Module 1 Z3osk5n engineering
Module 1 Z3osk5n engineering
For a thermometer calibrated between 200°C to 500°C, the range is 200°C to 500°C
but the span is 500-200=300°C.
• Linearity: If the input and output varies in a proportional manner system is linear. It is
closeness to which curve approximates straight line.
Drift: undesired change in the output input relationship over a period of
time. No drift means that with given input the measured value do not
vary with time.
a) zero drift:
If the whole calibration gradually shifts due to slippage, permanent
set, or due to undue warming up of electronic tube circuits, zero drift
sets in.
b) span drift or sensitivity drift
If there is proportional change in the indication all along the upward
scale, the drifts is called span drift or sensitivity drift.
c) Zonal drift:
In case the drift occurs only a portion of span of an instrument, it is
called zonal drift.
Drift
• The set of criteria defined for the instruments, which are changes
rapidly with time, is called dynamic characteristics.
• The various dynamic characteristics are:
i) Speed of response
ii) Measuring lag
iii) Fidelity
iv) Dynamic error
Speed of response - It is defined as the rapidity with which
a measurement system responds to changes in the measured
quantity.
Measuring lag - It is the retardation or delay in the
response of a measurement system to changes in the
measured quantity.
Dynamic error -It is the difference between the true value
of the quantity changing with time & the value indicated by
the measurement system if no static error is assumed. It is
also called measurement error
Fidelity - It is defined as the degree to which a
measurement system indicates changes in the measurand
quantity without dynamic error.
Measurement Standards
• International standards
• Primary standards
• Secondary standards
• Working standards
International standards
• These are defined on the basis of international agreement.
• They represent the units of measurements which are closest to
the possible accuracy attainable with present day technological
and scientific methods.
• International standards are checked and evaluated regularly
against absolute measurements in terms of the fundamental
units.
• These standards are maintained at the International Bureau of
Weights and Measures and are not available to the ordinary user
of measuring instruments for the purposes of calibration or
comparison.
• The international unit of length , the metre, was defined in
1960 in terms of the wavelength of Krypton-86.
• In 1983, the metre was redefined. The metre is now, the length
travelled by light in vaccum in a time interval of 1/299792458
sec.
Primary standards
• The Primary standards are the absolute standards which can be
used as the ultimate reference standards.
• These standards are maintained by National Standards
Laboratories in different parts of the world.
• The primary standards which represent the fundamental units
are independently calibrated by absolute measurements at each
of the national laboratories and the results of these
measurements are compared against each other, leading to a
world average figure for the primary standards.
• The primary standards are very few in number. These standards
have the highest possible accuracy and stability.
• The primary unit of mass is a prototype kilogram kept at
National Physical Laboratories of every country.
• This has an accuracy of 1 part in 108 and is occasionally verified
against the standard kept al the International Bureau
Secondary standards
• The secondary standards are the basic reference
standards used in industrial measurement laboratories.
• The responsibility of maintenance and calibration of these
standards lies with the particular industry involved
• They are sent periodically to the national standards
laboratories for calibration and comparison against
primary standards
• The secondary standards of mass are kept by industrial
laboratories. These standards have an accuracy of 1 ppm
and are checked against the primary standards.
Working standards
The origination of error may be in a variety of ways. They are categorized in three
main types.
• Gross error
• Random error
1. Gross Error
• The errors occur because of mistakes in observed readings, or using instruments and in
recording and calculating measurement results. These errors usually occur because of
human mistakes and these may be of any magnitude and cannot be subjected to
mathematical treatment.
2. Systematic Error: These are the errors that remain constant or change according to a definite
law on repeated measurement of the given quantity.
• Instrumental errors: Such errors are may be due to inherent short comings in the instrument,
misuse of the instruments ie failure due to adjust zero, poor initial adjustment etc, due to
loading effect of instruments.
• Environmental errors: These errors are introduced due to using an instrument in different
conditions than in which it was assembled and calibrated. These may be effects of
temperature, pressure , humidity, dust, magnetic or electrostatic fields etc
• Observational errors: Errors introduced by the observer. Eg parallax error due to observer not
having line of sight on pointer exactly at right angles to plane of scale
In tangent galvanometer, the magnitude of current passes through the coil determines by
the tangent of the angle of deflection of their coil, the horizontal component of the
earth magnetic field, radius and the number of turns of wire used. The most common
applications of this type of instrument are found in laboratories.
