This document provides guidance on optimizing liquid chromatography-mass spectrometry (LCMS) methods. It discusses various aspects of method development including mobile phase selection, column optimization, and parameter adjustment. The key steps outlined are selecting an appropriate mobile phase, adjusting column conditions for compatibility with mass spectrometry, confirming MS methods, and optimizing sample preparation processes. Parameters like flow rate, gradient time and range, column length, and diameter can be adjusted to improve method performance while maintaining selectivity. Proper method validation is also emphasized.
This document provides guidance on optimizing liquid chromatography-mass spectrometry (LCMS) methods. It discusses various aspects of method development including mobile phase selection, column optimization, and parameter adjustment. The key steps outlined are selecting an appropriate mobile phase, adjusting column conditions for compatibility with mass spectrometry, confirming MS methods, and optimizing sample preparation processes. Parameters like flow rate, gradient time and range, column length, and diameter can be adjusted to improve method performance while maintaining selectivity. Proper method validation is also emphasized.
TM TM TM 2 TM TM Met hod devel opMent & Va l i da t i on f or Ma ss spect r oMet r i c a na l ysi s b y N a v na t h Ja yb ha y e Bi oA na ly t i ca l Technol ogi es ( Ind i a ) Pv t Lt d 3 TM TM Agenda Method Development Pathway Method Optimization for LCMS New Method Development Method Validation Reporting the Sample Analysis Results References 4 TM TM Method Optimization for LCMS 5 TM TM Method Development Pathway 6 TM TM Four Steps Mobile Phase : MS- compatible mobile phase Analytical Column: Adjusting column conditions with respect to MS Confirming the MS methods Sample Preparation Process Optimization Methods Optimization 7 TM TM Mobile Phase Requirements Mobile Phase Solvent -Volatility, Amount of Water, ionization issues Mobile Phase Additives -Volatility, Concentration, ion suppression, buffering capacity Buffer/pH control -Formic acid, Ammonium formate, Ammonium acetate, Trifluroacetic acid 8 TM TM Mobile Phase Requirements Mobile phase degassing is an important step in the LC/MS experiment and can be accomplished via on-line membrane or vacuum devices, sonication,helium sparging. Degassing will eliminate pump, cavitation, ensure reproducible retention times and minimize possible sputering from ion the source. 9 TM TM Organic Components contd Acetonitrile and Methanol are almost exclusively chosen in LC/MS method as organic mobile-phase components. Methanol has: - Greater phase acidity, polarity, and volatility than acetonitrile. - Has been shown 10-50% better sensitivity than acetonitrile in positive ion mode. 10 TM TM Limitation there is small chance for methanol analyte adduct formation Less frequently used solvent: Ethanol, 2-Propanol, Tetrahydofluran Solvent used for Normal Phase Separation: Toluene, Hexane, Cyclohexane,Dichloromethane and Carbon Tetrachoride Organic Components contd 11 TM TM Aqueous Components Careful choice of an appropriate acid or buffer will help ensure success in the LCMS experiment Nonvolatile aqueous components greatly decrease and prevent the detection of ion Nonvolatile aqueous components can also foul ion sources and vacuum region of MS Nonvolatile phosphate and citrate buffer are strongly discouraged from both ionization and practical reasons Ion suppression and decreased sensitivity will be observed when non-volatile buffers are used 12 TM TM Buffer Consideration: Phosphate Buffers are commonly used in HPLC because of their strong buffering capacity and useful pH range unfortunately, phosphate cannot be used, because they will crystallize out and contaminate the ion source Aqueous Components contd 13 TM TM Acid and Bases Formic Acid or Acetic Acid concentration of 0.1 to 1%(V/V) are recommended in preparing low pH Mobile phases to enhance ionization in Electrospray Trifluoro-acetic acid is preferred for protein and peptide separation, but should be avoided when negative Ion mode is utilized Ammonium hydroxide, or, in rarer cases, Triethylamine, are recommended for high pH mobile phases 14 TM TM Recommended MP Additives and Buffers for LCMS 15 TM TM Isocratic Vs.Gradient Seperation Isocratic elution: A separation that employs a single solvent or solvent mixture of constant composition. Gradient elution: Here two or more solvent systems that differ significantly in polarity are employed. After elution is begun; the ratio of the solvents is varied in a programmed way, sometimes continuously and sometimes in a series of steps. Separation efficiency is greatly enhanced by gradient elution 16 TM TM Column Selection & Optimzation Column selection