0% found this document useful (0 votes)
48 views4 pages

Compression Testing Machine

The document outlines a calibration and verification checklist for compression testing machines and associated equipment in accordance with ASTM standards. It details the requirements, measurements, and pass/fail results for various components, including load cells, measuring tools, and environmental controls. The document also specifies responsibilities, procedures, and record-keeping practices to ensure compliance and accuracy in testing.

Uploaded by

khobarlab
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
48 views4 pages

Compression Testing Machine

The document outlines a calibration and verification checklist for compression testing machines and associated equipment in accordance with ASTM standards. It details the requirements, measurements, and pass/fail results for various components, including load cells, measuring tools, and environmental controls. The document also specifies responsibilities, procedures, and record-keeping practices to ensure compliance and accuracy in testing.

Uploaded by

khobarlab
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 4

Compression Testing Machine & Equipment – Calibration and

Verification Checklist

Standards Referenced:

 ASTM C39 / C39M

 ASTM C617 / C617M

SECTION A – Compression Testing Machine (ASTM C39)

Check Requirement Measured Pass/


Item Remarks
Parameter / Tolerance Value Fail

Annual
Machine Model & Serial
- CTM-001 ✅ calibration
ID No.
done

Calibrated with
0.98%
Load Cell Accuracy ±1% or better ✅ certified
error
weights

Load Display
±1% 0.99% ✅ Verified
Indicator Accuracy

Cylinder Within
Diameter 152 ±1 mm 151.8 mm ✅
Platens tolerance

Cylinder Measured with


Flatness ≤ 0.025 mm 0.020 mm ✅
Platens dial gauge

Test Speed Rate of Load 0.15 ±0.03 0.14


✅ Timer verified
Control Application MPa/sec MPa/sec

Visual,
Safety Integrity &
operational OK ✅ No defects
Guards Functionality
check

SECTION B – Measuring Tools (ASTM C39 & C617)


Requirement Measure Pass/
Equipment Parameter Remarks
/ Calibration d Value Fail

Calibrated, Calibration
Measurement
Calipers traceable Certified ✅ valid till
Accuracy
standard 2026

Measurement
Micrometers Calibrated Certified ✅
Accuracy

Extensometer Displacement Calibrated,


Certified ✅
(if used) Accuracy traceable

Timer / ±2 Digital timer


Accuracy Certified ✅
Stopwatch seconds/hour calibrated

SECTION C – Environmental Controls (For Testing Conditions)

Measured Pass/
Parameter Requirement Remarks
Value Fail

Testing Thermometer
23 ± 2 °C 22.8 °C ✅
Temperature calibrated

Testing ≥ 50% RH (as Hygrometer


55% RH ✅
Humidity applicable) calibrated

Document Control

 Date of Inspection: 2025-06-10

 Inspected By: John D. Tester

 Equipment ID(s): CTM-001, PLTN-001, CAL-005, EXT-007, TMR-003

 Next Due Date: 2026-06-10

SOP Integration: Calibration and Verification Procedure for


Compression Testing Equipment
Purpose: To verify and calibrate the compression testing machine and
associated equipment in accordance with ASTM C39 and C617 to maintain
test accuracy and ISO/IEC 17025 compliance.

Scope: Applies to load frames, load cells, platens, speed controls, measuring
tools, and environmental controls used in compressive strength testing of
concrete cylinders.

Responsibilities:

 Lab Technician: Perform daily/periodic equipment verification and


checks.

 QA Officer: Schedule and maintain calibration records.

 Technical Manager: Approve equipment for use based on calibration


results.

Procedures:

 Calibrate load cells with certified weights annually.

 Measure platen dimensions and flatness per ASTM C39.

 Verify load application speed with timers.

 Ensure measuring tools are calibrated and traceable.

 Maintain environmental conditions and verify measurement devices.

Records:

 Use checklist form after each verification.

 Maintain calibration certificates and tracking logs.

 Review records annually or as required.

Excel Tracking Sheet Layout Preview

Parame Next
Equip Equipm Meas Last
ter Stan Require Pass/ Calibr Rema
ment ent ured Calibr
Checke dard ment Fail ation rks
ID Type Value ation
d Due

CTM- Compre Load ASTM ±1% 0.98% ✅ 2025- 2026- Calibra


001 ssion Cell C39 accurac error 06-10 06-10 ted
Parame Next
Equip Equipm Meas Last
ter Stan Require Pass/ Calibr Rema
ment ent ured Calibr
Checke dard ment Fail ation rks
ID Type Value ation
d Due

with
certifie
Accurac
Machine y d
y
weight
s

152 ±1 Measu
151.8
Diamete mm, ≤ red
PLTN- ASTM mm, 2025- 2026-
Platen r& 0.025 ✅ with
001 C39 0.020 06-10 06-10
Flatness mm dial
mm
flatness gauge

Calibra
Measure Calibrat
ASTM tion
CAL- ment ed, Certifi 2025- 2026-
Calipers C39 & ✅ valid
005 Accurac traceabl ed 06-10 06-10
C617 till
y e
2026

Displace Calibrat
EXT- Extenso ment ASTM ed, Certifi 2025- 2026-

007 meter Accurac C39 traceabl ed 06-10 06-10
y e

Digital
Time ±2
TMR- ASTM Certifi 2025- 2026- timer
Timer Accurac seconds/ ✅
003 C39 ed 06-10 06-10 calibra
y hour
ted

You might also like