. 124 28 Part No. 124 29 Part No. 124 30 Part No. 124 31 Part No. 124 32 Part No. 124 33
55 C16
Connectable Sensors
THERMOVAC - TTR 211 - TTR 216 S - TTR 90 - TTR 91 - TTR 96 S - TTR 100 - TTR 100 S2 PENNINGVAC
M3 - 3,5 mm
91,5 106,0
3,5 mm
91,5 106,5
Dimensional drawing and panel cut-out for the DISPLAY TWO and THREE
C16 56
Technical Data
Number of measurement channels Display for measured values Display range Unit of measurement (selectable) Gas type correction (for PTR) Sensor connection Sensor power supply Electrical outputs Switching threshold Number Adjustment range Hysteresis Relay contact Load rating Ready relay Relay contact Load rating V DC mbar (Torr) 1x
DISPLAY TWO
2 digital, 7 segment LED, 4 digits 10-10 to 2000 (0.75 x 10-10 - 1500)
DISPLAY THREE
3 digital, 7 segment LED, 4 digits 1 x 10-10 to 2000 (0.75 x 10-10 - 1500) mbar, Torr, Pascal factor adjustable FCC68 (RJ45) 24 screw terminal
normally open contact 60 V, 1 A DC / 30 V, 1 A AC 0 - 10 V per measurement channel, output characteristic corresponds to the connected transmitter PENNINGVAC PTR: high voltage on
normally open contact 60 V, 1 A DC / 30 V, 1 A AC 0 - 10 V per measurement channel, output characteristic corresponds to the connected transmitter PENNINGVAC PTR: high voltage on 85 - 240 / 50 - 60 < 15 +5 to +50 1.4 (3.09) 40
Control input Mains connection Power consumption Nominal temperature range Weight Protection class V AC / Hz W C kg (lbs) IP
Ordering Information
EURO version, including mains cord US version, including mains cord THERMOVAC transmitter PENNINGVAC transmitter Connection cables for THERMOVAC and PENNINGVAC (Type A)
DISPLAY TWO
Part No. 230 024 Part No. 235 024 see section Active Sensors see section Active Sensors
DISPLAY THREE
Part No. 230 025 Part No. 235 025 see section Active Sensors see section Active Sensors
57 C16
CENTER ONE
Universal compact display and operating unit for the active sensors from the CERAVAC, THERMOVAC, PENNINGVAC and IONIVAC series.
Connectable Sensors
THERMOVAC - TTR 211 - TTR 216 S - TTR 90 - TTR 91 - TTR 96 S - TTR 100 - TTR 100 S2 PENNINGVAC - PTR 90 - PTR 225 - PTR 225 S
M3 - 3,2 mm
- CTR 90
91,2 105 106,5 204,5 285 287,5
C16 58
Technical Data
Number of measurement channels Display for measured values Display range Unit of measurement (selectable) Gas type correction Sensor connection Sensor power supply Electrical inputs and outputs Switching threshold Number Adjustment range Hysteresis Relay contact Load rating Error message Relay contact Load rating V DC mbar (Torr)
CENTER ONE
1 digital, 7 segment LED, 5 digits 1 x 10-10 to 2000 (0.75 x 10-10 to 1500) mbar, Torr, Pascal, Micron factor adjustable 15 way Sub-D socket FCC68 (RJ45) 24 9 way Sub-D plug
floating normally open contact 60 V, 0.5 A DC / 30 V, 0.5 A AC 0 - 10 Volt, characteristic corresponds to the connected transmitter PTR: high voltage on ITR 100: emission on 9 way Sub-D socket
Interface RS 232 C Mains connection Power consumption Weight Protection class V AC / Hz W kg (lbs) IP
Ordering Information
EURO version with 2 m EURO mains cord US version with 2 m US mains cord THERMOVAC, PENNINGVAC, CERAVAC and IONIVAC transmitters Connection cables for THERMOVAC and PENNINGVAC (Type A), CERAVAC (Type B analog, Type C digital) and IONIVAC (Type C) Adapter panel for installation in a 3 HU, 19 in. rack Screw terminal for the 9 way output socket
CENTER ONE
Part No. 230 002 Part No. 235 002
59 C16
Connectable Sensors
THERMOVAC - TTR 211 - TTR 216 S - TTR 90 - TTR 91 - TTR 96 S - TTR 100 - TTR 100 S2 PENNINGVAC - PTR 90 - PTR 225 - PTR 225 S - PTR 237 CERAVAC - CTR 90 - CTR 91
M3 - 3,2 mm
91,5 104,5
106,5
- ITR 200 SL
Dimensional drawing and panel cut-out for the CENTER TWO and THREE
C16 60
Technical Data
Number of measurement channels Display for measured values Display range Unit of measurement (selectable) Gas type correction Sensor connection Sensor power supply Electrical outputs Switching threshold Number Adjustment range Hysteresis Relay contact Load rating Error message Relay contact Load rating V DC mbar (Torr) 1x
CENTER TWO
2 digital, 7 segment LED, 5 digits 10-10 to 2000 (0.75 x 10-10 to 1500)
CENTER THREE
3 digital, 7 segment LED, 5 digits 1 x 10-10 to 2000 (0.75 x 10-10 to 1500) mbar, Torr, Pascal, Micron factor adjustable 15 way Sub-D socket FCC68 (RJ45) 24 25 way Sub-D socket independently assignable 6 sensor dependent adjustable floating changeover contact 60 V, 0.5 A DC / 30 V, 0.5 A AC
mbar, Torr, Pascal, Micron factor adjustable 15 way Sub-D socket FCC68 (RJ45) 24 25 way Sub-D socket independently assignable 4 sensor dependent adjustable floating changeover contact 60 V, 0.5 A DC / 30 V, 0.5 A AC
floating normally open contact 60 V, 0.5 A DC / 30 V, 0.5 A AC 0 - 10 V per measurement channel, output characteristic corresponds to the connected sensor, in addition one chart recorder output can be programmed PENNINGVAC PTR: high voltage on 9 way Sub-D socket
floating normally open contact 60 V, 0.5 A DC / 30 V, 0.5 A AC 0 - 10 V per measurement channel, output characteristic corresponds to the connected sensor, in addition one chart recorder output can be programmed PENNINGVAC PTR: high voltage on 9 way Sub-D socket 90 - 250 / 50 - 60 65 +5 to +50 1.2 (2.65) 20
Control input Interface RS 232 C Mains connection Power consumption Nominal temperature range Weight Protection class V AC / Hz W C kg (lbs) IP
Ordering Information
EURO version with 2 m EURO mains cord US version with 2 m US mains cord THERMOVAC, PENNINGVAC, CERAVAC and IONIVAC transmitters Connection cables for THERMOVAC and PENNINGVAC (Type A), CERAVAC (Type B) and IONIVAC (Type C) Screwed connection for 9 way Sub-D socket
CENTER TWO
Part No. 230 004 Part No. 235 004
CENTER THREE
Part No. 230 003 Part No. 235 003
61 C16
PIEZOVAC PV 20
This complete vacuum gauge offers accurate and cost-effective measurements in the range between 1 and 2000 mbar (0.75 and 1500 Torr), or 0.1 and 200 mbar (0.075 and 150 Torr).