will based on following requirements: Chromatographic separation required Resolution expected Asymmetric factor 17 TM TM Analytical Column Optimization Column selection will based on following information's Compound Specific information Nature of compound Solubility of compound Polarity of compound Analytical Column Information Column Length Column Chemistry (C8 or C18, ) Physical Properties of column (Pore size, Particle size & diameter) Sample Specific Information Origin of Sample Purity of sample Sample Preparation technique 18 TM TM Adjust Column Diameter 19 TM TM Eliminate Extra-Column Volume 20 TM TM Reduce Column Length 150 x 2.1 mm 0.2 mL/min 1500 psi 50 x 2.1 mm 0.2 mL/min 500 psi 0 2 4 6 8 10 Time (min) 0 2 4 6 8 10 Time (min) 21 TM TM Increase Flow Rate 50 x 2.1 mm 0.2 mL/min 500 psi 50 x 2.1 mm 0.6 mL/min 1500 psi 0 2 4 Time (min) 0 2 4 Time (min) 22 TM TM Isocratic vs. Gradient Retention Isocratic: log(k) = log(k 0 ) - Su k* ~ t G F Au V m S Gradient: Remember: k controls selectivity 23 TM TM Gradient Separation In gradient separation, k* controls selectivity. However, as opposed to isocratic separation, one needs to be careful to make compensating adjustments to maintain k* if one of these factors is changed: tG (gradient time), flow rate (F), gradient range (Df), or column volume (Vm). S is a constant, with a value of 5 suitable for small molecule calculations. Note that k* is not influenced by particle size. 24 TM TM Eliminate Gradient Waste 25 TM TM Shorten Column 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 Time (min) 0 2 4 6 8 10 Time (min) 150 x 4.6 mm, 5 m 1.5 mL/min 15-35%B / 10 min 30 x 4.6 mm, 3 m 1.5 mL/min 15-35%B / 2 min t G F Au V m S 26 TM TM Reduce Diameter 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 Time (min) 30 x 4.6 mm, 3 m 1.5 mL/min 15-35%B / 2 min 30 x 2.1 mm, 3 m 0.3 mL/min 15-35%B / 2 min 1.0 2.0 Time (min) t G F Au V m S 27 TM TM Other parameter consider in Method optimization Define specificity & selectivity Define LOD & LLOQ Optimize the sample Preparation Method Indentify Level of Matrix effect (ME) and system carry over (SCO) Re-optimization the Method to reduce ME& SCO Calibration curve building Pre Method validation Verification 28 TM TM Summary Use a volatile mobile phase Scaling isocratic runs (i.d., length, flow) Keep k* constant when scaling gradients Minimize extra-column effects Beware of dwell volume 29 TM TM Steps New Method development Mobile phase selection and Optimization Analytical column selection and optimization Detector selection and its parameter optimization Chromatographic optimization Define specificity & selectivity Define LOD & LLOQ Optimize the sample Preparation Method Indentify Level of Matrix effect (ME) and system carry over (SCO) Re-optimization the Method to reduce ME& SCO Calibration curve building Pre Method validation Verification 30 TM TM 30 Method Validation Parameters Validation of Analytical Method validation is the process used to confirm that the analytical procedure employed for a specific test is suitable for its intended use. The various validation parameters are: Accuracy, Precision (Repeatability and Reproducibility), Linearity and Range, Limit of detection (LOD) / Limit of quantitation (LOQ), Selectivity / Specificity, Robustness / Ruggedness, Stability and System suitability studies 31 TM TM Validation methods Selective and sensitive analytical methods for the quantitative determination of drugs and their metabolites (analytes) are critical for successful performance of PK and bioequivalence studies. Validation of analytical methods includes all the procedures recommended to demonstrate that a particular method, for a given matrix, is reliable and reproducible 32 TM TM Validation methods 1. A prior validation: Pre-study validation for analytical method development and method establishment 2. In-life validation (Routine validation) 33 TM TM Regulatory requirements G.L.P. (e.g.; bioequivalence, Toxicokinetics) S.O.P. (standard operating procedure) (from sample collection to reporting) Record keeping Chain of sample custody Sample preparation Analytical tools Procedures for quality control and verification of results 34 TM TM A prior validation makes sure the method is suitable for its intended use 35 TM TM A prior validation: criteria to be validated 1. Calibration curve 2. Accuracy 3. Precision (repeatability, reproducibility) 4. Limit of quantification (LOQ) 5. Limit of detection (LOD) 6. Sensitivity 7. Specificity/selectivity 8. Stability of the analyte in the matrix under study 9. Others (ruggedness, agreement,) 36 TM TM 1. Calibration curve 37 TM TM Definition It is the relationship between known concentrations and experimental response values Goal Determine the unknown concentration of a sample Calibration curve 38 TM TM Response: dependent variable (peak,area ..) Y
