DF55 Technical Parameters and Configuration List
DF55 Technical Parameters and Configuration List
1. Detection principle: Coulter's principle to detect the number and volume distribution of red blood cells and platelets; use colorimetry
to measure hemoglobin concentration; use semiconductor laser flow cytometry to obtain five classification results of white blood cells
and white blood cell/basophil count results .
2. Detection parameters: 27 parameters, including 23 reporting parameters: WBC,Neu#, Lym#, Mon#, Eos#, Bas#,Neu% , Lym%,
Mon%, Eos%, Bas%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PDW, PCT; 4 study parameters: ALY#,
ALY %, LIC#, LIC%.
4. Histogram: 3 histograms.
7. Blood sample mode: venous whole blood, peripheral whole blood, pre-diluted blood.
10. Calibration mode: with calibrator calibration, fresh blood calibration and manual calibration mode.
11. Quality control mode: There are two quality control modes, LJ and XB, which can automatically draw quality control charts.
13. Input method: Support the input of patient information and hospital information in Chinese and English.
14. Output method: external printer, Chinese and English report format, can print histogram, scatter plot, reference range, etc.,
including a variety of print modes and user-defined templates.
15. Data storage management: Automatically record specimen results, instrument status, logs, etc., and support information review
and query, 50,000 sample records (including scattergrams, histograms, and patient information).
16. Unit system selection: It supports 7 unit systems of Custom, China, International, United States, Canada, Netherlands, and United
Kingdom, which can be set freely.
17. Anti-kill: The sampling needle has the function of anti-kill, which can reduce hole blockage and improve sample suction accuracy.
18. Networking function: It supports two-way LIS and has the function of communicating with the laboratory information system.
19. Blockage removal function: The instrument has the function of automatic detection of blocked holes and automatic blockage
removal; it has various blocking methods such as positive and recoil flushing and burning.
20. Cleaning function: Manual whole machine cleaning, manual WBC pool, RBC pool and sampling needle cleaning can be performed
on the instrument maintenance interface. At the same time, automatic sampling needle cleaning and automatic machine cleaning are
supported.
22. Alarm function: It has FLAG alarm function for abnormal samples, supports custom alarm rules; supports classification or
abnormal shape alarm; has reagent alarm function and fault alarm function.
23. Maintainability: It has a visual reagent management interface, state detection of key components of the instrument, and self-check
of the instrument.
24. Sleep mode: with automatic sleep function, the sleep time can be set manually.
26. Quality control: supporting high, medium and low quality control can be provided.
HCT 0~80%
28. Repeatability
Project Repeatability
WBC ≤2.0%
RBC ≤1.5%
HGB ≤1.5%
MCV ≤1.0%
PLT ≤4.0%
HCT 0 - 67%
HGB ≤1g/L
HCT ≤ 0.5 %
9
PLT ≤ 1010 /L
WBC ≤0.5%
RBC ≤0.5%
HGB ≤0.5%
HCT ≤0.5%
PLT ≤1.0%
Packing List
4 manual 1 PCS
7 certificate 1 PCS