Typical Applications
- General pressure measurements in the fine and rough vacuum range, in connection with vacuum furnaces and analytical instruments - Monitoring the operation of backing pumps and vacuum systems - Replacement of mercury manometers in the lab and in production lines
Sensors
The PIEZOVAC PV 20 is delivered complete with a DI 2000 or DI 200 sensor. The sensor is supplied fully aligned for operation. For information on the THERMOVAC sensors, see chapter Activ Sensors.
C16 62
Technical Data
Display range Display mbar (Torr)
PIEZOVAC PV 20
0.1 to 2000 (0.075 to 1500) digital: 7 segment LCD, analog: LCD bar
Measurement uncertainty with DI 200/201/2000 Sensor Unit of measurement (selectable) Chart recorder output (Ra 2,5 k)
% FS
0.2 mbar, Torr, Pa, Micron 0 to 10 V, linear divisions for absolute pressure sensors, about 10.5 V during faults 0 to 10 V, logarithmic divisions for THERMOVAC sensors (1.67 V/decade)
Main connection 50/60 Hz (selectable) Power consumption Storage temperature range Nominal temperature range Max. rel. humidity Weight, approx. Dimensions (W x H x D) Cable length, max. Connectable sensors
V VA C C
90 to 130 / 180 to 250 9 -40 to +60 0 to 40 80 2 (4.4) 106.5 x 128.5 x 172 100 TR 211/212/216, DI 200/201/2000/2001
% n.c. kg (lbs) mm m
Ordering Information
Complete with 2 m long main cord and DI 2000 sensor, with 5 m long cable, 230 V AC, Measurement range 1 - 2000 mbar (0.75 - 1500 Torr) Complete with 2 m long main cord and DI 200 sensor, with 5 m long cable, 230 V AC, Measurement range 0.1 - 200 mbar (0.075 - 150 Torr) Calibration
PIEZOVAC PV 20
63 C16
Connectable Sensors
- Compact benchtop enclosure (1/2 19 in., 3 HU) - RS 232 C interface provided as standard - Simple software updates possible through the RS 232 interface - Profibus interface (optional) - CE mark - Bayard-Alpert sensor IE 414 - Extractor sensor IE 514 (see Chapter Additional Sensors) combined with - THERMOVAC TTR 211, TTR 216 S, TTR 90, TTR 91 and TTR 96 S - CERAVAC CTR 90, CTR 91 and CTR 100 (see Chapter Active Sensors) Two passive sensors working with ionization technology (IE 414 and/or IE 514) could be connected simuntaneously to the IONIVAC IM 540 while only one is in operation. A pressure dependent emission control of these sensors is possible if a THERMOVAC TTR or CERAVAC CTR 100/CTR 91 of suitable range overlap is connected.
Typical Applications
- Pressure measurement and control in the UHV range - Measurement of ultimate pressure in UHV systems - Checking of ultimate pressure in semiconductor production - Total pressure measurements in the area of cryo technology - Total pressure measurements in calibration systems
Front panel cut out (left) and dimensional drawing (right) for the IONIVAC IM 540
C16 64
Technical Data
Number of measurement channels Bayard-Alpert / Extractor THERMOVAC / CERAVAC Display range Display range Extractor Display range Bayard-Alpert Measurement range switching Units of measurement (selectable) Measurement uncertainty Trend indication Measurement value display rate % mbar (Torr) mbar (Torr) mbar (Torr)
IONIVAC IM 540
3 Channel 1 or 2 Channel 3 and 4 1 x 10-12 to 1100 (0.75 x 10-12 to 825) 1 x 10-12 to 1 x 10-4 (0.75 x 10-12 to 0.75 x 10-4) 1 x 10-11 to 1 x 10-2 (0.75 x 10-11 to 0.75 x 10-2) automatic or decade pre-select mbar, Torr, Pa, Micron 10 of the value displayed bargraph to 1 x 10-2 mbar, 5 s-1 1x -12 to 1 x 10-10 mbar, 0.5 s-1 1 x 10 10-10
mA mA
1.6 0.1 to 10; automatic control p > 1 x 10-2 mbar, broken cathode, short-circuit, interruption of the electric circuit
20 / 40 anode Extractor / Bayard-Alpert: 220 V, cathode Extractor / Bayard-Alpert: 100 V/80 V, Reflector Extractor: 205 V Bayard-Alpert and Extractor - single operation is possible 2 x Bayard-Alpert or Extractor (redundant operation) automatically automatically, pressure dependent, error dependent logarithmic 0 to 10 V (1 V / dec.) or linear 0 to 10 Volt, error indication U > 10.5 V RS 232 C / Profibus
Sensor connections
Measurement system detection Measurement system switchover Chart recorder outputs for Extractor/Bayard-Alpert (Ra 2.5 k) Interface (standard / optional) Switching thresholds (single operation or interval) Mains connection Storage temperature range Nominal temperature range V / Hz C C
65 C16
Ordering Information
IONIVAC IM 540 with 2 m long mains cord (EURO and US) Options Profibus DP interface DKD calibration IONIVAC sensors IE 414, DN 40 CF Replacement cathode IE 414 IE 514, DN 40 CF Replacement cathode IE 514 Gauge head cables for IE 414/514 05 m 10 m 05 m, bakeable up to 200 C Extension cables for IE 414/514 10 m 20 m THERMOVAC Transmitter TTR Connection cable for THERMOVAC (Type A) CERAVAC Transmitter CTR Connection cable for CERAVAC (Type B)
IONIVAC IM 540
66 63 67 61
Part No. 245 002 Part No. 200 02 937 see section Active Sensors see section Connection Cables for Active Sensors see section Active Sensors see section Connection cables for active sensors
C16 66
Notes
67 C16
COMBIVAC CM 51
The COMBIVAC CM 51 covers the complete pressure range between 10-9 and 1000 mbar by combining two measurement principles THERMOVAC and PENNINGVAC providing both monitoring and control functions.