( o b s e r v e d ) Y X y = ax + b Independent variable: exactly known concentrations Calibration curve Calibration curve x1 y1 xn Yn 39 TM TM Response: dependent variable Y
( o b s e r v e d ) Y X y = ax + b Independent variable: X estimated concentration ^ Calibration curve Calibration curve x1 y1 xn Yn 40 TM TM Calibration curve Construction 5 to 8 points over the analytical domain replicates are required to test linearity 3 to 5 replicates per levels 41 TM TM Calibration curve The calibration curve should be prepared in the same biological matrix (e.g. plasma ) as the sample in the intended study by spiking with known concentration of the analyte (or by serial dilution). 42 TM TM Reference Standard Calibration standards and quality control samples (QC) Authenticated analytical reference standard should be used to prepare (separately) solution of known concentration certified reference standards Never from a marketed drug formulation commercially supplied reference standards other material of documented purity 43 TM TM Building the calibration curve: a regression problem 44 TM TM Statistical requirements to build a calibration curve 45 TM TM Statistical requirements to build a calibration curve 1. Standard concentration (X) are known without error 2. Variance of response (Y) should be constant over the analytical domain (homoscedasticity hypothesis); this equivalent to say that the random errors i are homoscedastic i.e., they all have the same variance. 3. The random errors i have expected value 0. 4. The random errors i should be independent from Y and are uncorrelated. These assumptions imply that least-squares estimates of the parameters are optimal in a certain sense 46 TM TM Regression can be used for prediction 47 TM TM Regression can be used for prediction These uses of regression (calibration curve) rely heavily on the model assumptions being satisfied. Calibration curve is misused for these purposes where the appropriate assumptions cannot be verified to hold The misuse of regression is due to the fact that it take considerably more knowledge and experience to critique a model than to fit a model with a software. 48 TM TM Assessing the calibration curve the calibration curve (here a statistical model ) should be checked for two different things: 1. Whether the assumptions of least-squares are fulfilled Analysis (inspection) of residuals 2. Whether the model is valid and useful Test of linearity Back calculations 49 TM TM Calibration curve : Linearity Calibration curve : Linearity - Specific tests of linearity should be used - The coefficient of correlation (r) cannot assess linearity except for r = 1 e.g.: r = 0.999 can be associated with a calibration curve which is not a straight line 50 TM TM R e s p o n s e
Y X Calibration curve: linearity Calibration curve: linearity Concentration Test of linearity : Coefficient of correlation r = 0.99 does not prove linearity 51 TM TM Freeze/thaw stability Avoid freeze and thaw cycles Enough aliquot samples should be to be prepared 52 TM TM Accuracy and precision 53 TM TM Origin of the error : Accuracy and precision Systematic (not random) bias impossible to be corrected accuracy Random can be evaluated by statistics precision 54 TM TM Good Precision Good Accuracy Poor Precision Good Accuracy Good Precision Poor Accuracy Poor Precision Poor Accuracy Gold Standard Silver Standard Off-Base Model Hit or Miss Model Bias and precision 55 TM TM Accuracy Closeness of determined value to the true value. The acceptance criteria is mean value s 15% deviation from true value. At LOQ, 20% deviation is acceptable. 56 TM TM Accuracy Accuracy The accuracy is calculated using the following equation : Accuracy (%) = 100 x Found value - Theoretical value Theoretical value The accuracy at each concentration level must be lower than 15% except a LOQ (20%) 57 TM TM Accuracy Determination by replicate analysis of the sample containing known amount of analyte 5 samples for at least 3 levels The mean value should be within 15% of the actual value except at LOQ where it should not deviate by more than 20% 58 TM TM Precision The closeness of replicate determinations of a sample by an assay. The acceptance criteria is s 15% CV. At LOQ, 20% deviation is acceptable. 