Secondary instruments
• In this type, the quantity under measurement can only be measured
by observing the output of the instrument.
• The secondary instruments should be calibrated by comparing with an
absolute instrument or another secondary instrument which has
already been calibrated against an absolute instrument.
• e.g:- Voltmeter, pressure gauge.
Null Type Instrument
• In this instrument, the zero or null deflection indicates the
magnitude of the measured quantity.
• The instrument has high accuracy and sensitivity. When the
value of the known and the unknown measuring quantities are
equal, the pointer shows the zero or null deflection.
Deflection Type Instrument
• The instrument in which the value of measuring quantity is
determined through the deflection of the pointer is known as
the deflection type instrument.
• The measuring quantity deflects the pointer of the moving
system of the instrument which is fixed on the calibrated scale.
Thus, the magnitude of the measured quantity is known
Electrical measuring instruments may be classified
according to their functions as;
• Indicating instruments
• Integrating instruments
• Recording instruments
Indicating Instruments-
It indicate the magnitude of an electrical quantity at
the time when it is being measured. The indications
are given by a pointer moving over a graduated dial.
e.g:- Ammeter, voltmeter, wattmeter.
Recording Instruments -
• The instruments which keep a continuous record of
the variations of the magnitude of an electrical
quantity to be observed over a defined period of time.
• The variations of the quantity being measured· are
recorded by a pen (attached to the moving system of
the instrument ; the moving system is operated by the
quantity being measured) on a sheet of paper carried
by a rotating drum
• E.g.- a recording ·voltmeter in a substation which
keeps record of the variations of supply voltage during
the day.
Integrating Instruments -
• Totalize the events over a specified period of time.
• The instruments which measure the total amount of
electrical energy supplied over a period of time. E.g.-
Energy meters.
ESSENTIALS OF INDICATING INSTRUMENTS OR OPERATING FORCES
coil
• Due to movement of Aluminum former an emf is induced
resulting in eddy current which opposes the motion thereby
provides a damping torque
PRINCIPLE OF OPERATION
When DC supply is given to moving coil, DC current flows through it.
Figure
Voltmeter Multipliers
• The combination of the meter movement and the
multiplier (series resistance ) is put across the
circuit whose voltage is to be measured.
• The multiplier limits the current through the meter
so that it does not exceed the value for full scale
deflection.
Moving Iron (MI) instruments
• One of the most accurate instrument used for both AC and DC
measurement is moving iron instrument.
• There are two types of moving iron instrument.
• Attraction type
• Repulsion type
Attraction type M.I. instrument
• A plate or vane of soft iron or of high permeability steel
forms the moving element of the system.
• This iron vane is so situated that it can move in a
magnetic field produced by a stationary coil.
• The coil is excited by the current or voltage under
measurement.
• When the coil is excited, it becomes an electromagnet
and the iron vane moves in such a way so as to occupy
a position of minimum reluctance.
• Thus the force (or torque) produced is always in such a
direction so as to increase the inductance of the coil
(As inductance increases as reluctance decreases)
• Hence the moving iron moves from the weaker
field outside the coil to the stronger field inside it
or the moving iron is attracted in.
• The controlling torque is provided by springs
• Gravity control can be
used for panel type of
instruments which are
vertically mounted
• Damping is provided
by air friction
Repulsion type MI Instruments
• In the repulsion type, there are two vanes inside the coil,
one fixed and other movable.
• These are similarly magnetised when the current flows
through the coil and there is a force of repulsion between
the two vanes resulting in the movement of the moving
vane.
(i) Radial Vane Type - In this type, the vanes are radial
strips of iron. The strips are placed within the coil as shown
in Fig. The fixed vane is attached to the coil and the
movable one to the spindle of the instrument
(Ii) Co-axial Vane Type - In this type of instrument, the fixed
and moving vanes are sections of co-axial cylinders
Torque Equation
Electrical energy supplied = Increase in stored energy
+ mechanical work done
• Let ‘ θ ’ be the deflection corresponding to a current of ‘i’ amp
• Let the current increases by di in dt seconds, the corresponding
deflection is θ + dθ
• There is change in inductance since the position of moving iron
changes w.r.t the fixed electromagnets
• Let the new inductance value be ‘L+dL’.
• Let the emf induced in the coil be ‘e’ volt.