M3 - 3,5 mm
105
91,5 106,0
3,5 mm
91,5 106,5
162 164,5
105
Front panel cut out (left) and dimensional drawing (right) for the CM 51
C16 68
Typical Applications
- Universal monitoring the operation of high vacuum pump systems like: Turbomolecular pump systems Diffusion pump systems Cryogenic pump systems - Annealing, melting, brazing and hardening furnaces - Coating systems - Analytical instrumentation - Deployment in thermal radiation resistant and degassable systems is possible - Particle accelerators
Connectable Sensors
THERMOVAC - TR 211 - TR 211 NPT - TR 212 - TR 216 PENNINGVAC - PR 25 - PR 26 - PR 27 - PR 28
69 C16
Technical Data
Number of measurement channels Display range Channels 1, 2 (THERMOVAC) Channel 3 (PENNINGVAC) Unit of measurement (selectable) Measurement uncertainty THERMOVAC PENNINGVAC Measured value display Type of gas (selectable) Switching thresholds Operating mode Adjustable switching thresholds THERMOVAC PENNINGVAC Switching relay hysteresis Relay contact load rating Chart recorder output (default) THERMOVAC PENNINGVAC Interface Mains connection 50/60 Hz Power consumption Storage temperature range Nominal temperature range Max. rel. atmospheric humidity Weight Dimensions (W x H x D) Installation depth Protection class V W C C % n.c. kg mm mm IP mbar mbar
COMBIVAC CM 51
3
mbar mbar
20% of the measured value in the range 10-3 to 10-2 mbar / 15% of the measured value in the range 10-2 to 102 mbar 30% of the measured value in the range 10-8 to 10-4 mbar digital, 7 segment LCD, 4 digits factor adjustable 2 per channel single, interval-trigger 5 x 10-3 to 500 1 x 10-8 to 9.9 x 10-3 10% of the trigger value (default), freely adjustable for THERMOVAC and PENNINGVAC AC/DC, max. 30 V / 1 A 0 to 10 V, log. divisions linear: 3 decades, approximately 10.5 V in case of a failure, logarithmic: (1 x 10-3 mbar), 1.67 V/decade logarithmic: (1 x 10-9 mbar), 1.43 V/ decade RS 232 C, RS 485 and Profibus 100 - 240 < 10 -20 to +60 +5 to +50 80 1.4 106.4 x 128.5 x 164.5 approx. 220 40
C16 70
Ordering Information
Operating unit COMBIVAC CM 51 including EURO and US mains cord, 2 m with RS 232 C/ 485 with Profibus DB Cable adapter CM 31 - CM 51 DKD calibration Options 19 in. installation frame 1/4 19 in. blank panel THERMOVAC sensors for CM 51 TR 211, DN 16 KF TR 211, 1/8" NPT TR 212, DN 16 KF TR 212, DN 16 CF TR 216, DN 16 KF Gauge head cables for TR sensors 005 m 010 m 015 m 020 m 030 m 040 m 050 m 075 m 100 m PENNINGVAC sensors for CM 51 PR 25, DN 25 KF PR 26, DN 40 KF PR 27, DN 40 CF PR 37, DN 40 CF, degassable Gauge cables for PR sensors 005 m 010 m 015 m 020 m 030 m 040 m 050 m 075 m 100 m
COMBIVAC CM 51
Part No. 230 110 Part No. 230 111 upon request see chapter Oerlikon Leybold Calibration Service
85 33 52 86 87
26 27 34 28 35 36 37 38 39
52 46 47 48
88 89 49 56 50 51 52 53 54
71 C16
IE 414
- Bayard-Alpert sensing system - Measurement range to 2 x 10-11 mbar (1.5 x 10-11 Torr) - Protection shied welded in place
IE 514
- Extractor sensing system - Reliable to 1 x 10-12 mbar (0.75 x 10-12 Torr) - Significant reduction of X-ray and ion desorption effects
33 69,4
33 69,4
DN 40 CF 25 12,8
55 DN 40 CF 12,8
66 58 90 98
C16 72
Technical Data
Measurement range X-ray limit Operating temperature range Degassing temperature at the flange, max. Material Cathode Feedthrough pins Anode Vacuum connection Adjustment data Ion detector potential Cathode potential Anode potential Emission current Heating current for the hot cathode Heating voltage for the hot cathode Sensitivity for nitrogen Bake out operation Electron bombardment Operating units DN mbar (Torr) mbar (Torr) C
IE 414
2 x 10-11 to 10-2 (1.5 x 10-11 to 10-2) 10-11 ( 10-11)
IE 514
10-12 to 1 x 10-4 (10-12 to 7.5 x 10-5) 10-12 ( 10-12)
0 to +80
0 to +80
200 1) / 350 2)
200 1) / 350 2)
V V V mA A V mbar-1 V mA
0 100 220 1.6 1.4 3.7 6.6 700 30 IM 520, 510, 540
Ordering Information
IONIVAC sensors Replacement cathode
1) 2)
IE 414
Part No. 158 66 Part No. 158 63
IE 514
Part No. 158 67 Part No. 158 61
With bakeable gauge head cable With gauge head cable detached
73 C16
TR 211
- Aluminum sensing cell with tungsten filament - Improved temperature compensation
