Compression Testing Machine
Compression Testing Machine
Verification Checklist
Standards Referenced:
Annual
Machine Model & Serial
- CTM-001 ✅ calibration
ID No.
done
Calibrated with
0.98%
Load Cell Accuracy ±1% or better ✅ certified
error
weights
Load Display
±1% 0.99% ✅ Verified
Indicator Accuracy
Cylinder Within
Diameter 152 ±1 mm 151.8 mm ✅
Platens tolerance
Visual,
Safety Integrity &
operational OK ✅ No defects
Guards Functionality
check
Calibrated, Calibration
Measurement
Calipers traceable Certified ✅ valid till
Accuracy
standard 2026
Measurement
Micrometers Calibrated Certified ✅
Accuracy
Measured Pass/
Parameter Requirement Remarks
Value Fail
Testing Thermometer
23 ± 2 °C 22.8 °C ✅
Temperature calibrated
Document Control
Scope: Applies to load frames, load cells, platens, speed controls, measuring
tools, and environmental controls used in compressive strength testing of
concrete cylinders.
Responsibilities:
Procedures:
Records:
Parame Next
Equip Equipm Meas Last
ter Stan Require Pass/ Calibr Rema
ment ent ured Calibr
Checke dard ment Fail ation rks
ID Type Value ation
d Due
with
certifie
Accurac
Machine y d
y
weight
s
152 ±1 Measu
151.8
Diamete mm, ≤ red
PLTN- ASTM mm, 2025- 2026-
Platen r& 0.025 ✅ with
001 C39 0.020 06-10 06-10
Flatness mm dial
mm
flatness gauge
Calibra
Measure Calibrat
ASTM tion
CAL- ment ed, Certifi 2025- 2026-
Calipers C39 & ✅ valid
005 Accurac traceabl ed 06-10 06-10
C617 till
y e
2026
Displace Calibrat
EXT- Extenso ment ASTM ed, Certifi 2025- 2026-
✅
007 meter Accurac C39 traceabl ed 06-10 06-10
y e
Digital
Time ±2
TMR- ASTM Certifi 2025- 2026- timer
Timer Accurac seconds/ ✅
003 C39 ed 06-10 06-10 calibra
y hour
ted