59 TM TM Repeatability (r) Agreement between successive measurements on the same sample under the same conditions Reproducibility (R) The closeness of agreement between results obtained with the same method under different conditions Precision Precision 60 TM TM PrecisionConsidered at 3 Levels Repeatability Intermediate Precision Reproducibility 61 TM TM Repeatability Express the precision under the same operating conditions over a short interval of time. Also referred to as Intra-assay precision (within day) 62 TM TM Intermediate Precision Express within-laboratory variations. Between days variability Known as part of Ruggedness in USP 63 TM TM Reproducibility Definition: Ability reproduce data within the predefined precision Repeatability test at two different labs 64 TM TM Precision: measurement Should be measured using a minimum of 5 determinations per concentration A minimum of 3 concentrations in the range of expected concentrations The precision at each concentration should not exceed 15% except for the LOQ (20%) 65 TM TM Precision: measurement for a single measurement : CV(%) for intra-day and inter-day precision ANOVA 66 TM TM Precision: data analysis Single level of concentration with repetition e.g. 12, 13, 12, 14, 13, 14 g/mL mean : 13.0 g/mL SD: 0.8944 g/mL CV% = SD/mean * 100 = 6.88% CV% is also known as the relative standard deviation or RSD 67 TM TM Precision: data analysis Several levels of concentration and several days day 1 levels (g/mL) 0.5 5 20 Repetitions 0.4 5.2 20.5 0.5 5.1 21.0 0.4 4.9 19.8 0.6 5.2 18.8 day 2 and 3 : same protocol ANOVA 68 TM TM Precision: the statistical model The statistical model (for each concentration level) Y = + day + c : general mean day: an effect (day, technician, or any factor = inter ) c: error-random = intra 69 TM TM ANOVA Allows an estimation of the 2 variance terms inter-day mean square (BMS) intra-day mean square (WMS) 70 TM TM Repeatability and reproducibility SD for repeatability o r = Var (e) SD for reproducibility o R = o (day) + o (r) variance for reproducibility is the sum of the variance for repeatability and the inter-day variance Inter-day intra-day 71 TM TM Precision: ANOVA CV intra: 5% CV inter: 8% CV inter > CV intra 72 TM TM The limit of quantification (LOQ) LOQ is the lowest amount of analytes in a sample which can be determined with defined precision and accuracy LOQ : 20% 73 TM TM Limit of quantification (LOQ) The lowest standard on the calibration curve is the LOQ if: no interference is present in the blanks at retention time of the analyte for this concentration the response (analyte peak) has a precision of 20% and accuracy 80-120% 74 TM TM Estimation of chromatographic baseline noise Estimation of chromatographic baseline noise N p-p N p W : Peak width 1 Baseline noise Largest variation of the baseline noise (N ) p-p Most important deviation (N ) p Sample chromatogram Blanc chromatogram (a) (b) 75 TM TM Three analytical areas Three analytical areas 1 2 3 Xb not detected Area of detection Area of quantification or CV<20% LOD LOQ 76 TM TM The recovery of an analyte in an assay is the detector response obtained from an amount of the analyte added to and extracted from the biological matrix, compared to the detector response obtained for the true concentration of the pure authentic standard The recovery allows to determine the percent of lost drug during sample preparation Minimal extraction ratio required to ensure a good repeatability Recovery: definition 77 TM TM Absolute recovery is evaluated using low, medium, and high QC samples and at least three times for each level The extraction recovery of the analyte (s) and internal standard(s) should be higher than 70%, precise, and reproducible. Recovery: Determination Recovery: Determination 78 TM TM Recommended to be a close analog of the analyte of interest Advantages and limits Recovery: Internal standard Recovery: Internal standard 79 TM TM Recovery ) _ tan , / _ _ _( _ ) _ , / _ _ _( _ 100 cov Re solution dard s mL ng x of area Peak extract plasma mL ng x of area Peak ery = 80 TM TM Specificity / Selectivity (1) Specificity : for an analyte ability of the method to produce a response for a single analyte metabolites enantiomers Selectivity: for a matrix 81 TM TM Specificity / Selectivity (2) Analyses of blank samples from different subjects (n=6) Blanks should be tested for interference using the proposed extraction procedure and other chromatographic conditions Results should be compared with those obtained with aqueous solution of the analyte at a concentration near the LOQ Blank plasma and pre-dose samples should be without interference 82 TM TM Specificity / Selectivity (3) If more than 10% of the blank samples exhibit significant interference, the method should be changed to eliminate interference 83 TM TM Definition The drug must keep all its properties during the investigations Stability at room temperature An experiment should cover 6 to 24h Stability in frozen biological samples : (-20C or -80C) Stability sample should allow assay from day 0 to day 20 Stability during a freeze / thaw cycle Samples should be frozen and submitted to three freeze / thaw cycles Aliquotage is better than repeated freeze / thaw cycles Stability Stability 84 TM TM To see this guidance 85 TM TM To see this guidance 86 TM TM Quest ions please?...