TR 216
- Stainless steel sensing cell with platinum filament and ceramics feed through - Well suited for corrosive processes and water vapor atmospheres
35
135 (144)
133 (122)
20
40
Dimensional drawing for the TR 211, TR 212 and TR 216; TR 211 NPT in brackets
C16 74
Technical Data
Measurement range
TR 211
TR 211 NPT
TR 212
TR 216
mbar 5 x 10-4 to 1000 5 x 10-4 to 1000 5 x 10-4 to 1000 5 x 10-4 to 1000 -4 to 7.5 x 102) (3.8 x 10-4 to 7.5 x 102) (3.8 x 10-4 to 7.5 x 102) (3.8 x 10-4 to 7.5 x 102) (Torr) (3.8 x 10
Operating temperature range (compensated) Storage temperature range, max. Filament Filament temperature Permissible overload (abs.), max. Volume of the sensing cell, approx. Vacuum connection Materials in contact with the medium
C C
0 to +40 80 tungsten
C bar cm3 DN
stainless steel, Vacon, stainless steel, Vacon, stainless steel 1.4301 glass, tungsten glass, tungsten (SS 304), CrNi 8020 CrNi 8020 Al2O3 ceramics, epoxy cement CrNi 8020, THERMOVAC TM 21, 22, 23 / COMBIVAC CM 31, 32, 33, 51 / PIEZOVAC PV 20 THERMOVAC TM 21, 22, 23 / COMBIVAC CM 31, 32, 33, 51 / PIEZOVAC PV 20 THERMOVAC TM 21, 22, 23 / COMBIVAC CM 31, 32, 33, 51 / PIEZOVAC PV 20
Operating units
Ordering Information
THERMOVAC sensors Series 200 DN 16 KF DN 16 CF DN 1/8" NPT Replacement sensing cell
TR 211
TR 211 NPT
TR 212
TR 216
75 C16
Option
For protection of the PENNINGVAC sensors against contamination, radiation and other disturbing factors the in-stallation of a baffle is recommmended.
50
50
132 94
147 107
DN 25 KF / DN 40 KF
DN 40 CF
C16 76
Technical Data
Measurement range mbar (Torr) kV C C C bar cm3
PR 25
1 x 10-9 to 10-2 (0.75 x 10-9 to 10-2) +3.3/+1.6 -25 to +80 0 to +80 6 1) 21 stainless steel, nichrome, ceramics, titanium
PR 26
1 x 10-9 to 10-2 (0.75 x 10-9 to 10-2) +3.3/+1.6 -25 to +80 0 to +80 6 1) 21 stainless steel, nichrome, ceramics, titanium 0.75 (1.66) 40 KF
PR 27
1 x 10-9 to 10-2 (0.75 x 10-9 to 10-2) +3.3/+1.6 -25 to +80 0 to +80 6 1) 21 stainless steel, nichrome, ceramics, titanium 0.8 (1.66) 40 CF
PR 28
1 x 10-9 to 10-2 (0.75 x 10-9 to 10-2) +3.3/+1.6 -25 to +80 0 to +200 350 6 1) 21 stainless steel, nichrome, ceramics, titanium 0.8 (1.66) 40 CF
High voltage supply (anode potential) Storage temperature range Nominal temperature range Bake out temperature (flange) Permissible overload (abs.) Dead volume Materials in contact with the medium
kg (lbs) DN
0.75 (1.66) 25 KF
COMBIVAC COMBIVAC COMBIVAC COMBIVAC CM 31, 32, 33 / CM 31, 32, 33 / CM 31, 32, 33 / CM 31, 32, 33 / PENNINGVAC PM 31 PENNINGVAC PM 31 PENNINGVAC PM 31 PENNINGVAC PM 31
Ordering Information
PENNINGVAC sensors Replacement cathode plate, titanium (5 pcs., incl. 5 ceramics discs) Replacement anode ring Baffle, with centering ring
1)
PR 25
Part No. 157 52
PR 26
Part No. 136 46
PR 27
Part No. 136 47
PR 28
Part No. 136 48
Part No. EK 162 91 Part No. EK 162 91 Part No. EK 162 91 Part No. EK 162 91 Part No. 200 28 711 Part No. 200 28 711 Part No. 200 28 711 Part No. 200 28 711 Part No. 230 078 Part No. 230 079
77 C16
Typical Applications
- Pressure measurements in the rough vacuum range, and for corrosive media - Chemical process engineering - Vacuum packaging - Drying processes - Measurement of operating and filling pressure, during the production of lamps - Filling systems for brake fluids (DI 201/DI 2001) - Filling systems for refrigerants - Measurement of pressure relative to atmospheric pressure (DI 2001 rel)
80 58
48,5
DN 16 KF
Dimensional drawing for the sensors DI 200/DI 201/DI 2000/DI 2001/DI 2001 rel
C16 78
Technical Data
Measurement range mbar (Torr)
DI 200
0.1 to 200 (0.075 to 150)
DI 201
0.1 to 200 (0.075 to 150)
DI 2000
1 to 2000 (0.75 to 1500)
DI 2001
1 to 2000 (0.75 to 1500)
DI 2001 rel
-1000 to +1000 (-750 to +750) relative pressure 10 0 to +60
Overload range, max. Nominal temperature range Measurement uncertainty 1) ( temperature error) Resolution Reproducibility Linearity Temperature error Zero drift Sensitivity drift Principle of measurement Sensing head supply Output signal Supply voltage Operating range % % % %
bar C
5 0 to +60
5 0 to +60
10 0 to +60
10 0 to +60
FS FS FS FS
% FS/10K % FS/10K
