Lecture 3 (Introduction To Pure Tone Audio Me Try)
Lecture 3 (Introduction To Pure Tone Audio Me Try)
Lecture 3 (Introduction To Pure Tone Audio Me Try)
Lecture 3
References
Audiology Diagnosis Chapter 11 (Pure Tone Tests) Lecture Notes
Topics
Pure tone audiometry Definition and purposes Audiometer Definition & types Audiometer components Calibration of audiometers Calibration schedules Audiogram and audiometric symbols
Definition:
The procedure by which sensitivity thresholds for each ear are obtained for pure tone stimuli of different frequencies (250, 500, 1000, 2000, 4000 & 8000 Hz)
The lowest intensity level at which the patient is able to respond to the stimulus 50% of the time
Stimulus: anything that can elicit or evoke a response in an excitable receptor i.e. stimulus = signal (pure tone or noise) excitable receptor = ear / hearing system
50 % means the patient must respond to two out of four or three out of six trials
Purposes:
Results from pure tone audiometry are used to make the initial diagnosis of hearing impairment (normal/abnormal hearing sensitivity)
In the case of abnormal hearing sensitivity, the results would help in the diagnosis of the type and degree of hearing impairment
It assist in the determination of (re)habilitation services needed to assist hearing impaired persons
The instrument used in pure tone testing Electronic instrument used to quantify hearing sensitivity Have frequencies with a scale based on the octave and half-octave frequencies of 125, 250, 500, 750, 1000, 1500, 2000, 3000, 4000, 6000, and 8000 Hz.
Standard Pure Tone and Automatic/Microprocessor Audiometers Diagnostic Audiometers Extended High-Frequency Audiometers Free-Field Equivalent Audiometers
Audiometers Types
Standard Pure Tone and Automatic/Microprocessor Audiometers: Used for screening or basic threshold testing
Commonly used for mass screening (i.e. industrial hearing conservation programs) to save time Have limited value in testing children and in diagnostic testing
Audiometers Types
Diagnostic Audiometers
Includes a wide variety of features; bone conduction testing, speech testing through monitored live voice or recorded presentations, masking stimuli and site of lesion testing
To measure hearing thresholds from 8000 Hz to 16,000 Hz Application: to monitor onset of possible high frequency HL from ototoxic medications
Audiometers Types
Free-Field Equivalent Audiometers
Calibrated so the output of the earphones or bone oscillator can be expressed in terms of equivalent hearing as if the sound was presented at 0 degrees azimuth in the sound field
The output from free field equivalent audiometers is expressed in SPL (sound pressure level)
Good to comparison of performance for hearing aids
Audiometer Components
Calibration Of Audiometers
Calibration Of Audiometers
Calibration is necessary to ensure that the audiometer is producing a pure tone at the specified frequency and intensity The stimulus is present only in the earphone to which it is directed The stimulus is free from unwanted noise, the interference, and distortion
Calibration Schedules
Four types of audiometer check/calibration schedules:
In the morning after warm up (10 min) Signal is checked at various intensities and at all frequencies Look for transient clicks or distortion (undesired acoustic signal) Ensure that signal is delivered to the correct earphone No cross talk is present (cross talk: when the signal intended for exclusive routing to one earphone is also routed to the other)
This process is done either with: A. A. Psychoacoustic (biologic) check OR Electroacoustical calibration check with a sound level meter
Obtain baseline thresholds on 3-5 normal hearing individuals who will be available for comparison testing throughout the year If on the monthly check, a threshold difference > 5 dB HL is found for one of the individuals for any test frequency between 500-6000 Hz, the other subjects should be checked If a shift > 5 dB HL in the same direction is confirmed, an electric calibration of the audiometer is required Monthly results should recorded in a form
Sound level meter: an electronic instrument designed to measure sound intensity in dB in accordance with an accepted standard
ANSI (American National Standards Institute): association of specialists, manufacturers, and consumers that determines standards for measuring instruments, including audiometers
At least one each year To make sure that the audiometer is meeting the minimum standards defined by ANSI S3.6-1996
Every 5 years
More comprehensive than the periodic electronic calibration Includes testing of all settings on the frequency and intensity dials Replacing switches, cords, earphone drivers and cushions
The Audiogram
Audiogram form either records both right and left ear information on one
The Audiogram
Audiometric Symbols
Air conduction: Response in right ear (red) unmasked: O Masked : Response in left ear (Blue) Unmasked: X Masked: Sound Field (Speakers): S (sound field) OR F (free field)
Audiometric Symbols
After all thresholds are plotted on the audiogram, the symbols are connected with a: Solid red line for the right ear Solid blue line for the left ear
Audiograms Forms
Right ear (RE) & left ear (LE) results plotted on the same audiogram Right ear (RE) & left ear (LE) results plotted on two separate audiograms