Agilent 8697 Headspace Sampler Installation Checklist
Introduction
Customer Responsibilities
• Customers should leave the instrument shipment for the engineer to unpack.
• Customers should provide all necessary operating supplies upon request of the engineer.
• A customer representative should be available to the engineer while performing the installation.
• Being present at some installation tasks will be beneficial to you – refer to sections in this checklist
that are noted with (Customer Presence Recommended).
System Information
❑ Check this box if an instrument configuration report is attached instead of completing the table.
2.
3.
4.
5.
6.
7.
8.
9.
10.
❑ Unpack/verify the condition and completeness of shipment. For discrepancies, use the following
table.
❑ Before starting, discuss any specific questions or issues with the customer.
❑ Review with the customer the location in the lab for the instrument system and areas near the
instrument where consumables, accessories and tools will be stored.
❑ Before starting, discuss any configuration options with the customer.
❑ Check for required service note applicability and firmware update requirements.
❑ Upgrades only – Ensure with customer that instrument control settings, data, methods, etc. have
been properly saved or archived before starting any installation procedures.
Installation Procedure
HS Installation Tasks
❑ Place the HS on the bench.
❑ Verify that the customer site line voltage is compatible with the configuration plug and power
cord that was shipped with the unit.
❑ Remove the tray from the packing box.
❑ Remove the shipping screws and clamps from the tray using the T-20 screwdriver provided in the
instrument ship kit.
❑ 120-Vial tray only: Lift the two clips from the rail
❑ Install the tray on top of the HS unit and connect the cable from the tray to the HS unit.
❑ Install the two vial racks for 8697 HS 48-vial tray and five vial racks for 120-vial tray
❑ Remove the transferline from the packing box.
❑ Install the transferline to the HS unit and connect the heater/sensor cable to the HS unit.
Service Review
❑ Attach available reports/printouts to this documentation.
❑ Record the time/date of installation or upgrade completion in the customer’s records/logbook.
❑ Complete the following Service Engineer comments section if there are additional comments.
❑ Review the installation/upgrade with the customer.
❑ Explain Agilent warranty for instruments.
❑ Explain how to use manuals, guides, and online help.
❑ Explain how to get self-help, and FAQs online.
❑ Explain how to log an instrument service call and support services that are available.
❑ Advise customer of additional instrument training options.
❑ If the instrument firmware was updated, record the details of the change in the service engineer’s
comments box or if necessary, in the customer’s IQ records.
❑ Provide the customer a copy of the Smart Alerts flyer.
❑ Describe Smart Alerts to the customer.
❑ Install Smart Alerts if requested.
If there are any specific points you wish to note as part of performing the installation or other items of
interest for the customer, please write include them in this box.
Service Completion
Service request number Date service completed