Two-wire system Two-wire system Two-wire system Two-wire system Two-wire system mA V DC V 4 to 20 +24 typ. 12 to 30, ripple 1 Vpp 3 16 KF 0.55 (1.2) 44 Stainless Steel 1.4305 Al2O3 (96 %) Ceramics FPM (FKM) 4 to 20 +24 typ. 12 to 30, ripple 1 Vpp 3 16 KF 0.55 (1.2) 44 Stainless Steel 1.4305 Al2O3 (96 %) Ceramics EPDM 4 to 20 +24 typ. 12 to 30, ripple 1 Vpp 3 16 KF 0.55 (1.2) 44 Stainless Steel 1.4305 Al2O3 (96 %) Ceramics FPM (FKM) 4 to 20 +24 typ. 12 to 30, ripple 1 Vpp 3 16 KF 0.55 (1.2) 44 Stainless Steel, 1.4305 Al2O3 (96 %) Ceramics, EPDM 4 to 20 +24 typ. 12 to 30, ripple 1 Vpp 3 16 KF 0.55 (1.2) 44 Stainless Steel 1.4305 Al2O3 (96 %) Ceramics EPDM
Dead volume Vacuum connection Weight, approx. Protection class Materials in contact with the medium
cm3 DN kg (lbs) IP
Operating units
MEMBRANOVAC MEMBRANOVAC MEMBRANOVAC MEMBRANOVAC MEMBRANOVAC DM 11, DM 12 / DM 11, DM 12 / DM 11, DM 12 / DM 11, DM 12 / DM 11, DM 12 / PIEZOVAC PIEZOVAC PIEZOVAC COMBIVAC COMBIVAC PV 20 / PV 20 / PV 20 / CM 32, 2T CM 32 COMBIVAC COMBIVAC COMBIVAC CM 32, 2T CM 32, 2T CM 32, 2T
Ordering Information
Linear absolute pressure sensor, complete with 5 m long connection cable and connecting plug Extension cable 10 m 20 m
1)
DI 200
DI 201
DI 2000
DI 2001
DI 2001 rel
Part No. 245 000 Part No. 200 04 112 200 02 645
79 C16
Type
Replacement spare sensor IE 10 Replacement spare sensor IE 20 Replacement spare sensor IE 21 Replacement spare sensor IE 100 DN 25 KF DN 40 CF Replacement spare cathode Replacement spare sensor TR 301 Replacement sensing cell Replacement sensing cell Replacement sensing cell
1)
Ordering Information
Part No. 163 43 Part No. 163 14 Part No. 158 17
ITR 100 ITR 100 IE 220 TM 320, CM 350, IM 520, DM 11, DM 12, CM 32 TR 201/901 DN 16 KF TR 201/901 NPT TR 301
Part No. E 163 61 Part No. E 163 67 Part No. 158 58 1) Part No. 157 40 Part No. 162 09 1) Part No. 896 76 1) Part No. 157 43
to be discontinued
C16 80
Notes
81 C16
Process Chamber
PLC, PC
C16 82
49
200
17
100
60
60
(17) 33
6.5 8
DN 10 ISO-KF 18 OD 1/4"
Example of a gas flow curve (mean value of 500 sccm FS) at a differential pressure of Vp = 1 bar for Move X
152 90
30
DN 16 ISO-KF
14.5 OD 6 mm
35
14.5
1E-6 0 1 2 3 4 5 6 7 8 9 10
USTG [V]
83 C16
Technical Data
Vacuum connection
Move X
Threaded connection M 14 x 1 for accepting of: Flange connection DN 10 ISO-KF Pipe connection 1/4" Pipe connection 6 mm Any, preferably vertical 1 x 10-9 1 x 10-8 mbar to 1000 mbar (max. pressure in closing direction 2 bar) 10 sccm FS 2), 3) 50 sccm FS 2), 3) 100 sccm FS 2), 3) 500 sccm FS 2), 3) 1000 sccm FS 2), 3) 5000 sccm FS 2), 3)
Installation orientation Leak tightness Pressure range Gas flow 1) with filter, on the inlet side mbar x l/s
with filter, on the inlet and the vacuum side Power supply Supply voltage Power consumption Current consumption Actuation Digital Analog Type of protection Closing time / opening time Response time Ambient temperature Bake out temperature Valve body Actuator Material Valve body Valve needle, valve disc Filter Seals Dosing sleeve Armature disc Guide tube Weight
1)
mbar x l/s
5 x 10-6 to 1000
V DC VA mA
24 ( 10%) 12 500, 20 to 30 (closed circuit current) Step motor CMove or IMove 1250 0 to 10 40 3/4 +5 to +40
V DC IP s ms C
C C
80 60
Stainless steel 1.4435 Stainless steel 1.4301 Stainless steel 1.4404 FPM (FKM) Fluorplastomer kg (lbs) 0.5 (1.1)
Stainless steel 1.4301 FPM (FKM) Stainless steel 1.4510 Stainless steel 1.4105 0.08 (0.18)
2)
For air at a differential pressure of p = 1 bar mbar x l s 1 sccm = 1.69 x 10-2 Minimum flow: in each case 1% of FS (full-scale)
3)
C16 84
Part No. 230 220 Part No. 230 221 Part No. 230 222 Part No. 230 223 Part No. 230 224 Part No. 230 225
Part No. 230 210 Part No. 230 211 Part No. 230 212 Part No. 230 213 Part No. 230 214 Part No. 230 215
85 C16
Technical Applications
- Simple to operate - Self-explanatory LCD display and function keys - Analog/digital inputs/outputs and interfaces - Digital input valve control, switchover Pressure and gas flow adjustment Digital output valve control, error message - Adjustable PID control algorithms - 99 pre-programmed PI control algorithms for fast and simple operation - For installation or tabletop use (1/4 19 in.)
