DF55 Technical Parameters and Configuration List

The document outlines the technical specifications of a five-class blood cell analyzer, detailing its detection principles, parameters, and analysis modes. It supports various blood sample types, offers multiple calibration and quality control modes, and features advanced data management and networking capabilities. Additionally, it includes maintenance functions, alarm systems, and a comprehensive packing list for the device and its accessories.

Uploaded by

lindongmei1000
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
12 views2 pages

DF55 Technical Parameters and Configuration List

The document outlines the technical specifications of a five-class blood cell analyzer, detailing its detection principles, parameters, and analysis modes. It supports various blood sample types, offers multiple calibration and quality control modes, and features advanced data management and networking capabilities. Additionally, it includes maintenance functions, alarm systems, and a comprehensive packing list for the device and its accessories.

Uploaded by

lindongmei1000
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Technical Parameter

1. Detection principle: Coulter's principle to detect the number and volume distribution of red blood cells and platelets; use colorimetry
to measure hemoglobin concentration; use semiconductor laser flow cytometry to obtain five classification results of white blood cells
and white blood cell/basophil count results .

2. Detection parameters: 27 parameters, including 23 reporting parameters: WBC,Neu#, Lym#, Mon#, Eos#, Bas#,Neu% , Lym%,
Mon%, Eos%, Bas%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PDW, PCT; 4 study parameters: ALY#,
ALY %, LIC#, LIC%.

3. Scatter plots: 3 DIFF scatter plots and 1 BASO scatter plot.

4. Histogram: 3 histograms.

5. Analysis mode: CBC, CBC+DIFF.

6. Analysis speed: the maximum detection speed is 60 samples/hour.

7. Blood sample mode: venous whole blood, peripheral whole blood, pre-diluted blood.

8. Injection mode: Open injection.

9. Sample volume: 20 μL.

10. Calibration mode: with calibrator calibration, fresh blood calibration and manual calibration mode.

11. Quality control mode: There are two quality control modes, LJ and XB, which can automatically draw quality control charts.

12. Operating software: support Chinese and English software operation.

13. Input method: Support the input of patient information and hospital information in Chinese and English.

14. Output method: external printer, Chinese and English report format, can print histogram, scatter plot, reference range, etc.,
including a variety of print modes and user-defined templates.

15. Data storage management: Automatically record specimen results, instrument status, logs, etc., and support information review
and query, 50,000 sample records (including scattergrams, histograms, and patient information).

16. Unit system selection: It supports 7 unit systems of Custom, China, International, United States, Canada, Netherlands, and United
Kingdom, which can be set freely.

17. Anti-kill: The sampling needle has the function of anti-kill, which can reduce hole blockage and improve sample suction accuracy.

18. Networking function: It supports two-way LIS and has the function of communicating with the laboratory information system.

19. Blockage removal function: The instrument has the function of automatic detection of blocked holes and automatic blockage
removal; it has various blocking methods such as positive and recoil flushing and burning.

20. Cleaning function: Manual whole machine cleaning, manual WBC pool, RBC pool and sampling needle cleaning can be performed
on the instrument maintenance interface. At the same time, automatic sampling needle cleaning and automatic machine cleaning are
supported.

21. Maintenance function: with one-key fault elimination function.

22. Alarm function: It has FLAG alarm function for abnormal samples, supports custom alarm rules; supports classification or
abnormal shape alarm; has reagent alarm function and fault alarm function.

23. Maintainability: It has a visual reagent management interface, state detection of key components of the instrument, and self-check
of the instrument.

24. Sleep mode: with automatic sleep function, the sleep time can be set manually.

25. Reagents: Original supporting reagents can be provided.

26. Quality control: supporting high, medium and low quality control can be provided.

27. Main parameter display range:

Measurement Item Main Parameter Display Range


9
WBC 0.00–999×10 /L
12
RBC 0.00- 18.00 × 10 /L

HGB 0–300 g/L


9
PLT 0–5000 × 10 /L

HCT 0~80%
28. Repeatability

Project Repeatability

WBC ≤2.0%

RBC ≤1.5%

HGB ≤1.5%

MCV ≤1.0%

PLT ≤4.0%

29. Linear range

Measurement Item Linear Measuring Range


9
WBC 0.00- 50010 /L
12
RBC 0.00- 8.510 /L

HGB 0- 250 g/L


9
PLT 0-800010 /L

HCT 0 - 67%

30. Background requirements

Parameter Background Requirements for Daily Use


9
WBC ≤ 0.210 /L
12
RBC ≤ 0.0210 /L

HGB ≤1g/L

HCT ≤ 0.5 %
9
PLT ≤ 1010 /L

31. Carry-over rate

Project Carry-over Rate

WBC ≤0.5%

RBC ≤0.5%

HGB ≤0.5%

HCT ≤0.5%

PLT ≤1.0%

Packing List

Serial Number Name Dosage Unit

1 Five-class blood cell analyzer complete machine 1 PCS

2 power cable 1 PCS

3 External ground wire 1 PCS

4 manual 1 PCS

5 Diluent takeover 1 PCS

6 Waste liquid float take over 1 PCS

7 certificate 1 PCS

8 waste tank 1 PCS

9 Centrifuge tube 20 PCS