CB Testing 2
CB Testing 2
Close time= close pulse Time(Until Energize only) + Close time of Mechanism of CB
open time= close pulse Time (Until Energize only) + open time of Mechanism of CB
Recommended That Follow FAT Test Values OF (Close Pulse - open pulse- Close
time- open time)
CB Timing Test
1-Opening Time Test
Why we recommend to take Pulse Time 30ms?
Because CB Close Time range (30-40 msc)
CB Open Time range (20-30 msc)
So 30msc Pulse time will in Case of Opening and
Closing will be enough Value to Energize Coil of
Tripping and Coil Of Closing
CB Timing Test
1-Closing Time Test
Why we recommend to take Pulse Time 30ms?
Because CB Close Time range (30-40 msc) Closing Pulse
CB Open Time range (20-30 msc)
So 30msc Pulse time will in Case of Opening and
Closing will be enough Value to Energize Coil of
Tripping and Coil Of Closing
CB Timing Test
1-Close – Open (C-O Time Test)
C-o test simulate Switch On to Fault Test
(SOTF)
Consist of : Close Pulse
Close to open
2-Close To open (Delay time) :50ms
Delay times (0.3 Min & 3 Min) Called Dead time that Allow :
• Inject 100A DC
• Perform Test For Each Pole (R-S-T) And Compare Results With FAT test
• Results Shall be
CB Insulation Resistance (IR test)
IR during Close Circuit Breaker:
1-Phase To Phase Test (R-Y & R-B & Y-B)
• Inject 5 KV For 1 Min For Each Phase
• Results Shall be >100 Mega-Ohm
CB Insulation Resistance (IR test)
IR during Close Circuit Breaker:
1-Phase To Earth Test (R-E & B-E & Y-E)
• Inject 5 KV For 1 Min For Each Phase
• Results Shall be > 100 Mega-Ohm
CB Insulation Resistance (IR test)
IR during Open Circuit Breaker:
1-Phase (R &Y & B ) to check no Contact Between Circuit Breaker Contacts
During open position .
• Inject 5 KV For 1 Min For Each Phase
• Results Shall be > 100 Mega-Ohm
THANKS