Typical Applications
The controller CMove in connection with the electric control valve Move 1250 or Move X is suited for applications in the areas like for example - Semiconductor industry - Analytical - Coating - Quality control - Surface treatment as well as in all applications where simple, fast and accurate pressure control is desired and necessary.
128
122.5
110
command
edit
C16 86
Technical Data
Power supply Mains voltage, 50/60 Hz Power consumption Control types
V AC VA
90 - 250 50
Control inaccuracy Display Display units (selectable) Pressure Flow Display range Pressure control CERAVAC THERMOVAC PENNINGVAC IONIVAC with 0 bis 10 V linear Flow control with MOVE 1250 with MOVE X with analog output AA 2 Measurement uncertainty Operating mode
Auto = (PI) adjustable in 1 to 99 selectable steps PID = PID is user adjustable 5% F.S. sensor LCD 64 x 128 Pixel
mV
0 to 10.000 CLOSED, 5.0 x 10-6 to 1.25 x 10+3 mbar x l/s CLOSED, 1.0 x 10-5 to 1.0 x 10+2 mbar x l/s 0 to 10 000 0.2 FS Sensor Gas flow regulation (upstream regulation) Pressure regulation (downstream regulation) Local operation or remote control Flow matching, opening/closing of external valves / operating mode flow/regulation / switching on of emission/degas Valve position indication; valve fault; sensor fault; status message for sensor and valve; pressure regulation upstream/downstream 0 to 10 V DC setpoint pressure/flow 0 to 10 V DC, Pressure sensor signal, valve signal, valve position MOVE 1250 RS 232 C, RS 485 C 1/4 19", for installation or tabletop use
mV %
Digital output
Serial interface Housing Weight Temperature Operation Storage Protection class kg (lbs)
1.65 (3.64)
C C IP
87 C16
Ordering Information
Controller CMove 1250 Control valve Move 1250 Move X with gas flow 0010 sccm 0050 sccm 0100 sccm 0500 sccm 1000 sccm 5000 sccm Connection cable CMove 1250 - Control valve Move 1250 03 m 05 m 10 m 15 m 20 m 25 m CMove 1250 - Control valve Move X 03 m 05 m 10 m 15 m 20 m 25 m Sensor cable TTR 90, TTR 100, TTR 211, TTR 216 PTR 225, PTR 237 005 m 015 m 020 m 030 m 040 m 050 m 075 m 100 m ITR 90, ITR 100, ITR 200 005 m 015 m 020 m 030 m CTR 90, CTR 91, CTR 100 005 m 010 m 015 m 020 m 030 m 040 m 050 m 075 m 100 m
Part Part Part Part Part Part Part Part Part Part Part Part
No. No. No. No. No. No. No. No. No. No. No. No.
230 230 230 230 230 230 230 230 230 230 230 230
220 221 222 223 224 225 210 211 212 213 214 215
Part Part Part Part Part Part Part Part Part Part Part Part Part Part Part Part Part Part Part Part Part
No. No. No. No. No. No. No. No. No. No. No. No. No. No. No. No. No. No. No. No. No.
124 124 124 124 124 124 124 124 124 124 124 124 230 230 230 230 230 230 230 230 230
C16 88
Typical Applications
The interface module IMove 1250 connects a RS 232C interface (for example from a computer or a PLC) to the digital interface of the Move 1250 control valve.
Technical Data
Temperature Operation Storage Protection class Operating voltage Current consumption IMove 1250 Move 1250 Provide upfront fusing Interface Mounting Support rails
C C IP V DC
mA mA
Weight
Ordering Information
Interface module IMove 1250
54
18
Computer or PLC
62
89 C16
Typical Applications
- Venting facilities - Safety shutdown of vacuum systems - Load locks - Increased switching capacity when using the switching amplifier SV 110
Technical Note
Due to the diaphragm material used (EPDM) the PS 113 A is not suited for applications in which the process gas contains large quantities of helium. Owing to helium permeation, the leak rate of the diaphragm for helium amounts to values 1 x 10-4 mbar x l/s.
