Using File Processing Mode To Optimize Particle Image Capture
Using File Processing Mode To Optimize Particle Image Capture
INTRODUCTION
3. Adjusting Capture Settings: Users iteratively adjust specific 3. After successfully autofocusing, flush out any remaining bead
capture settings prior to performing File Processing mode. solution, followed by at least 1 mL of particle-free water. Flush
all fluid out of the flow cell before proceeding.
4. Re-Analyzing Raw Images with File Processing Mode: The raw
images are re-segmented with the adjusted capture settings 4. From the main VisualSpreadsheet window, select Context and
from Step 3 and can then be qualitatively reevaluated. navigate to the Capture tab. For FlowCam 8000 Series instruments, we
recommend starting with the settings shown in Table 1 as a baseline.
Dark Pixels 17 17 20 12 12 12
Particles defined by
Light Pixels 17 17 20
dark pixels only EVALUATING PARTICLE IMAGES FROM INITIAL DATA
Table 1. Recommended starting capture settings for File Processing Mode by
CAPTURE
Application/Camera Type
Once you’ve completed a data run with raw image capture enabled,
select Open View to evaluate the particle images collected during
5. While still in the Capture tab, ensure that the Save Raw Camera the measurement. From the view window, toggle on the particle
Images checkbox is checked and Max Raw Camera Images is edge trace overlay (Show > Particle Edge Trace) to see where pixels
set to a high value (e.g., 5000), as shown in Figure 1. Click File > were detected in each image and assess the segmentation accuracy.
Save As and save this context as ObjX_General Settings, adding This step is significant as it helps you determine if further setting
the objective magnification in place of the Obj placeholder. optimization is required. If the edge trace qualitatively agrees with
the outer particle edge for most images, as shown in Figure 2, then
your initial data capture was successful.
Figure 2. The red line indicates the particle edge trace and aligns closely with the
outer particle edge. In this case, no setting optimization is necessary.
2
Recommended
Example Images Impact on Measurements Capture Setting
Adjustment
1. ↓ Decrease
Dark and/or Light
c) Holes are • Particles will be undersized because
Pixel Thresholds
holes are not included in calculations
present in the • Errors in other particle property
depending on which
segmentation pixels are omitted
measurements
2. ↑ Increase Close
Holes
1. ↓ Decrease Dark
• Particles will be undersized and
and/or Light Pixel
overcounted as a result of fragmenting
Thresholds
a single particle into multiple smaller
d) Fragmentation pieces
2. ↑ Increase Close
Holes
• It will be challenging to classify
3. ↑ Increase Distance
particles appropriately
to Nearest Neighbor
• ↓ Decrease
f) Missing Dark and/or Light
Pixel Thresholds
particles from • Specific particle types will be
depending on which
raw camera undercounted
particles are ignored
images • Check Capture Filter
settings
Table 2. Examples of sub-optimal capture settings, their impact on measurements, and suggested capture setting adjustments. In a), blank background images are being
detected. Compare the edge trace in b) to Figure 2 and note the extensive haloing occurring here. Compare the edge trace in c) to Figure 2, and note the holes present here.
In d), a single Fragilaria diatom has been segmented into multiple image fragments, and in e), multiple particles are being counted as one particle. In f), the small particle in
the lower right was not segmented from the raw camera image and thus not counted.
3
ADJUSTING THE CAPTURE SETTINGS 2. Click the File Processing Mode button from the operations bar in
the main window of VisualSpreadsheet (FlowCam version) or click
To optimize capture settings using File Processing Mode most
Analyze > File Processing Mode (Satellite version) (Figure 4).
efficiently, we recommend making modifications in the order and
increments shown in Table 3, after starting with the context settings
previously shown in Table 1.
# to raise or low-
Adjustment
Property er the value in
order
each iteration
1 Dark Pixel Threshold 2
2 Light Pixel Threshold 5 Figure 4. Enabling File Processing Mode on FlowCam (top) and on a satellite
computer (Bottom)
3 Close Holes (Iterations) 1
Distance to Nearest Neighbor
4 1
(µm) 3. A File Explorer Window will appear. Navigate to the location
Table 3. Order and increment in which the capture settings should be adjusted of your saved raw image data—the default location is
during File Processing Mode DocumentsVisualSpreadsheet6 data. Select the first raw image
file, which will be named rawfile_000000, and click Open (Figure 5).
4
7. Once the dark pixel threshold has been optimized, continue
to move through the settings in Table 3 until the particle edge
trace closely aligns with the true particle edge and you don’t see
the artifacts described in Table 2. Once your settings have been
optimized, make sure to save your new context file so it can be
used for future data acquisition.
SUMMARY
FlowCam instruments are flexible platforms that can accurately
analyze a variety of samples and particle types in wide-ranging
applications. File Processing Mode is an easy-to-use technique for
method development that allows users to capitalize on this flexibility
Figure 6. Comparison of images captured from original run (top) to those where
and efficiently optimize capture settings for specific sample and
Dark Pixel Threshold was increased to 14 (bottom). In this case, the particle edge
trace is more accurate for the original images than for the re-processed images particle types without the need to re-run samples. Imaging particles
where portions of the exterior and interior regions of particles were omitted (yellow with capture settings that have been optimized through file
arrows). Note the decrease in Diameter ABD shown below the images because of
these missing regions. processing mode ensures that users are collecting accurate sizing,
morphological, and concentration data.
Get in touch
1 Tell us about your needs and how we can
support you