72 50
48,5
DN 16 KF
C16 90
Technical Data
Switching pressure Return switching pressure Switching inaccuracy mbar (Torr) mbar (Torr) mbar (Torr)
Max. permissible operating pressure (abs.) mbar (Torr) Storage temperature range Nominal temperature range Switching contact Contact life Switching capacity C C
2000 (1500) -25 to +85 0 to +85 Changeover contacts, gold-plated, for prog. controls > 105 switching cycles 100 mA / 24 V AC 30 mA / 24 V DC 6.3 mm flat plug
Electrical connection Vacuum connection Helium permeation Internal volume Materials in contact with the medium DN mbar x l/s cm3
16 KF 1 x 10-4 2 Stainless steel 1.4305, Stainless steel 1.4310, Stainless steel 1.4300 PTFE coated, EPDM
g (lbs) IP
315 (0.70) 44
Ordering Information
Low pressure safety switch PS 113 A, DN 16 KF; complete with 3 m long cable
91 C16
21 133 105 19 51
14 18 DN 16 KF 85
C16 92
Technical Data
Switching range Overload limit Sensitivity Switching hysteresis Temperature coefficient Nominal temperature range briefly (max. 8 h) continuous Switching contact Switching voltage Switching current (max.) Contact resistance, max. Electrical connection Protection class Vacuum connection Materials in contact with the medium Measurement chamber Reference chamber Volume of the measurement chamber Volume of the reference chamber Weight cm3 cm3 kg (lbs) V mA k mbar (Torr) mbar (Torr) mbar (Torr) mbar (Torr) %/K
Pressure Switch
0.5 to 2000 (0.375 to 1500) 3000 (2250) 0.1 (0.75) 0.5 (0.375) 0.4 of the switching value
C C
120 0 to +90 Normally closed, gold-plated, for prog. controls 24 10 1 Plug (DIN 43 650) 65 16 KF
IP DN
Stainless steel 1.4301; 1.4401; 1.4310; 1.3541; FPM (FKM) Stainless steel 1.4301; 1.4401; 1.3541; glass; gold 4 20 1.3 (2.87)
Ordering Information
Pressure Switch PS 115, DN 16 KF Pressure switch adjustment For floating installations without SV 110, Clamping ring DN 16 KF, plastic Centering ring, DN 16 KF, plastic Option Differential pressure adapter, DN 16 KF, for connection to the PS 115 Spare parts kit PS 115 SV 110 switching amplifier
Pressure Switch
Part No. 160 04 Part No. 160 05
93 C16
55,5 15 4,2
121 109
80 Pg 9 Pg 9 21,5 163
21 50,5
C16 94
Technical Data
Mains supply, 50/60 Hz Power consumption Output relay Switching voltage / current Switching power, max. Response time Relaise time Control circuit Ambient temperature, max. Weight, approx VA
Switching Amplifier
110/130/220/240 V, selectable 3
V/A VA ms ms V / mA C kg (lbs)
Ordering Information
Switching amplifier SV 110
Switching Amplifier
Part No. 160 78
95 C16
Typical Applications
- Distillation processes of all kinds - Solvent recovery - Drying processes - Temperature control on bath cryostats - Degassing of liquids and plastics
135
98 120 58,5
39
67 130
92,5 165
C16 96
Technical Data
Throughput Nominal temperature range Storage temperature range Temperature coefficient Settling time Permissible overload for brief periods Diaphragm material Housing material Installation orientation Dimensions Vacuum connection Measurement connection Weight, approx.
Ordering Information
Diaphragm pressure regulator MR 16, DN 16 KF MR 50, DN 40 KF Options Stainless steel measurement flange, DN 16 KF, for connection to a reference and/or process chamber or pumping stud KALREZ diaphragm Spare Parts EPDM diaphragm and seal kit Viton diaphragm and seal kit Seal kit MR 50, incl. EPDM and Viton diaphragms Adjustment screw for the adjustable valve, complete with seal
97 C16
Typical Applications
Calibrated vacuum gauges are used under the following conditions: - If the requirements concerning reproducibility and comparability of experiment runs are high
- Production of pharmaceuticals and herbicides - Sputtering systems - Aircraft and space industry - Manufacture of lamp
C16 98
Technical Data
Calibration range mbar to 10-3
DKD Calibration
to 10-5 to 10-9
Ordering Information
DKD calibration Part No. 157 12
DKD Calibration
Part No. 157 13 Part No. 157 14
Technical Data
Calibration range mbar to 10-3
Factory Calibration
to 10-5 to 10-8
Ordering Information
Factory calibration Part No. 154 22
Factory Calibration
Part No. 154 23 Part No. 154 24
99 C16
Europe
Belgium Oerlikon Leybold Vacuum Nederland B.V. Belgisch bijkantoor Leuvensesteenweg 542-9A B-1930 Zaventem Sales: Phone: +32-2-711 00 83 Fax: +32-2-720 83 38 [email protected] Service: Phone: +32-2-711 00 82 Fax: +32-2-720 83 38 [email protected] France Oerlikon Leybold Vacuum France S.A. 7, Avenue du Qubec Z.A. de Courtaboeuf 1 - B.P. 42 F-91942 Courtaboeuf Cedex Sales and Service: Phone: +33-1-69 82 48 00 Fax: +33-1-69 07 57 38 [email protected] [email protected] Oerlikon Leybold Vacuum France S.A. Valence Factory 640, Rue A. Bergs B.P. 107 640 F-26501 Bourg-ls-Valence Cedex Service: Phone: +33-4-75 82 33 00 Fax: +33-4-75 82 92 69 [email protected] Great Britain Oerlikon Leybold Vacuum UK LTD. Silverglade Business Park Leatherhead Road Unit 2 KT9 2QL Chessington, Surrey (London) Sales: Phone: +44-13-7273 7300 Fax: +44-13-7273 7301 [email protected] Service: Phone: +44-20-8971 7030 Fax: +44-20-8971 7003 [email protected] Italy Oerlikon Leybold Vacuum Italia S.r.l. Via Trasimeno 8 I-20128 Milano Sales: Phone: +39-02-27 22 31 Fax: +39-02-27 20 96 41 [email protected] Service: Phone: +39-02-27 22 31 Fax: +39-02-27 22 32 17 [email protected] Netherlands Oerlikon Leybold Vacuum Nederland B.V. Proostwetering 24N NL-3543 AE Utrecht Sales and Service: Phone: +31-(30) 242 6330 Fax: +31-(30) 242 6331 [email protected] [email protected] Spain Oerlikon Leybold Vacuum Spain, S.A. C/ Huelva 7 E-08940 Cornell de Llobregat (Barcelona) Sales: Phone: +34-93-666 43 11 Fax: +34-93-666 43 70 [email protected] Service: Phone: +34-93-666 46 16 Fax: +34-93-685 43 70 [email protected] Sweden Oerlikon Leybold Vacuum Scandinavia AB Box 9084 SE-40092 Gteborg Sales and Service: Phone: +46-31-68 84 70 Fax: +46-31-68 39 39 [email protected] [email protected] Visiting/delivery address: Datavgen 57B SE-43632 Askim Switzerland Oerlikon Leybold Vacuum Schweiz AG Leutschenbachstrasse 55 CH-8050 Zrich Sales: Phone: +41-44-308 40 50 Fax: +41-44-302 43 73 [email protected] Service: Phone: +41-44-308 40 62 Fax: +41-44-308 40 60 [email protected]
Asia
P.R. China Oerlikon Leybold Vacuum (Tianjin) International Trade Co. Ltd. Beichen Economic Development Area (BEDA), No.8 Western Shuangchen Road Tianjin 300400 China Sales and Service: Phone: +86-22-2697 0808 Fax: +86-22-2697 4061 Fax: +86-22-2697 2017 [email protected] [email protected] [email protected] Oerlikon Leybold Vacuum (Tianjin) Co. Ltd. Beichen Economic Development Area (BEDA), No.8 Western Shuangchen Road Tianjin 300400 China Sales and Service: Phone: +86-22-2697 0808 Fax: +86-22-2697 4061 [email protected] [email protected] [email protected] Oerlikon Leybold Vacuum (Tianjin) International Trade Co. Ltd. Shanghai Branch: No.33 76 Fu Te Dong San Road Waigaoqiao Free Trade Zone Shanghai 200131 China Sales and Service: Phone: +86-21-5064-4666 Fax: +86-21-5064-4668 [email protected] [email protected] [email protected] Oerlikon Leybold Vacuum (Tianjin) International Trade Co. Ltd. Guangzhou Office and Service Center 1st F, Main Building Science City Plaza, No.111 Science Revenue, Guangzhou Science City (GZSC) 510663, Guangzhou, China Sales: Phone: +86-20-223 23 980 Fax:+86-20-223 23 990 [email protected] [email protected] [email protected] Oerlikon Leybold Vacuum (Tianjin) International Trade Co. Ltd. Beijing Branch: 1-908, Beijing Landmark Towers 8 North Dongsanhuan Road Beijing 100004 China Sales: Phone: +86-10-6590-7622 Fax: +86-10-6590-7607 [email protected] [email protected] India Oerlikon Leybold Vacuum India Pvt Ltd. EL 22, J-Block MIDC Bhosari Pune 411026 India Sales and Service: Phone: +91-20-3061 6000 Fax: +91-20-2712 1571 [email protected] [email protected] Japan Oerlikon Leybold Vacuum Japan Co., Ltd. Headquarter 23-3, Shin-Yokohama 3-chome Tobu A.K. Bldg. 4th Floor Kohoku-ku Yokohama-shi 222-0033 Sales: Phone: +81-45-471-3330 Fax: +81-45-471-3323 [email protected] [email protected] Oerlikon Leybold Vacuum Japan Co., Ltd. Osaka Sales Office 3F, Shin-Osaka Terasaki No.3 Bldg. 1-5-28 Nishi-Miyahara Yodogawa-ku, Osaka-shi Osaka 532-0004 Phone: +81-6-6399-6271 Fax: +81-6-6399-6273 [email protected] [email protected] Oerlikon Leybold Vacuum Japan Co., Ltd. Tsukuba Technical Service Center Kogyo Danchi 21, Kasuminosato, Ami-machi, Inashiki-gun Ibaraki-ken, 300-0315 Service: Phone: +81-298 89 2841 Fax: +81-298 89 2838 [email protected] [email protected] South Korea Oerlikon Leybold Vacuum Korea Ltd. 3F. Jellzone 2 Tower Jeongja-dong 159-4 Bundang-gu Sungnam-si Gyeonggi-do Bundang 463-384, Korea Sales: Phone: +82-31 785 1367 Fax: +82-31 785 1359 [email protected] Service: 623-7, Upsung-Dong Cheonan-Si Chungcheongnam-Do Korea 330-290 Phone: +82-41 589 3035 Fax: +82-41 588 0166 [email protected] Singapore Oerlikon Leybold Vacuum Singapore Pte Ltd. 1 Science Park Road Singapore Science Park 2 #02-12, Capricorn Building Singapore 117528 Sales and Service: Phone: +65-6303 7000 Fax: +65-6773 0039 [email protected] [email protected] Taiwan Oerlikon Leybold Vacuum Taiwan Ltd. No 416-1, Sec. 3 Chunghsin Road., Chutung Hsinchu County 310 Taiwan, R.O.C. Sales and Service: Phone: +886-3-500 1688 Fax: +886-3-583 3999 [email protected] [email protected]
America
USA Oerlikon Leybold Vacuum USA Inc. 5700 Mellon Road USA-Export, PA 15632 Phone: +1-724-327-5700 Fax: +1-724-325-3577 [email protected] Sales: Eastern & Central time zones Phone: +1-724-327-5700 Fax: +1-724-333-1217 Pacific, Mountain, Alaskan & Hawaiian time zones Phone: +1-408-436-2828 Fax: +1-408-436-2849 Service: Phone: +1-724-327-5700 Fax: +1-724-325-3577 Latin America Oerlikon Leybold Vacuum USA Inc. Brazilian Office Av. Dria 360 cj. 12 Campo Belo 04635-070 So Paulo SP BRAZIL Sales: Phone: +55 11-3554 3117 Fax: +55 11-3554 3117 [email protected]
08.09
Oerlikon Leybold Vacuum GmbH Bonner Strasse 498 D-50968 Cologne Phone: +49-(0)221-347 0 Fax: +49-(0)221-347 1250 [email protected]
LV_10029_2009
www.oerlikon.com/ leyboldvacuum